manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. Tektronix
  6. •
  7. Measuring Instrument
  8. •
  9. Tektronix 3177 User manual

Tektronix 3177 User manual

Other Tektronix Measuring Instrument manuals

Tektronix Keithley 2601B-PULSE User manual

Tektronix

Tektronix Keithley 2601B-PULSE User manual

Tektronix PA4000 Series User manual

Tektronix

Tektronix PA4000 Series User manual

Tektronix LumaColor II J18 User manual

Tektronix

Tektronix LumaColor II J18 User manual

Tektronix WVR4000 Use and care manual

Tektronix

Tektronix WVR4000 Use and care manual

Tektronix MSO56B Instruction manual

Tektronix

Tektronix MSO56B Instruction manual

Tektronix 1340 User manual

Tektronix

Tektronix 1340 User manual

Tektronix RSA6100A Series User manual

Tektronix

Tektronix RSA6100A Series User manual

Tektronix Series Manual

Tektronix

Tektronix Series Manual

Tektronix 495 User manual

Tektronix

Tektronix 495 User manual

Tektronix 7D20 User manual

Tektronix

Tektronix 7D20 User manual

Tektronix P6417 User manual

Tektronix

Tektronix P6417 User manual

Tektronix P6563A User manual

Tektronix

Tektronix P6563A User manual

Tektronix MTS4000 MPEG User manual

Tektronix

Tektronix MTS4000 MPEG User manual

Tektronix P6205 User manual

Tektronix

Tektronix P6205 User manual

Tektronix Keithley 4200A-SCS User manual

Tektronix

Tektronix Keithley 4200A-SCS User manual

Tektronix RSA7100A Manual

Tektronix

Tektronix RSA7100A Manual

Tektronix BPA100 User manual

Tektronix

Tektronix BPA100 User manual

Tektronix WFM8300 Administrator Guide

Tektronix

Tektronix WFM8300 Administrator Guide

Tektronix RSA306 Use and care manual

Tektronix

Tektronix RSA306 Use and care manual

Tektronix PRISM MPI2-10 User manual

Tektronix

Tektronix PRISM MPI2-10 User manual

Tektronix RSA5106B Installation and maintenance instructions

Tektronix

Tektronix RSA5106B Installation and maintenance instructions

Tektronix OI tek P5200A Series Manual

Tektronix

Tektronix OI tek P5200A Series Manual

Tektronix DC 501 User manual

Tektronix

Tektronix DC 501 User manual

Tektronix A622 User manual

Tektronix

Tektronix A622 User manual

Popular Measuring Instrument manuals by other brands

Powerfix Profi 278296 Operation and safety notes

Powerfix Profi

Powerfix Profi 278296 Operation and safety notes

Test Equipment Depot GVT-427B user manual

Test Equipment Depot

Test Equipment Depot GVT-427B user manual

Fieldpiece ACH Operator's manual

Fieldpiece

Fieldpiece ACH Operator's manual

FLYSURFER VIRON3 user manual

FLYSURFER

FLYSURFER VIRON3 user manual

GMW TG uni 1 operating manual

GMW

GMW TG uni 1 operating manual

Downeaster Wind & Weather Medallion Series instruction manual

Downeaster

Downeaster Wind & Weather Medallion Series instruction manual

Hanna Instruments HI96725C instruction manual

Hanna Instruments

Hanna Instruments HI96725C instruction manual

Nokeval KMR260 quick guide

Nokeval

Nokeval KMR260 quick guide

HOKUYO AUTOMATIC UBG-05LN instruction manual

HOKUYO AUTOMATIC

HOKUYO AUTOMATIC UBG-05LN instruction manual

Fluke 96000 Series Operator's manual

Fluke

Fluke 96000 Series Operator's manual

Test Products International SP565 user manual

Test Products International

Test Products International SP565 user manual

General Sleep Zmachine Insight+ DT-200 Service manual

General Sleep

General Sleep Zmachine Insight+ DT-200 Service manual

Sensa Core Lacto Spark user manual

Sensa Core

Sensa Core Lacto Spark user manual

VOLTCRAFT UTS-1980 operating instructions

VOLTCRAFT

VOLTCRAFT UTS-1980 operating instructions

EBCHQ 94915 Operation manual

EBCHQ

EBCHQ 94915 Operation manual

Balluff BIP LD2-T017-01-EP-S4 Series user guide

Balluff

Balluff BIP LD2-T017-01-EP-S4 Series user guide

Pulsar IMP Lite Series instruction manual

Pulsar

Pulsar IMP Lite Series instruction manual

ETCR 6800 user manual

ETCR

ETCR 6800 user manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.

