U-Line SCOUT H-9560 User manual

PAGE 1 OF 17 0921 IH-9560
1-800-295-5510
uline.com
SCOUT®JEWELRY SCALE
H-9560
DEFINITION OF SIGNAL WARNINGS AND SYMBOLS
SAFETY
These instructions contain installation, operation and
maintenance instructions for the Scout®SJX/E Series
scale. Please read these instructions completely before
using the scale.
Safety notes are marked with signal words and warning
symbols. These show safety issues and warnings.
Ignoring the safety notes may lead to personal injury,
instrument damage, malfunctions or false results.
SIGNAL WORDS
WARNING – For a hazardous situation with medium to
high risk, possibly resulting in injuries or death if not
avoided.
CAUTION – For a hazardous situation with low risk,
resulting in damage to the device or the property, loss
of data or injuries if not avoided.
ATTENTION – For important information about the
product.
NOTE – For useful information about the product.
WARNING SYMBOLS
General Hazard
Electric Shock Hazard
CAUTION: Read all safety warnings before
installing, making connections or servicing
this equipment. Failure to comply with these
warnings could result in personal injury
and/or property damage. Retain all
instructions for future reference.
• Verify that the local AC power supply is within
the input voltage range printed on the AC
adapter's data label.
• Only connect the AC adapter to a
compatible grounded electrical outlet.
• Do not position the scale such that it is
difficult to disconnect the AC adapter from
the power receptacle.
• Ensure the power cord does not pose a
potential obstacle or tripping hazard.
• This scale is intended for indoor use and
should only be operated in dry locations.
• Operate the unit only under ambient
conditions specified in these instructions.
• Do not operate the scale in hazardous or
unstable environments.
• Do not drop loads on the pan.
• Only use approved accessories and
peripherals.
• Disconnect power from the scale before
cleaning or servicing.
• Service should only be performed by
authorized personnel.

PAGE 2 OF 17 0921 IH-9560
SAFETY CONTINUED
INSTALLATION
INTENDED USE
This scale is intended for use in laboratories, schools,
businesses and light industry. It must only be used
for measuring the parameters described in these
instructions. Any other type of use and operation
beyond the limits of technical specifications, without
written consent, is considered not intended. This scale
complies with current industry standards and the
recognized safety regulations, however, it can constitute
a hazard in use. If the instrument is not used according
to these instructions, the intended protection provided
by the instrument may be impaired.
INSTALLING COMPONENTS
Refer to the illustrations and instructions below to
identify and assemble scale with its components. All
components must be assembled before using the scale.
RELEASING TRANSPORTATION LOCK
Release the red transportation lock by turning the red
pointer 90 degrees counterclockwise to unlock.
(See Figure 1)
INSTALLING THE WEIGHING PAN
For scales with a rectangular pan, place pan into
subplatform and rotate counter clockwise until it locks.
For scales with round a pan, place pan straight down
on subplatform. (See Figure 2)
CAPACITY LABEL
Attach capacity label above the screen. (See Figure 3)
SECURITY SLOT
This is provided at rear of scale, allowing it to be secured
by optional cable and lock accessory. (See Figure 4)
SELECTING LOCATION
For best performance, scale should be used in a clean,
stable environment. Do not use in environments with
excessive drafts, with rapid temperature changes, near
magnetic fields or near equipment that generates
magnetic fields or vibrations.
Figure 1
Unlocked
Position
Transportation
Lock Pointer
Subplatform
Locked
Position
Figure 2
Function
Mode
Back
Print
Units
No
Tare
Menu
Exit
Zero
Yes
Figure 3 Capacity Label
Security
Slot
Figure 4

