
HMP45A and HMP45D
U274en-1.2 Operating Manual
3
the ambient relative humidity and temperature level with the reference probe
head, and then connect the HMP45A probe head on the HMH45 handle and
check the output readings.
Prepare the HMI41 indicator for measurements with the HMP45A probe head
as follows: turn the HMI41 indicator on and within 1-2 seconds press
simultaneously buttons ENTER and MODE until the text SEtUP appears on
the display.
Wait a few seconds, and the text Unit appears; press ENTER repeatedly until
a text similar to the following appears on the display:
set
Select PROBE TYPE 3with buttons ▲(number up) or ▼(number down) and
turn the indicator off. Turn it on again by pressing the ON/OFF button. After a
few seconds, the relative humidity and temperature readings appear
automatically on the display. The HMP45D cannot be used with the HMI41
indicator as the passive temperature signal results in error messages on
the HMI41.
3.2 Humidity calibration
For a high-accuracy two-point calibration use a Vaisala HMK15 or HMK13B
calibrator and saturated salt solutions as described in the respective manuals.
Leave the calibrator, the HMI41 and the probe head in the same space for at
least four hours so that their temperatures have time to equalize. Unscrew the
plastic grid of the probe.
The calibration is done first for the dry end and then for the wet end by
adjusting trimmer potentiometers marked “D” (dry, <50 %RH) and “W” (wet,
>50 %RH). The potentiometers are located under a protective plug; see Figure
2. Use a ceramic screw driver with 2.5 mm blade for adjusting the
potentiometers. Note: if zero point is calibrated in Nitrogen (N2), the
minimum output signal of 0.008 V corresponds to a relative humidity of 0.8
%RH.
Greenspan’s calibration table
Temperature °C 15 20 25 30 35
LiCl %RH *) 11.3 11.3 11.3 11.3
NaCl %RH 75.6 75.5 75.3 75.1 74.9
K2SO4%RH 97.9 97.6 97.3 97.0 96.7
*) Do not use or store the LiCl solution in temperatures below +18°C as its humidity
equilibrium may change permanently