52521 R01 Whitney Safety Platform Walker
Operator’s Manual
Page 1 of 20
WHITNEY
Safety Platform Walker
OPERATOR’S MANUAL
WHITNEY
SAFETY PLATFORM WALKER
Model No. 52500

800.255.3126 | wyeastmedical.com
52521 R01 Whitney Safety Platform Walker
Operator’s Manual
Page 2 of 20
WHITNEY
Safety Platform Walker
WARRANTY
Wy’East Medical Corporation warrants to the original user that the Whitney Safety Platform
Walker manufactured by the company is free from defects in material and workmanship under
normal use and service. The term of warranty from the date of purchase/shipment is 2 years
on the mechanical lift components, 1 year on the electronic components, and 90 days for the
battery. During this term Wy’East will provide to the user replacement components or repaired
components.
In order to initiate a warranty claim the original user must provide prompt notice to Wy’East
Medical Corporation within the applicable warranty period specifying the nature of the claim.
Wy’East will issue instructions as to the procedure to follow. Wy’East reserves the right to confirm
the existence of any claimed defect and to determine whether any damage was caused by
neglect, misuse, unauthorized alteration or failure to follow maintenance instructions. With respect
to a claim covered by this warranty, the defective part will be replaced or repaired at no charge,
provided the defective part is returned to Wy’East Medical. The user has the responsibility for
shipping charges and installation. Assistance is available by calling (503) 657-3101 in Oregon.
This is the exclusive remedy for a claim under this warranty and in no event shall Wy’East be
liable for consequential damages. THIS WARRANTY IS IN LIEU OF ALL OTHER WARRANTIES
EXPRESSED OR IMPLIED.
PLEASE READ
Please read this entire manual before operating the Whitney Safety Platform Walker.
This warranty applies to the Wy’East Whitney Safety Platform Walker only.
For questions or information regarding other Wy’East products, contact:
Wy’East Medical Corporation
P.O. Box 1625
Clackamas, Oregon, USA
Telephone: (503) 657-3101
(800) 255-3126
FAX: (503) 657-6901
www.wyeastmed.com
contact@wyeastmed.com

800.255.3126 | wyeastmedical.com
52521 R01 Whitney Safety Platform Walker
Operator’s Manual
Page 3 of 20
WHITNEY
Safety Platform Walker
Contents
HOW TO USE THIS MANUAL:_________________________________________________ 4
DEFINITIONS: ______________________________________________________________ 4
SAFETY CONSIDERATIONS FOR ALL PROCEDURES: ____________________________ 6
CLEANING AND DISINFECTION: ______________________________________________ 6
PRIOR TO FIRST USE:_______________________________________________________ 7
BEFORE AND AFTER EVERY USE: ____________________________________________ 7
FEATURES: ________________________________________________________________ 8
Raising/lowering the Frame: _______________________________________________ 8
Legs Open/Close Function:________________________________________________ 8
Left and right adjustment of armrests: _______________________________________ 8
Locking/unlocking casters: ________________________________________________ 8
Adjusting the shin supports: _______________________________________________ 9
Adjusting the handles: ____________________________________________________ 9
Handset: _______________________________________________________________ 9
Control Box Wiring _______________________________________________________ 10
Battery Charging ________________________________________________________ 10
LEDs on LINAK Control Box and Battery* ____________________________________ 11
OPERATING THE WHITNEY SAFETY PLATFORM WALKER: ________________________ 12
TECH SPECIFICATIONS: _____________________________________________________ 13
GENERAL MAINTENANCE ADVICE: ___________________________________________ 14
TROUBLESHOOTING: ______________________________________________________ 14
BELT CLEANING: ___________________________________________________________ 16
ELECTRONICS: ____________________________________________________________ 17
BATTERY CARE AND CAUTIONS: __________________________________________ 17
ELECTRONICS ENVIRONMENTAL CONDITIONS: _____________________________ 18
OPERATING:_________________________________________________________ 18
STORAGE: __________________________________________________________ 18
TRANSPORT: ________________________________________________________ 18
FACTORY ASSISTANCE: _____________________________________________________ 19
TROUBLESHOOTING ASSISTANCE: ________________________________________ 19
RETURNING ITEMS: _____________________________________________________ 19
NOTES: ___________________________________________________________________ 20
CONTACT INFORMATION ____________________________________________________ 21
MAINTENANCE
& ASSISTANCE
INTRODUCTION

