Transmille 3000 Series Use and care manual

3000SERIES
MULTI PRODUCT CALIBRATORS
CALIBRATION GUIDE
Ü

3000 Series Service Manual.doc Page 17
CALIBRATION / VERIFICATION OVERVIEW
To verify the 3000 Series calibrators, it is necessary to measure the outputs from each
range and compare them to the published specifications. Linearity checks should also be
performed.
A basic verification procedure would be typically as little as 60 tests, although a full
procedure may be as many as 400 tests. Please see www.transmille.com for an example
3000 Series certificate. When using Transmille PROCAL calibration software, a fully
automated verification & calibration procedure is available for approved service centres.
Adjustment can be made using two methods – either direct front panel adjustment or
adjustment using a PC based Virtual Front Panel software package (optional) with the
calibrator connected to the PC RS232 interface.
WARNING : RISK OF SHOCK
THIS PROCEDURE SHOULD ONLY BE ATTEMPTED BY QUALIFIED PERSONNEL
To prevent unauthorised use of the VFP software, a password is required before access is
granted. Adjustment can be completed without disassembly of the calibrator.
Each function e.g. DC voltage, AC Current, Resistance etc. has several ranges. Each
range has one or more calibration constants. See table below.
3000 Series adjustment allows any calibration constant to be adjusted independently of
any other, therefore it is possible to adjust a single range without needing to adjust any
other points. Altering the calibration constants directly changes the calibrators output.
Adjusting the calibrator simply involves changing the constant until the output reads
correctly.
DC Voltage : Zero : + Full Scale : - Full Scale
AC Voltage : Zero : Full Scale @ 206Hz : Frequency Response
DC Current : Zero : + Full Scale : - Full Scale
AC Current : Zero : Full Scale @ 206Hz : Frequency Response
Resistance : 2 Wire & 4 Wire value for each resistance
Capacitance : Value for each Capacitor
Inductance : Value for each Inductor
Linearity is inherent within the design of the D to A in the calibrator and does not need to
be adjusted.
ADJUSTMENT : EQUIPMENT REQUIRED
Precision 8 ½ Digital Multimeter. E.g. Hewlett Packard HP3458A or Wavetek 1281.
Capacitance / Inductance bridge. E.g. Wayne Kerr B905.
Frequency counter.
Shunt resistors for measurement of 2A and 20A.
Low thermal test leads with 4mm plug terminations.
Shrouded test leads suitable for 1000V AC measurements.
1m BNC to BNC cable with 2off BNC to 4mm adapters.
Computer with RS232 interface running Transmille virtual front panel program.
RS232 cable.

3000 Series Service Manual.doc Page 18
ADJUSTMENT OVERVIEW –
USING 3000 SERIES VIRTUAL FRONT PANEL SOFTWARE
1) Install virtual front panel software.
2) Connect 30xx to computer RS232 port
3) Allow all equipment to stabilise for at least 4 hours.
4) Run virtual front panel program.
5) Select range & output to be adjusted using the virtual front panel program.
6) Enter calibration control mode. (Password required).
7) Press ‘Start’ to enable adjustment. A ‘C’ will appear on the calibrator display.
8) Adjust calibration constant until the output of the calibrator is correct. The constants for
each range must be adjusted in the correct sequence. See following pages for
details.
9) Press the store button to save the constant.
(Changing range will also store the constant.)
Press the ‘abort’ button to abandon calibration of the range being adjusted.
10) Select next range to be adjusted.
11) Close calibration control panel and exit virtual front panel program
Starting the virtual control program
1) Install the Virtual front panel program onto computer from CD supplied.
The CD will auto-run. Select ‘utilities’ and follow installation instructions.
2) Connect RS232 cable between computer and calibrator.
3) Run the Virtual front panel program.
4) Select Comm port
5) Click the ‘Show Calibration Control Button’

3000 Series Service Manual.doc Page 19
7) The main calibration screen is now displayed
Show
Calibration
Control
Enter Password

3000 Series Service Manual.doc Page 20
DC Low Voltage Adjustment
Settings and connections.
Adjustment sequence for DC 200mV to 20V ranges.
1) Zero 2) + full scale 3)- full scale
ZERO ADJUSTMENT
POSITIVE FULL SCALE ADJUSTMENT
Test Leads Low Thermal
HP3458 Setting DCV, NPLC 30, NDIG 6, ARANGE
30xx Terminals Voltage
Notes NULL DMM before test and re-check NULL after 200mV range adsustment
2) Click START
button
3) Click up/down
buttons until DMM
reads 0.0000mV
4) Click +FS
button
5) Enter reading on
DMM here and click
CALC button. DMM
should now read
200.0000mV. Fine
adjustment may be
made using the up/down
buttons
1) Click 200mV range
button