INSTRUCTION
MIAN
UAL
TYPE
3177
SAMPLING
SWEEP
UNIT
Tektronix,
Inc.
S.W,
Millikan
Way
@
P.O.Box
500
@
Beaverton,
Oregon
@
Phone
MI
4-0161
@
Cables:
Tektronix
070-333
763
Type
3177
WARRANTY
All
Tektronix
instruments
are
warranted
against
defective
materials
and
workman-
ship
for
one
year.
Tektronix
transformers,
manufactured
in
our
own
plant,
are
war-
ranted
for
the
life
of
the
instrument.
Any
questions
with
respect
to
the
war-
ranty
mentioned
above
should
be
taken
up
with
your
Tektronix
Field
Engineer.
Tektronix
repair
and
replacement-part
service
is
geared
directly
to
the
field,
there-
fore
all
requests
for
repairs
and
replace-
ment
parts
should
be
directed
to
the
Tek-
tronix
Field
Office
or
Representative
in
your
area,
This
procedure
will
assure
you
the
fastest
possible
service.
Please
include
the
instrument
Type
and
Serial
number
with
all
requests
for
parts
or
service.
Specifications
and
price
change
priv-
ileges
reserved.
Copyright
©)
1963
by
Tektronix,
Inc.,
Beaverton,
Oregon.
Printed
in
the
United
States
of
America.
All
rights
reserved.
Contents
of
this
publication
may
not
be
re-
produced
in
any
form
without
permission
of
the
copyright
owner.
http://manoman.sghill.com
Type
3177
TYPE
3777
SAMPLING
SWEEP
nes
VARIABLE
SITION
TIME/DIV.
oO
ADJ
5
uSEC
2
1
HORIZ.
DOTS
UNCAI
MAG
PER
DIV,
@
x10
é
TRIGGER
ee
=
sensiriviry
|
MANUAL
SCAN
OR
EXT
ATTEN,
SWEEP
MODE
NORMAL
pds
han
MANUAL
RECOVERY
TIME
SINGLE
DISPLAY
RESET
SWEEP
OUTPUT
™’
iv,
OV
(10
KA)
The
Type
3T77
Sampling
Sweep
Unit.
SECTION
1
CHARACTERISTICS
General
Information
The
Type
3177
Sampling
Sweep
Plug-In
Unit
is
designed
for
use
with
all
Tektronix
Type
560-Series
Oscilloscopes
except
the
Type
560
and
Type
565.
The
Type
3177
is
equipped
to
drive
a
digital
unit
in
the
Type
567
Readout
Oscilloscope.
The
Type
3177
must
be
inserted
into the
right-hand
plug-
in
compartment
of
the
oscilloscope
and
a
vertical
sampling
plug-in
unit
(such
as
the
Type
3876)
must
be
inserted
into
the
left-hand
plug-in
compartment
to
complete
the
composite
sampling
system.
Equivalent
Sweep
Rates
Variable
in
fifteen
steps
from
0.2
nsec/div
to
10
usec/div
(0.02
nsec/div
to
1
«sec/div
with
HORIZ.
MAG.
switch
at
X10).
Accuracy
typically
within
1%,
and
in
all
cases
within
3%
of
the
indicated
sweep
rate
with
HORIZ.
MAG.
switch
at
X1.
With
HORIZ.
MAG.
switch
at
X10,
accuracy
is
within
5%.
The
sweep
rate
at
any
given
setting
of
the
TIME/DIV.
switch
can
be
increased
by
about
3
times
with
the
VARIABLE
control.
Samples
Per
Division
10/div
or
100/div.
External
Triggering
Pulse
Repetition
Rate:
Up
to
3
&
10°
pulses/second
(300
megacycles}.
Trigger
circuitry
counts
down
to
a
maximum
sampling
rate
of
about
100
kilocycles/second.
Minimum
Pulse
Amplitude
and
Width:
10
millivolts,
peak-
to-peak,
with
at
least
a
2-nanosecond
width.
With
larger
pulse
amplitudes
(up
to
800
millivolts)
minimum
pulse
width
decreases.
Overload
damage
occurs
at
5
volts
and
above.
Sinusoidal
Frequency
Range:
100
kilocycles
to
300
mega-
cycles;
10
to
800
millivolts
amplitude.
Low
Frequency
Response
(to
