Walkeraid 4200-100 User manual

A PRODUCT
Model 4200-100
A Computer-Based DME Device
Follow this instruction manual to ensure
safe use of the Walker Alarm Device.
6
Instruction Manual for Rehabilitation Facilities

START UP
CHARGING
IMPORTANT: Charge the Walker Alarm at least
6 hours; but not more than 24 hours.
➊ Turn switch to OFF position
➋ Plug in charging cable to the wall outlet and plug charging
cable into the side charging port on the
Walker Alarm.
A YELLOW LED lights during charging.
A GREEN LED indicates a full charge.
The battery manufacturer recommends not to leave
the battery in a charge function over 24 hours.
LOW BATTERY INDICATION
After 8 hours or more of continuous normal use the YELLOW
Low Battery Light will start to ash indicating 10-15 left of
battery charge left. After the remaining 10-15 minutes is up,
the Walker Alarm will CHIRP and ash YELLOW for about 2
seconds and then shut off.
Note: Please read the information on lithium polymer batteries on page 10.
CHARGING
CABLE
ON/OFF
SWITCH
2

SUGGESTED BEST PRACTICE
Select, measure and adjust walker to patient.
16
16
STEP 1:
MEASURE FROM
WRIST TO FLOOR.
STEP 2:
WALKER IS
ADJUSTED TO
MEASUREMENT
FROM STEP 1
Follow the manufacturer’s suggested procedures to t each
style of walker.
SWITCH SET-UP
➊ Slide Cover open by
lightly pressing down and
forward on the rear of the
cover. If needed, gently lift
front with ngernail.
➋ To get the feel of the Dip
switches, move a switch
up or down with a tip of a
ball point pen, feel when
the switch engages into
the up or down position.
1 Vibrate On/Off
2 Beep Alarm On/Off
3 Color Glow On/Off
4 Distance Code Switch
5 Distance Code Switch
6 Distance Code Switch
7 Sleep Mode On/Off
8 Diagnostic Mode On/Off
ONLY FOR USE BY Administrators or Caregivers
For Proper Walker Alarm Setup and Working Conditions
WALKER ALARM, LLC WARRANTY DIRECTIVE
Switch Position
Top View Lid Removed
24
UP
DOWN
1
Up
Down
2 3 4 5 6 7 8
9
3
➌ Leave switches 7& 8 in the down position to maintain
factory settings.
SENORS

MATCHING WALKER ALARM TO PATIENT NEEDS
The Walker Alarm device as shipped from the factory comes
with 3 biofeedback controls TURNED ON or in the UP position:
Select which feature, and the number of features to be used
based on impairment and assessment.
NOTE: The GREEN GLOW give the person the instant gratification
that they are using the walker correctly.
SWITCH 1: Vibration Control.
The ON position activates the
VIBRATION sensory feature.
SWITCH 2: Sound Control.
The ON position activates the
BEEP audio alarm feature.
SWITCH 3: Light Control.
The ON position activates the
LED light visual feature that
ashes WHITE then RED or
Glows GREEN.
26
1
Up
Down
Factory Switch Setting
2 3 4 5 6 7 8
21
1
Up
Down
Switch 1: Vibrate On/Off
2345678
1
Up
Down
Switch 2: Beep Alarm On/Off
2345678
1
Up
Down
Switch 3: Color Glow On/Off
2345678
VIBRATE, ALARM, GLOW, DIP SWITCH SETTINGS:
21
1
Up
Down
Switch 1: Vibrate On/Off
2 3 4 5 6 7 8
1
Up
Down
Switch 2: Beep Alarm On/Off
2 3 4 5 6 7 8
1
Up
Down
Switch 3: Color Glow On/Off
2 3 4 5 6 7 8
VIBRATE, ALARM, GLOW, DIP SWITCH SETTINGS:
21
1
Up
Down
Switch 1: Vibrate On/Off
2 3 4 5 6 7 8 1
Up
Down
Switch 2: Beep Alarm On/Off
2 3 4 5 6 7 8
1
Up
Down
Switch 3: Color Glow On/Off
2 3 4 5 6 7 8
VIBRATE, ALARM, GLOW, DIP SWITCH SETTINGS:
4

