UniPix ATMOS Nano User manual

UniPix&&&&&Atmos&ATMOS&Nano&Operation&manual&&&&&Rev.&1.01&page&1/1&
ATMOS Nano
Pipette Accuracy Tester
P1 – P20
!
!
Operation!Manual!Rev.!1.01!
!
!!!!!!!!!!!!!!!!!! !
!
!
Designed&in&Switzerland&
Assembled&in&Switzerland&
!
!
UniPix,!Switzerland!
Web:!www.unipix-lab.com! ! Email:!info@unipix-lab.com!
! !

UniPix&&&&&Atmos&ATMOS&Nano&Operation&manual&&&&&Rev.&1.01&page&2/6&
Disclaimer
• The! Pipette! Accuracy! Tester!ATMOS!is!
designed! for! the! specific! use! of! testing! and!
calibrating!air!displacement!laboratory!pipette!
and!no!other!use!is!recommended.!
• The! user! must! read! and! understand! this!
manual!before!operation!
• This!instrument!creates!items!that!should!only!
be!used!in!a!laboratory!environment.!
• Do! not! open! or! attempt! to! repair! the!
instrument! without! expressed! and! explicit!
instructions!from!UniPix.!!
• Don't! use! the! instrument! in! an! atmosphere!
with!a!danger!of!explosion.!
• Do! not! allow! unauthorized! and! or! untrained!
operatives!to!use!this!device.!
• Any!misuse!will!be!the!sole!responsibility!of!the!
user/owner!and!UniPix!assumes!no!implied!or!
inferred! liability! for! direct! or! consequential!
damages!from!this!instrument!if!it!is!operated!
or! used! in! any! way! other! than! for! which! it! is!
designed.!
!
!
!
Function
The! UniPix! Atmos! Pipette! Accuracy! Tester!is! an!
instrument! for! precise! measurement! air!
displacement! pipette,! without! need! of! liquid.! It! is!
based!on!comparison!with!a!calibrated/internal/
reference/ cavity!in! the! instrument.! It! also!
measures!the!leak!rate!of!pipettes.!!
!
!
Never/use/the/
instrument/with/liquid/
in/the/pipette/tip./!
This/will/permanently/damage/the/instrument./
Installation
The!package!contains!the!UniPix!Atmos!unit,!the!AC!
adapter! and! the! USB! cable! and! the! card!with! QR!
code!to!access!the!UniPix!ATMOS!web!page.!
!
!
!
!
!
Operation
Press! the! main! switch! on! the! back!of! the!
instrument!to!power!on.!
It! is! recommended! to! verify! that! the! instrument!
battery! is! well! charged! or! use/ the/ instrument/
with/ the/ external/ power/ supply!connected! via!
the!USB!cable.!
The! instrument! is! powered! by! an! internal! Li-ion!
battery!which! allows! continuously! operating!
during!about!1!hour.!
After!powering!on,!the!instrument!runs!a!self-test!
procedure! for! about! 10! seconds.!The! internal!
ventilator!turns!on!and!create!a!small!noise.!
!
Note:&The&instrument&goes&on&sleep&mode&if¬&used&
during&more&than&30&second.&Any&touch&to&the&screen&
reactivates&the&instrument.&
!
!