trigger
pulses}:
300
kilo-
cycles
(3-db
down
point).
Jitter:
50
picoseconds
or
0.001%
fast
ramp
duration,
whichever
is
greater
(for
50-millivolt
amplitude,
2-nano-
second
width
pulses
with
a
repetition
rate
less
than
10
megacycles},
Jitter
increases
as
pulse
amplitude
and/or
width
decreases
when
repetition
rates
exceed
10
mega-
cycles.
Internal
Triggering
Same
characteristics
as
External
Triggering
but
modified
by
vertical
plug-in
unit
used.
When
used
with
the
Type
3876,
all
characteristics
are
the
same
except
five
times
more
amplitude
is
required
at
the
Type
3876
INPUT
A
or
INPUT
B
connectors
and
the
low-frequency
response
is
3-db
down
at
450
kilocycles.
External
Sweep
Input
Sensitivity:
Adjustable
from
5
to
25
volts
per
horizontal
division
(50
volts
required
for
a
full
10-division
display,
250
volts
maximum).
Input
Resistance:
28
to
100
kilohms,
depending
on
setting
of
EXT.
ATTEN.
control.
Sweep
Output
Amplitude:
1
volt/div
from
a
source
impedance
of
10
kilohms.
Delay
Variable
through
100
nsec.
Mechanical
Construction:
Aluminum
alloy
chassis.
Finish:
Photo-etched,
anodized
front
panel.
Weight:
5
pounds.
Accessories
Supplied
With
the
Type
3177
Tektronix
Part
Number
2
500,
10-nsec
cables
(017-501)
RG-58/AU
with
General
Radio
connectors.
1
Adapter,
BNC-to-GR
(017-025)
1
Adapter,
male
BNC-to-female
(103-032)
UHF
2
Attenuators,
1OXT,
509
with
(017-044)
GR
connectors
2
Instruction
manuals
(070-333}
1-1
Operating
Instructions
—
Type
3177
f
TYPE
31777
SAMPLING
SWEEP
VARIABLE
Beet
eye
TIME/DIV.
GAIN
ADJ.
5
nSEC
2
1
HORIZ.
DOTS
UNCAL
MAG.
PER
DIV.
.
2
@
X10
|
10
7—-
TRIGGERS
SENSITIVITY
x1
MANUAL
SCAN
OR
EXT.
ATTEN.
SWEEP
MODE
MANUAL NORMAL
=
100
nSEc
RECOVERY
TIME
SINGLE
DISPLAY
]
RESET
/
=
y
_
+
EXT.
@
‘)
SWEEP
SWEEP
~
j
INPUT
OUTPUT
iv/DIV.
(10
Kn)
:
.
j
J
SERIAL
7
'
TEKTRONIX.
INC.
PORTLAND,
OREGON,
U.S.A.
Fig.
2-1.
Front
panel
of
the
Type
3177.
2-0
SECTION
2
OPERATING
INSTRUCTIONS
Introduction
The
Type
3177
Sampling
Sweep
Plug-In
Unit
(with
a
ver-
tical
sampling
plug-in
unit)
equips
any
Type
561A,
RM561A,
564, 567,
or
RM567
Oscilloscope
for
sampling-type
opera-
tion.
Front-panel
operation
of
the
Type
3177
resembles
that
of
non-sampling
(real
time}
time-base
units.
This
section
of
the
manual
covers
the
operation
of
the
front-pane!l
controls
and
connectors
(see
Fig.
2-1).
Function
of
Front-Panel
Controls
and
Connectors
POSITION
Conirol
Positions
the
display
horizontally.
HORIZ.
MAG.
Switch
Selects
X1
or
X10
horizontal
display
magnification.
DOTS
PER
DIV.
Switch
Selects
either
10
or
100
samples
per
division.
SWEEP
MODE
Switch
NORMAL
position:
Permits
automatic
dot-by-dot
advance-
ment
through
the
oscilloscope
display.
SINGLE
DISPLAY
position:
Permits
one
display
after
the
RESET
button
is
pressed.
(Useful
for
photographing
the
display.)
-++EXT.
SWEEP
INPUT
position:
Permits
control
of
the
scanning
function
with
an
external
voltage.
MANUAL
position:
Permits
manual
dot-by-dot
advance-
ment
through
the
display
by
turning
the
MANUAL
SCAN
control.
(For
recorder
applications.)
MANUAL
SCAN
OR
EXT.