3000 Series Service Manual.doc Page 21
Negative Full Scale Adjustment
The 2V and 20V ranges are adjusted using the technique as per the 200mV range.
Adjustment points
TEST TITLE TEST VALUE RANGE
200mV Range 0mV 200mV
200mV Range 200mV 200mV
200mV Range -200mV 200mV
2V Range 0V 2V
2V Range 2V 2V
2V Range -2V 2V
20V Range 0V 20V
20V Range 20V 20V
20V Range -20V 20V
7) Enter reading on
DMM here and click
CALC button. DMM
should now read
-200.0000mV. Fine
adjustment may be
made using the
up/down buttons
8) Click the
STORE button

3000 Series Service Manual.doc Page 22
DC High Voltage Adjustment
200V Range
1) Connect shrouded test leads between 30xx Voltage terminals and DMM Voltage input.
2) Set HP3458A to DCV, NDIG 6, NPLC 30, 1000V RANGE
3) Click 200V range button on 30xx VFP
ZERO ADJUSTMENT
The full scales are calibrated in the same way as low voltage DC
The 1000V range is calibrated using the technique as the 200V range except that the zero
is set up at 50V. See table below for adjustment points.
Adjustment points
TEST TITLE TEST VALUE RANGE
200V Range 5V 200V
200V Range 200V 200V
200V Range -200V 200V
1kV Range 50V 1kV
1kV Range 1000V 1kV
1kV Range -1000V 1kV
5) Enter ‘5’ in
this box and
press return key
4) Click 200V range
button
6) Click
up/down
buttons until
DMM reads
0.000V

3000 Series Service Manual.doc Page 23
After calibrating the DC Voltage ranges, Click the HIDE CALIBRATION CONTROL button
to return to the ‘function selection screen’.
FUNCTION SELECT SCREEN.
HIDE
CALIBRARATION
CONTROL

3000 Series Service Manual.doc Page 24
DC Current Adjustment
1) Connect the shrouded test leads between 30xx Current terminals and DMM Current
input.
2) Set HP3458A to DCI, NPLC 30, NDIG 6, AUTORANGE.
3) Open circuit test leads at calibrator end and select MATH NULL on DMM
4) Re-connect test leads to 30xx
5) Select DCI on FUNCTION SELECTION SCREEN.
6) Click Show Calibration Control Button
7) Click 200uA Button
The adjustment procedure is the same as DC Voltage, adjust zero, positive full scale and
negative full scale as shown in the table below.
For 2A range adjustment, connect a 1 Ohm standard resistor to the 30xx output and
measure voltage on the V terminals of the resistor with the DMM on the 2V DC range.
For 20A range adjustment, use a 0.1 Ohm standard resistor.
Adjustment points
TEST TITLE TEST VALUE RANGE
200uA Range 0uA 200uA
200uA Range 200uA 200uA
200uA Range -200uA 200uA
2mA Range 0mA 2mA
2mA Range 2mA 2mA
2mA Range -2mA 2mA
20mA Range 0mA 20mA
20mA Range 20mA 20mA
20mA Range -20mA 20mA
200mA Range 0mA 200mA
200mA Range 200mA 200mA
200mA Range -200mA 200mA
2A Range 0A 2A
2A Range 2A 2A
2A Range -2A 2A
20A Range 0A 20A
20A Range 20A 20A
20A Range -10A 20A

3000 Series Service Manual.doc Page 25
AC Voltage Adjustment
AC voltage is calibrated by adjusting the output at 206Hz and then adjusting the frequency
response at other frequencies found in the drop down box. The 206Hz output must always
be set up first.
1) Connect screened test leads between 30xx Voltage terminals and DMM Voltage input.
2) Set HP3458A to ACV, SETACV SYNC, LFILTER 1, NPLC 30, MANUAL RANGE &
LFREQ LINE
3) Select MATH OFF on DMM
4) Click ACV on FUNCTION SELECT screen.
5) Click 200mV range button on 30xx VFP
6) Click ZERO
button
7) Click START
button
8) Click
up/down buttons
until DMM
reads 50.00mV
9) Click FS
Button
10) Enter reading on
DMM here and click
CALC button. DMM
should now read
200.0000mV. Fine
adjustment may be
made using the
up/down buttons
ZERO CAL
FS CAL
FREQ SET
POINT CAL
11) Double click frequency selection
box and select lowest frequency
available for this range. Look at the
user manual supplied with the
instrument to find the max/min
fre
q
uencies for this ran
g
e.
12) Enter reading
on DMM here and
click CALC button.
DMM should now
read 200.0000mV.
Fine adjustment
may be made using
the up/down buttons

3000 Series Service Manual.doc Page 26
The adjustment procedure is the same for the other AC Voltage ranges. Note the 1000V
range full scale is adjusted at 700V because of the voltage input limit of the DMM. See
table below.
13) Adjust 30xx at all
frequencies on this list
which are available on the
selected range as above
14) Click the
STORE button
when all the
frequency tests
are complete.

