Vapotherm 2000i Instruction Manual

2000i
Operating Instruction Manual

Section 1 Indications and Contraindications
Primary Indications:
Used to warm and humidify breathing gases, generally prescribed during oxygen therapy where concentrations
of oxygen greater than ambient air are utilized to treat symptoms and manifestations of hypoxia including:
• Documentedhypoxemia:decreasedPa02in blood below normal range
• Acutecareinwhichhypoxemiaissuspected
• Severetrauma
• Acutemyocardialinfarction
Secondary Indications:
• Managinghypothermia
• Treatingbronchospasmcausedbycoldair
Contraindications:
General:
• Anysituationsinwhichhumidificationiscontra-indicated(seeAARCClinicalPracticeGuidelines)
SpecifictoNasalCannula:
• Patientswithoccludedordefectivenaresshouldnotusethesystem.
Section 2 Definitions,WarningsandCautions
2.1 Definitions
AWARNINGindicatesthatapotentiallyharmfulsituationmayoccur.
ACAUTION indicates a condition that may lead to equipment damage, malfunction, or
inaccurateoperation.
ANOTEindicatesapointofemphasistomakeoperationmoreefficientorconvenient.
ASEPTIC TECHNIQUE is practices and procedures performed under carefully controlled conditions with
thegoalofminimizingcontaminationbypathogens.Specificallywithrespiratoryequipment,especially
withreferencetotheVapotherm2000i,thisincludesproperhandwashingandavoidingdirecthand
contactwithconnectionpoints.
Pleasetakethetimetofamiliarizeyourselfwiththedefinitions,warnings,cautionsandnoteslistedinthis
manual.Theycoversafetyconsiderations,specialrequirements,andregulations.
Theuserofthisproductassumessoleresponsibilityforanymalfunctionduetooperationormaintenance
performedbyanyonenottrainedbyVapotherm™stafforofficialtrainingdocumentation.
WhenhandlinganypartoftheVapotherm®2000i,alwaysfollowhospitalinfectioncontrolguidelinesand
StandardPrecautions.Vapothermrecommendsthatusersfollowthedisinfectionprocedurefoundinthis
manual.VapothermalsorecommendsthatusersfollowtheCentersforDiseaseControl(CDC)publications:
Guidelines for Maintenance of In-Use Respiratory Therapy Equipment and Guidelines for Prevention of
Nosocomial Pneumonia.
2.2 General Warnings
• FederalLaw(U.S.)restrictsthesaleofthisdeviceto,orbytheorderofanyphysician.
• Thisisahumidificationdevicegenerallyusedforprovidingcontinuousflowsofbreathinggas.
TheVapotherm®2000iisnotaventilatorydeviceandshouldnotbeusedaslifesupport.
• Thisdevicewillnotoperatewithoutflow.
Operating Instruction Manual Page 3Operating Instruction Manual Page 2
Contents Page
Section 1 Indications and Contraindications 3
Section 2 Definitions, Warnings and Cautions 3
Section 3 About Vapotherm™2000i Product Description 5
Section 4 General 2000i Set up and instructions
4.1 SetUp–The2000iUnit 7
4.2 SetUp–SelectingtheVaporTransferCartridge 7
4.3 SetUp–InsertingtheVaporTransferCartridge 8
4:4 SetUp–InsertingthePatientDeliveryTube 8
4:5 SetUp–VapothermSpikeSet(VSS-1) 9
4:6 SetUp–ConnecttoaGasSource 9
Section 5 Operations
5:1 Operations–PrepareforActivatingtheUnit 10
5:2 Operations–TurnonFlow 10
5:3 Operations–PrimingtheUnit 10
5:4 Operations–ActivatetheUnit 10
5:5 Operations–SettingtheTemperatureandWarm-Up 11
5:6 Operations–ConnectingtoaPatient 11
5.7 Operations–GeneralGuidelines 12
Section 6 Alarms and Troubleshooting
6.1 General 13
6.2 AlarmandTroubleshootingTable 13–14
6.3 ComponentChange-outs 15
6.3.1 ReplacingtheVaporTransferCartridge 15
6.3.2 ReplacingthePatientDeliveryTube 15
6.3.3 ReplacingtheVSS-1SpikeSet 15
Section 7 Removing From Patient and System Shut Down 16
Section 8 Routine Disinfection Procedure
8.1 DisinfectionSupplies 17
8.2 Pre-cleaningProcess 17
8.3 Pre-disinfectionCleaningandDecontamination 18
8.4 Set-up 18–19
8.5 DisinfectGasandWaterCircuits 20
8.6 Drying 21
Section 9 Specifications 22
9.2 Definitionsandsymbols 22
Section 10 Warranty Information 23
Section 11 Reference and Bibliography 24
AppendixA–DisinfectionSolution 25–26
AppendixB–SampleDisinfectionLog 27

