Section 1—7B10 Operators
GENERAL INFORMATION
INTRODUCTION
OPERATORS MANUAL
The Operators Manual contains information necessary to
effectively operate the 7B10 Time Base and is divided
into three sections: Section 1provides abasic description
of the 7B10 with instrument specifications and
accessories, section 2contains operating information for
the instrument, Instrument Option information is located
in section 3of the manual.
INSTRUCTION MANUAL
The Instruction Manual provides both operating and
servicing information for the 7B10 Time Base. The
Instruction Manual is divided into nine sections.
Operating information is covered in the first two sections;
servicing information for use by qualified service
personnel is contained in the remaining seven sections
of the manual. Schematic diagrams are located at the
rear of the manual and can be unfolded for reference
while reading other parts of the manual. The reference
designators and symbols used on the schematics are
defined on the first page of the Diagrams and Circuit
Board Illustrations section. All abbreviations used in this
manual, with the exception of the parts list and
schematic diagrams, comply with the American National
Institute Y1. 1-1 972 publication. The parts lists are
computer printouts and use computer-supplied
abbreviations. Instrument Option information is located
in section 6of the Instruction Manual.
INSTALLATION
The time-base unit is designed to operate in the
horizontal plug-in compartment of the mainframe. This
instrument can also be installed in avertical plug-in
compartment to provide avertical sweep on the crt.
However, when used in this manner, there are no
internal triggering or retrace blanking provisions, and the
unit may not meet specifications.
To install the unit in aplug-in compartment, push it in
until it fits firmly into the compartment. The front panel
of the unit should be flush with the front panel of the
mainframe. Even though the gain of the mainframe is
standardized, the sweep calibration of the unit should be
checked when installed. The procedure for checking the
unit is given under Sweep Functions in the Operators
Checkout procedure in section 2.
To remove the unit, pull the release latch (see Fig. 1-1 )to
disengage the unit from the mainframe, and pull it out of
the plug-in compartment.
INSTRUMENT PACKAGING
If this instrument is to be shipped for long distances by
commercial means of transportation, it is recommended
that it be repackaged in the original manner for
maximum protection. The original shipping carton can be
saved and used for this purpose. If more information is
needed, contact your local Tektronix Field Office or
representative.
Fig. 1-1. Location of release latch.
1-1