3000 Series Service Manual.doc Page 27
AC Voltage Frequency Response Calibration Test Points For 3050
TEST TITLE TEST VALUE RANGE
200mV Range : 206Hz 50mV 200mV
200mV Range : 206Hz 200mV 200mV
200mV Range : 10Hz 200mV 200mV
200mV Range :40Hz 200mV 200mV
200mV Range : 56Hz 200mV 200mV
200mV Range: 106Hz 200mV 200mV
200mV Range: 596Hz 200mV 200mV
200mV Range : 1000Hz 200mV 200mV
200mV Range : 2000Hz 200mV 200mV
200mV Range : 3500Hz 200mV 200mV
200mV Range : 5000Hz 200mV 200mV
200mV Range : 7500Hz 200mV 200mV
200mV Range : 10000Hz 200mV 200mV
200mV Range : 15000Hz 200mV 200mV
200mV Range : 20000Hz 200mV 200mV
2V Range : 206Hz 0.5V 2V
2V Range : 206Hz 2V 2V
2V Range : 10Hz 2V 2V
2V Range : 40Hz 2V 2V
2V Range : 56Hz 2V 2V
2V Range: 106Hz 2V 2V
2V Range: 596Hz 2V 2V
2V Range : 1000Hz 2V 2V
2V Range : 2000Hz 2V 2V
2V Range : 3500Hz 2V 2V
2V Range : 5000Hz 2V 2V
2V Range : 7500Hz 2V 2V
2V Range : 10000Hz 2V 2V
2V Range : 15000Hz 2V 2V
2V Range : 20000Hz 2V 2V
2V Range : 30000Hz 2V 2V
2V Range : 40000Hz 2V 2V
2V Range : 50000Hz 2V 2V
2V Range : 60000Hz 2V 2V
2V Range : 80000Hz 2V 2V
2V Range : 100000Hz 2V 2V
TEST TITLE TEST VALUE RANGE
20V Range : 206Hz 5V 20V
20V Range : 206Hz 20V 20V
20V Range : 10Hz 20V 20V
20V Range : 40Hz 20V 20V
20V Range : 56Hz 20V 20V
20V Range: 106Hz 20V 20V
20V Range: 596Hz 20V 20V
20V Range : 1000Hz 20V 20V
20V Range : 2000Hz 20V 20V
20V Range : 3500Hz 20V 20V
20V Range : 5000Hz 20V 20V
20V Range : 7500Hz 20V 20V
20V Range : 10000Hz 20V 20V
20V Range : 15000Hz 20V 20V
20V Range : 20000Hz 20V 20V
20V Range : 30000Hz 20V 20V
20V Range : 40000Hz 20V 20V
20V Range : 50000Hz 20V 20V
20V Range : 60000Hz 20V 20V
20V Range : 80000Hz 20V 20V
20V Range : 100000Hz 20V 20V
200V Range : 206Hz 50V 200V
200V Range : 206Hz 200V 200V
200V Range : 40Hz 200V 200V
200V Range : 56Hz 200V 200V
200V Range: 106Hz 200V 200V
200V Range: 596Hz 200V 200V
200V Range : 1000Hz 200V 200V
200V Range : 2000Hz 200V 200V
200V Range : 3500Hz 200V 200V
200V Range : 5000Hz 200V 200V
200V Range : 7500Hz 200V 200V
200V Range : 10000Hz 200V 200V
200V Range : 15000Hz 200V 200V
200V Range : 20000Hz 200V 200V
1000V Range : 206Hz 250V 1000V
1000V Range : 206Hz 700V 1000V
1000V Range : 40Hz 700V 1000V
1000V Range : 56Hz 700V 1000V
1000V Range: 106Hz 700V 1000V
1000V Range: 596Hz 700V 1000V
1000V Range : 1000Hz 700V 1000V
1000V Range : 2000Hz 700V 1000V
1000V Range : 3500Hz 700V 1000V
1000V Range : 5000Hz 700V 1000V
1000V Range : 7500Hz 700V 1000V
1000V Range : 10000Hz 700V 1000V

