Warnings
The 7161 manifold is designed for use with pressure controllers, calibrators and regulated
pneumatic pressure systems. These systems control a supplied pressure source and output
them to a unit under test. The 7161 is connected between the pressure source and the UUT.
Please observe all warnings and safety instructions when dealing with high pressure
supplies and outputs. Please ensure that all users of the instrumentation have fully read and
understood operating instructions and safety procedures for handling high pressure systems.
Observe all warnings before operating pressure calibration systems.
The 7161 shall only be used with clean dry air or nitrogen.
Never use hazardous media as pressure medium.
Do not exceed 200 bar (3000 psi) input to the manifold.
Before pressurisation of the 7161, check all components and connections are in good condition,
fully functional and that all screw fittings are firmly. All connections should be suitable for the
applied maximum pressure.
In the case of an error in connections or use, a high pressure or vacuum may be on the input
and output connections. Ensure supplies are vented before working on pressure connections.
During operation, do not disconnect the 7161 when pressure is being applied.
When working on and with the device wear safety glasses.
Use the manifold in the correct orientation on a solid work surface. The bleed valve vents
downwards so exercise with caution when using.
Connections and UUTs damaged from overpressure can cause high velocity shrapnel.
Do not over-tighten the input or output connectors as this can damage the threads.
The 7161 uses a stainless steel valve to vent. This means that the vent valve is fully open or
closed several turns from the end of travel. Over tightening does not increase the effectiveness
of the seal.
Note that this manual refers to the Unit Under Test as the UUT. The UUT should be
connected to the 7161 directly or by using suitable fittings ensuring they are rated for at least
the maximum pressure available from the pressure source.
Time Electronics Ltd do not assume liability for damages that arise from incorrect use of the
instrument device or from disregard of the information contained in this manual.