
P5100A Performance Verification
P5100A Performance Verification
Use these performance verification procedures to verify that the probe meets the warranted specifications. These
procedures should be performed in the order given using a DPO4104B oscilloscope with the equipment recommended in
the following table. (See Table 1.)
WARNING. The following instructions are for use by qualified service personnel only. To avoid electrical shock, do not
disassemble or maintain the probe while it is connected to a signal source other than those specified in these procedures.
NOTE. The probe must be low-frequency compensated before performance can be verified.
If your probe fails to qualify under these conditions, service-level adjustments (DC gain and high-frequency compensation)
may be required. These adjustments must be done by a trained service person.
Test Equipment
Table 1: Recommended test equipment
Equipment Minimum requirements Recommended test equipment 1
Oscilloscope ≥500 MHz BW, 1 MΩ±1% input R Tektronix DPO4104B
Calibration generator ≤0.5% accuracy, 10 Hz square wave,
5V
p–p output
Fluke 9500B with 9530 active head
Adapter, BNC Female-to-female barrel 103-0028-xx
Adapter, probe tip-to-BNC Probe tip-to-male BNC 013-0291-xx
Pulse generator 2≤500 ps rise time, ≥100 ns pulse width,
≥3V
p–p into 50 Ω
Picosecond Labs 2600C or
Fluke 9500B with 9530 active head
Termination, 50 Ωfeedthrough 250 Ωmale-to-female BNC 011-0049-xx
BNC cable, precision 250 Ω,36in 012-0482-xx
Adjustment tool 2Insulated, straight edge 003-1433-xx
Screwdriver 2Flat blade
Knife 2Razor edge
1Nine digit part numbers (xxx-xxxx-xx) are Tektronix part numbers
2Required for adjustment procedures only
Test Setup
1. Turn on the oscilloscope.
2. Connect the probe to any channel (1–4) of the oscilloscope.
3. Turn on the generators and let the probe and test equipment warm up for 20 minutes.
4. Make a copy of the test record to tabulate the test results. (See Table 2 on page 4.)
P5100A & P5150 High Voltage Probes Technical Reference 1