PAGE 3 OF 17 0921 IH-9560
INSTALLATION CONTINUED
LEVELING THE SCALE
Unit has an illuminated level indicator as a reminder
that the scale should be leveled for accurate weighing.
There is a level bubble in a small round window on
the front of the scale. To level, adjust the feet at each
corner until the bubble is centered in the circle. Be sure
the equipment is level each time its location is changed.
(See Figure 5)
CONNECTING POWER
AC ADAPTER INSTALLATION
NOTE: AC power is used to power the scale
when battery power is not needed.
Connect the AC adapter (included) to the AC adapter
input jack at the rear of the scale, and connect the AC
plug to an electrical outlet. (See Figure 5)
BATTERY INSTALLATION
Install the four AA batteries with polarity as shown in the
battery compartment. (See Figure 6)
NOTE: After powering on the unit, it is
recommended to let it warm for at least five
minutes before using.
INITIAL CALIBRATION
When the scale is first installed, and when it is moved
to another location, it must be calibrated to ensure
accurate weighing results.
Scale must be calibrated by using the external
calibration weights.
Have the appropriate calibration masses available
before beginning calibration. Refer to
Calibration Operation section on page 6 for
masses and calibration procedure.
Figure 6
Interface
Connection Battery
Compartment
Cover
AC Adapter
Input Jack
Function
Mode
Back
Tar e
Menu
Exit
Figure 5
Level
Level Bubble
Not Level

PAGE 4 OF 17 0921 IH-9560
OPERATION
CONTROLS
Function
Mode
Back
Print
Units
No
Tare
Menu
Exit
Zero
Yes
NET
%dwt
g/mol
hk H
sg
tw
GPct
N
lb:ozt
kgrn
0
PT
BUTTON Zero
Yes
Print
Units
No
Function
Mode
Back
Tare
Menu
Exit
Primary Function
(Short Press)
Zero/On
Turns on the scale.
If scale is on, it sets to
Zero.
Print
Sends the current
value to the selected
COM ports if
AUTOPRINT is set to Off.
Function
Initiates an
application mode.
Tare
Enter/clear a tare
value.
Secondary Function
(Long Press)
(Extended Press)
Zero/Off
Turns off the scale.
None
Units
Changes the weighing
unit.
None
Mode
Allows changing the
application mode.
None
Menu
Enter the User menu.
View the preset tare
value.
Menu Function
(Short Press)
Yes
Accepts the current
setting on the display.
No
Advances to the next
menu or menu item.
Rejects the current
setting on the display
and advances to the
next available setting.
Back
Moves back to
previous menu item.
Exit
Exits the User menu.
Aborts the calibration
in progress.

PAGE 5 OF 17 0921 IH-9560
OPERATION CONTINUED
NET
%dwt
g/mol
hk H
sg
tw
GPct
N
lb:ozt
kgrn
0
PT
2
1
3
4
5
6
7 8
9
10
11
12
13
ITEM DESCRIPTION
1Battery charge
2Stable weight
3Negative
4Center of Zero
5NET
6Preset Tare
7Pointers
8Kilogram, grain
9Check Weighing
10 Pound, Ounce, Pound:Ounce Symbols
11 Percent, dwt, g/mol
12 Gravity, Pieces, t hk, t sg, t tw, Hold
13 Newton, Totalization, Dynamic, Bluetooth