Section 2 WarningsandCautions
General Warnings (cont.)
• TheVapotherm®2000imustbedisinfectedbetweeneachpatientuseandafter30daysofuseonthe
samepatient.
• Gasflowdeliveredbythisdeviceislimitedto40litersperminute.Maximumoperationalflowrate
shouldnotbeexceeded.
• PriortousemachineshouldbepositionedandsecuredtoasturdyIVpole.
• Thesystemincludesseveraldisposableelementsthatarelabeledassinglepatientuseonly:donot
attempttosterilizeorreuse.Followalllocalandfederalregulationsfordisposal.
• Oxygensupportscombustion;thisdeviceshouldnotbeusednearoraroundopenflames,oil,grease
oranyflammablesoranesthetics.
• Performanceverificationmustbeperformedpriortouse.
• Serviceonthedeviceshouldbeperformedbyqualified,certifiedservicetechniciansONLY.
• TopreventinjurydonotattempttoperformanyservicetotheVapotherm2000iwhileapatientis
connectedtothedevice.
• Ifthedeviceisdamagedornotworkingproperlydonotuse.ContactVapothermoryourauthorized
Vapothermrepresentative.
• Donotoperateifpowercordisdamaged.
• Thedeviceshouldnotbeturnedonandleftunattended.
• DonotusetheVapotherm2000iinoraroundwaterotherthanthewaterbagthatfeedsthesystem.
• Failuretoutilizesterilewatersupplyorcleangassupplymayincreaseriskofbacterialcontamination.
• TheVapotherm2000iutilizeswarmedwaterandcanposeariskforcolonizationofbacteriaand
patientinfectionifdisinfectionproceduresarenotfollowed.
• GasflowisexternaltotheVapotherm,butthecaregivershouldconfirmtheintegrityofallrespiratory
gasesutilizedtoensuretheyarefreeofcontamination.
• Gassupplymustbemadeofcleandrymedicalgradegastopreventharmtothepatientandprevent
damagetotheVapotherm2000i.
• Anoxygenanalyzerwithalarmsmustbeusedwhenthedeliveredconcentrationleveliscritical.The
Vapotherm2000idoesnotprovideoxygenconcentrationanalysiscapability.
• Toreduceanypotentialtransmissionofcontaminatedwaterfromthesystem,allassemblyand/or
disassemblyoftheunitshouldtakeplaceoutsidetheprimarycareareas.
• The2000iisnotaContinuousPositiveAirwayDevice(CPAP).Therearenocontrolstodeliveror
monitorairwaypressure.The2000ihumidifiesbreathinggasesthataredeliveredexternallythrough
standardair/oxygenblenderandflowmeter.The2000ishouldnotbeusedtodeliverpressureina
closedsystem.
2.3 Cautions
• Verifythatthepowersourceiscompatiblewiththeelectricalspecificationsshownoneachcomponent.
Forpropergroundingreliability,connectthe2000ipowercordonlytoaproperlymarkedhospitalgrade
receptacle.DONOTUSEEXTENSIONCORDS.Ifanydoubtexistsastothegroundingconnection,DONOT
operatethedevice.
• DonotimmersetheVapotherm2000iinwater.DonotsteamorgassterilizetheVapotherm2000i.
• Readandunderstandthismanualpriortooperatingthesystem.
• TheVapotherm2000imustbedisinfectedifthewatercircuitisopenedupbyremovingorreplacinga
component.
• Aseptictechniques(includingproperhandwashingandavoidingdirecthandcontactwithconnection
points)andStandardPrecautionsshouldalwaysbefollowedwhenhandlingmedicalequipment.
• StandardPrecautionsshouldalwaysbefollowedwhencomingincontactwithpatients.
Operating Instruction Manual Page 4
Section 2 WarningsandCautions
2.4 General Inspection
WhenunpackingtheVapotherm2000isystem,ensurethattheunitisinspectedfordamagebeforeuse.
ReportanydamageormissingpartsimmediatelytoyourauthorizedVapothermdistributor.
WhenrentingaVapotherm2000i,customersshouldrequiretherentalservicetoprovideacertification
thatthemachinehasbeendisinfectedbeforeacceptingdelivery.
Section 3 AbouttheVapotherm2000iand2000h
TheVapotherm™2000hforhomeuseconsistsofthecombinationofa2000iunitandthe“HomeCare
CompressorKit(PartnumberHCK200-M)”.TheHCK-200KitconsistsofanHS-100Stand,a5060A
RoomAirCompressorandFlowmeterassembly.
1.AVapothermauthorizedDurableMedicalEquipment(DME)Supplierisresponsibleforthefollowing:
Operating Instruction Manual Page 5
2.Theuser/homecareproviderisresponsibleforthefollowing:
• Assemblingthe2000iunitandthekitcomponents
• Instructing the user in their responsibilities for operating the system
• Providing1000mlsterilewaterbagsasneeded
• Routineservicingofthesystem
• Removing,replacing,anddisposingofthedisposablecomponents(VaporTransferCartridge,
VapothermSpikeSet-1&DeliveryTube)every30days
• Disinfectingofthe2000iuniteverythirtydaysfollowingtheprocedurefoundin
Section8.1–8.6ofthe2000iOperatingInstructionManual
• Changenasalcannulaswhensoiledorexcessivelywetfromsecretions.Disposeofproperly.
• Installing and replacing the sterile water bag
• Avoidliquidspillsonthecomponents
• Donotattemptanyrepairs.IfyouhaveanyproblemswithyourVapotherm2000idevice
notifytheDMEproviderimmediately

Operating Instruction Manual Page 7
Mounting pole
Flowmeter
Air line Oxygen line
Vapotherm
2000i
Delivery tube
Sterile
Water
Bag
Simple, easy
to use touch
control panel
and information
screen.
Vapor transfer
cartridge
Note: System specification
and front panel layout
may vary.
VSS-1
Spike set
Air/Oxygen
Blender
Section 4 Set-up
4.1 The 2000i Unit
1. Thebackofthe2000ihasanIVpoleclampthat
enablesIVpoleattachment.
2. TheunitshouldbemountedonasturdyIVpole
approximately two feet from the top of the
pole to facilitate ease of access and proper
flowfromsterilewatersystem.
3. Ifusinganoxygenblender,mounttheblender
abovetheVapotherm2000iontheIVpole.
4. Connectblenderhosesintobothairand
oxygenwallconnections.
5. PlugVapothermpowercordintoahospital
wallpoweroutlet.
WARNINGS:
Aseptic technique (including proper hand washing
and avoiding direct hand contact with connection
points) should always be followed when setting up
and operating the Vapotherm 2000i.
The medical gas source is external to the Vapotherm
2000i. Always verify the integrity of the medical gas
source and utilize bacterial filters if necessary.
Closed system components (VSS-1, Vapor Transfer
Cartridge and Patient Delivery Tube) should not be
opened in patient care area.
4.2 Selecting the Vapor Transfer
Cartridge
Vapothermprovidesbothahighflowcartridge
(VT01-AS)andalowflowcartridge(VT01-BS).
• Thehighflowcartridge(VT01-AS)shouldbe
used with the pediatric cannula with an
operationalflowrangeof5 – 20litersperminute
(lpm)orwithadultsizedcannulawithan
operationalflowrangeof8– 40lpm.
• Thelowflowcartridge(VT01-BS)shouldbe
used with the neonatal, premature, infant, or
intermediate sized cannula with an operational
flowrangeof1– 8lpm.
WARNINGS:
Do not exceed maximum operational flow rates of 40 lpm
for the high flow cartridge (VT01-AS) and 8 lpm for the low flow cartridge (VT01-BS).
Ensure that the correct cartridge is inserted before operating.
Section 3 AbouttheVapotherm®2000i
TheVapotherm2000iwarmsandhumidifiesflowsofair,oxygenormedicalgasblendsfordeliverytoa
patient,bynasalcannulaorVapothermapprovedinterface.Warmingandhumidificationofbreathinggas
occursinaVaporTransferCartridge,whereairandwaterareseparatedbyamembranepermeableto
watervapor.Themembraneconsistsofmicrotubulesconstructedofpolysulfonematerial.Themembrane
meetsHIMA(HealthIndustryManufacturersAssociation)standardsonfiltersforsterilizingliquidsandhas
beenshowntoeffectivelyexcludebacteriafromcrossingfromthewatercirculationtothegasflow.
ThewarmedhumidifiedgasstreamreachesthepatientviaapatentedtriplelumenPatientDeliveryTube.
Humidifiedgasflowisdeliveredthroughthecenterlumen.Theouterlumenscontainwaterwhichis
warmedviaaninternalheaterandpropelledthroughthesystemviaaninternalpump(See schematic,
fig. 1).Thismaintainsbreathinggastemperatureandminimizescondensation.Thefinalpatientinterface
isaVapothermnasalcannulaorapprovedinterfaceconfiguredtominimizeresistanceandheatloss.
NOTE: The water circuit and gas circuit of the Vapotherm 2000i do not come in contact with each other.
RespiratorygasesaresuppliedtotheVapotherm™2000ifromanexternalgassupply,typicallythrougha
standardwall-mountedflowmeterconnectedtothehospitalmedicalgassupply.Gasflowrateiscontrolled
bytheexternalflowmeterormedicalgasblender.TherearenoflowcontrolsontheVapotherm™2000i.
Connectionsforgasflowandwateraremadeviatherearpanel.AllVapotherm2000icontrolsareonthe
frontpanelofthedevice.
WARNING: Use of patient interfaces not recommended by Vapotherm may cause safety concerns or
affect the performance of the device.
Operating Instruction Manual Page 6
Fig. 1 Simplified System Diagram
Water Reservoir
Pump
Heater
Heated
Delivery
tube
Breathing Gas to Patient
Breathing Gas
Inlet
Vapor
Tr ansfer
Cartridge