800.255.3126 | wyeastmedical.com
52521 R01 Whitney Safety Platform Walker
Operator’s Manual
Page 4 of 20
WHITNEY
Safety Platform Walker
INTRODUCTION
HOW TO USE THIS MANUAL:
This manual is designed to familiarize you with the operation of the Whitney Safety Platform Walker. Read it
thoroughly before using or servicing the equipment. Keep the manual for future reference and for new members
of your staff.
DO NOT ATTEMPT TO USE THIS EQUIPMENT WITHOUT UNDERSTANDING THIS MANUAL
To ensure safe operation, read carefully the entire manual, before installing, operating, or servicing this
equipment.
If anything is not completely understood, please contact your supplier for more details. Failure to comply with
warnings in this manual may result in injury.
Keep this manual with the Whitney Safety Platform Walker and refer to it as required. Contents of this manual are
subjected to change without prior notice to users.
DEFINITIONS:
The words WARNING, CAUTION, and NOTE carry specific meanings and should be carefully reviewed.
WARNING: The personal safety of the patient or user may be involved. Disregarding this information could result
in injury to the patient or user.
CAUTION: These instructions point out special procedures or precautions that must be followed to avoid
damaging the Whitney walker or it’s components.
NOTE: This provides special information to make maintenance easier or important instructions more clear.
MEANING OF SYMBOLS
Caution Safe working load
Consult instructions for use Do not step on surface
Important
Consistent with our policy of continuing
product improvement, WY’EAST MEDICAL
CORPORATION reserves the right to change
these specications without notice or obligation.

800.255.3126 | wyeastmedical.com
52521 R01 Whitney Safety Platform Walker
Operator’s Manual
Page 5 of 20
WHITNEY
Safety Platform Walker
The Whitney Safety Platform Walker is a conventional, electrically operated walker that is enhanced
with a power rise function. Not only does it act as a walker, it also promotes mobility and safety with
its unique power rise function and safety harness.
The Whitney can be used by anyone who regularly uses a walker. The user can remain active while
the power rise function is in use.
The Whitney consists of a walker and its associated harness and straps. The harness is secured to
the two attachment straps (one on each side), which are in turn attached to the walker.
When the patient stands the electric power rise function is employed and the smart attachment
reduces the distance between the patient and the walker until the patient is standing upright.
The Whitney has an electrically driven base width adjustment, which makes it easier to get closer to
the user.
The harness is available in different sizes (M, L) and has two rear straps. NOTE: that the lower part
of the harness attached around the patient’s bottom can be opened as required (i.e. toileting).
Intended for use in acute care, long-term care and outpatient care. It is not
recommended for use in a shower.
Prior to using the Whitney the patient must be assessed by a qualified professional to
determine the suitability of using it with that patient.

800.255.3126 | wyeastmedical.com
52521 R01 Whitney Safety Platform Walker
Operator’s Manual
Page 6 of 20
WHITNEY
Safety Platform Walker
SAFETY CONSIDERATIONS FOR ALL PROCEDURES:
1. The Whitney Walker is to be used ONLY on level surfaces.
2. ALWAYS lock the wheels while lifting and lowering the patient.
3. Prior to using the Whitney Walker, the patient must be assessed by a qualified professional to
determine the suitability of using the Whitney Walker with that patient.
4. Never leave the patient unattended while using the Whitney Walker.
5. Never exceed the safe working load.
6. Warranty applies only if repairs or alterations are done by an authorized technician.
CLEANING AND DISINFECTION:
1. Do not power wash the equipment.
2. Painted and plastic surfaces (other than electronics): For cleaning and disinfection, use a
quaternary or bleach-based solution on painted and plastic surfaces.
3. Electronics: Ensure actuators are at their minimum stroke before washing. Organic solvents
like halogenide/aromatic hydrocarbons and ketones must not be used. Observe the following
regarding detergents and disinfectants on electronic components:
a. Do not use strong alkaline or acidic compounds (pH value 6-8)
b. Compounds must not contain corrosive or caustic material.