ATTEN.
Control
Provides
an
internal
semi-intergated,
variable
voltage
for
scanning
when
the
SWEEP
MODE
switch
is
in
the
MANUAL
position,
and
serves
as
a
variable
attenuator
when
the
SWEEP
MODE
switch
is
in
the
+EXT.
SWEEP
INPUT
position.
+EXT.
SWEEP
INPUT
Connector
For
applying
an
external
scanning
voltage.
Sensitivity
variable
from
5
to
25
volts/div;
input
impedance
variable
from
28
to
100
kilohms.
Maximum
input
voltage
250
volts,
peak.
SWEEP
OUTPUT
Connector
For
monitoring
the
sweep
voltage.
(Output
variable
trom
about
3.5vde
to
about
15
vde;
I-volt
change
equals
1
division
of
horizontal
deflection
with
10-kilohm
source
impedance.)
TRIG.
OUT
Connector
For
externally
monitoring
a
positive
trigger
pulse
after
each
trigger
event.
The
pulse
width
is
at
least
0.4
micro-
second
at
0.15-volt
amplitude.
®
EXT.
INPUT
Connector
For
applying
an
external
trigger.
Input
impedance:
50
ohms
shunted
by
12
microhenries.
INT.-EXT.
Switch
Selects
either
an
internal
trigger
(INT.
position)
from
the
vertical
plug-in
unit,
or
an
external
trigger
(EXT.
position)
from
the
EXT.
INPUT
connector,
and
determines
whether
triggering
takes
place
on
the
positive
(+)
or
negative
(—)
slope
of
the
input
signal.
DELAY
Control
Allows
the
start
of
the
display
to
be
varied
through
100
nanoseconds
with
respect
to
the
trigger
event.
TRIGGER
SENSITIVITY
Control
Varies
the
sensitivity
of
the
triggering
circuit.
Also
causes
the
trigger
circuit
to
free-run
when
turned
fully
clockwise.
RECOVERY
TIME
Control
Varies
the
holdoff
time
of
the
trigger
circuits
to
assure
stable
triggering.
TIME/DIV.
Switch
Sets
the
equivalent
sweep
rate
of
the
display.
VARIABLE
Control
Varies
the
sweep
rate
(uncalibrated)
between
TIME/DIV.
steps.
The
equivalent
sweep
rate
at
any
given
setting
of
the
TIME/DIV.
switch
can
be
increased
about
3
times.
GAIN
ADJ.
(a
front-panel
screwdriver
adjustment)
Adjusts
gain
to
match
oscilloscope
deflection
factor.
Installing
the
Type
3177
into
the
Oscilloscope
CAUTION
Turn
off
oscilloscope
power
while
inserting
or
removing
plug-in
units.
Otherwise,
power
supplies
in
the
oscilloscope
may
fail
to
regulate
momen-
tarily
as
plug-in
units
are
removed
or
replaced.
The
Type
3177
is
designed
to
drive
the
horizontal
de-
flection
plates
of
the
crt;
it
must
be
used
in
the
right-hand
plug-in
compartment.
When
inserting
the
Type
3177
into
the
plug-in
compartment,
first
check
that
the
latch
at
the
bottom
of
the
front
panel
is
in
a
horizontal
position.
Then
make
sure
the
interconnecting
plugs
are
properly
aligned.
The
Type
3177
should
then
slip
easily
into
the
compartment.
Once
the
plug-in
has
been
properly
seated,
turn
the
aluminum
knob
of
the
plug-in
unit
a
few
turns
clockwise
until
it
is
hand-tight.
To
remove
the
plug-in
unit,
turn
the
aluminum
knob
counterclockwise
as
far
as
it
will
go
and
pull
the
plug-in
unit
straight
out.
2-1
Operating
Instructions
—
Type
3177
Displaying
a
Signal
The
following
procedure
covers
first-time
operation
of
the
Type
3177.
It
will
enable
you
to
display
a
signal
on
the
ert.
Use
this
procedure
in