Operating Instruction Manual Page 9
Fig. 4
Fig. 5
Fig. 6
Section 4 Set-up
4.5 Vapotherm Spike Set (VSS-1): for
connecting Sterile Water Bag
1. HangasterilewaterbagfromIVPole.
2. ConnectVSS-1tothewaterinletportonbackofunitandmake
sureitlocksintoplace(Fig.5).
3. EnsuretheVSS-1isclampedthenremovespikecap.Wipespike
withdisinfectantwipes,70-90%isopropylalcohol.
4 Firmlyinsertspikeintosterilewaterbagwhileavoidingdirect
handcontactwiththespiketipandwaterbagseptum.
5. LeaveVSS-1spikesetclampeduntilreadytofillunit.
WARNINGS:
The VSS-1 is single patient use item and should be changed with
each patient.
If the VSS-1 is removed from the Vapotherm device for any reason
the Vapotherm 2000i should be disinfected following the routine
disinfection procedure before being returned to service.
CAUTION: Never leave the VSS-1 unclamped when the system is not
running.
NOTE: Removing an empty sterile water container does not constitute
opening the closed system. New sterile water containers can be spiked
using the same VSS-1 without removing the device from service
following the procedure above.
4.6 Connect To A Gas Source
1. Connectasourceofair,oxygenormedicalgasblendertothegas
inletportofVapotherm2000i(Fig. 6).Gasinletconnectionisa
hosebarbthatacceptsfemalefittingonastandard1/4”
(6.35mm)oxygentube.
NOTE: Vapotherm will not operate unless there is gas pressure at gas
inlet. With no flow/pressure sensed, a “System Failure” alarm will
sound.
Operating Instruction Manual Page 8
Fig. 1
Fig. 2
Fig. 3
Section 4 Set-up
4.3 Inserting the Vapor Transfer Cartridge
1. TheVaporTransferCartridge(VT01-ASorVT01-BS)attachesto
theunitbytwowaterandtwoairconnections.
2. Whenfacingtheunit,accessisviaahingedcoveronrightside.
Thecartridgemaybefittedineitherdirection.
3. Datethecartridge.
4. Removeprotectivecapsfromluersideportsofcartridge (Fig. 1).
5. AttachlowerairtubefromVapotherm2000itolowerendof
cartridge.
6. Insertprojectingsideportsintomatchingconnectionsinunit.
Presscartridgefirmlyintoplace(Fig. 2).
7. AttachupperairtubefromVapotherm2000itotopof
cartridge (Fig. 3).Makesuretubingisnotkinked.
8. Closehingedcover.Ifitdoesnotcloseeasily,checkthatcartridge
is pressed fully into place and that air tubes are not interfering
withcover.
WARNINGS:
The cartridge must be changed between patients and discarded
after each use.
If the cartridge is removed, the unit should be disinfected.
If the cartridge is dropped, it should be discarded.
NOTE: Do not remove cartridge from the Vapotherm 2000i
without first draining the machine.
4.4 Inserting the Patient Delivery Tube
1. InsertPatientDeliveryTubeintolowerportionoftheunitby
aligningbluetabsontubewithnotchesonbottomofunit.
2. Firmlypressintoplace(Fig. 4, see next page).Bluelipontube
mustbeflushwiththebottomofunit.
3. Rotate1/4turnclockwiseandpullslightlydownwardstolockinto
place(Fig.4,see next page).
WARNING: The Patient Delivery Tube is a single patient use item
and should be changed with each patient.
If the Patient Delivery Tube is removed from the device for any reason
the Vapotherm 2000i should be disinfected following the routine
disinfection procedure before returning to service.
CAUTION: Unit will not operate correctly if the Patient Delivery Tube
is inserted improperly or not locked into place.

