3 Operating Instructions
© 2020 Thorlabs 7
3.3.1 Operating Modes
The SPCM software offers five operating modes to be selected from a drop-down list:
-Manual Mode:The APD is biased beyond breakdown voltage internally. The counter is
started and stopped manually via the appropriate button in the software, at each "start"
command the counter is reset. No settings except Pulse Blind Time can be made. The
count result is updated 10 times per second and displayed numerically (Alignment display)
and graphically (tabs Graph and Bar).
-Free Running Timed Counter: The APD is biased beyond breakdown voltage at any time.
The counter is started / stopped by the internal timer. The timer is programmed via soft-
ware, this way defining the bin length and the waiting time between bins. The sum of bin
length and time between bins must not go below 0.020 ms. An additional Pulse Blind Time
can be set. After pushing the Start button, the timer begins to control the counter and inci-
dent photons are counted into bins until the Stop button is pressed. The results are dis-
played graphically and numerically, see Measurement Result Display .
-Externally Triggered Timed Counter: This operating mode is similar to the Free Running
Timed Counter mode, with the difference that the timer is started by an external trigger sig-
nal. The sensitive trigger edge (rising or falling) can be selected. The APD is biased bey-
ond breakdown voltage at any time. The counter is started / stopped by the internal timer.
The bin length is programmed via software. After pushing the Start button, the trigger is
armed and waits for the first pulse slope. When the selected trigger edge is detected, the
timer is started and photons are counted for a time, defined by the set bin length. After fin-
ishing the bin, the counter stops. With the next trigger slope, the count starts again. This
continues until the Stop button is pressed. The results are displayed graphically and numeri-
cally.
-Externally Triggered Counter: In this mode, the counter is not timed, but controlled (star-
ted and stopped) only by an external trigger signal. The sensitive trigger edge (rising or fall-
ing) can be selected. The APD is biased beyond breakdown voltage at any time. After push-
ing the Start button, the trigger is armed and waits for the first pulse slope. When the selec-
ted trigger edge is detected, the counter starts photon count until the next trigger pulse ar-
rives. As for this reason, the trigger period Tpdefines the bin lenght as well as the time
between subsequent bins. This counting process continues until the Stop button is pressed.
The results are displayed graphically and numerically.
-External Gating: The External Gating is similar to Externally Triggered Counter mode. The
significant difference is that the APD is not always biased beyond breakdown voltage, but
the APD bias is controlled by an external gating signal. The advantage of this active gating
is a lower probability to miss a photon: In previously described modes, the APD is biased
beyond breakdown voltage at any time. That means, it reacts on incident photons at any
time. This reaction includes also the activity of the quenching circuit; during the quenching
time (35 - 45ns) the APD is "blind" for incident photons. When the APD is gated actively, it
is armed with the arrival of the gating signal so that the first incoming photon can be coun-
ted. In External Gating mode, the counter is not triggered on a selectable edge, but enabled
9