manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. Thermo Scientific
  6. •
  7. Measuring Instrument
  8. •
  9. Thermo Scientific 1111101 User manual

Thermo Scientific 1111101 User manual

North America
166 Cummings Center
Beverly, MA 01915 USA
Toll Free: 1-800-225-1480
Tel: 1-978-232-6000
Dom. Fax: 1-978-232-6015
Int’l Fax: 978-232-6031
Europe
Denmark House, Angel Drove
Ely, Cambridgeshire
England, CB7 4ET
Tel: 44-1353-666111
Fax: 44-1353-666001
Asia Pacific
Blk 55, Ayer Rajah Crescent
#04-16/24, Singapore 139949
Tel: 65-6778-6876
Fax: 65-6773-0836
www.thermo.com/water
Environmental Instruments
Water Analysis Instruments
A
B
C
D
E
F
pH
©2008 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific Inc.
and its subsidiaries. ROSS and the COIL trade dress are trademarks of Thermo Fisher Scientific Inc.
Preparation
1. Power adapter – Select the appropriate wall outlet plug and slide the plug
plate into the groove on the back of the adapter. See A.
2. Prepare the electrode according to the directions in the electrode user guide.
In general, these steps include:
• Gently remove the protective cap, sleeve or storage bottle from the
electrode and save it for electrode storage.
• If the electrode is refillable, remove the fill hole cover and fill the electrode
with the appropriate electrode filling solution. See B.
• Soak the electrode in pH electrode storage solution, Cat. No. 910001, or
ROSS®pH electrode storage solution, Cat. No. 810001. See C.
3. Meter connections – Connect the power adapter to the meter and then to the
wall outlet. See D. Connect the pH electrode to the BNC input on the meter.
See E. Connect the ATC probe to the 8 pin miniDIN input on the meter. See F.
Note: All unused inputs on the meter should be covered with the black caps.
Meter Overview
1. To power on the meter, press the power key.
2. In the measurement mode, press the up arrow or down arrow key to change
the measurement units on the top display line to pH, mV, RmV or a blank line.
3. In the setup mode, press the line select key to choose the top, middle or
bottom display line. The icon will point to the selected line.
4. To escape out of any meter function, press and hold the measure key until the
meter returns to the measurement mode.
5. The 2-Star pH meter can be calibrated using one to three pH buffers.
Key Description
power up arrow line select
calibrate down arrow setup
measure digit view log
Meter Connections
For the complete Orion Star™and Star Plus meter user guide, visit www.thermo.com/water.
Thermo Scientific Orion
2-Star pH Benchtop Meter
Quick Start Guide
English
Pin Tip –
Reference
BNC – pH
or ORP
8 Pin MiniDIN
– ATC
RS232
Ground
Power
253168-001 Rev. B 08-08
pH
47
DI
DI
1
3a
4
6
8
10
2
3b
5
7
9
11
pH Automatic Calibration with
Two Buffers
1. Press the power key to turn on the meter. See 1.
2. Select two buffers that bracket the expected sample
pH and are one to four pH units apart. See 2. See
the Automatic Buffer Recognition Selection
section for the buffer sets that are automatically
recognized by the meter.
3. In the measurement mode, press the up arrow key
until the pH icon is shown on the top line. See 3a.
Press the calibrate key to begin the calibration.
See 3b.
4. Rinse the electrode (and ATC probe, if in use) with
distilled water and blot dry with a lint-free tissue.
See 4.
5. Insert the electrode into the first buffer and gently stir.
See 5.
6. Wait for the pH icon to stop flashing and the
icon to start flashing. The meter should display the
temperature-corrected pH buffer value. See 6.
If the displayed buffer value is incorrect or a custom
buffer is being used, press the digit key until the first
digit to be changed is flashing, press the up/down
arrow keys to change the value of the flashing digit
and continue to change the digits until the meter
displays the temperature-corrected pH buffer value.