800.255.3126 | wyeastmedical.com
52521 R01 Whitney Safety Platform Walker
Operator’s Manual
Page 7 of 20
WHITNEY
Safety Platform Walker
PRIOR TO FIRST USE:
The battery must be charged for 24 hours in order to reach proper function and prolong
the lifetime of the batteries.
Insure all cables are completely pushed in. A light film of silicon grease may be dabbed
onto oring to allow easier insertion.
The duty cycle on the control box product label must always be noted. Otherwise there
is a risk of damaging the products. Exceeding the duty cycle will result in a dramatic
reduction of the lifetime of the system. Unless specified otherwise on the product label
the duty cycle is max. 10% : Max. 2 minutes in the use followed by 18 minutes not in use.
If any individual parts are suspected to be damaged, do not operate the parts, but return
them for inspection/service.
BEFORE AND AFTER EVERY USE:
1. Always perform a risk assessment before starting the specific lifting session in order to make
sure the product model is correct and that the lifting procedure is fully understood.
2.Before using the sling, read this instruction manual and the instruction manual for the lifting
equipment being used.
3.When soiled or stained, the sling must be washed before use.
4. Warning: Check the sling for any fraying, loose stitching, attachments or any deteriorations.
5.After use, check the sling according to steps 3 and 4 and store in a dry area.

800.255.3126 | wyeastmedical.com
52521 R01 Whitney Safety Platform Walker
Operator’s Manual
Page 8 of 20
WHITNEY
Safety Platform Walker
FEATURES:
Raising/lowering the Frame:
Adjust to preferred height using the and arrows on
handset.
Move up and down faster
Legs Open/Close Function:
Adjust frame base using the
and buttons on
handset.
Left and right adjustment of armrests:
The armrests can be adjusted sideways for optimal support;
the whole arm can be moved in or out and locked with the
lever.
Locking/unlocking casters:
The walker is fitted with 2 brake casters (teal) and 2
directional locking caster (gray).
Depress the foot brake to set it and lock the caster.
To unlock, press down on the protruding part of the brake.
Locking affects rolling and swiveling.

800.255.3126 | wyeastmedical.com
52521 R01 Whitney Safety Platform Walker
Operator’s Manual
Page 9 of 20
WHITNEY
Safety Platform Walker
Adjusting the shin supports:
a & b - knobs for height adjustment
b - knob for locking shin supports in forward position
Adjusting the handles:
Turn the butterfly knob and adjust handles appropriately
Handset:
1. To Raise frame use
and to go faster
2. To Lower frame use
and to go faster
3. To Open Legs use
4. To Close Legs use
FEATURES:
Lower
frame
Fast
Up
Expand
Whitney
legs
Raise
frame
Fast
Down
Close
Whitney
legs
a
b

800.255.3126 | wyeastmedical.com
52521 R01 Whitney Safety Platform Walker
Operator’s Manual
Page 10 of 20
WHITNEY
Safety Platform Walker
FEATURES:
Control Box Wiring
Remove the control box cover to make the connections as
described below. Use a screwdriver or similar; refer to the
special labels on the control box.
NOTE! Make sure that plugs are connected to the correct
sockets otherwise the actuators can be damaged.
1. Pendant socket.
2. Battery terminal socket.
3. Socket for leg width adjustment actuator.
4. Socket for the platform height adjustment actuator.
Battery Charging
• Connect plug to 110V wall socket. Charging begins
automatically and the battery LED flashes green.
• Make sure the battery is fully charged before first
use. It takes around 24 hours to fully charge the
battery.
2. Battery
4. Platform
Actuator
1. Pendant
(behind
battery plug)
3. Leg
Actuator

800.255.3126 | wyeastmedical.com
52521 R01 Whitney Safety Platform Walker
Operator’s Manual
Page 11 of 20
WHITNEY
Safety Platform Walker
FEATURES:
* Information extracted from
page 102 of the LINAK CA30/CA40
MEDLINE CARELINE user manual
LEDs on LINAK Control Box and Battery*
Control box CA30/CA40 is equipped with a three color LED
for indication of mains or battery operation.
Connected to MAINS
• Green – On mains, not activated by hand or foot control.
The system is working ok and is ready for normal
operation
• Yellow – On mains, not activated by hand or foot control.
The system is defective and should not be operated
• Yellow – On mains, activated by hand or foot control.
The system is working
Not connected to mains but with BATTERY back-up
• Orange – On battery, activated by hand or foot control.
The system is working
• No LED – On battery, not activated by hand or foot
control or CA30/CA40 not connected to mains