TWO WHEEL WALKER
ROLLATOR WITH NARROW SEAT
POWER OFF BEFORE CHANGING DISTANCE SETTINGS
22
1
Up
Down
16" Distance Alarm
Switch 4 UP, Switch 5 UP, Switch 6 DOWN
2 3 4 5 6 7 8
1
Up
Down
18" Distance Alarm
2 3 4 5 6 7 8
Switch 4 DOWN, Switch 5 UP, Switch 6 DOWN
1
Up
Down
20" Distance Alarm
2 3 4 5 6 7 8
Switch 4 UP, Switch 5 DOWN, Switch 6 DOWN
1
Up
Down
22" Distance Alarm
2 3 4 5 6 7 8
Switch 4 DOWN, Switch 5 DOWN, Switch 6 DOWN
DISTANCE DIP SWITCH SETTINGS:
22
1
Up
Down
16" Distance Alarm
Switch 4 UP, Switch 5 UP, Switch 6 DOWN
2 3 4 5 6 7 8
1
Up
Down
18" Distance Alarm
2 3 4 5 6 7 8
Switch 4 DOWN, Switch 5 UP, Switch 6 DOWN
1
Up
Down
20" Distance Alarm
2 3 4 5 6 7 8
Switch 4 UP, Switch 5 DOWN, Switch 6 DOWN
1
Up
Down
22" Distance Alarm
2 3 4 5 6 7 8
Switch 4 DOWN, Switch 5 DOWN, Switch 6 DOWN
DISTANCE DIP SWITCH SETTINGS:
22
1
Up
Down
16" Distance Alarm
Switch 4 UP, Switch 5 UP, Switch 6 DOWN
2 3 4 5 6 7 8
1
Up
Down
18" Distance Alarm
2 3 4 5 6 7 8
Switch 4 DOWN, Switch 5 UP, Switch 6 DOWN
1
Up
Down
20" Distance Alarm
2 3 4 5 6 7 8
Switch 4 UP, Switch 5 DOWN, Switch 6 DOWN
1
Up
Down
22" Distance Alarm
2 3 4 5 6 7 8
Switch 4 DOWN, Switch 5 DOWN, Switch 6 DOWN
DISTANCE DIP SWITCH SETTINGS:
22
1
Up
Down
16" Distance Alarm
Switch 4 UP, Switch 5 UP, Switch 6 DOWN
2 3 4 5 6 7 8 1
Up
Down
18" Distance Alarm
2 3 4 5 6 7 8
Switch 4 DOWN, Switch 5 UP, Switch 6 DOWN
1
Up
Down
20" Distance Alarm
2 3 4 5 6 7 8
Switch 4 UP, Switch 5 DOWN, Switch 6 DOWN
1
Up
Down
22" Distance Alarm
2 3 4 5 6 7 8
Switch 4 DOWN, Switch 5 DOWN, Switch 6 DOWN
DISTANCE DIP SWITCH SETTINGS:
Suggested Dip Switch Settings
Suggested Dip Switch Settings
SLIDE BAR
MOUNT
LOCATION
SLIDE BAR
MOUNT
LOCATION
5