PAGE 6 OF 17 0921 IH-9560
OPERATION CONTINUED
TURNING ON/OFF
To turn on scale, press and hold the Zero button for
one second. The scale performs a display test and
momentarily displays the software version before
entering the active weighing mode.
To turn off the scale, press and hold the Zero button until
OFF is displayed.
CALIBRATION OPERATION
When the scale is operated for the first time, a span
calibration is recommended to ensure accurate
weighing results. Before performing the calibration, be
sure to have the appropriate calibration weights. Ensure
that the security switch is set to unlocked position.
Press and hold Menu until [MENU] is displayed. When the
button is released, the display will show [C.A.L]. Press Yes
to accept. [SPAN] will then be shown. Press Yes to begin
the span calibration.
[--C--] will be displayed while zero reading is stored.
Next, the display shows the calibration weight value.
Press the No key to toggle the value. Place the specified
calibration weight on the pan. [--C--] will be displayed
while the reading is stored. The display will show [donE]
if the calibration was successful. The scale will return to
the previous application mode and is ready for use.
WEIGHING MODE
This mode is the factory default setting.
1. If needed, press and hold Mode until [WEIGH] is
displayed.
2. If required, place an empty container on the pan
and press Tare.
3. Add sample to the pan or container. The display will
show the weight of the sample.
COUNTING MODE
This mode counts large numbers of items based on the
weight of a reference count.
1. If required, place an empty container on the pan
and press Tare.
2. Press and hold Mode until [Count] is displayed.
[C L r. P W ] (clear average piece weight, APW) will then
display. If no APW exists, the scale will display [PUt.10],
proceed to step 5.
3. Press No to use the stored APW. Proceed to step 7.
4. Press Yes to establish an APW.
5. The scale will then display the sample size, i.e.
[PUt 10]. Press No or Back to toggle the choices
(5, 10, 20, 50 or 100).
6. Put the indicated number of pieces on the pan and
press Yes to calculate the APW. The display will show
the piece count.
7. Add additional pieces until the desired count is
reached.
8. To clear the stored APW, press and hold Mode
until [Count] is displayed. Press Yes when [C l r. p W ] is
displayed.
NOTE: Press Function to view the current APW.
PERCENT MODE
This mode measures the weight of a sample as a
percentage of a reference weight.
1. If required, place an empty container on the pan
and press Tare.
2. Press and hold Mode until [PErcnt] is displayed.
[C L r. r E F ] will then display. If no reference weight
exists, the scale will display [Put.rEF]; proceed to
step 5.
3. Press No to use the stored reference weight and
proceed to step 6.
4. Press Yes to establish a new reference. Scale will now
display [Put.rEF].
5. Add the desired reference material to the pan or
container. Press Yes to store the reference weight.
The display will show 100%.
6. Replace the reference material with the sample
material. The display will show the percentage of the
sample compared to reference weight.
7. To clear the stored reference, press and hold Mode
until [PErcnt] is displayed. Press Yes when [C L r. r E F ] is
displayed.
NOTE: Press Function to view the current
reference weight.
CHECK MODE
Use this mode to compare the weight to a target weight
range. The scale supports positive, negative and zero-
check weighing.
CHECK WEIGHING
Use this mode to compare the weight of items to a
target weight range.

PAGE 7 OF 17 0921 IH-9560
OPERATION CONTINUED
1. Press and hold Mode until [CHECk] is displayed.
[C L r. r E F ] (clear check limits) will then display.
2. Press No to use the stored check limits and proceed
to step 5.
3. Press Yes to establish new check limits. The scale will
then display [SEt. Lo]. Press Yes to view the "Low" limit
value. Press Yes to accept or No to edit the "Low"
limit value. The stored value will display with the
first digit highlighted [000.000 kg]. Repeatedly press
No until the desired number appears. Press Yes to
accept and highlight the next digit. Repeat until all
the digits are correct. Press Yes to accept the "Low"
limit value, [SEt. Hi] will be displayed.
4. Repeat the same procedure to accept or edit the
"high" value.
5. If required, place an empty container on the pan
and press Tare.
6. Place sample material on the pan or in the
container. If the sample weight is under the target
weight range, the under icon ( ) will light. If the
sample is within the target weight range, the accept
symbol ( ) will light. If the sample is over the target
weight range, the over icon ( ) will light.
NOTE: Press Function to view the low and high
check limits.
POSITIVE CHECK
Positive check is used to determine when the material
added to the scale is within the target range. In this
case the "Under" and "Over" limits must be positive
values. (The "Over" limit must be greater than the
"Under" limit.)
Add material to the scale until it is within the
"Accept" range.
NEGATIVE CHECK
Negative check is used to determine when the material
removed from the scale is within the target range. In this
case, the "Under" and "Over" limits are both negative
values. (The "Under" limit must be greater than the "Over"
limit.)
Place the item to be weighed on the scale and press
Tare.
Remove a portion of the item until it is within the "Accept"
range.
ZERO CHECK
Zero check is used when comparing subsequent
samples to an initial reference sample. In this case, the
"Under" limit must be a negative value and the "Over"
limit must be a positive value.
Place the reference item on the scale and press Tare.
Remove the reference sample and place the item to be
compared on the scale to determine if it is within
the "Accept" range.
TOTALIZE MODE
This mode allows the user to store a series of weight
measurements. Totalize mode has been initiated
when the symbol is displayed and the current unit is
displayed.
NOTE: Only positive numbers are totalized.
1. Press and hold Mode until [totAL] is displayed.
[CLr.tot] will then be displayed.
2. Press Yes or No key to clear the current totalized
data or not. When a weight is added to the scale,
the value is displayed.
3. If required, place an empty container on the pan
and press Tare. Add the first item. Its weight will be
displayed. Press Function to store the weight. The
symbol will flash and the display will show the total
weight.
4. Press Tare (or remove the weight in previous
operation) and add the next item. The scale will
display its weight. Press Function to store its weight.
The symbol will flash and the new total weight will
be displayed.
5. Repeat step 4 for all of the items to be accumulated.
6. To clear the stored total, press and hold Mode until
[totAL] is displayed. When [CLr. tot] is displayed,
press Yes.
DENSITY MODE
This mode allows the user to calculate the specific
gravity of a sample when using the density kit. Density
mode has been initiated when "G" and the current unit
is displayed.
The sample is weighed in air suspended from the hook
and then weighed in water. The formula is
dry weight
dry weight - wet weight

