• 2 / 16 •
• Do not touch the bottom or sides of the metal tank, or a burn may result.
• Do not heat paran on a stove, hotplate, or in a microwave.
• Do use this appliance on a level, secure surface.
• Do not use any additives in the paran except as allowed by the manufacturer.
• Do not modify this equipment without authorization from the manufacturer.
• Care must be taken when operating this equipment around other equipment. Potential
electromagnetic or other interference could occur to this or the other equipment. Minimize this
interference by not using other equipment in conjunction with it.
Cautions
• Do not remelt used paran, reintroduced particles or chemicals may damage the tank.
• Never use sharp objects to remove hardened paran from the unit.
• Do not leave an empty unit plugged in.
• Do not pour melted paran from the tank.
NOTE: The emissions characteristics of this equipment make it suitable for use in industrial areas
and hospitals (CISPR 11 class A). If it is used in a residential environment (for which CISPR 11 class
B is normally required) this equipment might not oer adequate protection to radio-frequency
communication services. The user might need to take mitigation measures, such as relocating or re-
orienting the equipment.
Technical Information
• Power, Model TB6: 100–125V, 0.87–1.3A, 50–60Hz
• Power, Model TB7: 220–240V, 0.55–0.65A, 50–60Hz
• Heater: TB6: 87–162 watts, TB7: 121-156 watts
• Operating temperature: 126º–134º F. (52.2º–56.7º C.)
• 14.3” L x 8.2” W x 7.5” H (36.32 cm L x 20.83 cm W x 19.1 cm H)
• Capacity: 9 lbs. (4.08 kg.)
• Melt time: 6-8 hours
• Duty cycle: continuous operation
• Degree of protection against moisture ingress: IPX0
• 220V unit must be used with a hospital grade cord of at least the maximum ratings: 10 A, 220 VAC.
• Meets IEC/EN 60601-1-2:2014 electromagnetic compatibility interference safety standard.
Environmental Ratings
Operating Transport and Storage
Temperature: 65° to 85° F (18.3° to 29.4° C) -40° to 130° F (-40° to 54.4° C)
Relative Humidity: 10% to 95% 10% to 95%
Atmospheric Pressure: 70 to 160 kPa 70 to 160 kPa
NOTE: “Hospital Grade” grounding
reliability can only be achieved
when the equipment is connected
to an equivalent receptacle marked
“Hospital Only” or “Hospital Grade”
NOTE: Current and wattage must be
measured at operating temperature.
No. Test/Standard Port (enclosure,
AC, DC, I/O)
Emissions Class and Group /
Immunity Test Level
1CISPR 11, Radiated Emissions Enclosure Class A; Group 1
2IEC 61000-4-2 Enclosure ±8kV contact, ±2kV, ±4kV, ±8kV, ±15kV air
3IEC 61000-4-3 Radiated RF
Immunity
Enclosure 3V/m, 80-2700MHz, 80% 1kHz AM
4IEC 61000-4-3 Proximity elds
from RF wireless equipment
Enclosure Section 8.10 of the IEC 60601-1-2 standard
5IEC 61000-4-4 AC 2kV, 100kHz repetition rate
6IEC 61000-4-5 AC Line-to-Line: ±0.5 kV, ±1 kV
Line-to-Ground: ±0.5 kV, ±1 kV, ±2 kV
7IEC 61000-4-6 AC 3V, 0.15-80MHz, 80% 1kHz AM, 6V in ISM Band
within 0.15-80MHz, 80% 1kHz AM
8IEC 61000-4-8 Enclosure 30A/m, 50Hz and 60Hz¬
9IEC 61000-4-11 Voltage Dips AC 0% (100% reduction), 0.5 cycle; 0% (100%
reduction), 1 cycle; 70% (30% reduction) UT, 0.5
sec
10 IEC 61000-4-11 Voltage
Interruptions
AC 0% (100% reduction), 5 sec