3000 Series Service Manual.doc Page 28
AC VOLTAGE FREQUENCY RESPONSE CALIBRATION TEST POINTS FOR 3041
20V Range : 10Hz 20V 20V
20V Range : 30Hz 20V 20V
20V Range : 56Hz 20V 20V
20V Range: 106Hz 20V 20V
20V Range: 596Hz 20V 20V
20V Range : 1000Hz 20V 20V
20V Range : 2000Hz 20V 20V
20V Range : 3500Hz 20V 20V
20V Range : 5000Hz 20V 20V
20V Range : 7500Hz 20V 20V
20V Range : 10000Hz 20V 20V
20V Range : 15000Hz 20V 20V
20V Range : 20000Hz 20V 20V
20V Range : 30000Hz 20V 20V
20V Range : 40000Hz 20V 20V
20V Range : 50000Hz 20V 20V
20V Range : 60000Hz 20V 20V
20V Range : 80000Hz 20V 20V
20V Range : 100000Hz 20V 20V
200V Range : 206Hz 50V 200V
200V Range : 206Hz 200V 200V
200V Range : 30Hz 200V 200V
200V Range : 56Hz 200V 200V
200V Range: 106Hz 200V 200V
200V Range: 596Hz 200V 200V
200V Range : 1000Hz 200V 200V
200V Range : 2000Hz 200V 200V
200V Range : 3500Hz 200V 200V
200V Range : 5000Hz 200V 200V
200V Range : 7500Hz 200V 200V
200V Range : 10000Hz 200V 200V
200V Range : 15000Hz 200V 200V
200V Range : 20000Hz 200V 200V
1000V Range : 206Hz 250V 1000V
1000V Range : 206Hz 700V 1000V
1000V Range : 30Hz 700V 1000V
1000V Range : 56Hz 700V 1000V
1000V Range: 106Hz 700V 1000V
1000V Range: 596Hz 700V 1000V
1000V Range : 1000Hz 700V 1000V
1000V Range : 2000Hz 700V 1000V
1000V Range : 3500Hz 700V 1000V
1000V Range : 5000Hz 700V 1000V
1000V Range : 7500Hz 700V 1000V
1000V Range : 10000Hz 700V 1000V
TEST TITLE TEST VALUE RANGE
200mV Range : 206Hz 50mV 200mV
200mV Range : 206Hz 200mV 200mV
200mV Range : 10Hz 200mV 200mV
200mV Range :30Hz 200mV 200mV
200mV Range : 56Hz 200mV 200mV
200mV Range: 106Hz 200mV 200mV
200mV Range: 596Hz 200mV 200mV
200mV Range : 1000Hz 200mV 200mV
200mV Range : 2000Hz 200mV 200mV
200mV Range : 3500Hz 200mV 200mV
200mV Range : 5000Hz 200mV 200mV
200mV Range : 7500Hz 200mV 200mV
200mV Range : 10000Hz 200mV 200mV
200mV Range : 20000Hz 200mV 200mV
200mV Range : 30000Hz 200mV 200mV
200mV Range : 40000Hz 200mV 200mV
200mV Range : 50000Hz 200mV 200mV
200mV Range : 60000Hz 200mV 200mV
200mV Range : 80000Hz 200mV 200mV
200mV Range : 100000Hz 200mV 200mV
200mV Range : 200000Hz 200mV 200mV
200mV Range : 400000Hz 200mV 200mV
200mV Range : 500000Hz 200mV 200mV
2V Range : 206Hz 0.5V 2V
2V Range : 206Hz 2V 2V
2V Range : 10Hz 2V 2V
2V Range : 30Hz 2V 2V
2V Range : 56Hz 2V 2V
2V Range: 106Hz 2V 2V
2V Range: 596Hz 2V 2V
2V Range : 1000Hz 2V 2V
2V Range : 2000Hz 2V 2V
2V Range : 3500Hz 2V 2V
2V Range : 5000Hz 2V 2V
2V Range : 7500Hz 2V 2V
2V Range : 10000Hz 2V 2V
2V Range : 15000Hz 2V 2V
2V Range : 20000Hz 2V 2V
2V Range : 30000Hz 2V 2V
2V Range : 40000Hz 2V 2V
2V Range : 50000Hz 2V 2V
2V Range : 60000Hz 2V 2V
2V Range : 80000Hz 2V 2V
2V Range : 100000Hz 2V 2V
2V Range : 200000Hz 2V 2V
2V Range : 400000Hz 2V 2V
2V Range : 500000Hz 2V 2V