conjunction
with
the
vertical
plug-in
unit
instructions.
1.
Set
the
Type
3777
front-panel
controls
as
follows:
POSITION
Midrange
TIME/DIV.
5
nSEC
VARIABLE
CALIB.
DELAY
Fully
counterclockwise
TRIGGER
SENSITIVITY
Fully
counterclockwise
HORIZ.
MAG.
Xl
DOTS
PER
DIV.
100
SWEEP
MODE
NORMAL
INT.-EXT.
INT.
{Set
to
polarity
of
signal
you
wish
to
observe)
RECOVERY
TIME
2.
Insert
the
Type
3177
into
the
right-hand
plug-in
com-
partment
of
the
oscilloscope
and
the
vertical
sampling
plug-
in
unit
into
the
left-hand
compartment.
Turn
on
the
oscillo-
scope
power
and
allow
a
few
minutes
for
warm-up.
Fully
counterclockwise
3.
Apply
the
signal
you
wish
to
observe
to
the
input
connector
of
the
vertical
sampling
plug-in
unit.
(Note:
Make
sure
the
signal
meets
the
triggering
requirements
described
in
Section
1.}
4.
Slowly
advance
the
TRIGGER
SENSITIVITY
control
for
a
stable
display.
The
RECOVERY
TIME
control
may
also
help
stabilize
the
display.
5.
Set
the
TIME/DIV.
switch
to
the
position
where
the
displayed
signal
covers
the
desired
amount
of
horizontal
graticule
divisions.
6.
With
the
POSITION
control,
move
the
display
hori-
zontally
to
the
desired
point
on
the
graticule.
7.
Turn
the
DELAY
control
and
notice
its
effect
on
the
display.
The
action
of
the
DELAY
control
is
most
significant
at
faster
sweep
rates.
The
DELAY
control
varies
the
position
of
the
displayed
pulse
with
respect
to
the
start
of
the
trace.
The
position
of
the
pulse
with
respect
to
the
start
of
the
trace
can
be
varied
100
nanoseconds
with
the
DELAY
control.
8.
Set
the
SWEEP
MODE
switch
to
MANUAL
and
turn
the
MANUAL
SCAN
control.
Note
the
horizontal
scanning
of
the
electron
beam
on
the
crt.
9.
Set the
SWEEP
MODE
switch
to
SINGLE
DISPLAY.
Press
the
RESET
button.
After
pressing
the
RESET
button,
the
Type
3177
allows
one
complete
scan
of
the
electron
beam
across
the
crt.
This
is
particularly
useful
for
photo-
graphing
displays
at
slow
pulse
repetition
rates.
Gain
Adjust
The
basic
oscilloscope
crt
deflection
factor
varies
slightly
from
one
oscilloscope
to
another.
For
this
reason,
the
GAIN
ADJ.
(a
front-panel
screwdriver
adjustment)
should
be
checked
and
adjusted
as
necessary
each
time
the
Type
3177
is
used
in
a
different
oscilloscope.
Also,
check
the
2-2
GAIN
ADJ.
occasionally
during
regular
use
of
the
instru-
ment.
Checking
or
setting
the
GAIN
ADJ.
requires
the use
of
an
accurate
frequency
source,
such
as
the
Tektronix
Type
180A
Time-Mark
Generator.
The
frequency
standard
used
must
have
a
frequency
output
of
at
least
100
kilocycles,
and
preferably
above
1
megacycle.
To
check
or
adjust
the
Type
3T77
GAIN
ADJ.
control,
proceed
as
follows:
1.
Insert
the
Type
3T77 and
the
associated
vertical
sampling
plug-in
unit into
the
oscilloscope,
turn
on
the
power
and
allow
the
instrument
to
warm
up
for
at
least
2
minutes
before
proceeding.
2.
Set
the
front-panel
controls
of
the
Type
3177
as
follows:
POSITION
Midrange
DOTS
PER
DIV.
100
TRIGGER
SENSITIVITY
Fully
clockwise
VARIABLE
CALIB.
SWEEP
MODE
NORMAL
HORIZ.
MAG.
X]
INT.-EXT.
+
INT.
Other
controls
may
be
set