UniPix&&&&&Atmos&ATMOS&Nano&Operation&manual&&&&&Rev.&1.01&page&3/6&
Pipette nominal size selection:
ATMOS! Nano! is! designed! for! testing! fixed! and!
variable!volume!pipettes!form!P1!to!P20!size.!
!
Enter! the! tested! dispense! volume! with! the! touch!
screen!and!then!press!"OK".!
The!"C"!button!erases!the!entered!number.!
!
!
!
Notes:&
• It&is&possible&to&run&the&test&without&entering&
a&volume.&
&
• If&the&instrument&is&used&test&volumes&larger&
than&20&µl,&the&measurement&accuracy&is¬&
valid.&
!
!
Tip insertion:
Insert!the!pipette!tip!in!the!center!of!the!tip!port!of!
the!instrument.!Hold!the!pipette!vertical!and!firmly!
press!the!tip!on!the!port.!!
!
!!
When!the!tip!is!well!inserted!on!the!port,!press!de!
GO!arrow,!without!moving!the!pipette.!
Notes:&
• Only&use&new&and&clean&tips&
• Never&use&tips&that&contains&a&liquid.&
• Don't&move&the&pipette&piston&
• Keep& the& pipette& in& stable& position& during&
the&testing&
!
The! arrow! "ID"!at! the!bottom! right! of! the! screen!
allows!to!enter!date!and!pipette!ID!for!USB!transfer!
of!calibration!data!(see!"interfacing"!section).!
!
!
!
!

UniPix&&&&&Atmos&ATMOS&Nano&Operation&manual&&&&&Rev.&1.01&page&4/6&
Measurement:
The!measurement!is!done!in!two!phases:!!
1. piston!pressed!(keep!about!5!seconds)!
2. piston!released!
!
!
!
1: Press phase:
Press!on! the! pipette! piston,! while! firmly! keeping!
the!tip!inserted!in!the!port.!The!displacement!of!the!
should!be!done!in!about!1!to!2!seconds.!
!
The!progress/bar!at!the!bottom!indicates!that!the!
instrument! is! measuring.! The/ pipette/ and/ the/
piston/ should/ not/ move/ during/ the/
measurement.!The!measurement!time!is!typically!
five!seconds.!
2: Release phase:
On! the! display! of! this! new! screen,! release!the!
piston,!but!don’t!remove!the!tip.!
!
Results:
The!calibration!result!is!displayed!at!the!end!of!the!
measurement!process.!
Displayed!parameters:!
Volume:!!nominal!pipette!volume!
Temperature:!!internal!temperature!
Value:!! ! measured!displacement!volume!
Leak:!! ! pipette!leak!rate!
Nb!Meas.:!!number!of!conducted!calibrations!!
!
!
!
Should!something!go!wrong!with!the!
measurement,!the!message!"BAD/READING"!
appears!on!the!screen.!By!pressing!continue,!you!
restart!the!measurement.!
!
!
!
The!reasons!for!bad!reading!can!be:!
• Bad!insertion!of!the!tip!into!the!port!
• User!not!following!instructions!during!the!
calibration!process!
• A!totally!nonfunctional!pipette!
• A!very!leaky!pipette!or!tip!connection!
!
Release
Bad reading
Please repeat
Continue

UniPix&&&&&Atmos&ATMOS&Nano&Operation&manual&&&&&Rev.&1.01&page&5/6&
Statistics:
The!measurements!can!be!repeated!several!times!
in!order!to!refine!statistical!accuracy.!!
Press! NEXT!for! repeating! the! measurement.! The!
previous!results!are! stored! in! the! memory! of! the!
instrument! for! later! statistical! calculations.! The!
repeat!number!is!indicated!on!the!screen.!
Press! END!to! go! the! statistics.! The! statistics! is!
calculated! only! with! a! minimum! of! 3! consecutive!
measurements.!
!
!!
!
Press! RESTART!for! testing! another! dispense!
volume!or!start!with!another!pipette!size.!
!
!
!
!
ISO& 8655-2& tolerances& for& variable& volume& air&
displacement&pipettes&P1&to&P10&:&
&
!
Systematic!error!(a)!
Random!error!(b)!
P10!
±!0.12!µl!
±0!08!µl!
P5!
±!0.125!µl!
±!0.075!µl!
P2!
±!0.08!µl!
±!0.04!µl!
P1!
±!0.05!µl!
±!0.05!µl!
&
(a)& Expressed& as& the& deviation& of& the& mean& of& a& tenfold&
measurement&from&selected&volume&
(b)& Expressed& as& the& coefficient& of& variation& of& a& tenfold&
measurement&
!
!
!
!
!
Notes!about!calibration:!
!
I. It! is! recommended! to! take! at! least! 5!
measurements!in! order! to! have! good!
statistics!for!error!and!standard!deviation.!
!
II. After! moving! ATMOS! or! if! you! bring!
pipettes! from! another! room,! wait! 5!
minutes!before!making!the!calibration.!The!
system! is! sensitive! to! quick! temperature!
changes.!
!
III. ATMOS! in! compensated! for! temperature.!
The! calibration! values! correspond! to! the!
equivalent! gravimetric! measurements!at!
20°C!.!
!
IV. ATMOS! is! optimized! for! standard! conical!
tips.! If! you! perform! gravimetry!
measurements! with! long! tip! or! other!
special!shape,!the!values!related!to!ATMOS!
are!no!longer!as!accurate.!
!
!

