Alarms for Clinicians
For additional information and
guidelines, refer to the
HeartMate
III Instructions for Use.
Thoratec Corporation
6035 Stoneridge Drive, Pleasanton, CA 94588
Tel (925) 847-8600 | Fax (925) 847-8574
HeartLine (800) 456-1477
www.thoratec.com
European Authorized Representative
St. Jude Medical
Coordination Center BVBA. The Corporate Village
Da Vincilaan 11 Box F1. 1935 Zaventem, Belgium
Tel +32 2 774 68 11 | 24-Hr Tel +32(0)2 200 66 45
Thoratec Corporation continually strives to provide the highest quality
products for mechanical circulatory support. Specifications may change
without notice. Therefore, please refer to the HeartMate III Instructions for
Use shipped with the HeartMate III for the most current information regarding
indications, contraindications, warnings, and cautions. Thoratec, the Thoratec
logo, and HeartMate are registered trademarks, and HeartMate III, Mobile
Power Unit, and HeartLine are trademarks of Thoratec Corporation.
©2017 Thoratec Corporation. 10009740.A 10/2017
Alarm Means
IMPORTANT! The Pump Running ( ) light is always green (ON) when the pump is running.
1Check if the fixed speed setting
is below 4,000 rpm AND the
System Controller’s backup battery
is not installed. Under these
conditions, the pump can only
be started from the System Monitor’s
Clinical or Settings screen by
pressing the Pump Start button.
Otherwise, press any button on
the System Controller to attempt
pump start.
1Ensure that the Driveline is
connected to the System Controller.
2Ensure that a power
source is connected to
the System Controller.
3Clinically evaluate the patient.
+
Low flow, flow is
less than 2.5 lpm
Driveline is
disconnected.
The Pump Running
symbol is black.
Both power cables
are disconnected.
System Controller
Hardware Fault
(Microcontroller
Failure)
Pump is off.
The Pump Running
symbol is black.
3If the alarm still persists, check
if the fixed speed setting is below
4,000 rpm AND the System
Controller’s backup battery is not
installed. Under these conditions,
from the System Monitor’s Clinical
or Settings screen by pressing
the Pump Start button.
4If the Driveline is connected and
the alarm persists, replace the
System Controller with a configured
Immediately connect to a working power source
(Power Module, Mobile Power Unit, or two fully-charged
HeartMate 14 Volt Lithium-Ion batteries).
Immediately connect to a working power source
(Power Module, Mobile Power Unit™, or two HeartMate®
14 Volt Lithium-Ion batteries).
Priority
System Controller
ScreenActive
SymbolsTo Resolve Alarm
HAZARD
+
Low Battery,
Power Input is extremely
low with less than 5 min.
remaining.
1Immediately reconnect the
Driveline to the System Controller
and move the Driveline safety
lock on the System Controller
to the locked position. Also check
that the Modular In-Line connector
is secure.
2If the alarm persists after
reconnecting the Driveline,
press any button on the System
Controller to attempt pump start.
1Immediately switch to the backup System Controller.
2 Provide the patient with a new System Controller.
No active symbols (constant audio tone).
2Switch to the backup
System Controller and
attempt to restart the pump.
3Clinically evaluate the patient.
+
+
+
+
+