Section 5 OperationofTheVapotherm™2000i
5.5 Setting the Temperature and Warm-Up
1. TheVapotherm2000idisplaystheactualtemperatureofthecirculatingwater.Pressandreleasethe
upordownarrowonthefrontoftheunittodisplaytemperaturesettingfor3seconds.
2. ToadjustthetemperaturesettingoftheVapotherm2000i,pressandholdtheupordownarrow
untilthedesiredtemperatureisdisplayedintheLED.
NOTES:
The Vapotherm 2000i always defaults to previous set temperature at power up.
The temperature can be set between 33 and 43˚C.
5.6 Connecting to Patient
1. WaitfordesiredoperatingtemperaturetobereachedBEFORE placing the cannula on the end of
thePatientDeliveryTube.
2. Checkwaterlevel,temperaturedisplayandgasflowrate.
3. Sizecannulatopatientbyensuringthatnasalprongsdonotfittightlyintonares.
4. Attachproperlysizedcannulathatisdesignedtofunctionwiththecartridgeinstalledinthemachine
ontothedeliverytube.Adjusttheflowtothedesiredrateandplacethecannulaonthepatient.
5. Somecondensationofmoisturearoundnoseispossible.Inaddition,highmoisturelevelmay
mobilizemucusfromnoseandsinuses.Makesurepatienthasasupplyoftissues.
Cartridge Cannula Type Operational Flow Rates
High Flow (VT01-AS) Adult 8 – 40 lpm
High Flow (VT01-AS) Pediatric 5 – 20 lpm
Low Flow (VT01-BS) Premature, neonatal, infant, intermediate 1 – 8 lpm
WARNINGS:
Always follow aseptic technique (including proper hand washing and avoiding direct hand contact with
connection points) when setting up the Vapotherm 2000i and Standard Precautions when placing on
a patient.
Cannula should not obstruct the nares of the patient.
Change nasal cannulas when soiled.
After start-up, during the normal purging of the Patient Delivery Tube, air will release and appear as
bubbles in the bubble trap of the VSS-1. If the Patient Delivery Tube is filled and a stream of continuous
bubbles appear in the bubble trap, it may indicate a problem with the cartridge or Patient Delivery Tube
and both should be checked or changed.
NOTE:
If using a low flow cartridge (VT01-BS) the flow cannot be decreased below 5 lpm until an appropriate
cannula has been attached to the delivery tube
Droplets of condensation may appear at the end of Patient Delivery Tube while unit is warming up. This
is normal and will stop within a few minutes when temperature is reached. If this condition continues
refer to trouble shooting section. If the system operates while not connected to a patient, condensation
is likely to develop.
Operating Instruction Manual Page 11
Operating Instruction Manual Page 10
S
C
Mute
Powe
r
Fig. 2
Fig. 3
Section 5 OperationofTheVapotherm™2000i
5.1 Prepare for Activating the Unit
1. EnsurethattheVapotherm2000ipowercordispluggedintoa
hospitalelectricalwalloutlet.
2. UnclamptheVapothermSpikeSet(VSS-1)(Fig. 1).
5.2 Turn on Flow
1. Ifusingthehighflowcartridge(VT01-AS)flowshouldbe
startedatleast8lpmforwarm-up.
2. Ifusingthelowflowcartridge(VT01-BS)flowshouldbeset
toatleast5lpmforwarm-up.
3. Turnonflow.
5.3 Priming the Unit
1. UnitshouldbestartedinCLEANINGMODE“CL”toprimeanew
PatientDeliveryTube.
2 ToplaceunitinCLEANINGMODE,pressboththepoweronand
theALARMSILENCE/MUTEbuttonssimultaneously(Fig. 2).
3. Thedisplaywillshow“CL”andLEDnexttocleaningicon
S
willilluminate.
4. WaterwillbegintocirculateandfillthePatientDeliveryTube.
5. OperateinCLEANINGMODEuntilPatientDeliveryTubehasbeen
purgedofairbubbles.
6. Whenairhasbeenpurged,pressPOWERtostopsystem
7. Waituntildisplayblanks.
NOTES:
Pressing both the power and alarm mute buttons will set unit
in CLEANING MODE, pressing only the power button sets unit in
NORMAL mode.
Water is not being heated in CLEANING MODE: the purpose
of this mode is to fill the outer lumens of the Patient Delivery Tube
with water.
Gas flow is highly recommended (but not mandatory) during priming.
5.4 Activate the Unit
1. PressthePowerbuttononly,tostartinNORMALMODE(Fig. 3).
2. IfnotusingthePatientDeliveryTubewithintegratedcannula,
donotplacecannulaontheendoftubeuntilwarmupiscomplete.
Fig. 1

Operating Instruction Manual Page 13
Alarm Cause Action
indication
NOTE: To restart after a system failure, the unit must be reset by a momentary pressure on the Power
button. Do not hold the Power button. The alarm will shut off after a delay of about a second, and
the unit can then be restarted normally.
S
S
Waterisnotfillingsystemproperly
LowWaterPressure
MalfunctioningWaterorGas
PressureSensor
MakesureVSS-1spikesetisopen,and
the tube is not kinked or blocked by
airbubbles.
Make sure gas and water connections are
open, gas can flow to unit, and air has
purged from water system: if not, run
‘CLEANINGMODE’.
Sendinforservice.
Water low
Insufficientgasorwaterpressure.
Malfunctioning WaterorGas
PressureSensor
Make sure the gas and water circuits are
open and functional and air has purged
from the water circuit if the unit is in
normaloperatingmode.
Runin“CLEANINGMODE”.
Make sure there is correct flow for
cartridgeflowrates.Ifusing<5lpmwitha
low flow cartridge, a nasal cannula must
beattached.
If a component failure return the unit
forservice.
System Failure
Cartridge Waterdropsinthecircuitwill
causeacartridgealarm;thisdoes
not necessarily mean the cartridge
needstobereplaced.
Section 6Alarms,TroubleShootingandComponentChange-Outs
6.1 General
1. Periodicallycheckforalarmconditions.
2. Unitwillshutdownifthereisnogasflow.However,flowwillnotbeinterruptedifunit
shutsdownormalfunctionsforanyotherreason.
3 Unitwillshutdowniftemperaturesafetylimitsareexceeded,orifwaterlevelislowfor
morethan4mins.However,unheatedgasflowwillcontinue.
NOTE: Should a malfunction occur, indicators on the front panel will light and an alarm will sound.
If the actions listed here do not correct the problem causing the alarm, the unit should be returned to
an approved facility for service.
6.2 Alarms and Troubleshooting
Section 5 OperationofTheVapotherm™2000i
5.7 Operations – General Guidelines
1. CheckthatwaterisproperlycirculatingthroughthemachinebymakingsurethePatientDelivery
Tubeiswarmacrosstheentirelength.
2. Ifgoodcirculationcannotbeconfirmed,checkthatthewaterflowisnotobstructedbyairbubbles.
3. Takeprecautionstominimizecoolingoftheunheatedcannulabytryingtomaintaincontactwith
thepatient’sskinandinsulatingtheexposedportionofthecannulawithbedding.
4. Cartridgedoorshouldbeclosedduringoperation.
NOTE: Condensation in the cannula may occur at low flow rates. To minimize condensation, these
general guidelines should be followed:
• If using flow rates less than 5 lpm, do not set the temperature higher than 34ºC.
• TheVapothermunitshouldnotbeinapositionwhereitiscooled(eg.byanairconditioningoutlet).
CAUTION: DO NOT EXCEED flows of 8 lpm for VT01-BS and 40 lpm for VT01-AS cartridge.
DO NOT SET flows below 1 lpm for VT01-BS.
WARNING:
Never occlude the nares with cannula.
NOTES:
It may become necessary to disconnect the cannula from the Patient Delivery Tube for short periods, such as
when moving a patient out of a radiant warmer. At flow rates less than 5 lpm, cannula disconnection will
activate a system failure alarm, requiring a reset. To avoid this alarm, briefly turn off unit by pressing the power
key once. The display will show two bars. Disconnect the cannula from Patient Delivery Tube and move the
patient, reconnect the cannula, then press the power key once more to restart the unit.
A SYSTEM FAIL (88) alarm will activate if there is insufficient gas pressure in the manifold. If no cannula is fitted,
flow rate at startup should be at least 5 lpm. The minimum flow rate for operation is 1 lpm if a neonate, infant
or premature-sized cannula is fitted, 5 lpm with a pediatric cannula, and 8 lpm with an adult cannula.
Cannulas are single use patient items, dispose of as necessary or according to your institution’s guidelines or
when visibily soiled or excessively wet from secretions. Weekly change out is recommended to avoid any
hardening of nasal prongs.
An air lock can develop at the pump, preventing normal water flow. Try restarting the unit in
cleaning mode.
Operating Instruction Manual Page 12

