Priority
System
Controller
Screen
Active
Symbols Alarm Means To Resolve Alarm
+
Pump is o.
The pump
running symbol
( ) is black.
1. Check if the xed speed setting is below 8,000 rpm AND the System
Controller’s backup battery is not installed. Under these conditions, the
pump can only be started from the System Monitor’s Clinical or Settings
screen by pressing the Pump Start button. Otherwise, press any button on
the System Controller to attempt pump start.
2. Switch to backup System Controller and attempt to restart pump.
3. Clinically evaluate patient.
Driveline is
disconnected.
The pump
running symbol
( ) is black.
1. Immediately reconnect the driveline to System Controller and move the
driveline safety tab on the System Controller to the locked position.
2. If alarm persists after reconnecting the driveline, press any button on the
System Controller to attempt pump start. Otherwise, check if the xed
speed setting is below 8,000 rpm AND the System Controller’s backup
battery is not installed. Under these conditions, the pump can only be
started from the System Monitor’s Clinical or Settings screen by pressing
the Pump Start button.
3. If driveline is connected and alarm persists, replace System Controller with
pre-programmed backup System Controller.
+Both power
cables are
disconnected
Immediately connect to a working power source
(Power Module or two fully-charged HeartMate 14 Volt Lithium-Ion batteries).
+Low ow,
ow is less
than 2.5 lpm
1. Ensure that the driveline is connected to System Controller.
2. Ensure that a power source is connected to System Controller.
3. Clinically evaluate patient.
+
Low Battery,
Power input
is extremely
low with less
than 5 min.
remaining
Immediately connect to a working power source
(Power Module or two fully-charged HeartMate 14 Volt Lithium-Ion batteries).
Laser % Live Trim Bleed Color Fold
11 x 8.5 .125 process+satin 3.66x8.5
Designer AE Creative Traffic Production Proofer
Russ
Alarms for Clinicians
For additional information
and guidelines, please refer to the
HeartMate II Instructions for Use.
H A Z A R D
Important! The pump running ( ) symbol is
always lit green when the pump is running.
Thoratec Corporation 6035 Stoneridge Drive, Pleasanton, CA
94588 Tel: (925) 847-8600, Fax: (925) 847-8574, HeartLine:
(800) 456-1477, www.thoratec.com
Thoratec Europe Limited Burnett House, 3 Lakeview Court,
Ermine Business Park, Huntingdon, Cambridgeshire, PE29 6UA
United Kingdom, Tel: +44 (0) 1480 455200, Fax: +44 (0) 1480
454126, 24-Hr Tel: +44 (0) 7659 877901
Thoratec Corporation continually strives to provide the highest quality products for mechanical circulatory support.
Specifications may change without notice. Therefore, please refer to the HeartMateII Left Ventricular Assist System IFU shipped
with the HeartMate II for the most current information regarding indications, contraindications, warnings, and cautions.
Thoratec, the Thoratec logo, HeartMate, and HeartMate II are registered trademarks and HeartLine is a trademark of Thoratec
Corporation ©2013 Thoratec Corporation.
107526.D
HMII_Pocket Guide_Alarms_Clinicians_WW_107526D_ENGLISH-R1.indd 1HMII_Pocket Guide_Alarms_Clinicians_WW_107526D_ENGLISH-R1.indd 1 4/16/2013 1:54:09 PM4/16/2013 1:54:09 PM