Environmental Chamber ________________________________________________________________Installation and Set-Up
2 - 3
2.5 Start-Up
When the humidification system is operational, the incuba-
tor may be started. Preset the controls as follows:
Overtemp Safety Thermostat . . . . . . . . . . . . . Fully Clockwise
Undertemp Safety Thermostat . . . . . . Fully Counterclockwise
Main Power Switch . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .ON
Humidity Controller . . . . . . . . . . . . . . . . . . . .Desired Setpoint
Temperature Controller . . . . . . . . . . . . . . . . . .Desired Setpoint
Door Heater . . . . . . . . . . . . . . . . . . . . . . . . . .40% (factory set)
For best overall performance of the incubator, the refrigera-
tion switch should be turned on for most applications. When
running Low or No humidity at high temperatures, the refriger-
ation switch may be turned off.
The defrost switch must be turned on when the
temperature setpoint is 10°C, or below.
2.6 Setting the Overtemp Safety Thermostat
Allow the chamber temperature and humidity to stabilize
then set the overtemp safety thermostat as follows:
1. Turn the overtemp control knob slowly counterclock-
wise until the audible alarm sounds and the overtemp
indicator lights.
2. Turn the overtemp control knob clockwise at least 2
degrees. The alarm should be silenced and the overtemp
indicator light should go out. The overtemp safety ther-
mostat is now set a few degrees above the control tem-
perature setpoint. When the chamber temperature rises
to the overtemp control point, the alarm system will acti-
vate, power to the heaters will shut off, and the chamber
temperature will be maintained at the overtemp control
point.
When an overtemp condition occurs, the cause must be
determined and corrected before normal operation under the
main temperature controller can be resumed.
Note: When the chamber temperature control setpoint is
changed, the overtemp safety thermostat must be reset to
accommodate the change.
Note: The overtemp control is not directly calibrated. The num-
bers on the dial are for reference only.
2.7 Setting the Undertemp Safety Thermostat
Allow the chamber temperature and humidity to stabilize,
then set the undertemp safety thermostat as follows:
1. Turn the undertemp control knob slowly clockwise until
the audible alarm sounds and the undertemp indicator
lights.
2. Turn the undertemp control knob counterclockwise at
least 2° on the scale. The alarm will silence and the
undertemp indicator light go out.
The undertemp safety thermostat is now set a few degrees
below the control temperature setpoint. When the chamber tem-
perature drops to the undertemp control point, the alarm system
activates, power to the compressor shuts off, and the chamber
temperature is maintained at the undertemp control point.
When an undertemp condition occurs, the cause must be
determined and corrected before normal operation under the
main temperature controller can be resumed.
Note: When the chamber temperature control setpoint is
changed, the undertemp safety thermostat must be reset to
accommodate the change.
Note: The undertemp control is not directly calibrated. The
numbers on the dial are for reference only.
2.8 Preparing the (Optional) CoBex Recorder
a. Installing the Battery
The seven-day circular chart recorder is located on the
front of the incubator cabinet and is protected by a glass door.
To prepare the recorder for operation, open the glass door
and snap the connector onto the 9-volt battery (Figure 2-4). If
the unit is operating, the green LED lights steady. If the unit is
not turned on, the LED blinks.
If the battery is weak or not connected, the green LED will
flash. If power is lost to the cabinet, the LED will also flash.
When replacing the 9V battery, use only an alkaline style bat-
tery. Dispose of the old battery following established environ-
mental practices.
Figure 2-4