Operating Instruction Manual Page 15
Fig. 1
Fig. 3
Section 6Alarms,TroubleShootingandComponentChange-Outs
6.3 Component Change Outs
WARNINGS:
The vapor transfer cartridge, patient delivery tube, and VSS-1
spike set are all single use only and should be discarded after
removal from the Vapotherm 2000i.
The Patient Delivery Tube, VSS-1 and Vapor Transfer Cartridge
should not be changed or replaced in the patient care area.
The system must be disinfected any time the Vapor Transfer
Cartridge, VSS-1 or Patient Delivery Tube are removed.
NOTE: The cannula and sterile water source can be replaced without
disinfecting the system. As with all respiratory equipment, proper
hand washing techniques should be followed before contacting
or replacing any patient interfaces.
6.3.1 Replacing Vapor Transfer Cartridge
1. Poweroffunit.Disconnectgasflow.
2. ClosecliponVSS-1.(Fig. 1)
3. Openhingedcover.
4. Disconnectairtubesfromcartridgeendsbypressingtubingaway
fromcartridge.
5. Removecartridgebypullingstraightoutwards.(Fig. 2)
6. ProceedtoSection8.0anddisinfecttheVapotherm®2000idevice
beforereturningthedevicetoservice.
7. Forset-uppleaserefertoSection4.3ofthemanual.
CAUTION: Do not grip cartridge tubing with sharp instruments.
6.3.2 Replacing the Patient Delivery Tube
1. Powerunitoff.Disconnectgasflow.
2. Toremovetube,pushbaseoftubeupwards,rotate1/4turn
counterclockwiseandpulldownward.(Fig. 3)
3. ProceedtoSection8.0anddisinfecttheVapotherm®2000idevice
beforereturningthedevicetoservice.
4. Forset-uppleaserefertoSection4.4ofthemanual.
6.3.3 Replacing the VSS-1 Spike Set
1. Poweroffunit.Disconnectgasflow.
2. ClampVSS-1andremoveVSS-1SpikeSetfromthewaterinlet
portonthebackoftheVapotherm2000ibyreleasingthe
quickconnectonthewaterinletport.
3. ProceedtoSection8.0anddisinfecttheVapotherm®2000idevice
beforereturningthedevicetoservice.
4. Forset-uppleaserefertoSection4.5ofthemanual.
Fig. 2
Alarm Cause Action
indication
Operating Instruction Manual Page 14
If further assistance is needed please call your clinical product specialist or local distributor representative.
S
S
IftheCartridgeAlarmiscontinuous
and air bubbles are rising into the
VSS-1bubbletraporifaflowor
waterisvisibleinthetubebelow
the cartridge, then the cartridge
hasfailed.
If cartridge alarm is intermittent
and there are no bubbles in the
VSS-1bubbletrapornoobvious
water flow below the cartridge there
maybecondensationinthesystem.
Firstdisconnectthepatientfromtheunit,
shut down unit, drain unit, disinfect unit,
replacecartridge,VSS-1anddeliverytube,
andfollowsetupinstructions.
Occasional brief alarms due to
condensationarenotacauseforconcern.
Trybrieflypinchingandreleasingtube
undercartridgetodislodgethedropsand/
ordecreasesettemperature.
Cartridge
MalfunctionofTemperature
ControlSystem.
Shutdownsystemandreturnforservice.
NOTE: A momentary High Temperature
alarm may occur when the unit has
been switched off and on again. If the
temperature then stabilizes, no action
is needed.
High
Temperature
Alarm
Highwaterorairpressuredueto
high resistance in water circulation
or air outlet: or malfunctioning
pressuresensor.
Blockedtubealarmduetohigh
WATERpressurewillcausea
continuous or intermittent tone
andalarmlight.Theflowof
breathing gas continues, but is
nolongerheated.
Blockedtubealarmduetohigh
GASpressurewillcausea5second
alarmtone.Iftheobstruction
persists the system will continue
toalarmin5secepisodes.Water
circulation continues but the heater
shutsoff.
Checkthatdeliverytubeiscorrectly
positioned, rotated clockwise, and pulled
intolockedposition.Checkthatwateris
circulatingwithindeliverytube.Ifalarm
persistsreplacedeliverytubeand/or
cartridge.Disinfectunitpriortoreplacing
components.
Findandcorrectthecauseofobstruction.
Themostcommoncauseisakinkinthe
nasalcannualorintheprong.Attempting
toruntheVapotherm2000iatveryhigh
flow through a patient interface not
approvedbyVapothermmayalsoraisethe
internal pressure sufficiently to trigger a
BlockedTubeAlarm.
Blocked Tube
Alarm
S
Section 6Alarms,TroubleShootingandComponentChange-Outs

Operating Instruction Manual Page 17
Fig. 1
Section 8RoutineDisinfectingProtocol
8.1 Disinfection Supplies
1. DK-301(a-eincludedinkit) a. DisinfectionBagA b. DisinfectionBagB
c. DisinfectionTube d. CartridgeBypassTubes
e. Y-SpikeAssembly
2. Gloves
3. SafetyGlasses
4. Disinfectantwipes,70-90%isopropylalcohol
5. ApprovedDisinfectant
6. 1000mlSterileWaterBag
7. MedicalGradeAirSource
8. Standardadultflowmeterwithoxygen7fttubingattached
WARNINGS:
Vapotherm should be disinfected after each patient or every 30 days on a single patient. Do not disinfect in
an open patient care area.
The DK-301 disinfection kit is a single use item. Operators should open a new disinfection kit for each
disinfection procedure and discard the components at the end of the procedure.
Disinfection Procedure should be performed in a well ventilated area. Use Standard Precautions and
aseptic techniques during this procedure.
The Vapor Transfer Cartridge SHOULD NOT be in place when disinfecting the unit.
The Vapor Transfer Cartridge is a single use disposable and must be discarded after each patient use.
DK-301 IS NOT designed to disinfect Vapor Transfer Cartridges.
8.2 Pre-Cleaning Process
1. Afterpatientuse,itisrecommendedthattheVapotherm2000iSystemremainattachedtotheIV
polewithallthecomponentpartsintact.
2. MovetheVapotherm2000iSystemtoahospitalapproved
reprocessingareaoutsidethepatientcarearea.
WARNING: The water circuit of the Vapotherm 2000i System is not
sterile and can potentially have bacterial contamination. The water
circuit of the device should never be opened in a patient care area.
Transport the Vapotherm 2000i System to an appropriate area for
draining, cleaning and disinfection.
3. Washhandsandputongloves.
4. DraintheVapotherm2000iSysteminareceptaclebycuttingthe
deliverytube.(Fig. 1)
5. Removeanddisposeofthedeliverytube,cannula,VSS-1spike
set,andsterilewatersource.
6. Removeanddisposeofthevaportransfercartridge.
WARNING: The water circuit of the Vapotherm 2000i system has the potential for bacterial growth so
standard precautions should be used to open the water circuit. Disposable components should be disposed of
in accordance with hospital guidelines and operators should wash their hands after breaking down the device.
Operating Instruction Manual Page 16
Fig. 1
Fig. 2
Section 7RemovingFromPatientandSystemShutDown
1. Removecannulaorotherinterfacefrompatient.
2. Pressandreleasepowerswitch(Fig. 1).Displaywillshow“--”
3. ClosecliponVSS-1(Fig. 2).
NOTE: The system’s pump continues to run for 1 minute to allow
heater to cool down.
4. After1minute,waterpumpshutsoffandnumericdisplayis
blank.Unitmaynowbedisconnectedfrompoweroutlet.
5. Removeunitfrompatientcareareaandproceedtodisinfection
process.
CAUTIONS:
Avoid disconnecting from power or gas sources while machine is
operating.
Do not unplug from power source until display is blank.