Once the buffer value is set, press the digit key until
the decimal point is in the correct location.
7. Press the calibrate key to proceed to the next
calibration point. See 7.
8. Rinse the electrode (and ATC probe, if in use) with
distilled water and blot dry with a lint-free tissue.
See 8.
9. Insert the electrode into the second buffer and gently
stir. See 9.
10. Wait for the pH icon to stop flashing and the
icon to start flashing. The meter should display the
temperature-corrected pH buffer value. See 10.
11. Press the measure key to save and end the
calibration. See 11. The slope will be displayed and
the meter will proceed to the measurement mode.
Automatic Buffer Recognition
Selection
1. In the measurement mode, press the setup key.
2. Press the up arrow key until PH is displayed on the
top line.
3. Press the line select key to move the icon to the
middle line. Press the up arrow key until bUF is
displayed on the middle line.
4. Press the line select key to move the icon to the
bottom line. Press the up arrow key to select USA
for the pH 1.68, 4.01, 7.00, 10.01 and 12.46 buffer
set or EUr0 for the pH 1.68, 4.01, 6.86 and 9.18
buffer set. USA is the default setting.
5. Press the line select key to move the icon to the
top line and press the measure key.
pH Measurement Resolution
1. In the measurement mode, press the setup key.
2. Press the up arrow key until PH is displayed on the
top line.
3. Press the line select key to move the icon to the
middle line. Press the up arrow key until rES is
displayed on the middle line.
4. Press the line select key to move the icon to the
bottom line. Press the up arrow key to select 0.01
or 0.1 for the resolution. 0.01 is the default setting.
5. Press the line select key to move the icon to the
top line and press the measure key.
Measurement Mode Selection
1. In the measurement mode, press the setup key.
2. Press the up arrow key until rEAd is displayed on
the top line.
3. Press the line select key to move the icon to the
middle line. Press the up arrow key until tYPE is
displayed on the middle line.
4. Press the line select key to move the icon to the
bottom line. Press the up arrow key to select AUt0
for the AUTO-READ™ mode, tImE for the timed
mode or C0nt for the continuous mode. AUt0 is the
default setting.
5. Press the line select key to move the icon to the
top line and press the measure key.
2-Star pH Benchtop Meter Quick Start Guide
pH Measurements
1. Rinse the electrode (and ATC probe, if in use)
with distilled water, blot dry with a lint-free
tissue and insert into the sample.
2. If the meter is in the AUTO-READ measurement
mode, press the measure key to start a reading.
Once the reading is stable, the AR icon will stop
flashing and the display will freeze. Press the
measure key to take a new measurement.
If the meter is in the continuous or timed
measurement mode, the meter will immediately
start taking readings and continuously update the
display. Once the reading is stable, the pH icon
will stop flashing.
3. Remove the electrode (and ATC probe, if in use)
from the sample, rinse with distilled water, blot
dry with a lint-free tissue, insert into the next
sample and repeat step 2 or see the electrode
user guide for recommended storage.
Printing Data
In the AUTO-READ and continuous measurement
modes, every time the measure key is pressed the
meter logs the data on the display. In the timed
measurement mode, the meter logs the data at the
predetermined time interval.
1. Connect the meter to a printer or computer and
verify the baud rate and output settings in the
meter setup menu.
2. In the measurement mode, press the log
view key.
3. Press the up arrow key to select SEnd to print
the datalog or CALO to print the calibration log.
4. Press the line select key to send the selected
data to the printer or computer.
English