3000 Series Service Manual.doc Page 29
AC VOLTAGE FREQUENCY RESPONSE CALIBRATION TEST POINTS FOR 3010
20V Range : 10Hz 20V 20V
20V Range : 30Hz 20V 20V
20V Range : 56Hz 20V 20V
20V Range: 106Hz 20V 20V
20V Range: 596Hz 20V 20V
20V Range : 1000Hz 20V 20V
20V Range : 2000Hz 20V 20V
20V Range : 3500Hz 20V 20V
20V Range : 5000Hz 20V 20V
20V Range : 7500Hz 20V 20V
20V Range : 10000Hz 20V 20V
20V Range : 15000Hz 20V 20V
20V Range : 20000Hz 20V 20V
20V Range : 30000Hz 20V 20V
20V Range : 40000Hz 20V 20V
20V Range : 50000Hz 20V 20V
20V Range : 60000Hz 20V 20V
20V Range : 80000Hz 20V 20V
20V Range : 100000Hz 20V 20V
200V Range : 206Hz 50V 200V
200V Range : 206Hz 200V 200V
200V Range : 30Hz 200V 200V
200V Range : 56Hz 200V 200V
200V Range: 106Hz 200V 200V
200V Range: 596Hz 200V 200V
200V Range : 1000Hz 200V 200V
200V Range : 2000Hz 200V 200V
200V Range : 3500Hz 200V 200V
200V Range : 5000Hz 200V 200V
200V Range : 7500Hz 200V 200V
200V Range : 10000Hz 200V 200V
200V Range : 15000Hz 200V 200V
200V Range : 20000Hz 200V 200V
200V Range : 30000Hz 200V 200V
200V Range : 40000Hz 200V 200V
1000V Range : 206Hz 250V 1000V
1000V Range : 206Hz 700V 1000V
1000V Range : 30Hz 700V 1000V
1000V Range : 56Hz 700V 1000V
1000V Range: 106Hz 700V 1000V
1000V Range: 596Hz 700V 1000V
1000V Range : 1000Hz 700V 1000V
1000V Range : 2000Hz 700V 1000V
1000V Range : 3500Hz 700V 1000V
1000V Range : 5000Hz 700V 1000V
1000V Range : 7500Hz 700V 1000V
1000V Range : 10000Hz 700V 1000V
TEST TITLE TEST VALUE RANGE
200mV Range : 206Hz 50mV 200mV
200mV Range : 206Hz 200mV 200mV
200mV Range : 10Hz 200mV 200mV
200mV Range :30Hz 200mV 200mV
200mV Range : 56Hz 200mV 200mV
200mV Range: 106Hz 200mV 200mV
200mV Range: 596Hz 200mV 200mV
200mV Range : 1000Hz 200mV 200mV
200mV Range : 2000Hz 200mV 200mV
200mV Range : 3500Hz 200mV 200mV
200mV Range : 5000Hz 200mV 200mV
200mV Range : 7500Hz 200mV 200mV
200mV Range : 10000Hz 200mV 200mV
200mV Range : 20000Hz 200mV 200mV
200mV Range : 30000Hz 200mV 200mV
200mV Range : 40000Hz 200mV 200mV
200mV Range : 50000Hz 200mV 200mV
200mV Range : 60000Hz 200mV 200mV
200mV Range : 80000Hz 200mV 200mV
200mV Range : 100000Hz 200mV 200mV
200mV Range : 200000Hz 200mV 200mV
200mV Range : 400000Hz 200mV 200mV
200mV Range : 500000Hz 200mV 200mV
2V Range : 206Hz 0.5V 2V
2V Range : 206Hz 2V 2V
2V Range : 10Hz 2V 2V
2V Range : 30Hz 2V 2V
2V Range : 56Hz 2V 2V
2V Range: 106Hz 2V 2V
2V Range: 596Hz 2V 2V
2V Range : 1000Hz 2V 2V
2V Range : 2000Hz 2V 2V
2V Range : 3500Hz 2V 2V
2V Range : 5000Hz 2V 2V
2V Range : 7500Hz 2V 2V
2V Range : 10000Hz 2V 2V
2V Range : 15000Hz 2V 2V
2V Range : 20000Hz 2V 2V
2V Range : 30000Hz 2V 2V
2V Range : 40000Hz 2V 2V
2V Range : 50000Hz 2V 2V
2V Range : 60000Hz 2V 2V
2V Range : 80000Hz 2V 2V
2V Range : 100000Hz 2V 2V
2V Range : 200000Hz 2V 2V
2V Range : 400000Hz 2V 2V
2V Range : 500000Hz 2V 2V

3000 Series Service Manual.doc Page 30
AC Current Adjustment
1) Connect the screened test leads between 30xx Current terminals and DMM Current
input.
2) Set HP3458A to ACI, NPLC 30.SETACV SYNC, LFILTER 1, LFREQ LINE
3) Select MATH OFF on DMM
4) Select ACI on FUNCTION SELECTION SCREEN.
5) Click Show Calibration Control Button
6) Click 200uA Button
The adjustment procedure is the same as AC Voltage, calibrate zero, positive full scale
and frequency points as shown in the table below.
For 2A range adjustment, connect a 1 Ohm standard resistor to the 30xx output and
measure voltage on the V terminals of the resistor with the DMM on the 2V AC range
For 20/30A range adjustment, use a 0.1 Ohm standard resistor.