PAGE 8 OF 17 0921 IH-9560
OPERATION CONTINUED
PREPARATION
Remove the pan, insert the hook and install the beaker
stand (beaker is not included). Press the Zero button to
turn on the scale. (See Figure 6)
DENSITY SETUP
1. Press and hold Mode until [SPEC.Gr] is displayed.
[Air.Wt] will then be displayed.
2. Place the sample on the hook as shown in Figure 6
and press Yes to store the dry weight value. [W.Wt]
will then be displayed.
3. Suspend the sample in water and press Yes to store
the wet weight value. The specific gravity value will
then be displayed and the Gravity icon will blink
while the unit icon is turned off.
NOTE: Specific gravity has no unit of measure.
For large samples, it is recommended to use
the "weighing below" feature instead of the
hook. (See Weight Below Hook on page 12)
HOLD MODE
There are two modes for the display hold:
• Peak Hold: Allows the user to capture and store the
highest stable weight value (>=5d).
• Display Hold (default): Allows the user to capture
and store the first stable weight value (>=5d).
START
If no weight value is held on the display, press Function
key to begin. [rEAdY] will be displayed until a weight is
added on the pan.
When the stable value is being held on the display, the
Hold icon will blink and the displayed weight will not
change.
RESET
If the pan is empty and a weight value was held on the
display, a single short press of the Function key will clear
the held value and show the new weight on the pan.
1. Press and hold Mode until [HoLd] is displayed.
2. If required, place an empty container on the pan
and press Tare. Zero value will then display.
3. Press Function key to begin. [rEAdY] will be displayed.
4. Place samples to be weighed on the pan.
5. The stable value will be held on the display. The
Hold icon will blink.
Figure 6
Beaker Stand
Preparation
Air Weight Wet Weight

PAGE 9 OF 17 0921 IH-9560
MENU SETTINGS
The User Menu allows the customizing of scale settings. NOTE: Additional sub-menus may be available
if Interface Options are installed.
MENU NAVIGATION
SUB-MENUS C.A.L S.E.T.U.P M.O.D.E U.N.I.T* E.N.D
Tare
Menu
Exit
Yes
No/Back
Exit
Long Press –
Enter Menu
Short press –
Accept
Short press –
Go Next/
Prev. menu
Short press –
Exit menu
Span
Lin
Reset
Yes/No
Auxiliary
Graduation*
On/Off
Filter
Low/Med/High
AZT
0.5/1/3/Off
StableRange
0.5/1/2//5
Backlight
Off/On/Auto
Auto Tare
Off/On/On-acc
Auto Off
Off/1/5/10
Reset
Yes/No
Weighing
On/Off
Counting
On/Off
Percent
On/Off
Check
On/Off
Totalize
On/Off
Density
On/Off
Hold
Disp/Peak/Off
Reset
g
kg
ct
oz
ozt
dwt
lb
grn
hkt
sgt
twt
t
tola/tical
End End End End End End
*Available units may vary by local regulations.
**Available settings may vary by local regulations.
When the security switch is in locked position
(See Figure 7), the menu settings are affected as follows:
• External calibration function is hidden.
• The readability will be reduced by a factor of 10
or brackets will be displayed around the last digit,
depending on the model.
• The auxiliary graduation menu will be locked.
• Units are locked to the current setting.
• Stable range locked to 1d.
• AZT locked to 0.5d.
• Zero range forced to 2%.
• Filter level is locked to current setting.
• Stable only when Print menu locked ON.
• Continuous/Interval print cannot be selected.
Figure 7
Unlocked Locked

