TLD-Irradiator is built in a portable condition housing.
The front side of the housing contains the following modules from on the left of to the right:
-the electronic tax and control unit
-the thrust subject
-the calibration factor announcement.
On the back is a cold equipment installation plug in combination with the in out rocker
switch, the safeguard subject and the voltage switch as slidegate valve.
On the front side of the tax and control unit are LED displays:
-One also,,power” designated green LED, which indicates whether the
equipment is switched on
.
-One also,,more shutter” designated white 3-Farben-LED-Anzeige to
optical control of the Device status.
-A tracer with the designation , is started ,,start” which procedure the
calibration.
On the right side of the equipment are the thrust subject and the counter.
The thrust subject can be equipped with carrier disks. There are 3 different types:
-Combination disk for 50 chips and rods
-Disk for 50 micro cubes
-Disk for 5 TLD maps
On the front of the counter are keys for adjusting the counter value.
Example: During a mains voltage of 230V/50Hz the disk turns in the Irradiator
with a speed of 1 revolution/minute.
If 90 revolutions are necessary for the irradiation, the counter desired value must
be adjusted to 90.
The counter counts the revolutions by a pulse generator and increases its actual value around
1 by each impulse
.
If the counter resembling and the output contact are the same the calibration procedure
switches off. The carrier disk in the thrust subject is located in basic position, the LED
display, shutter shines green and an acoustic signal sounds.
The counter desired value of the input field should indicate at least the value to 1.
The carrier disk can be taken off after opening the thrust subject. After pressing of the key,
,,reset”at the counter the equipment is ready for a further calibration procedure.
4
2. Technical manual