SPECIFICATIONS
The Tektronix Type 502 O cillo cope provide linear dual-beam
di play with a wide range of weep rate combined with high in
put en itivity. In addition, the Type 502 may be u ed to provide
dual-beam X-Y di play at medium en itivitie , and ingle-beam
X-Y di play at high en itivitie . Vertical amplifier for both
beam may be operated with ingle-ended input for conventional
operation, or with differential input for cancellation of common
mode ignal . The wide range of operational mode available,
make the Type 502 adaptable to a great many indu trial and
cientific application .
VERTICAL- DEFLECTION SYSTEM
Characteri tic at each input terminal
Direct connection— 1 megohm paralleled by
47 /a/aT With P510A probe— 10 megohm
paralleled by 14^/Af.
Deflection factor
Sixteen calibrated deflection factor from 200
fxv/cm to 20v/cm accurate within 3%.
Frequency re pon e
Frequency re pon e characteri tic of the Type
502 vary with etting of the SENSITIVITY con-
Typical reading are a follow : X
200 /Av/cm 100 kc W-
5 mv/cm 200 kc
50 mv/cm 400 kc
.2 v/cm 1 me . X
Differential input rejection ratio
The rejection ratio pecified below apply if
the ignal voltage at the Input connector
doe not exceed pecified limit . The ignal
limit for dc-coupling are ± 2 volt with re pect
to ground for etting of the SENSITIVITY con
trol from 200/Av/cm to .2v/cm, =±=20 volt
with re pect to ground for etting from
.5v/cm to 2v/cm, and ±2 0 0 volt with re
pect to ground for etting from 5v/cm to 20
v/cm. For ac coupling, the ignal limit are 2
volt peak-to-peak from 200/xv/cm to .2 v/cm,
20 volt peak-to-peak from .5 v/cm to 2 v/cm,
and 200 volt peak-to-peak from 5 v/cm to 20
v/cm. From 200/xv/cm to 1 mv/cm, the rejec
tion ratio i approximately 1000-to-l, decrea
ing to 100-to-1 at 0.2 v/cm, and to 25-to-1 in
the range from .5 v/cm to 20 v/cm.
HORIZONTAL-DEFLECTION SYSTEM
Sweep rate
Twenty-one calibrated weep rate from 1
/A ec/cm to 5 ec/cm. Accuracy typically with
in 1% of full cale; in all ca e , within 3% of
full cale.
Magnifier
Expand weep 2, 5, 10, or 20 time . Calibra
tion of magnified weep rate accurate within
3% for weep peed which do not exceed
the maximum calibrated rate of 1 [xsec/cm.
Unblanking
DC coupled.
Triggering ignal requirement
Internal— a ignal producing 2 mm vertical de
flection on either the upper or lower
beam.
SPECIFICATIONS — TYPE 502 1-1