4
Measurement Frequency
• The meter can provide 5 frequency testing points, namely, 100Hz/120Hz/1kHz/
10kHz/100kHz. On powering on the default frequency is 1K.
• Press FREQ button to select different frequency points for measurement.
Note: DC impedance is measured under AUTO OCR mode and measurement
frequency is automatically set.
Calibration function
• This can be used to reduce interference of distribution parameters introduced by
the test leads. Calibration includes both short and open circuit.
• Short circuit calibration reduces the inuence of the test leads and contact
resistance when measuring low impedance elements.
• Open circuit calibration reduces the inuence of the capacitance and distributed
resistance when testing high impedance elements.
• Press and hold CAL button to enter open circuit calibration and the display shows
OPEN. Press CAL again and the screen shows a countdown from 30 to 0 after
which it displays PASS.
Sorting Measurement
• Sorting mode is used to quickly sort elements of which the main parameter is within
a certain scope.
• The sorting range can be set to ±0.25%, ±0.5%, ±1%, ±2%, ±5%, ±10%, ±20%, and
+80%~-20%. Default is set to ±1%.
• Press SETUP button, RANGE displays on the LCD.
• Press ENTER to conrm then use ▼ or ▲ to decrease or increase the sort value or
press ► or ◄ for to adjust the value of the main parameter.
• Press ENTER to conrm the sieving mode value settings.
• Press FUNC button to select required mode AUTO L, AUTO C, AUTO R or AUTO
DCR.
• Connect the test leads have been connected to the measured element and press
FREQ to enter sieving mode.
• The main display will show PASS and the auxiliary display shows the main value of
measured elements if within the typed nominal value and the buzzer sounds.
• The main display will show FAIL and the auxiliary display shows the main value of
measured elements if outside the typed nominal value.
Deviation ratio measurement
• Deviation measurement is used to compare with deviation the ratio of 2 elements.
• The main LCD display can be typed in automatically as the nominal value.
• Percentage display scope: -99.9%~99.9%.
• Display percentage: REL%=(DCUR DREF) / DREF x 100%.
• DCUR: Main parameter of measured elements.
• DREF: Typed nominal value.
• Auxiliary display is OL% and the main display is the main parameter of measured
elements if DCUR >DREF or 2DCUR < DREF .
• To select deviation measurement press FUNC button to select required mode
AUTO L, AUTO C, AUTO R or AUTO DCR.
• Connect the test leads have been connected to the measured element and press
REL to enter deviation proportion measurement mode. displays on the LCD.
• The main display will show the main parameter of the measured element and the
auxiliary display shows the deviation as a percentage.
• Press and hold REL button to exit deviation measurement.