to
any
position.
3.
From
an
accurate
frequency
source,
apply
a
signal
to
the
Input
connector
of
the
vertical
sampling
plug-in
unit
and
adjust
for
a
vertical
deflection
of
2
to
6
divisions.
4.
Determine
the
time
duration
of
one
cycle
of
the
signal
from
the
frequency
source
{time
duration
of
one
cycle
(in
seconds)
is
the
reciprocal
of
the
frequency
{in
cycles
per
second)
).
5.
Set
the
TIME/DIV.
switch
for
a
sweep
rate
that
will
display
one-half
to
one
cycle
of
the
applied
signal
per
division
of
deflection.
6.
Set
the
TRIGGER
SENSITIVITY
and
RECOVERY
TIME
controls
for
a
stable
display.
7.
Check
for
the
proper
number
of
cycles
per
division.
If
the
number
of
cycles
per
division
does
not
exactly
agree
with
the
setting
of
the
TIME/DIV.
switch,
set
the
GAIN
ADJ.
for
the
proper
timing.
Use
the
POSITION
control
to
align
the
display
with
the
graticule
markings.
Triggering
the
Type
3177
The
Type
3T77
can
be
triggered
either
internally
or
externally.
Internal
triggering
requires
no
signal
connections
to
the
Type
3777
since
the
triggering
signal
is
coupled
internally
from
the
vertical
sampling
plug-in
unit.
However,
external
triggering
is
more
sensitive
(see
Section
1).
External
triggering
is
also
independent
of
the
displayed
waveform.
Thus,
when
signals
of
different
amplitudes
are
applied
to
the
vertical
plug-in
unit,
the
triggering
controls
do
not
require
resetting
for
a
stable
display.
However,
an
external
triggering
signal
must
be
related
in
time
to
the
displayed
signal
to
maintain
stable
triggering.
The
+
and
—
positions
of
the
INT.-EXT.
switch
determine
whether
triggering
takes
place
on
the
positive-
or
negative-
@l
going
slope
of
the
triggering
signal.
The
INT.
or
EXT.
posi-
tions
determine
whether
the
triggering
signal
comes
from
the
vertical
sampling
plug-in
unit
or
from
the
EXT.
INPUT
connector.
Selecting
the
Equivalent
Sweep
Rate
The
Type
31777
TIME/DIV.
switch
selects
equivalent
sweep
rates
from
0.2
nanosecond
per
division
to
10
micoseconds
per
division.
These
rates,
in
turn,
provide
an
equivalent
10-
division
display
width
from
2
nanoseconds
to
100
micro-
seconds.
Setting
the
HORIZ.
MAG.
switch
to
X10 increases
the
equivalent
sweep
rate
10
times
at
any
setting
of
the
TIME/DIV.
switch.
This
expands
the
display
in
both
direc-
tions
from
the
center
graticule
division
to
cover
the
entire
graticule
(horizontally).
The
VARIABLE
control
increases
the
equivalent
sweep
rate
at
any
setting
of
the
TIME/DIV.
switch
about
3
times
(uncalibrated)
when
the
control
is
fully
clockwise.
Selection
of
sweep
rate
depends
on
the
duration
of
the
applied
signal
and
the
specific
portion
of
the
signal
you
wish
to
observe.
The
DELAY
control
aids
in
observing
a
specific
portion
of
the
applied
signal.
The
equivalent
sweep
rates
of
the
Type
3777
are
accurate
within
3%
of
the
TIME/DIV.
control
setting
when
the
VARI-
ABLE
control
is
set
to
CALIB.
and
the
HORIZ.
MAG.
switch
is
set
to
X1.
This
permits
accurate
time
measurements
di-
rectly
from
the
oscilloscope
display.
Selecting
the
Dots
Per
Division
In
a
sampling