ROLLATOR WITH STANDARD SEAT
U-STEP WALKER
22
1
Up
Down
16" Distance Alarm
Switch 4 UP, Switch 5 UP, Switch 6 DOWN
2 3 4 5 6 7 8 1
Up
Down
18" Distance Alarm
2 3 4 5 6 7 8
Switch 4 DOWN, Switch 5 UP, Switch 6 DOWN
1
Up
Down
20" Distance Alarm
2 3 4 5 6 7 8
Switch 4
UP
, Switch 5
DOWN
, Switch 6
DOWN
1
Up
Down
22" Distance Alarm
2 3 4 5 6 7 8
Switch 4 DOWN, Switch 5 DOWN, Switch 6 DOWN
DISTANCE DIP SWITCH SETTINGS:
22
1
Up
Down
16" Distance Alarm
Switch 4 UP, Switch 5 UP, Switch 6 DOWN
2 3 4 5 6 7 8 1
Up
Down
18" Distance Alarm
2 3 4 5 6 7 8
Switch 4 DOWN, Switch 5 UP, Switch 6 DOWN
1
Up
Down
20" Distance Alarm
2 3 4 5 6 7 8
Switch 4 UP, Switch 5 DOWN, Switch 6 DOWN
1
Up
Down
22" Distance Alarm
2 3 4 5 6 7 8
Switch 4
DOWN
, Switch 5
DOWN
, Switch 6
DOWN
DISTANCE DIP SWITCH SETTINGS:
22
1
Up
Down
16" Distance Alarm
Switch 4 UP, Switch 5 UP, Switch 6 DOWN
2 3 4 5 6 7 8 1
Up
Down
18" Distance Alarm
2 3 4 5 6 7 8
Switch 4 DOWN, Switch 5 UP, Switch 6 DOWN
1
Up
Down
20" Distance Alarm
2 3 4 5 6 7 8
Switch 4 UP, Switch 5 DOWN, Switch 6 DOWN
1
Up
Down
22" Distance Alarm
2 3 4 5 6 7 8
Switch 4 DOWN, Switch 5 DOWN, Switch 6
DOWN
DISTANCE DIP SWITCH SETTINGS:
22
1
Up
Down
16" Distance Alarm
Switch 4 UP, Switch 5 UP, Switch 6 DOWN
2 3 4 5 6 7 8 1
Up
Down
18" Distance Alarm
2 3 4 5 6 7 8
Switch 4 DOWN, Switch 5 UP, Switch 6 DOWN
1
Up
Down
20" Distance Alarm
2 3 4 5 6 7 8
Switch 4 UP, Switch 5 DOWN, Switch 6 DOWN
1
Up
Down
22" Distance Alarm
2 3 4 5 6 7 8
Switch 4 DOWN, Switch 5 DOWN, Switch 6
DOWN
DISTANCE DIP SWITCH SETTINGS:
23
1
Up
Down
24" Distance Alarm
Switch 4 UP, Switch 5 UP, Switch 6 UP
2 3 4 5 6 7 8
1
Up
Down
26" Distance Alarm
2 3 4 5 6 7 8
Switch 4 DOWN, Switch 5 UP, Switch 6 UP
DISTANCE DIP SWITCH SETTINGS:
Suggested Dip Switch Settings
Suggested Dip Switch Settings
POWER OFF BEFORE CHANGING DISTANCE SETTINGS
SLIDE BAR
MOUNT
LOCATION
SLIDE BAR
MOUNT
LOCATION
6

TRAINING ON THE USE OF THE WALKER ALARM
11
Alarm
Safe Zone
Alarm
5"ALARM ZONE:
This is the area between the patient and up to 5 inches away
from the face of the Walker Alarm. Alarm will activate: The
White LED ashes, then the Red LED ashes, the unit vibrates,
then there is a BEEP Alarm. This alarm cycle will repeat until
the patient moves back into the Safe Green Zone.
SAFE ZONE:
Walk with the walker keeping the Green LED light on. THIS IS
YOUR TARGET – STAY IN THE GREEN SAFE ZONE.
16"OR GREATER DISTANCE ALARM ZONE:
This area is 16” inches OR GREATER away from the face of
the Walker Alarm. Alarm will activate: The White LED ashes,
then the Red LED ashes, the unit vibrates, then there is a
BEEP Alarm. This alarm cycle will repeat until the patient
moves forward into the Safe Green Zone.
5 INCH ALARM ZONE 16"OR GREATER ALARM ZONE
7

SUGGESTED BEST PRACTICE
Training or refreshing training patient in the use
of the walker.
SUGGESTED BEST PRACTICE
Training or refreshing training patient in the use
of the walker.
15
Follow this suggested procedure or similar for all types of
walkers.
If the patient has previous training in the use of the walker do
an prociency assessment to verify the patient is using the
walker SAFELY.
CORRECT:
One foot inside
the walker.
INCORRECT:
Both feet are outside
the walker.
8

THERAPY SESSION TIME – GOOD TRAINING
2
SLIDE BAR MOUNT
Slide the Walker Alarm unit
onto the slide bar mount that
is installed on the walker you
intend to use.
The sensors must be facing the patient.
Turn on the Walker Alarm using the ON/OFF switch on the
side of the device.
28
• GREEN Glow
•Safe Zone
• GREEN Glow
• Safe Zone
• Flashing Red and White
• Vibration
• ALARM Beep
ON/OFF
SWITCH
SENSORS
9

