Section1—2465B/2467BService
SPECIFICATION
INTRODUCTION
TheTEKTRONIX2465Band2467BOscilloscopesare
portable400-MHzbandwidthinstrumentshavingfour-
channelverticaldeflectionsystems.Channel1 andChan-
nel2 providecalibrateddeflectionfactorsfrom2 mVper
divisionto5 V perdivision.Foreachofthesechannels,
inputimpedanceisselectablebetweentwovalues:either
1MQinparallelwith15pF,or50Q internaltermination.
Input-signalcouplingwith1 Mfiimpedancecanbeselected
aseitherACorDC.Channel3 andChannel4 have
deflectionfactorsofeither0.1V or0.5V perdivision.
Eachofthesechannelshasaninputimpedanceof
1
Mfiin
parallelwith15pF,withDCinput-signalcoupling.
Thetriggersystemworksautomaticallyformost
sig-
nals.
Theyoperate invariousmodes,fromanychannel,
withcouplingsfora widerangeofsignals.Thetrigger
systemgivesstabledisplaysfromdcto500MHz.
Thehorizontaldeflectionsystemprovidescalibrated
sweepspeedsfrom1.5s perdivisionto500psper
divi-
sion,
includingtheeffectsoftheX10magnifierandthe
calibratedvariablebetweenthe1-2-5steps.Horizontal
displaysincludeA-Sweep,B-Sweep(delayed),A alternated
withB,andCH1 (forX/Ydisplays).
TheAUTO,SAVE,andRECALLfeaturessavetimeand
preventerrors.PressingtheAUTOSetupbuttongivesa
workablesetupforalmostanysignal.Forrepetitivemea-
surements,theSaveandRecallfunctionsrecordand
immediatelyorsequentiallyrestoreasmanyas30instru-
mentsetups.TheSETUPbuttonsoperate allinstrument
functions,includingtheextendedfunctionoptions.
Direct,on-screenreadoutsoftimemeasurements,volt-
agemeasurements,scalefactors,triggerlevels,and
auxiliaryinformationalsosavetimeandimproveoperator
confidence.
The2467Byields4 divisions/nsvisualwritingrate.This
isabout100timesfasterthanconventional,
high-
performanceoscilloscopes.The2467B visiblydisplaysany
signal,
atanyrepetition-rate,atanysweepspeed,in
typi-
calroomlight.Visiblesingle-shotsinclude1 nsstepsat
500ps/division.
Theinstruments areshippedwiththefollowing
stan-
dardaccessories:
2Probepackages(2465B)
4Probepackages(2467B)
1Snap-lockaccessoriespouch
1Zip-lockaccessoriespouch
1Operatorsmanual
1Powercord(installed)
12-A,250-Vfuse
1ClearplasticCRTfilter
1BlueplasticCRTfilter(installed)
1Front-panelcover
1Operatorspocketreferencecard
Forpartnumbersandfurtherinformationaboutboth
standardandoptionalaccessories,referto"Optionsand
Accessories"(Section7)oftheinstruments Operators
manualortheAccessoriesinformationattherearofthis
manual.
YourTektronixrepresentativeorlocalTektronix
FieldOfficecanalsoprovideaccessoriesinformationand
orderingassistance.
PERFORMANCECONDITIONS
Thefollowingelectricalcharacteristics(Table1-1)are
validfortheinstrumentwhenithasbeenadjustedatan
ambienttemperaturebetween+20°Cand+30°C,has
hada warm-upperiodofatleast20minutes,andis
operatingatanambienttemperaturebetween
—
15°Cand
+55°C(unlessotherwisenoted).
Itemslistedinthe"PerformanceRequirements"column
definethemeasurementcapabilitiesoftheinstruments.
Supplementarymeasurementconditionsmayalsobelisted
inthe"PerformanceRequirement"column.
Mechanicalcharacteristics arelistedinTables1-6and
1-7.
Environmentalcharacteristics aregiveninTable1-8.
Theoscilloscopemeetstheenvironmentalrequirementsof
MIL-T-28800CforTypeIII,Class3,StyleC equipment,
withthehumidityandtemperaturerequirementsdefinedin
paragraphs3.9.2.2,3.9.2.3,and3.9.2.4.
1-1