Operating Instructions—TG 501
Remove U100 before using external clocks. Interface
connector pin 24B is used for External Clock Input with
the associated ground on pin 25B. Fig. 1-2 shows the
location on the main circuit board of the jumpers that must
be installed to use the TG 501 with the various clock
choices. The required jumpers are:
Standard Internal Reference
1MHz Crystal No jumpers required
Buffered Output of Internal
Clocks
External 1MHz clock
(Remove U100)
External 5MHz clock
(Remove U100, Add U50)
External 10 MHz clock
(Remove U100, Add U50)
Install J2
Install J1, J5
Install J1, J4, J7
Install J1, J4, J6
OPERATING MODES
Calibrated Mode
When center-button of VARIABLE TIMING Control is
pushed in, the MARKER Control selects calibrated marker
intervals from 5sto 10 ns indicated by the lighted knob
skirt. To select 5, 2or 1ns modified sine wave markers, set
MARKER Control to 52 IN position and depress desired
pushbutton.
Variable Timing Mode
When center button of VARIABLE TIMING Control is
out, output timing is faster or slower than the calibrated
marker to the percentage indicated by the VARIABLE
TIMING READOUT Display and the FAST-SLOW Indica-
tor lights. 5, 2and 1ns push-button controlled modified
sine waves are not variable and Readout Display is turned
off when MARKER Control is in 521N position.
Readout Display Test
To check the operation of the display LED's (Light-
Emmiting Diodes), fully depress the center button of the
VARIABLE TIMING Control (push in firmly). All LED
segments should light, displaying 8.8 in the Readout.
TYPICAL APPLICATIONS
The TG 501 can be used for many applications not
described in the manual. Use the following examples to
become familiar with the TG 501's VARIABLE TIMING
READOUT. Contact your Tektronix Field Office or rep-
resentative for making specific measurements with this
instrument.
Time Base Measurements
To accurately determine the timing error of atime base:
1. Connect the MARKER OUT of the TG 501 to the
Vertical Input of the oscilloscope. Use a50 coaxial cable
terminated into 50 SI.
2. connect the TRIGGER OUT of the TG 501 to the
External Trigger Input of the time base. Use a50 S2 coaxial
cable terminated into 50 S2.
3. Select external trigger source on the time base.
4. Select, for example, 1jjis sweep rate of the time base.
5.
Set MARKER Control rotary switch of TG 501 to
1/is.
6.
Make sure center button of VARIABLE TIMING
Control, concentric with MARKER Control, is out (Vari-
able Timing Mode).
7.
Use Horizontal Position control of time base to make
first time mark coincide with first graticule line on the
oscilloscope. See Fig. 1-3.
8.
Rotate VARIABLE TIMING knob until all time
markers coincide with graticule lines on the oscilloscope, as
shown in Fig. 1-4.
1-3