PAGE 10 OF 17 0921 IH-9560
MENU SETTINGS CONTINUED
CALIBRATION MENU
ENTER THIS MENU TO PERFORM CALIBRATIONS.
Internal Calibration Perform
Calibration Adjustment -100…0…100
Span: Perform
Linearity: Perform
End Calibration: Exit menu
SPAN [SPAN]
Initiates a span calibration procedure (zero and span).
LIN [LIN]
Initiates a linearity calibration procedure (zero,
mid-point and span).
END CAL [End]
Advance to the next menu or return to the top of the
current menu.
SETUP MENU
ENTER THIS MENU TO SET SCALE PARAMETERS.
Reset no, yes
Auxiliary Graduation On, Off
Filter Low, Med, High
Auto Zero Tracking off, 0.5d, 1d, 3d
Stable 0.5d, 1d, 2d, 5d
Backlight off, on, auto
Auto Tare off, on, on-acc
Auto Off off, 1, 5, 10
End Setup Exit menu
NOTE: Bold always represents factory default.
RESET [rESEt]
Reset the setup menu to factory defaults.
• NO = Not reset
• YES = Reset
AUXILIARY GRADUATION [Au.GrAd]
Set whether the weight is displayed with the auxiliary
graduation when the security switch is in the locked
position.
• ON = The auxiliary graduation is displayed within
brackets.
• OFF = The auxiliary graduation is not displayed.
NOTE: If the scale will be used in direct sales to
the public, the OFF setting must be selected.
FILTER
Set the amount of signal filtering.
• LOW = Less stability, faster stabilization time
• MED = Normal stability, stabilization time
• HI = Greater stability, slower stabilization time
AUTO ZERO TRACKING
Set the automatic zero tracking functionality.
• Off = Disabled
• 0.5d = The display will maintain zero until a change
of 0.5 divisions per second has been exceeded.
• 1d = The display will maintain zero until a change of
one division per second has been exceeded.
• 3d = The display will maintain zero until a change of
three divisions per second has been exceeded.
STABLE RANGE
Set the amount the reading can vary while the stability
symbol remains on.
• 0.5d = 0.5 scale division
• 1d = 1 scale division
• 2d = 2 scale division
• 5d = 5 scale division
BACKLIGHT
Sets backlight functionality.
• Off = Always off
• On = Always on
• Auto = Turns on when a button is pressed or the
displayed weight changes.
NOTE: When powered by AC adapter, the
backlight is always on.
AUTO TARE
Sets the automatic tare functionality.
• Off = Automatic tare is disabled.
• On = The first stable gross weight is tared.
• On-acc = Stable gross loads within the accept limits
are tared (in Check Weighing mode).

PAGE 11 OF 17 0921 IH-9560
MENU SETTINGS CONTINUED
AUTO OFF
Sets the automatic shutoff functionality.
• Off = Disabled
• 1 = Powers off after one minute of no activity.
• 5 = Powers off after five minutes of no activity.
• 10 = Powers off after 10 minutes of no activity.
END SETUP
Advances to the next menu or return to top of current
menu.
MODE MENU
THIS MENU ACTIVATES MODES SO THEY WILL BE
AVAILABLE FOR USE WITH THE MODE BUTTON.
Reset no, yes
Weigh off, on
Count off, on
Percent off, on
Check off, on
Totalization off, on
Density off, on
Hold Disp, Peak, Off
End Mode Exit menu
RESET [rESEt]
Reset the Mode menu to factory defaults.
• No = Not reset
• Yes = Reset
WEIGH [WEIGH]
Set the status.
• Off = Disabled
• On = Enabled
COUNT [Count]
Set the status.
• Off = Disabled
• On = Enabled
PERCENT [PErcnt]
Set the status.
• Off = Disabled
• On = Enabled
CHECK [CHECk]
Set the status.
• Off = Disabled
• On = Enabled
TOTALIZATION [totAL]
Set the status.
• Off = Disabled
• On = Enabled
DENSITY [SPEC.Gr]
Set the status.
• Off = Disabled
• On = Enabled
HOLD [HoLd]
Set the sub-mode.
• Off = Disabled
• Peak Hold = Allows the user to capture and store the
highest stable weight value (>=5d).
• Display Hold = Allows the user to capture and store
the first stable weight value (>=5d).
When the stable value is being held on the display,
the "Hold" icon will blink and the displayed weight will
not change. Advance to the next menu or return to the
top of the current menu.

