Operating CALsys -15/110 Autocal
Note:
Metal Insert should be of required hole size so that the hole size is suitable for sensor under calibration.
Always use a master sensor (master RTD/TC) for comparison calibration method.
Place the metal insert first in all sensors, and then set the controller at desired set point.
Check that all sensors (test and master) are immersed at same depth in insert (metal block).
If sensors are not immersed at same depth than there will be an error in temperature reading.
Always use a metal block calibrator (insert). Metal block calibrator heat transfer characteristics should
match with those in the normal measuring situations.
The temperature sensors should be long enough to immerse completely in the calibration bath. Errors
due to stem conduction can be ignored.
Do not change the set temperature value too often. For example, you set the temperate initially at
100°C, and when it reaches at 80°C, you change the temperature to 30°C. Or, you set the temperature
initially at -10°C, and when it reaches close to -5°C, you change the temperature to 50°C. Doing so
may affect unit's performance.
Plan the calibration point by gradually increasing temperature, then set the controller accordingly. This
method will save the time.
Take the reading of sensors (master and test) at stable temperature with the help of digital multi-
meter, or by using Tempsens' TEMPMET 08 or TEMPMET 09 for reading RTD or TC sensor.
3.4 Cooling Down the Source
Before transporting the metal block, ensure that the temperature of bath has cooled sufficiently.
If you require metal block to cool quickly, set the temperature to the room temperature (ambient
temperature).
Always cool the heated instrument to ambient temperature before disconnecting it from the mains,
switching it OFF, or removing the temperature sensor or test item.
Always keep the heated instrument supervised and under observation until it cools down sufficiently.
Page | 19