3000 Series Service Manual.doc Page 31
AC CURRENT FREQUENCY RESPONSE CALIBRATION POINTS FOR 3050
TEST TITLE TEST VALUE RANGE
200uA Range: 206Hz 50uA 200uA
200uA Range: 206Hz 200uA 200uA
200uA Range: 10Hz 200uA 200uA
200uA Range: 40Hz 200uA 200uA
200uA Range: 56Hz 200uA 200uA
200uA Range: 106Hz 200uA 200uA
200uA Range: 596Hz 200uA 200uA
200uA Range: 1000Hz 200uA 200uA
200uA Range: 2000Hz 200uA 200uA
200uA Range: 3500Hz 200uA 200uA
200uA Range: 5000Hz 200uA 200uA
200uA Range: 7500Hz 200uA 200uA
200uA Range: 10000Hz 200uA 200uA
2mA Range: 206Hz 0.5mA 2mA
2mA Range: 206Hz 2mA 2mA
2mA Range: 10Hz 2mA 2mA
2mA Range: 40Hz 2mA 2mA
2mA Range: 56Hz 2mA 2mA
2mA Range: 106Hz 2mA 2mA
2mA Range: 596Hz 2mA 2mA
2mA Range: 1000Hz 2mA 2mA
2mA Range: 2000Hz 2mA 2mA
2mA Range: 3500Hz 2mA 2mA
2mA Range: 5000Hz 2mA 2mA
2mA Range: 7500Hz 2mA 2mA
2mA Range: 10000Hz 2mA 2mA
20mA Range: 206Hz 5mA 20mA
20mA Range: 206Hz 20mA 20mA
20mA Range: 10Hz 20mA 20mA
20mA Range: 40Hz 20mA 20mA
20mA Range: 56Hz 20mA 20mA
20mA Range: 106Hz 20mA 20mA
20mA Range: 596Hz 20mA 20mA
20mA Range: 1000Hz 20mA 20mA
20mA Range: 2000Hz 20mA 20mA
20mA Range: 3500Hz 20mA 20mA
20mA Range: 5000Hz 20mA 20mA
20mA Range: 7500Hz 20mA 20mA
20mA Range: 10000Hz 20mA 20mA
TEST TITLE TEST VALUE RANGE
200mA Range: 206Hz 50m
A
200m
A
200mA Range: 206Hz 200m
A
200m
A
200mA Range: 10Hz 200m
A
200m
A
200mA Range: 40Hz 200m
A
200m
A
200mA Range: 56Hz 200m
A
200m
A
200mA Range: 106Hz 200m
A
200m
A
200mA Range: 596Hz 200m
A
200m
A
200mA Range: 1000Hz 200m
A
200m
A
200mA Range: 2000Hz 200m
A
200m
A
200mA Range: 3500Hz 200m
A
200m
A
200mA Range: 5000Hz 200m
A
200m
A
200mA Range: 7500Hz 200m
A
200m
A
200mA Range: 10000Hz 200m
A
200m
A
2A Range: 206Hz 500m
A
2
A
2A Range: 206Hz 2
A
2
A
2A Range: 10Hz 2
A
2
A
2A Range: 40Hz 2
A
2
A
2A Range: 56Hz 2
A
2
A
2A Range: 106Hz 2
A
2
A
2A Range: 596Hz 2
A
2
A
2A Range: 1000Hz 2
A
2
A
2A Range: 2000Hz 2
A
2
A
20A Range: 206Hz 3
A
20
A
20A Range: 206Hz 12
A
20
A
20A Range: 10Hz 12
A
20
A
20A Range: 40Hz 12
A
20
A
20A Range: 56Hz 12
A
20
A
20A Range: 106Hz 12
A
20
A
20A Range: 596Hz 12
A
20
A
20A Range: 1000Hz 12
A
20
A
20A Range: 2000Hz 12
A
20
A

