4.2. Making the Safety Interlock Connections
The LDM Series is equipped with a remote interlock connector located on the end
cap. In order to enable the laser driver, a short circuit must be applied across the
terminals of the remote interlock connector. In practice this connection is made
available to allow the user to connect a remote actuated switch to the connector
(i.e. an open door indicator). The switch (which must be normally open) has to be
closed in order for the unit to be enabled. Once the switch is in an open state, the
laser diode must automatically shutdown.
All units shipped from Thorlabs are configured with a shorting device installed in
the Interlock connector. If you are not going to use this feature then you can leave
the shorting device installed and the unit will operate normally as described in the
procedures in this manual. If you wish to make use of the Interlock feature you will
need to acquire the appropriate connector mate and wire it your remote interlock
switch. Next, remove the shorting device by unscrewing it from the output and
install the connector into the Interlock output.
The Interlock output only accepts a 2.5 mm mono phono jack. This connector is
readily available at most electronic supply stores.
The electrical specifications for the Interlock output are as follows:
Type of Mating Connector: 2.5 mm mono phono jack
Connector Polarity: Tip is positive, barrel is ground
Interlock Switch Requirements: Must be N.O. dry contacts
Figure 4 Remote Interlock Connector
CAUTION!
Under no circumstances should any external voltages be applied to the
Interlock output