PAGE 12 OF 17 0921 IH-9560
UNIT MENU
This menu activates units so they will be accessible with
the Units button. The units in the menu must be turned
"On" to be active.
NOTE: Available units vary by model and local
regulations.
goff, on
kg off, on
ct off, on
oz off, on
ozt off, on
dwt off, on
lb off, on
gm off, on
hkt off, on
sgt off, on
twt off, on
toff, tola, tical
End Unit Exit menu
MENU SETTINGS CONTINUED
ADDITIONAL FEATURES
WEIGH BELOW HOOK
Unit is equipped with a weigh below hook for weighing
below the scale.
The weigh below hook is located at the reverse side of
the battery cover as shown below). To use this feature,
remove the red protective cover on underside of scale
for the weigh below opening. (See Figure 8)
ATTENTION! Before turning the scale over,
remove the pan and pan support (if present)
and set the transportation lock to "locked"
position.
The scale can be supported using lab jacks or any other
convenient method. Ensure the scale is level and secure
and that the transportation lock has been released.
Power on the scale and use a string or wire to attach
items to be weighed.
CONNECTING THE INTERFACE
Use an optional interface connectivity kit to connect the
scale to a computer, printer or OHAUS auxiliary display.
Interface kit accessories are available:
RS232, USB Host, USB Device, Ethernet, Bluetooth®*.
Interface kits may vary according to local regulations.
The Bluetooth® word mark and logos are registered
trademarks owned by Bluetooth SIG, Inc. and any use of
such marks by OHAUS is under license.
NOTE: Bluetooth kit is only available in certain
regions according to the local regulations.
Figure 8
Weigh Below
Protective Cover Weigh Below Hook

PAGE 13 OF 17 0921 IH-9560
MAINTENANCE
CLEANING
WARNING! Electric shock hazard. Disconnect
the equipment from power supply before
cleaning.
The housing may be cleaned with a cloth dampened
with a mild detergent if necessary.
ATTENTION! Do not use solvents, chemicals,
alcohol, ammonia or abrasives to clean the
housing or control panel.
TROUBLESHOOTING
SYMPTOM POSSIBLE CAUSE
Cannot turn on No power to scale.
Poor accuracy Improper calibration.
Unstable environment.
Cannot calibrate Security switch in locked position.
Err 8.1 Weight exceeds power on zero range.
Err 8.2 Weight below power on zero range.
Err 8.3 Overload (weight exceeds rated capacity).
Err 8.4 Underload (pan removed).
Err 8.5 Tare out of range.
Err 8.6 Displayed value exceeds 999999 (possible
in Totalization mode).
Err 8.7 Unknown internal weight position.
rEF.Err Parts counting or percentage error –
sample weight <1d. Scale shows error, then
exits parts counting or goes to [Clr.APU].
Lo.rEF Percent Reference weight or APW is too low
for accurate results.
CAL E Failure to calibrate.
USb.Err Cannot find menu or app file in U-disk.
SERVICE INFORMATION
If the troubleshooting section does not resolve or
describe problem, contact Uline Customer Service at
1-800-295-5510.
ACCESSORIES
DESCRIPTION ITEM NUMBER
RS232 kit 30268982
USB Host kit 30268983
USB Device Kit 30268984
Bluetooth Kit 30268985
Ethernet Kit 30268986
Stacking Kit, x6 30268987
Stacking Kit, x1 30268988
Specific Gravity kit 30269020
Auxiliary Display Kit 30269019
Carrying Case 30269021
In-Use Cover 30269022
Printers and Cables Contact Uline