3000 Series Service Manual.doc Page 32
AC CURRENT FREQUENCY RESPONSE CALIBRATION POINTS FOR 3041
TEST TITLE TEST VALUE RANGE
200uA Range: 206Hz 50uA 200uA
200uA Range: 206Hz 200uA 200uA
200uA Range: 10Hz 200uA 200uA
200uA Range: 30Hz 200uA 200uA
200uA Range: 56Hz 200uA 200uA
200uA Range: 106Hz 200uA 200uA
200uA Range: 596Hz 200uA 200uA
200uA Range: 1000Hz 200uA 200uA
200uA Range: 2000Hz 200uA 200uA
200uA Range: 3500Hz 200uA 200uA
200uA Range: 5000Hz 200uA 200uA
200uA Range: 7500Hz 200uA 200uA
200uA Range: 10000Hz 200uA 200uA
2mA Range: 206Hz 0.5mA 2mA
2mA Range: 206Hz 2mA 2mA
2mA Range: 10Hz 2mA 2mA
2mA Range: 30Hz 2mA 2mA
2mA Range: 56Hz 2mA 2mA
2mA Range: 106Hz 2mA 2mA
2mA Range: 596Hz 2mA 2mA
2mA Range: 1000Hz 2mA 2mA
2mA Range: 2000Hz 2mA 2mA
2mA Range: 3500Hz 2mA 2mA
2mA Range: 5000Hz 2mA 2mA
2mA Range: 7500Hz 2mA 2mA
2mA Range: 10000Hz 2mA 2mA
20mA Range: 206Hz 5mA 20mA
20mA Range: 206Hz 20mA 20mA
20mA Range: 10Hz 20mA 20mA
20mA Range: 30Hz 20mA 20mA
20mA Range: 56Hz 20mA 20mA
20mA Range: 106Hz 20mA 20mA
20mA Range: 596Hz 20mA 20mA
20mA Range: 1000Hz 20mA 20mA
20mA Range: 2000Hz 20mA 20mA
20mA Range: 3500Hz 20mA 20mA
20mA Range: 5000Hz 20mA 20mA
20mA Range: 7500Hz 20mA 20mA
20mA Range: 10000Hz 20mA 20mA
TEST TITLE TEST VALUE RANGE
200mA Range: 206Hz 50m
A
200m
A
200mA Range: 206Hz 200m
A
200m
A
200mA Range: 10Hz 200m
A
200m
A
200mA Range: 30Hz 200m
A
200m
A
200mA Range: 56Hz 200m
A
200m
A
200mA Range: 106Hz 200m
A
200m
A
200mA Range: 596Hz 200m
A
200m
A
200mA Range: 1000Hz 200m
A
200m
A
200mA Range: 2000Hz 200m
A
200m
A
200mA Range: 3500Hz 200m
A
200m
A
200mA Range: 5000Hz 200m
A
200m
A
200mA Range: 7500Hz 200m
A
200m
A
200mA Range: 10000Hz 200m
A
200m
A
2A Range: 206Hz 500m
A
2
A
2A Range: 206Hz 2
A
2
A
2A Range: 10Hz 2
A
2
A
2A Range: 30Hz 2
A
2
A
2A Range: 56Hz 2
A
2
A
2A Range: 106Hz 2
A
2
A
2A Range: 596Hz 2
A
2
A
2A Range: 1000Hz 2
A
2
A
2A Range: 2000Hz 2
A
2
A
2A Range: 3500Hz 2
A
2
A
2A Range: 5000Hz 2
A
2
A
30A Range: 206Hz 3
A
30
A
30A Range: 206Hz 20
A
30
A
30A Range: 10Hz 20
A
30
A
30A Range: 30Hz 20
A
30
A
30A Range: 56Hz 20
A
30
A
30A Range: 106Hz 20
A
30
A
30A Range: 596Hz 20
A
30
A
30A Range: 1000Hz 20
A
30
A