UniPix&&&&&Atmos&ATMOS&Nano&Operation&manual&&&&&Rev.&1.01&page&6/6&
Interfacing ATMOS:
ATMOS! was! conceived! to! allow! transfer! of! the!
captured! results! to! any! computer! (Window,! Mac,!
Linux)! using! the! appropriate! software.!ATMOS!
microUSB! is! connected! to! the! industry! standard!
CP2102!from!Silicon!labs.!
!
Connection!is! read! as! a! serial! port! and!
communication!is!done!at!115200!bauds!using!the!
standard!ASCII!code.!
Commands!are!sent!as!a!single!ASCII!character!and!
will!be!interpreted!by!the!ATMOS!internal!parser.!
Upon! reception! of! a! meaningful! command! the!
device!will!issue!a!reply!followed!by!<CR/LF>!
!
Command/summary:/
The! first! set! of! commands! is! for! downloading!
measurement!data!from!Atmos!:!
l!Atmos! sends! a! string! of! characters! listing!
all! measurements! data! stored! in! the!
memory!
w!Erase!all!data!stored!in!the!memory!
z!Toggle! the! verbose! mode! on! and! off.! The!
verbose!mode!automatically!returns!all!the!
measurement! data! after! a! calibration!
sequence!
h!Help!command:!this!will!send!the!list!of!all!
available!commands!
!
The! second! set! of! commands! is! for! remote!
controlling! Atmos! for! operation! in! a! robotic!
platform!:!
a!After! receiving! this! command,! Atmos! will!
start!a!calibration!measurement!sequence!
with!the!stored!sequence!parameters.!The!
sequence!parameters!can!be!changed!with!
the!command!below.!(note:&this&function&is&
not&yet&fully&implemented)!
c!Enter!the!calibration!mode!
Soon!after!issuing!the!“c”!command,!Atmos!
will!send!the!single!character!“<”!indicating!
that! it! is! ready! to! read! the! pressure!
increase.! The! calibration! process! takes!
place!and!when!completed!Atmos!send!the!
single! character! “>”! indicating! that! the!
pressure!can!be!released.!
t!After! receiving! this! command,! Atmos! will!
read! a! string! of! character! terminated! by!
<CR/LF>! used! to! set! the! timing! of! the!
calibration! sequence! (values! should! be!
between!2-10)!
v!Set! the! nominal! displacement! volume! (in!
microliters)! that! is! planned! to! be! tested.!
Enter!a!valid!integer!followed!by!<CR/LF>!
p!After! receiving! this! command,! Atmos! will!
read! a! string! of! character! terminated! by!
<CR/LF>! used! for! the! identification!
number!(max.!8!digits)!of!the!pipette!to!be!
calibrated!
d!After! receiving! this! command,! Atmos! will!
read! a! string! of! character! terminated! by!
<CR/LF>!used!to!set!the!date!in!format!dd-
mm-yy!(i.e.!!24-02-2019).!(note:&the&validity&
of&the&entry&is¬&checked).!
Maintenance
Cleaning
The! instrument! can! be! cleaned! with! a! dry! cloth.!
Never!spill!liquid!on!the!instrument.!
Dust! present! on! the! tip! port! might! be! source! of!
leakage!and!lead!to!improper!measurements.!The!
dust! should! be! carefully! removed! with! a! slightly!
humid!cloth.!Never!use!a!tweezer!because!this!can!
damage!permanently!the!port.!
Replacing the joint of the tip port
The!joint!of!the!tip!port!is!designed!to!work!several!
years!without!maintenance.!However,!in!the!case!of!
damage,!the!join!can!be!replaced!by!the!user.!Please!
carefully!follow!these!instructions:!
• Before! changing! the! joint,! order! the!
original!replacement!part!from!UniPix!web!
site!
• Unscrew!the!port!
• Remove!the!elastomer!joint!which!is!inside!
• Replace!by!the!new!joint!
• Tighten! the! port! (recommended! torque:!
0.2!Nm)!
!
ONLY!USE!the!elastomer!joint!provided!by!
UniPix!
A!video!tutorial!is!available!on!UniPix!web!site.!!
Complementary! information! on! maintenance! and!
FAQ!can!be!found!on!the!UniPix!web!site:!
www.unipix-lab.com!
!