Other Thermo Scientific Measuring Instrument manuals

Thermo Scientific Orion Versa Star Pro User manual

Thermo Scientific

Thermo Scientific Orion Versa Star Pro User manual

Thermo Scientific Orion Star User manual

Thermo Scientific

Thermo Scientific Orion Star User manual

Thermo Scientific Eutech Expert pH User manual

Thermo Scientific

Thermo Scientific Eutech Expert pH User manual

Thermo Scientific HAAKE RotoVisco 1 User manual

Thermo Scientific

Thermo Scientific HAAKE RotoVisco 1 User manual

Thermo Scientific Orion Star A215 User manual

Thermo Scientific

Thermo Scientific Orion Star A215 User manual

Thermo Scientific Orion Star A210 Series User manual

Thermo Scientific

Thermo Scientific Orion Star A210 Series User manual

Thermo Scientific XL2 User manual

Thermo Scientific

Thermo Scientific XL2 User manual

Thermo Scientific Star A221 User manual

Thermo Scientific

Thermo Scientific Star A221 User manual

Thermo Scientific NanoDrop Lite Instruction Manual

Thermo Scientific

Thermo Scientific NanoDrop Lite Instruction Manual

Thermo Scientific Orion Star A213 User manual

Thermo Scientific

Thermo Scientific Orion Star A213 User manual

Thermo Scientific LevelPRO User manual

Thermo Scientific

Thermo Scientific LevelPRO User manual

Thermo Scientific 20-35-NM-F User manual

Thermo Scientific

Thermo Scientific 20-35-NM-F User manual

Thermo Scientific Star A223 User manual

Thermo Scientific

Thermo Scientific Star A223 User manual

Thermo Scientific 42i User manual

Thermo Scientific

Thermo Scientific 42i User manual

Thermo Scientific Orion 2120XP User manual

Thermo Scientific

Thermo Scientific Orion 2120XP User manual

Thermo Scientific TruNarc User manual

Thermo Scientific

Thermo Scientific TruNarc User manual

Thermo Scientific Orion Star User manual

Thermo Scientific

Thermo Scientific Orion Star User manual

Thermo Scientific Orion Star User manual

Thermo Scientific

Thermo Scientific Orion Star User manual

Thermo Scientific AutoPILOT PRO User manual

Thermo Scientific

Thermo Scientific AutoPILOT PRO User manual

Thermo Scientific Orion Star A200 Series Technical manual

Thermo Scientific

Thermo Scientific Orion Star A200 Series Technical manual

Thermo Scientific Nicolet Summit User manual

Thermo Scientific

Thermo Scientific Nicolet Summit User manual

Thermo Scientific AquaPro User manual

Thermo Scientific

Thermo Scientific AquaPro User manual

Thermo Scientific AQUAfast AQ3140 COD User manual

Thermo Scientific

Thermo Scientific AQUAfast AQ3140 COD User manual

Thermo Scientific Prima PRO User manual

Thermo Scientific

Thermo Scientific Prima PRO User manual

Popular Measuring Instrument manuals by other brands

PCE Health and Fitness PCE-PHM 12 user manual

PCE Health and Fitness

PCE Health and Fitness PCE-PHM 12 user manual

Keller ITS CellaPort PT 11 Series operating manual

Keller

Keller ITS CellaPort PT 11 Series operating manual

CPN 503 ELITE Hydroprobe operating manual

CPN

CPN 503 ELITE Hydroprobe operating manual

Agilent Technologies 1670G Series Service guide

Agilent Technologies

Agilent Technologies 1670G Series Service guide

Peak Atlas SCR100 user guide

Peak

Peak Atlas SCR100 user guide

Pulsar FLOWPULSE instruction manual

Pulsar

Pulsar FLOWPULSE instruction manual

RND 365-00004 user manual

RND

RND 365-00004 user manual

Emerson Rosemount 2230 quick start guide

Emerson

Emerson Rosemount 2230 quick start guide

Dostmann Electronic P700-EX Operation manual

Dostmann Electronic

Dostmann Electronic P700-EX Operation manual

Pfeiffer Vacuum HPT 200 operating instructions

Pfeiffer Vacuum

Pfeiffer Vacuum HPT 200 operating instructions

Draminski GRAIN MOISTURE METER instruction manual

Draminski

Draminski GRAIN MOISTURE METER instruction manual

ATIM Cloud Wireless MR4 user guide

ATIM

ATIM Cloud Wireless MR4 user guide

Agilent Technologies 490 Micro GC user manual

Agilent Technologies

Agilent Technologies 490 Micro GC user manual

Ditel KOSMOS JUNIOR-C user manual

Ditel

Ditel KOSMOS JUNIOR-C user manual

Myzone Inbody 230 Setup instructions

Myzone

Myzone Inbody 230 Setup instructions

Armada Pro65 user guide

Armada

Armada Pro65 user guide

Fluke 433 Getting started

Fluke

Fluke 433 Getting started

PeakTech 2275 Operation manual

PeakTech

PeakTech 2275 Operation manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.