800.255.3126 | wyeastmedical.com
52521 R01 Whitney Safety Platform Walker
Operator’s Manual
Page 12 of 20
WHITNEY
Safety Platform Walker
OPERATING THE WHITNEY SAFETY PLATFORM WALKER:
1. Fit the harness as low as possible across the hips and bottom with the broad upper part around the
hips and the narrow lower part around the bottom.
2. Place waist strap around the user and tighten using the adjustable safety belt. (Note that the lower
part of the harness attached around the patient’s bottom can be opened as required (i.e. toileting).
3. Place the user’s legs at a 90 degree angle with their feet on the floor or platform. Adjust the shin
supports to fit the user’s shins. (Remember to avoid pressure on the patellae and patellar tendons)
4. Maximize the base frame width to get close to the user.
5. Make sure the belts are threaded completely through the locking devices. (Note: Always check that
the belts are not damaged, worn, or frayed and that the holder and straps are not bent or crooked).
6. Connect the power rise harness and attachment belts and tighten them using the loops; the belts
must be taught on both sides. Slide the loop back once the straps have been tightened.
7. Swing the forearm supports to the side; the user must now grasp the handles without support for the
forearms.
8. Lock the two front casters.
9. Before the user raises to a standing position, the caregiver must stand in front of the user ensuring
stability during sit-to-stand raising. After the user is standing, check that the harness pulls evenly on
both sides; if not, adjust immediately.
10. In most cases the patient will not achieve a fully upright standing position; encourage the patient to
achieve a final upright posture independently or assist the patient by tightening the adjustable harness
straps.
11. Lower the walker to its individually adjusted height and swing the forearm supports back into position
to provide support for the user. (Note: if the platform is not being used, swing the shin supports away
for increased walking room).
12. Use directional locks for setting direction of travel. For example, you may consider locking just one
side to help correct for a gait drift. You can lock either, both, or unlock directional as therapeutically
appropriate.

800.255.3126 | wyeastmedical.com
52521 R01 Whitney Safety Platform Walker
Operator’s Manual
Page 13 of 20
WHITNEY
Safety Platform Walker
5.66”
MAINTENANCE & ASSISTANCE
TECH SPECIFICATIONS:
PARAMETER SPEC
• Maximum User Weight 330 lb
• Part No. 52500
• Product Width 25.98” - 41.73”
• Length 39.76”
• Height 37.4” - 53.15”
• Weight 88.18 lb
• Armrest/Internal Dimension 13.39” - 25.59”
• Wheels 3” Diameter
• Leg Under bed/Stretcher Clearance 4.99” or 5.66”
39.76”25.98” - 41.73”
13.39” - 25.59”
37.4” - 53.15”
4.99”
Low Base Option

800.255.3126 | wyeastmedical.com
52521 R01 Whitney Safety Platform Walker
Operator’s Manual
Page 14 of 20
WHITNEY
Safety Platform Walker
GENERAL MAINTENANCE ADVICE:
Recommended at least once a year to ensure trouble-free operation.
Not recommended for use in shower.
Do not pressure-wash the unit.
• Before use, check that the belt holders are not bent.
• Before use, check that the harness is not more than 3 years old (see labeling).
• Wipe the walker, handle grips and its arm rests clean as necessary; use soap solution/surface
disinfectant. Please note that the product is not made out of stainless material – frequent use of
water/detergents may therefore shorten the life of the product.
• Check the walker for loosened parts.
• Check that the walker is level and its castors are in contact with the floor.
• Check to make sure all cables are securely connected.
• Check for mechanical damage.
• Make sure the height adjustment clamps lock.
• Verify platform raising and lowering are functional.
• Check that the casters are secure, and freely roll.
• Replace damaged parts immediately; must be done by experienced personnel.
• The product is only intended for indoor use.
TROUBLESHOOTING:
SYMPTOM POSSIBLE CAUSE ACTION
Rattling Parts have become loose Tighten bolts/nuts as necessary
Handle grips damaged Use or misuse Replace handle grips
Handle adjustment not
functioning Fit lock knobs/plates as required
Arm rests won’t widen Replace lock knob
No power Battery is not fully charged. Charge battery
Battery is damaged. Replace battery.
Mains cable is plugged into
wall outlet. Unplug mains cable.
Battery, actuators or control
box are loose Tighten or replace fasteners
Battery will not charge Battery is damaged. Replace battery.