system,
the
applied
signal
is
displayed
as
a
series
of
dots
distributed
across
the
ert.
With
a
greater
number
of
dots
in
the
display,
the
trace
appears
more
con-
tinuous.
With
less
dots
in
the
display,
there
is
less
continu-
ity
in
the
trace
and
the
individual
dots
become
more
apparent
(see
Fig.
2-2).
The
DOTS
PER
DIV.
switch
selects
either
10
dots
per
division
or
100
dots
per
division.
Proper
setting
of
the
DOTS
PER
DIV.
switch
is
a
choice
between
good
trace
continuity
and
flicker.
Flicker
will
become
ap-
parent
with
signals
of
low
repetition
rate
when
the
DOTS
PER
DIV.
switch
is
set
to
100.
By
changing
the
setting
of
the
DOTS
PER
DIV.
switch
to
10,
flicker
will
become
less
apparent.
However,
with
10
dots
per
division,
make
sure
the
SMOOTH-NORMAL
switch
of
the
vertical
sampling
plug-in
unit
is
set
to
NORMAL.
Otherwise,
the
display
may
be
distorted.
Signals
with
fast
repetition
rates
cause
less
flicker;
thus
the
DOTS
PER
DIV.
switch
may
be
set
to
100
for
best
trace
continuity.
When
making
time
measurements
from
the
crt,
the
indi-
vidual
dots
can
serve
as
time
markers.
For
example,
suppose
the
Type
3177
TIME/DIV.
switch
is
set
to
1
nSEC
and
the
DOTS
PER
DIV.
switch
is
set
to
10.
In
this
case,
each
dot
@i
Operating
Instructions
——
Type
3177
4
+
+
*e,
*,
+
.
>
.
.
*
+
.
,
ji
aa
.
eee eee
SSS
ES
SESS
SS
SSSR
SEER
SER
SESE
Eee
eee
TrvTyrrer
TT
Tere
Rrergrrre
TUT
TTT
rTrTrypserre
. .
7
hal
.
+
. :
>
°:
‘
es
y
os
ere
een'
(a)
=
TIT
TTT
Tyr
tt
SEE
Ree
SaaS
ea
eeee
TTT
+
tipi
THT
+
a:
a
+
N
arn
nenie
A
\
4)
(b)
Fig,
2-2.
(a)
A
typical
display
with
the
DOTS
PER
DIV.
switch
set
to
10.
(b)
The
same
signal
with
the
DOTS
PER
DIV.
switch
set
to
100.
represents
0.1
nanosecond
in
equivalent
time
(1
nanosecond
per
division
divided
by
the
10
dots
per
division).
This
application
is
easiest
when
the
display
contains
10
dots
per
division
(i.e.,
the
DOTS
PER
DIV.
switch
must
be
set
to
10
or
the
HORIZ.
MAG.
switch
must
be
set
to
X10).
Under
certain
conditions
it
is
possible
to
get
a
false
display
when
using
a
sampling-type
system.
This
happens
when
certain
relationships
exist
between
the
frequency
of
the
applied
signal
and
the
equivalent
time
between
dots.
A
false
display
can
be
detected
by
changing
the
setting
of
the
DOTS
PER
DIV.
switch
or
the
TIME/DIV.
switch.
With
a
valid
display,
a
change
in
the
equivalent
time
between
dots
should
not
affect
the
displayed
waveform.
However,
if
the
display
is
false,
it
will
change
in
apparent
repetition
rate
or
disappear
when
the
equivalent
time
per
dot
changes.
2-3
SECTION
3
APPLICATIONS
The
procedures
in
this
section
describe
some
basic
applications
for
the
Type
3177.
Time
and
Frequency
Measurements
The
Type
3177
is
accurately
calibrated
to
indicate
equiva-
lent
time
per
division
along
the
horizontal
axis
of
the
oscilloscope
display.
Thus,
the
following
procedures
concern
measuring
time
and
frequency
of
electrical
events.
Time
Duration.
To
measure
the
time
duration
of
an
electrical
event,
proceed
as
follows:
1.
Obtain
a
stable
display
of
the
event
you
wish
to
measure.
2.