Section 8RoutineDisinfectingProtocol
8.4 Set-up (cont.)
6. HangBagAwith200mlofapproveddisinfectingsolutionfromIV
polehook.
7. Wipeblueendofdisinfectiontubewithanapproveddisinfectant
wipe.InsertintobottomportofVapotherm™2000isystem.Press
firmlyintoplace,rotate1/4turnclockwise,andpulldownslightly
tolockinposition.(Fig. 5)
8. Wipetheotherendofthedisinfectiontubewithanapproved
disinfectionwipeandattachittothebottomoutletofBagA.
Lockintoplace.(Fig. 6)
9. SuspendBagBonaseparatehookonIVpole.Capshouldbe
firmlyclosedtominimizepotentialspilling.
10. AttachY-SpikeAssemblyby:
a. DisinfectwipemalecolderfittingendofY-SpikeAssembly
and insert it into water inlet port on the back of the
Vapotherm2000isystem.Lockintoplace.(Fig. 7)
b. DisinfectwipetheO2bushing connector end of the
Y-SpikeAssembly.Attachconnectortotheairinletconnector
portonthebackoftheVapotherm2000isystem. (Fig. 8)
c. DisinfectwipethespikeendoftheY-SpikeAssemblyandinsert
intobottomoutletofBagBuntilitcomestoastop.
WARNING: If spike is not properly inserted, it may cause disinfectant
to leak creating a potential safety risk. The spike should be inserted
into Bag B up to the ridge at the bottom of the spike. Do not insert
the spike past the ridge at the
bottom of the spike.
NOTE: If the spike is inserted too
far into Bag B it will be difficult to
remove, do not insert the spike
past the ridge at the bottom of
the spike. If the spike is hard to
remove from Bag B rotate the
spike while removing it to make
it easier.
Operating Instruction Manual Page 19
Fig. 5
Fig. 8Fig. 7
Fig. 6
Operating Instruction Manual Page 18
Fig. 2
Fig. 3
Fig. 4 Tube B
Section 8 RoutineDisinfectingProtocol
8.3 Pre-Disinfection Cleaning and Decontamination
1. Washhandsandputonnewgloves.
WARNING: Always use standard precautions when cleaning and
disinfecting the Vapotherm 2000i system.
Always use individually wrapped 70-90% isopropyl alcohol disinfectant
wipes when wiping down the Vapotherm 2000i and disinfection kit
components.
Always use a new disinfectant wipe taken directly from the package or
container and ensure that the disinfectant wipe has not dried out
before using it on the Vapotherm 2000i device.
2. Wipeexteriorcasingincludinginsidehingedcoverwithan
approveddisinfectantwipe.(Fig. 2)
3. Wipeinsideandoutsideofthefollowingconnectionswithan
approveddisinfectantwipe:
a. Fourcartridgeconnectionportsinsidehingedcover
(Upperandlowercartridgeairtubesandthewater
connectionports)
b. Waterinletandairinletconnectorsonrearofunit
c. Deliverytubeportonthebottomofunit.
8.4 Set-up
1. Takethe4inchbypasstubewith90°barbfittings(BypassTubeA)
andwipetheendswithanapproveddisinfectantwipe.Pressfirmly
intotheinnercartridgeconnectionports(watercircuit).(Fig. 3)
2. Takethe6inchbypasstubewithstraightbarbfittings(BypassTube
B)andwipetheendswithanapproveddisinfectantwipe.Insert
firmlyintoouterupperandlowercartridgeports(aircircuit).(Fig. 4)
WARNING: The By-pass Tubing has been designed to optimize the
flow of solutions through the Vapotherm 2000i system. Failure to use
the By-Pass Tubing supplied by Vapotherm™, Inc. could lead to
improper disinfection and or drying.
3. Closehingedcover.
4. Prepare200mlofapproveddisinfectantsolutionandadditto
BagAwithslideclampclosed(seeAppendixA–“Disinfection
Solutions”forapproveddisinfectionsolutions,appropriate
concentrations,andrequiredholdtimes).
WARNINGS:
Disinfection solutions, concentrations and hold times in Appendix A
have been verified by independent laboratory testing to adequately
disinfect the Vapotherm™2000i machine when following these
instructions. Modifying this procedure or using an alternative disinfection solution, concentration, or hold
time could result in inadequate disinfection thereby increasing the risk of contamination.
Always wear gloves when handling disinfectant solutions, work in a well ventilated area, and use an
accurate measuring device to ensure the proper concentration of disinfection solution and water.