3000 Series Service Manual.doc Page 33
AC CURRENT FREQUENCY RESPONSE CALIBRATION POINTS FOR 3010
TEST TITLE TEST VALUE RANGE
200uA Range: 206Hz 50uA 200uA
200uA Range: 206Hz 200uA 200uA
200uA Range: 10Hz 200uA 200uA
200uA Range: 30Hz 200uA 200uA
200uA Range: 56Hz 200uA 200uA
200uA Range: 106Hz 200uA 200uA
200uA Range: 596Hz 200uA 200uA
200uA Range: 1000Hz 200uA 200uA
200uA Range: 2000Hz 200uA 200uA
200uA Range: 3500Hz 200uA 200uA
200uA Range: 5000Hz 200uA 200uA
200uA Range: 7500Hz 200uA 200uA
200uA Range: 10000Hz 200uA 200uA
2mA Range: 206Hz 0.5mA 2mA
2mA Range: 206Hz 2mA 2mA
2mA Range: 10Hz 2mA 2mA
2mA Range: 30Hz 2mA 2mA
2mA Range: 56Hz 2mA 2mA
2mA Range: 106Hz 2mA 2mA
2mA Range: 596Hz 2mA 2mA
2mA Range: 1000Hz 2mA 2mA
2mA Range: 2000Hz 2mA 2mA
2mA Range: 3500Hz 2mA 2mA
2mA Range: 5000Hz 2mA 2mA
2mA Range: 7500Hz 2mA 2mA
2mA Range: 10000Hz 2mA 2mA
20mA Range: 206Hz 5mA 20mA
20mA Range: 206Hz 20mA 20mA
20mA Range: 10Hz 20mA 20mA
20mA Range: 30Hz 20mA 20mA
20mA Range: 56Hz 20mA 20mA
20mA Range: 106Hz 20mA 20mA
20mA Range: 596Hz 20mA 20mA
20mA Range: 1000Hz 20mA 20mA
20mA Range: 2000Hz 20mA 20mA
20mA Range: 3500Hz 20mA 20mA
20mA Range: 5000Hz 20mA 20mA
20mA Range: 7500Hz 20mA 20mA
20mA Range: 10000Hz 20mA 20mA
TEST TITLE TEST VALUE RANGE
200mA Range: 206Hz 50m
A
200m
A
200mA Range: 206Hz 200m
A
200m
A
200mA Range: 10Hz 200m
A
200m
A
200mA Range: 30Hz 200m
A
200m
A
200mA Range: 56Hz 200m
A
200m
A
200mA Range: 106Hz 200m
A
200m
A
200mA Range: 596Hz 200m
A
200m
A
200mA Range: 1000Hz 200m
A
200m
A
200mA Range: 2000Hz 200m
A
200m
A
200mA Range: 3500Hz 200m
A
200m
A
200mA Range: 5000Hz 200m
A
200m
A
200mA Range: 7500Hz 200m
A
200m
A
200mA Range: 10000Hz 200m
A
200m
A
2A Range: 206Hz 500m
A
2
A
2A Range: 206Hz 2
A
2
A
2A Range: 10Hz 2
A
2
A
2A Range: 30Hz 2
A
2
A
2A Range: 56Hz 2
A
2
A
2A Range: 106Hz 2
A
2
A
2A Range: 596Hz 2
A
2
A
2A Range: 1000Hz 2
A
2
A
2A Range: 2000Hz 2
A
2
A
2A Range: 3500Hz 2
A
2
A
2A Range: 5000Hz 2
A
2
A
30A Range: 206Hz 3
A
30
A
30A Range: 206Hz 20
A
30
A
30A Range: 10Hz 20
A
30
A
30A Range: 30Hz 20
A
30
A
30A Range: 56Hz 20
A
30
A
30A Range: 106Hz 20
A
30
A
30A Range: 596Hz 20
A
30
A
30A Range: 1000Hz 20
A
30
A

3000 Series Service Manual.doc Page 34
Resistance Adjustment - 2 wire
0 Ohms to 10 kOhm ranges
1) Select 2 wire Ohms function on ‘function selection screen’. Click ‘show calibration
control’..
2) Set HP3458A to OHMS4, NPLC 30, AUTO RANGE
3) Use 2 sets of shrouded test leads connected as shown above
4) Select 0 Ohms and note reading on DMM
Repeat this procedure for resistance ranges from 100mOhms to 10kOhms.
6) Type the DMM
reading in this box and
click the ‘send cal factor’
button.
7) The reading here &
on the 30xx should
now be the same as
the reading on the
DMM. If not, enter the
DMM reading again
as in step 6

3000 Series Service Manual.doc Page 35
100kOhms to 1GOhm ranges
1) Connect screened test leads between 30xx Voltage terminals and DMM Voltage input.
2) Set HP3458A to OHMS2, NPLC 30, AUTO RANGE
3) Select 2 wire Ohms from FUNCTION SELECTION SCREEN
4) Select 100k Ohms and note reading on DMM
Adjust value as in steps 6 & 7 on previous page.
Repeat for 1MOhm, 10MOhm, 100MOhm and 1GOhm. (Note that 2040/2005 calibrators
do not have a 1GOhm range)
Other manuals for 3000 Series
1
This manual suits for next models
1
Table of contents
Other Transmille Test Equipment manuals

Transmille
Transmille 3200A Series User manual

Transmille
Transmille 2100 Use and care manual

Transmille
Transmille 1000 Series Use and care manual

Transmille
Transmille 9000 Series User manual

Transmille
Transmille 3000HR User manual

Transmille
Transmille 4610M Owner's manual

Transmille
Transmille 1000 Series User manual

Transmille
Transmille 3000A Series Use and care manual

Transmille
Transmille 4610M User manual

Transmille
Transmille 2100 User manual