Set
the
TIME/DIV.
switch
so
the
distance
between
the
two
points
you
wish
to
measure
covers
a
large
portion
of
the
graticule.
3.
To
get
the
time
duration,
count
the
number
of
grati-
cule
divisions
between
the
two
points,
multiply
by
the
TIME/
DIV.
switch
setting,
then
divide
the
result
by
the
HORIZ.
MAG.
switch
setting.
To
illustrate
this
procedure,
assume
the
TIME/DIV.
switch
is
set
to
5nSEC
and
the
HORIZ.
MAG.
switch
to
X1.
Then,
assume
the
distance
between
the
two
points
on
the
display
is
4.8
major
graticule
divisions.
Thus,
the
time
duration
would
be
4.8
divisions
multiplied
by
5
nanoseconds/divi-
sion
(TIME/DIV.
switch
setting),
or
24
nanoseconds,
Risetime.
The
risetime
of
a
pulse
is,
by
definition,
the
time
required
for
the
pulse
to
rise
from
10%
to
90%
of
2.2
div.
or
.44
nsec
—
!
"TT
90%
f
point
TTTETT
TTT
TAT
TY
Ty
10%
point
+t
ttt
HH
i
Pitisi
ris
tig
4
4+
tt
4+
+
+
.2
nsec/div.
Fig.
3-1.
Determining
pulse
risetime.
®
its
maximum
amplitude.
To
measure
pulse
risetime
or
fall-
time
proceed
as
follows:
1.
Set
the
TIME/DIV.
switch
and
the
DELAY
control
to
display
the
leading
edge
of
the
pulse.
(Or,
for
falltime
measurement,
display
the
falling
edge.)
For
best
accuracy,
the
rising
portion
of
the
pulse
should
span
at
last
two
horizontal
graticule
divisions.
2.
Set
the
vertical
plug-in
unit
so
the
display
covers
two
to
eight
vertical
graticule
divisions.
3.
Measure
the
horizontal
distance
between
the
10%
and
90%
amplitude
points
on
the
waveform.
See
Fig.
3-1.
4.
The
risetime
is
the
number
of
divisions
in
step
3
multi-
plied
by
the
setting
of
the
TIME/DIV.
switch
(divided
by
the
HORIZ.
MAG.
switch
setting).
Frequency
or
Repetition
Rate.
Frequency
or
repetition
rate
is
the
number
of
complete
electrical
events
occurring
in
a
second.
To
measure
the
frequency
or
repetition
rate
of
the
displayed
signal
in
cycles
(or
pulses}
per
second,
pro-
ceed
as
follows:
1.
Find
the
time
duration
of
a
complete
event
as
described
under
‘Time
Duration’.
2.
The
frequency
or
repetition
rate
of
the
applied
signal
is
the
reciprocal
of
step
1.
To
illustrate
this
procedure,
assume
the
TIME/DIV.
switch
is
set
to
20nSEC
and
the
HORIZ.
MAG.
switch
to
X10.
Then,
assume
one
event
covers
6.4
major
graticule
divisions.
Thus,
the
frequency
or
repetition
rate
would
be
6.4
divisions
multiplied
by
20
nanoseconds/division
(TIME/DIV.
switch
setting)
divided
by
10
{the
HORIZ.
MAG.
switch
setting),
or
12.8
nanoseconds.
The
frequency
is
then
the
reciprocal
of
12.8
nanoseconds,
or
about
78.1
megacycles.
An
alternative
method
for
measuring
frequency
or
repeti-
tion
rate
is
as
follows:
1.
Set
the
TIME/DIV.
switch
for
a
2-
to
10-cycle
display
of
the
input
signal.
2.
Count
the
exact
number
of
cycles
(including
fractional
cycles)
occurring
in
10
major
graticule
divisions.
3.
Multiply
the
number
of
cycles
obtained
in
step
2
by
the
Frequency
Multiplier
value
opposite
the
appropriate
TIME/DIV.
switch
setting
in
Table
3-1.
TABLE
3-1
TIME/DIV.
Setting
|
Frequency
Multiplier
1
nSEC
100mc™
10
nSEC
10
mc
1
SEC
1
me
1
»SEC
100
ke
10
wSEC
10
ke