Operating Instruction Manual Page 21
Fig. 12
Section 8RoutineDisinfectingProtocol
8.5 Disinfect Gas and Water Circuits(cont.)
14. LeaveallDK-301tubingconnectedtotheVapothermsystem.
15. DisposecontentsofBagAinaccordancewithallapplicableregulationsandinstitutionalguidelines.
8.6 Drying
1. EnsurethattheDisinfectantTube,CartridgeBypassTubesandthe
Y-SpikeAssemblyareallinplace.
CAUTION: In order to dry the Vapotherm 2000i system, the
Disinfectant Tube, Bypass Tubes and the Y-Spike Assembly must all be
connected to the Vapotherm 2000i system.
2. Removethespikefromtheemptysterilewaterbagandwipewith
anapproveddisinfectantwipe.
3. InsertspikefromY-SpikeAssemblyintostandardoxygentubing.
4. Setflowmeterto15lpm.
5. TaketheVapotherm2000isystemandpositionthesystemflaton
itssideoppositethecartridgedoorfor2minutes.(Fig.12)
6. After2minutesattachtheVapotherm®2000idevicebackontheIVPoleandcontinuetodryat
15lpmfor25minutes.
7. Afteraminimumof25minutes,disconnectthedisinfectiontube,thecartridgeBy-Passtubingand
Y-spikeassemblyandclosethedoortothecartridgeareaforstorage.Discardallcomponentsofthe
DK-301kitinaccordancewithallapplicableregulationsandinstitutionalguidelines.
8. Wipedownexteriorcasingwithdisinfectantwipe.
9. Placeastickeroverthecartridgeaccessdoortocertifythatthedevicehasbeendisinfected.
10. LogthedisinfectionprocedureonaVapotherm2000iDisinfectionLogSheetorinasimilarlog
approvedbyyourinstitution.AppendixBhasaSampleDisinfectionLog.TheDisinfectionLogcanbe
accessedandprintedoutatwww.vtherm.com.
11. Placethesysteminacleanplasticcoverandsealtheendbytyingaknotoraclip.
12. Thesystemisnowreadyforuseorstorage.
CAUTION: Do not set the flowmeter above 35 lpm or start the drying process without the disinfecting
tube in place. This can cause damage to the pressure transducers in the Vapotherm 2000i system.
WARNING:
Gram (-) bacteria can grow in moist environments. The Vapotherm 2000i system should not be stored
with visible water remaining in the device.
Vapotherm should be disinfected after each patient or every 30 days on a single patient.
Do not disinfect in an open patient care area.
The DK-301 disinfection kit is a single use item and must be discarded after the disinfection procedure.
Section 8RoutineDisinfectingProtocol
8.5 Disinfect Gas and Water Circuits
1. OpenclamponBagA.DisinfectantwillstarttodrainfromBagA
throughtheunitandintoBagB.(Fig. 9)
2. WhendisinfectanthasstoppeddrainingintoBagB,clampBagA.
NOTE: Not all contents from Bag A will drain into Bag B.
3. StarttheVapotherm2000isystemincleaningmodebypressing
theMuteandPowerbuttonsatthesametime(Fig. 10).
Disinfectantwillcirculatethroughgasandwatercircuits.
Rununitincleaningmodefortherequiredholdtimegivenin
AppendixAforthedisinfectionsolutionused.
WARNING: Operating the Vapotherm 2000i system in cleaning mode
for less than the required hold time may not adequately disinfect the
machine and could lead to contamination of the air and water circuits
thereby increasing the risk of infection.
4. Aftercirculatingdisinfectionsolutionthroughthemachineforthe
appropriateholdtime,turntheVapotherm2000ioffbypressingthe
powerbutton.
5. UnclampandlowerBagAandhangontoVapothermi.v.pole
clampknob. (Fig. 11)
6. ThedisinfectantthatwascontainedinreservoirBagB,has
circulated through the unit, and the circulated disinfectant
solutiondrainsintoBagAforcollectionanddisposal.
7. Hang1000mlofpre-packagedsterilewateroni.v.pole.
8. LoosencaponBagAtoallowairtoventoutofthebagassterile
waterfillsthebag.
WARNING: The disinfection procedure has been specifically designed
to use 200 ml of disinfection solution and 1000 ml of sterile water.
Bag A is designed to hold 1200 ml of solution. Using larger than the
recommended volumes during disinfection or rinsing could cause a
spill of diluted disinfection solution.
9. ClosetheclamponBagB,removethespike,anddiscardtheBagBin
accordancewithallapplicableregulationsandinstitutionalguidelines.
10. Wipespikewithanapproveddisinfectantwipeandfirmlyinsert
intothespikeportofaprepackagedsterilewaterbag.Confirm
thatwaterisflowing.
11. ImmediatelystarttheVapotherm2000iincleaningmodeby
pressingtheMuteandPowerbuttonsatthesametime.(Fig.10)
12. RunthecleaningmodetocirculatethesterilewaterthroughtheVapotherm2000isystemuntilall
thewaterhasdrainedfromthesterilewaterbag.Immediatelyturnoffthesystem.
CAUTION: Running the system dry can damage the water pump.
13. ClamptubingandclosecaponBagAanddisconnectthebagfromthedisinfectanttube.
WARNING: Failure to clamp off or close the cap of Bag A firmly before disconnecting it from the
disinfection tube could cause diluted disinfection solution to spill.
Operating Instruction Manual Page 20
A
Fig. 9
Fig. 10
Fig. 11

Section 10 Warranty
Vapotherm,IncwarrantsthattheVapotherm™2000ishallbefreeofdefectsofworkmanshipand
materials and will perform in accordance with the product specifications for a period of one year from
thedateofsalebyVapotherm,Inc.Iftheproductfailstoperforminaccordancewiththeproduct
specifications,Vapotherm,Inc.willrepair,orreplace,atitsoption,thedefectivematerialsorpart.This
warrantydoesnotcoverdamagecausedbyaccident,misuse,abuse,alterationandotherdefectsnot
relatedtomaterialorworkmanship.
VAPOTHERM,INC.DISCLAIMSALLLIABILITYFORECONOMICLOSS,LOSSORPROFITS,OVERHEADOR
CONSEQUENTIALDAMAGESWHICHMAYBECLAIMEDTOARISEFROMANYSALEORUSEOFTHIS
PRODUCT.THISWARRANTYISGIVENINLIEUOFALLOTHEREXPRESSWARRANTIES.
Operating Instruction Manual Page 23
Operating Instruction Manual Page 22
Attention
Consult Manual
Silence
Alarms
Power
On/Off
Alternating
Current
Type BF
Class 1
Single
Patient Use
9.2 Definitions and symbols
0297
Section 9 General Information
9.1 Specifications
Dimensions: Height11”(280mm),width5.5”(140mm),depth4.5”
(114 mm)excludingIVpoleclamp.
Weight: Lessthan6lbs(2.7 kg)withoutwaterreservoir.
Vapotherm spike set: Workswithsterilewaterbagsupto2000 mlbag.
Circulating water volume <100 ml.(excludingPatientPatientDeliveryTube).
Mounting: RearmountedclampfitsstandardIVpoleorhanger.
Power: (US)115 V,60 Hz,250 VA(warmup),approximately80VA(continuous).
(Otherversions)220 –240 V, 50–60 Hz,250 VA(warmup),
approximately80 VA(continuous).
Gas source pressure: 4–50 psi.Athighpressures(e.g.hospitalwallsystem)theVapotherm2000i
mustbeconnectedtothegasoutletviaastandardmedicalflowmeterand
flowregulatorwithapprovedfittings.
Gas flow: Controlledbyexternalflowmeter.Operatingrange1–40 lpm, dependent
oncartridgetypeandpatientinterfaceused.
Output gas temperature: (US)33–43ºCatoutletofthedeliverytube,adjustablebyfrontpanel
settings.(Otherversions)33–41ºC.
Humidification: Vaporphase,bytranspirationthroughmicroporousmembrane.Output
isatleast95%relativehumidityatnasalcannulaataflowrateupto
20 lpm,atleast90%atflowratesfrom20–40 lpm,overthefullrange
ofoperatingconditions.