800.255.3126 | wyeastmedical.com
52521 R01 Whitney Safety Platform Walker
Operator’s Manual
Page 15 of 20
WHITNEY
Safety Platform Walker
SYMPTOM POSSIBLE CAUSE ACTION
Charging cable is not
securely connected
Unplug and replug the charging
cable to the control box
Armrest platform will not raise Too much weight on the lift. Check the weight of person
being lifted.
Actuator rod has been
damaged.
Check for visual signs of side
impact. Call for service.
Broken connection. Check that all cables completely
pushed into control box.
Cable damaged. Replace cable.
Leg adjustment won’t
respond Actuator is damaged.
Check for physical damage to
housing and/or actuator rod.
Call for service.
Broken connection. Check that all cables completely
pushed into control box.
Cable damaged. Replace cable.
Adjustments are moving
slowly Battery has a low charge. Charge battery.
Brakes aren’t working Caster brakes are broke Replace caster assembly
Handbrake has adjusted too
much or broken
Adjust the brake or fit new
casters
Unusual noise from actuator. Actuator is worn or damaged
or spindle is bent.
Replace actuator.
Call for service.
Standing belt is more than 3
years old or not intact Use or misuse Replace standing belt.
Available in S, M, & L
Standing belt slides off belt
holder Walker lockplate broken Replace lockplate
Standing belt holder is bent Product has been overloaded Replace frame before further use
TROUBLESHOOTING: (cont.)

800.255.3126 | wyeastmedical.com
52521 R01 Whitney Safety Platform Walker
Operator’s Manual
Page 16 of 20
WHITNEY
Safety Platform Walker
BELT CLEANING:
Laundering Instructions
Wipeable Cleaning Instructions
1. Using an approved disinfectant solution(see Table 1.0), either spray or wipe full surface of the belt
including all buckles and strap completely. Be sure to wear gloves and protect skin from chemical
exposure.
2. Soak all surfaces until damp, air dry.
3. Do not use on patient if still wet with chemicals. If soil unable to be removed in this method, launder
according to laundering instructions in user manual.
4. Use of chemical cleaning for disinfecting may cause material color to fade and sling to wear more
rapidly than with typical laundering.
Table 1.0 Approved Disinfecting Solutions for Whitney Belt (S 52530, M 52501, L 52502, XL 52529)
CHEMICAL CLASS ACTIVE INGREDIENT pH CLEANER/
DISINFECTANT MANUFACTURER
Quarternary ammonium
chloride
Didecyl dimethyl ammonium chloride = 8.704% Alkyl
dimethyl benzyl ammonium chloride = 8.19%
9.0-10.0
in use Virex II (256) Johnson/Diversey
Quarternary ammonium
chloride
Alkyl dimethyl benzyl ammonium chloride = 13.238%
Alkyl dimethyl ethylbenzyl ammonium chloride = 13.238% 9.5 in use HB Quat 25L 3M
Phenolic Ortho-Phenylphenol = 3.40% Ortho-Benxyl-para-
Chlorophenol = 3.03
3.1 +/-0.4
in use Wexcide Wexford Labs
Alcohol Isopropyl alcohol = 70% 5.0-7.0 Viraguard Veridien
Quaternary ammonium
n-Alkyl dimethyl benzyl ammonium chlorides= 0.105%
n-Alkyl dimethyl ethlybenzyl ammonium chlorides=
0.105%
11.5-12.5 CSI Central Solutions Inc
Benzyl-C12-18-
alkyldimethylammonium,
chlorides
Benzyl-C12-18-alkyldimethylammonium, chlorides (22%)
2-Phenoxyethanol (20%)Tridecylpolyethylenglycolether
(15%)Propan-2-ol (8%)
Approx.
8.6 in
use
Terralin Protect Shulke
Ethanol Adipic acid 10-30%Amorphous silica < 1%Sodium
Toluene sulphonate 5-10% Troclosene sodium 10-30% 7 Anioxy-Spray WS Anios

