GMM220 Series CO2 Transmitter Modules
M010115EN-D
4
TECHNICAL DATA
Measuring ranges
GMM221 0 … 2%, 0 … 3%, 0 … 5%, 0 … 10% , 0 … 20% CO2
GMM222 0 … 2000 ppm, 0 … 3000 ppm, 0 … 5000 ppm,
0 … 7000 ppm, 0 … 10 000 ppm
Accuracy at 25°C against
certified factory references
GMM221 ±[1.5 % of range + 2 % of reading]
GMM222 ±[1.5 % of range + 2 % of reading]
(including repeatability and calibration uncertainty)
Nonlinearity ±0.5 % FS
Temperature dependence
of output (typically) -0.1 %FS / °C1)
Pressure dependence (typic.) +0.15 of reading/hPa2)
1)When temperature rises, the output decreases, default 25 °C (77 °F).
2)When pressure rises, the output increases, default 1013 hPa (1 atm).
Long-term stability <±5 %FS / 2 years
Response time (63%) 20 seconds (GMM221)
30 seconds (GMM222)
Operating conditions
Operating temperature range -20…+60 °C
Humidity range
probe 0 … 100 %RH (non-condensing)
mother board 0 … 85 % RH (non-condensing)
General
Analog outputs 0 … 20 mA or 4 … 20 mA and
0 … 1 V or 0 … 2 V, 0 … 2.5 V, 0 … 5 V
Resolution of analog outputs 0.03 %FS
Recommended external load:
current output max. 200 Ω
voltage output min. 1kΩ
Power supply 11 … 20 VDC or 18 … 30 VDC
Serial output @ 5V level
Power consumption < 2.5 W
Warm-up time < 5 minutes
Probe cable length 0.6 m, 2 m, 6 m, 10 m
Mechanics
Probe:
housing material PC plastic
housing classification IP65/NEMA4
Weight:
GMM221(w/ 2m cable) max. 180 g
GMM222 (w/ 2m cable) max. 200 g
GUARANTEE
Vaisala issues a guarantee for the material
and workmanship of this product under
normal operating conditions for one (1)
year from the date of delivery.
Exceptional operating conditions, damage
due to careless handling and
misapplication will void the guarantee.
COMPLIANCE WITH THE EMC-
STANDARDS
EN 61326-1:1997 + Am1:1998, Electrical
equipment for measurement, control and
laboratory use - EMC requirements-
Generic environment.
Testing: CISPR16/22 class B, EN/IEC
61000-4-2, EN/IEC 61000-4-3, EN/IEC
61000-4-4, EN/IEC 61000-4-5, EN/IEC
61000-4-6.
Accessories
Order code Description
GMP221,
GMP222
Spare probe
25378GMSP Spare filter for GMP221
25879GMSP Spare filter for GMP222
25245GM Clips (2 pcs) for attaching
the probe
GM45168SP Protective sleeve for
GMP221
GM45237SP Protective sleeve for
GMP221
26150GM Field check adapter
GM45156 Mounting flange for the
probe
19040GM Serial COM adapter
VAISALA SERVICE CENTERS
NORTH AMERICAN SERVICE CENTER
Vaisala Inc., 100 Commerce Way, Woburn, MA 01801-1068, USA.
Phone: +1 781 933 4500, Fax +1 781 933 8029
EUROPEAN SERVICE CENTER
Vaisala SSD Service, Vanha Nurmijärventie 21 FIN-01670 Vantaa,
FINLAND.
Phone: +358 9 8949 2758, Fax +358 9 8949 2295
ASIAN SERVICE CENTER
Vaisala KK, 42 Kagurazaka 6-Chome, Shinjuku-Ku, Tokyo 162-0825,
JAPAN.
Phone: +81 3 3266 9611, Fax +81 3 3266 9610