Vapotherm2000IOperatingInstructionManual–AppendixA:DisinfectionSolutions Page 25
Vapotherm 2000i Operating Instruction Manual
Appendix A – Disinfection Solutions
Thisappendixlistsapproveddisinfectantsolutionsandtherequiredholdtimesnecessaryto
disinfecttheVapotherm2000imachineusingtheroutinedisinfectionprocedureoutlinedinthe
Vapotherm2000iManualRevB,Section8.Thefollowingdisinfectantshavebeenindependently
testedbyanISOcompliantFDAregisteredlabusing“goodlaboratorypractices”(GLP):
Manufacturer Active Ingredients Trade Name Concentration Hold Time*
Minntech Corporation
1460528thAvenueNorth
Minneapolis,MN55447
(800)328-3345
Hydrogen
Peroxide22%
andPeracetic
Acid4.5%
Minncare™ 1% 10minutes
at20˚C
MarilProductsInc.
320West6thStreet
Tustin,CA92780
(800)546-7711
DimethylBenzyl
AmmoniumChloride10%
andDimethylEthylBenzyl
Ammonium10%
Control 3™ 1% 10minutes
at20˚C
*Section8Step4oftheVapotherm2000iOperatingManualrequiresapproveddisinfection
solutionstobecirculatedthroughthedeviceforanappropriateholdtimeasoutlinedinthis
columnofDisinfectionAppendixA.
WARNINGS:
These disinfectant solutions are designed to be used to disinfect the Vapotherm 2000i
machine without the cartridge in place. These disinfectant solutions ARE NOT approved to
disinfect the cartridge.
Failure to properly prepare the disinfection solution or circulate the disinfection solution
throughout the machine for the appropriate hold time could result in inadequate disinfection.
Disinfectants must be used at proper concentrations. User must confirm solution has been
mixed according to disinfectant manufacturers instructions, or used in pre-diluted form.
Solutions must not be used past their expiration dates See disinfectant manufacturer’s
product labeling for instructions.
Section 11 References
• Bamford,OwenandLain,David.VerificationoftheBacteriologicalFiltrationPropertiesofthe
VapothermCartridge.RespiratoryCareNovember2004.Vol.49No.11
• CDC,Guidelineforpreventionofnosocomialpneumonia.MMWR1997:46(N0.RR-1)
• CDC,GuidanceforIsolationPrecautionsinHospitals.Garner,Julia.January1996.
RetrievedMay26,2006,fromhttp://www.cdc.gov/ncidod/dhqp/gl_isolation.html
• StandardPrecautions:CDC,ExcerptedfromGuidelineforIsolationPrecautionsinHospitals(January
1996)RetrievedMay26,2006fromhttp://www.cdc.gov/ncidod/dhqp/gl_isolation_standard.html
Operating Instruction Manual Page 24

Vapotherm2000IOperatingInstructionManual:AppendixB:SampleDisinfectionLogSheet Page 27
SAMPLE
Vapotherm 2000i Operating Instruction Manual
Appendix B – Sample Disinfection Log
Vapotherm 2000i Hospital ID Disinfection Disinfectant
Operator
Machine Serial No. (If Applicable) Date Solution &
Hold Time
Vapotherm2000IOperatingInstructionManual:AppendixADisinfectionSolutions Page 26
Vapotherm 2000i Operating Instruction Manual
Appendix A – Disinfection Solutions (cont.)
Dilution Instructions for preparing a 1% solution of Minncare®and Control III™
Supplies Needed:
• Abeaker,graduatedcylinder,orcontainersuitableformeasuringandmixing200mlof
disinfectant solution
WARNING: Components or instruments that are not part of the DK-301 Kit and are used to mix
or pour disinfectant solutions should be either single use components or glassware that has been
sterilized, high level disinfected, or pasteurized before use in accordance with your institution’s
specific disinfection and sterilization procedures.
• 10mlsyringe,orpipette,graduatedin1mlincrements
• ConcentrateDisinfectant
• Minncare®ColdSterilantEPARegNo.52252-4(22.00%HydrogenPeroxide,4.50%
PeroxyaceticAcid,73.50%InertIngredients)
• Control III®DisinfectantP/N10006(10%n-alkyldimethylbenzylammoniumchloride,
10%n-aykylethylbenzylammoniumchloride,80%inertingredients)
• 200 ml of sterile water
WARNINGS:
Always follow manufactures recommendations for handling disinfectant solutions.
Always wear Personal Protective Equipment when handing disinfectant solution.
Procedure:
1. DonappropriatePersonalProtectiveEquipment:Safetygoggles,gloves,splashapron.
2. Usinganappropriatesyringeorpipette,drawup2 ml of undiluted Minncare®
(EPARegNo.52252-4)orControlIII®(P/N10006)andexpelitinadeviceappropriatefor
measuring200 mlofsolution.
3. Makeup200 mlofsolutionbyaddingsterilewatertothemeasuringdevice.
4. ClosetheslideclamponDisinfectionBagAandpourdisinfectantsolutionintothebag.
5. ClosethecapofBagAfirmly,toavoidspillsandcontinuewithStep6inSection8.4ofthe
disinfectionprocedure(Page18).
6. AttachtoIVPoleandVapotherm2000iaccordingtoDisinfectionprocedureandcontinueto
followDisinfectionProtocol.

Vapotherm, Inc.
198 Log Canoe Circle, Stevensville MD 21666
T: (001) 410.604.3977 F: (001) 410.604.3978 www.vtherm.com PN 8-300068-00 Rev. E
bdc 8369
Suppliedby
Table of contents
Other Vapotherm Medical Equipment manuals

Vapotherm
Vapotherm Oxygen Assist Module User manual

Vapotherm
Vapotherm precision flow User manual

Vapotherm
Vapotherm precision flow Instruction Manual

Vapotherm
Vapotherm Precision Flow Hi-VNI User manual

Vapotherm
Vapotherm precision flow User manual

Vapotherm
Vapotherm Heliox Precision Flow User manual

Vapotherm
Vapotherm FELIX-1 User manual