DISCLAIMER
Walker Aid, LLC is not responsible for a patient fall while using
the walker with a Walker Alarm when:
• The Walker has not been tted properly to the patient
• The patient has not been trained or is not being properly
trained in the use of the walker.
• The patient has cognitive impairments
• The patient is too unsteady to walk unsupervised
• Any other circumstances which may effect an individuals
balance (dehydration or illness for example)
• Any change in medical, physical, or cognitive condition
LITHIUM-POLYMER BATTERY
Supplied with the Walker Alarm – a Lithium-Polymer Battery
WITH Protection Circuit Module that protects the battery pack
from over charging and over discharging, reducing the risk of
explosion, re and damage.
CHARGER AND CHARGING
All Walker Alarms will be supplied with UL approved chargers.
If available, PC USB ports/hubs are acceptable when used
with approved power supplies.
Just ensure the charger used is a good one and is rated at 5V
at 200mA or more.
A YELLOW/GREEN charging LED is near the USB charging
connector.
With the charger plugged in, the computer shows the charging
LED YELLOW anytime the battery is being charged, and
GREEN when there is no current being supplied, indicating
that the battery is fully charged. This is an important visual
safety feature showing what the charger is doing related
to the battery and is important that the responsible person
has knowledge of this.
10

TROUBLESHOOTING
Computer-based system alerts:
Incorrect Switch Position
If here is a continuous FLASHING RED LIGHT when turning on
the Walker Alarm, turn it off, check the switch 4, 5 or 6 settings
and turn it back on.
When rst turned on if the LED, vibrator, or buzzer doesn’t
work, turn the Walker Alarm off, check switches 1, 2, or 3
settings and turn it back on.
Incorrect Distance Reading
If a PURPLE LIGHT IS SEEN, IT IS DUE TO AN INCORRECT
DISTANCE READING. Turn the Walker Alarm off and back on
again, resetting the computer, correcting the error.
WARNING
• This is a COMPUTER BASED ELECTRONIC DEVICE.
Keep away from Water!
• DO NOT DROP OR JOLT!
• Operating Temperature 0°-70°C, 32°-158°F
• Always follow Walker Alarm instructions for operation.
SAFETY DEVICE | REMEMBER – IMPORTANT
THIS IS A SAFETY DEVICE WITH A HIGH-END, ELECTRONIC,
SHORT DISTANT RADAR TRACKER; TO KEEP YOU IN A SAFE
AREA WHEN WALKING WITH A WALKER. TAKE PROPER CARE
OF THE WALKER ALARM.
Toll Free: (844) 253-7624
Direct: (630) 728-1973
Email: [email protected]
REPAIR DEPOT:
Ellutions, LLC
c/o Walker Aid, LLC
1701 West Flintlock Way
Chandler, AZ 85286
EIN# Request As Needed
Certicate of Registration
Illinois: 4129-1328
Arizona B-2004167-8
Walker Aid, LLC, is owned by
William G. Trout.
©2016 All right reserved.
Owner of U.S. Patents:
7963294 and 8453662 11

PRODUCT SPECIFICATIONS
DURABLE MEDICAL EQUIPMENT DEVICE —
DME ( Life of 5 yr+ )
FDA Class I (Exempt)
Registration Card documented by Serial Number in case
of a recall. Mail in registration card or scan and email card email
address on registration card.
Factory Notication – User to report to Walker Aid, LLC, Health
or Equipment related incidents.
6 month Warranty | 15-Day Money Back Guarantee
May be under specic contractual agreements.
PRIMARY BENEFITS
May reduce the risk of falling. Will assist with Walker
Physical Therapy Training Sessions by having a Green Zone-
SAFE ZONE (box) to be seen by person in therapy, balance,
posture, and gait session.
FEATURES
CHARGER: UL approved charger (supplied with Walker
Alarm) or USB port/hub.
BATTERY: LP-523334 3.7v 500mAh with PCM (Protection
Circuit Module)
LITHIUM BATTERY: Polymer battery with PCM
protection circuit
PRODUCT SIZE & WEIGHTS
Size without slide bar mount:
(L x W x H): 2-1/2” x 2-3/8” x 1-5/8”
Walker Alarm, Slide Bar Mount, and Charger 7.0 oz.
Walker Alarm with Slide Bar Mount 4.6 oz.
Slide Bar Mount 1.9 oz.
12
Table of contents