PAGE 14 OF 17 0921 IH-9560
LEGAL FOR TRADE
TECHNICAL DATA
When the scale is used in trade or a legally controlled
application it must be set up, verified and sealed
in accordance with local weights and measures
regulations. It is the responsibility of the purchaser to
ensure that all pertinent legal requirements are met. As
the requirements vary by jurisdiction, the purchaser is
advised to contact local weights and measures office
for instructions about putting the scale into service.
SETTINGS
Before verification and sealing, perform the following
steps:
1. Verify that the menu settings meet the local weights
and measures regulations.
2. Units menu should be reviewed. Verify the units
turned on are permitted by the local weights and
measures regulations.
3. Perform a calibration.
4. Set the security switch to the locked position.
VERIFICATION
A weights and measures official must perform the
verification procedure. Contact the local weights and
measures office for more information.
SEALING
After the unit has been verified, it must be sealed by the
weights and measures official to prevent undetected
access to the legally controlled settings. Before sealing
the device, ensure that the security switch is in the
locked position. (See Figure 9)
If using a wire seal, pass the sealing wire through the
holes in the security switch and bottom housing.
(See Figure 9)
If using a paper seal, place the seal over the security
switch and bottom housing. (See Figure 9)
Figure 9
Unlocked
Locked With Paper Seal
Locked With Wire Seal
Equipment Ratings: Indoor use only
Altitude: Up to 6,560 ft
Operating temperature range: 41°F to 104°F
Specified temperature range: Refer to tables on
page 15
Relative humidity: 10% to 80% at 88°F, decreasing
linearly to 50% at 104°F, non-condensing
Power: AC power adapter input 100-240V 50/60 Hz and
output 5 V DC 1 A (for use with certified or approved
power supply, which must have a SELV and limited
energy circuit output), or 4 AA batteries
Main supply voltage fluctuations: Up to +/- 10% of the
nominal voltage
Pollution degree: 2
Installation category: II

PAGE 15 OF 17 0921 IH-9560
TECHINICAL DATA CONTINUED
NTEP AND MEASUREMENT CANADA APPROVED MODELS:
MODEL H-9560 H -9561
Capacity 620 g 1,500 g
Readability 0.1 g 0.01 g or 0.1 g
Verification Interval e 0.1 g 0.1 g
Class III II
Span Calibration Mass 300 g 1.5 kg
Linearity 300 g, 600 g 1 kg, 1.5 kg
Capacity x Readability (kg) / 1.5 x 0.00001 or 1.5 x 0.0001
Capacity x Readability (ct) 3,100 x 0.5 7,500 x 0.5
Capacity x Readability (oz) 21.870 x 0.005 52.910 x 0.005
Capacity x Readability (ozt) 19.935 x 0.005 48.225 x 0.005
Capacity x Readability (dwt) 398.7 x 0.1 964.52 x 0.01 or 964.5 x 0.1
Capacity x Readability (lb) 1.3670 x 0.0005 3.3070 x 0.0005
Capacity x Readability (grn) 9,568 x 2 23,148 x 2
Stabilization Time (s) 11.5
Construction ABS housing and stainless steel pan
Draftshield No
Calibration User-selectable external span or linearity calibration
Tare Range Full Capacity by subtraction
Weighing Units* g, kg, ct, oz, ozt, dwt, lb, grn
Application Modes* Weighing, Parts Counting, Percent Weighing, Check Weighing, Totalization,
Density Determination, Display Hold
ypical Battery Life 80 hours
Specified Temperature Range Class II: 50°F to 86°F
Class III: 50°F to 104°F
Storage Conditions -4°F to 131°F at 10% to 90% relative humidity, non-condensing
Communication RS232, USB Host, USB Device, Ethernet or Bluetooth**
(all available as accessory)
Display Type Backlit LCD: 6-digit 7-segment with white LED backlight
Display Size 0.78 in / 20 mm digits
Pan Size (W x D) Ø120 mm / 4.7 in 170 x 140 mm / 6.7 x 5.5 in
Scale Dimensions
(W x D x H) 202 x 224 x 54 mm / 8 x 8.8 x 2.1 in
Shipping Dimensions
(W x D x H) 300 x 250 x 86 mm / 11.8 x 9.8 x 3.4 in
Net Weight 2.2 lb / 1 kg
Shipping Weight 3.3 lb / 1.5 kg
* Available Weighing Units and Application Modes vary by local regulations.
** Bluetooth kit is only available in certain regions according to the local regulations.

PAGE 16 OF 17 0921 IH-9560
TECHINICAL DATA CONTINUED
DRAWINGS
MODEL A B C
H-9560 202 mm / 8.0 in. 222 mm / 8.7 in. 103 mm / 4.1 in.
H-9561 202 mm / 8.0 in. 224 mm / 8.8 in. 54 mm / 2.1 in.
B
A
B
A
CC
COMPLIANCE
Compliance to the following standards is indicated by the corresponding mark on the product.
MARK STANDARD
This product conforms to the applicable harmonized standards of EU Directives 2011/65/EU (RoHS),
2014/30/EU (EMC), 2014/35/EU (LVD) and 2014/31/EU (NAWI). The complete Declaration of Conformity is
available online at www.ohaus.com/ce.
This product complies with the EU Directives 2012/19/EU (WEEE) and 2006/66/EC (Batteries). Please
dispose of this product in accordance with local regulations at the collecting point specified for
electrical and electronic equipment. For disposal instructions in Europe, refer to
www.ohaus.com/weee.
AS/NZS 61000.6.1, AS/NZS 61000.6.3
C US
MC173467
CAN/CSA-C22.2 No. 61010-1, UL Std. No. 61010-1

PAGE 17 OF 17 0921 IH-9560
TECHNICAL DATA CONTINUED
Weighing Instruments verified at the place of
manufacture bear the following supplementary
metrology marking on the descriptive plate.
Weighing Instruments to be verified in two stages have
no supplementary metrology marking on the descriptive
plate. The second stage of conformity assessment must
be carried out by the applicable weights and measures
authorities.
If national regulations limit the validity period of the
verification, the user of the weighing instrument must
strictly observe the re-verification period and inform the
weights and measures authorities.
As verification requirements vary by jurisdiction, the
purchaser should contact local weights and measures
office if they are not familiar with the requirements.
FCC COMPLIANCE STATEMENT
NOTE: This equipment has been tested and
found to comply with the limits for a Class B
digital device, pursuant to Part 15 of the
FCC Rules. These limits are designed to
provide reasonable protection against harmful
interference in a residential installation.
This equipment generates, uses and can
radiate radio frequency energy and, if not
installed and used in accordance with the
instructions, may cause harmful interference
to radio communications. However, there is
no guarantee that interference will not occur
in a particular installation. If this equipment
does cause harmful interference to radio or
television reception, which can be determined
by turning the equipment off and on, the user
is encouraged to try to correct the interference
by one or more of the following measures:
• Reorient or relocate the receiving antenna.
• Increase the separation between the
equipment and receiver.
• Connect the equipment into an outlet on
a circuit different from that to which the
receiver is connected.
• Consult the dealer or an experienced
radio/TV technician for help.
Changes or modifications not expressly approved by
the party responsible for compliance could void the
user's authority to operate the equipment.
ISED CANADA COMPLIANCE STATEMENT
This Class B digital apparatus complies with Canadian
ICES-003.
ISO 9001 REGISTRATION
The management system governing the production of
this product is ISO 9001 certified.
MXX 1259
1
-800-295-551
0
uline.com
Table of contents
Other U-Line Scale manuals

U-Line
U-Line OHAUS VALOR 3000 User manual

U-Line
U-Line OHAUS CS H-2715 User manual

U-Line
U-Line H-1114 User manual

U-Line
U-Line H-2382 User manual

U-Line
U-Line H-9983 User manual

U-Line
U-Line OHAUS DEFENDER 5000 User manual

U-Line
U-Line H-1114 User manual

U-Line
U-Line H-5836 User manual

U-Line
U-Line H-2289 User manual

U-Line
U-Line H-479 User manual

U-Line
U-Line H-3677 User manual

U-Line
U-Line H-670 User manual

U-Line
U-Line OHAUS DEFENDER 3000 User manual

U-Line
U-Line H-4602 User manual

U-Line
U-Line H-1096 User manual

U-Line
U-Line H-3478 User manual

U-Line
U-Line H-1679 User manual

U-Line
U-Line H-2714 User manual

U-Line
U-Line H-5819 User manual

U-Line
U-Line OHAUS SCOUT H-9596 User manual