UniPix&&&&&Atmos&ATMOS&Nano&Operation&manual&&&&&Rev.&1.01&page&7/6&
Specifications
ATMOS Performance
!
Absolute&accuracy&(=&trueness)&
P20!
20!µl!
0.2!
P10!
10!µl!
±0.1µl!
P!5!
5!µl!
±0.1µl!
P2!
2!µl!
±0.06µl!
P1!
1!µl!
±0.05µl!
!
Repeatability&(=&precision)&
P20!
20!µl!
0.1!
P10!
10µl!
±0.06µl!
P!5!
5!µl!
±0.06µl!
P2!
2!µl!
±0.04µl!
P1!
1!µl!
±0.05µl!
!
Technical specs
Dimensions! ! 14.5!!x!!10!!x!!5.5!cm!
Weight!! ! 820!g!
Power!supply! ! 5V!DC,!2A!
Battery!! ! 2000!mAh,!Li-ion!
Operational!temp.!from!10!°C!to!40°C!
Altitude! ! from!sea!level!up!to!3500m!
Support
Technical!support! is!provided!by!UniPix!by! Email!
at!info@unipix-lab-com.!
!
!
!
!
!
!
!
Troubleshooting
Problem/
Possible/
cause/
Solution/
No!power!
Low!battery!
Connect!to!
power!
supply!
!
Instrument!on!
sleep!mode!
Touch!the!
screen!
Ventilator!
intermittent!
during!
measurements!
Battery!is!too!
low!
Connect!to!
power!
supply!
Not!
reproducible!
measurement!
Bad!
manipulation!
Refer!to!
user's!
manual!for!
proper!
operation!
!
Dirty!of!
damage!tip!
port!
Clean!or!
replace!tip!
port!joint!
!
Very!bad!
pipette!
Try!another!
pipette!
Warranty
UniPix!shall! not! be!retained!liable! for! any!
consequences!of!improper!handling,!use,!servicing,!
operating! or! unauthorized! repairs! of! the!
instrument! or! the! consequences! of! normal! wear!
and! tear! as! well! as! the! failure! to! follow! the!
instructions!of!the!operating!manual.!UniPix!shall!
not! liable! for! damage!resulting! from! any! actions!
other!than!those!described!in!the!operating!manual!
or! following! the! use! of!non-original! replacement!
parts.!
Disposal
The!adjoining!symbol!means!that!storage!batteries!
and! electronic! devices! must! be! disposed! of!
separately!from!household!trash!(mixed!municipal!
waste)!at!the!end!of!their!service!life.!
!