800.255.3126 | wyeastmedical.com
52521 R01 Whitney Safety Platform Walker
Operator’s Manual
Page 17 of 20
WHITNEY
Safety Platform Walker
ELECTRONICS:
BATTERY CARE AND CAUTIONS:
Battery packs may emit hydrogen gas. So do not bring fire or a heated object close to
the battery pack, and never use the battery near a spark, fuses and/or equipment that
emits sparks. Further, do not store the battery in a closed environment or incorporate
it into a closed structure of an enclosure. Doing so can cause an explosion, fire,
equipment damage and bodily injury.
Do not connect the positive terminal and the negative terminal of the battery packs with
a wire or other metals.
Short-circuiting the terminals of the battery can cause burn injuries, damage to the
storage battery or trigger explosions.
Never connect the battery packs directly to a power supply socket or an automobile’s
cigarette lighter without using a charger as a medium. Connecting the battery directly
can cause the battery to leak fluid, generate heat, explode, cause fires or bodily burns
and injuries.
Battery packs contain dilute sulphuric acid, which is a toxic substance. If the battery’s
internal fluid leaks out and gets onto your skin or clothing, make sure it is washed off
with clean water. Additionally, if the fluid gets in your eyes, wash them with clean water
immediately, and see a doctor. When dilute sulphuric acid gets in your eyes, it can
cause a loss of eyesight and when it gets on your skin, it can cause a burn on your
skin.
Do not use or store LINAK battery packs where the surrounding temperature exceed
50°C, such as inside a hot automobile, in direct sunlight, or in front of a stove or a
source of intense heat. Doing so can shorten battery life, lower its performance level,
cause the battery to leak fluid, be damaged or deformed.

800.255.3126 | wyeastmedical.com
52521 R01 Whitney Safety Platform Walker
Operator’s Manual
Page 18 of 20
WHITNEY
Safety Platform Walker
ELECTRONICS ENVIRONMENTAL CONDITIONS:
OPERATING:
Temperature_______________________________________________________ 5°C to 40°C
Relative humidity_____________________________ 20% to 90% @ 30°C – not condensing
Atmospheric pressure __________________________________________ 700 to 1060 hPa
STORAGE:
Temperature____________________________________________________ -10°C to +50°C
Relative humidity_____________________________ 20% to 90% @ 30°C – not condensing
Atmospheric pressure __________________________________________ 700 to 1060 hPa
TRANSPORT:
Temperature____________________________________________________ -10°C to +50°C
Relative humidity_____________________________ 20% to 90% @ 30°C – not condensing
Atmospheric pressure __________________________________________ 700 to 1060 hPa
If the actuator is assembled in the application and is exposed to push or pull during
transportation, the actuator can be damaged. Do not drop an actuator or otherwise
damage the housing during disassembly or transportation. We do not recommend using
an actuator which has been damaged.

800.255.3126 | wyeastmedical.com
52521 R01 Whitney Safety Platform Walker
Operator’s Manual
Page 19 of 20
WHITNEY
Safety Platform Walker
FACTORY ASSISTANCE:
TROUBLESHOOTING ASSISTANCE:
For assistance in troubleshooting, contact Wy’East Medical customer service or your Wy’East
Medical representative.
Include the following information with your request:
1. Serial and Model numbers of your Whitney Walker Safety Platform Walker.
2. Purchase date, if available.
RETURNING ITEMS:
To return parts or goods, contact Wy’East Medical for authorization.
Contact information:
Telephone: (503) 657-3101
Toll Free: (800) 255-3126
Fax: (503) 657-6901
Email: admin@wyeastmed.com
WARNING
No unauthorized modification of this equipment is allowed. Contact Wy’East Medical prior to any
modification of this equipment

800.255.3126 | wyeastmedical.com
52521 R01 Whitney Safety Platform Walker
Operator’s Manual
Page 20 of 20
WHITNEY
Safety Platform Walker
NOTES: