Yosemite Y600-A User manual

Y600-A
Multifunctional Handheld
Water Quantity Meter
Manual
Yosemite Technologies Co.,Ltd

- 1 -
CONTENTS
Preface.................................................................................................................................................... - 2 -
1 Overview............................................................................................................................................ - 3 -
1.1 Introduction ............................................................................................................................... - 3 -
1.2 Specifications............................................................................................................................. - 4 -
2 Setting-up........................................................................................................................................... - 5 -
2.1 Configuration............................................................................................................................. - 5 -
2.2 Installing the batteries............................................................................................................. - 5 -
2.3 Installing the sensor................................................................................................................. - 6 -
3 Operation ........................................................................................................................................... - 7 -
3.1 Date/Time................................................................................................................................... - 8 -
3.2 System......................................................................................................................................... - 8 -
Auto Power.................................................................................................................................. - 9 -
Air Pressure.................................................................................................................................. - 9 -
Salinity.........................................................................................................................................- 10 -
Probe Info ..................................................................................................................................- 10 -
Meter Info ..................................................................................................................................- 10 -
3.3 Calibration................................................................................................................................- 11 -
One-Point calibration ............................................................................................................- 12 -
Two-points calibration ..........................................................................................................- 13 -
Sensor Cap.................................................................................................................................- 14 -
3.4 Data Log....................................................................................................................................- 14 -
Data Store..................................................................................................................................- 14 -
Data View...................................................................................................................................- 15 -
Data Delete................................................................................................................................- 15 -
Cal......................................................................................................................................................- 16 -
Connecting to a Computer ..........................................................................................................- 16 -
4 Maintenance....................................................................................................................................- 17 -
4.1 The instrument maintenance............................................................................................- 17 -
4.2 DO maintenance ....................................................................................................................- 17 -
5 Common Errors and Trouble Shooting.......................................................................................- 19 -

- 2 -
Preface
Dear customer
Thank you for your Yosemite Technologies’ Y600-A instrument purchase. Reading the entire
manual before use is highly recommended for operation and maintenance the instrument and out
of unnecessary trouble.
Please observe the operating procedures and precautions in this manual.
To make sure the effective after-sales protection provided by the instrument, please do not use
any operation or maintenance other than which mentioned in the manual.
Due to non-compliance with the precautions specified in this manual, any fault and loss caused
shall not be covered by the warranty, and the manufacturer shall not bear any relevant
responsibility. If you have any questions, please contact our after-sales service department or
representative.
Carefully unpack the instrument and accessories from the shipping container, and inspect for
possible damage during shipping. Check received parts with items on the packing list. If any parts
or materials are damaged or missing, please contact Yosemite Technologies customer service or the
authorized distributor immediately.
Save all packing materials until you are sure that the instrument functions properly. Any damaged
or defective items must be returned in their original packaging materials.

- 3 -
1 Overview
The Y600-A instrument is independently developed by Yosemitech. It provides keyboard and
backlight with comprehensive functions, simple operation and brief interface. The instrument is
a portable device that can automatically recognize Yosemite's digital sensors including optical
dissolved oxygen, 4-electronic conductivity, turbidity, chlorophyll, BGA, OIW and TSS ,and can
measure, display, log. The microprocessor based meter can transmit USB signals and perform
sensor calibration easily on a single platform.
The dissolved oxygen sensor adopts the internationally leading fluorescence lifetime technology,
which is based on the quenching principle of specific substances to active fluorescence in physics.
The remarkable advantage of this fluorescence method in measuring dissolved oxygen is that it
does not consume oxygen in the measurement process, so there is no flow rate limit, no need to
preheat electrolyte, no need for maintenance and frequent calibration. The response time of
Yosemitech optical dissolved oxygen probe is as low as 30 seconds, making the measurement of
dissolved oxygen more accurate, more stable, faster and more convenient.
1.1 Introduction

- 4 -
1.2 Specifications
Y506-A Optical dissolved oxygen Sensor
Sensor Type
immiscible
Size
Φ16×180mm
Weight
51g
Cable length
5m default, 10m,15m and 30m optional
Min cable bending radius
15mm
Communication protocol
RS-485, MODBUS
Range
0-20mg/L or 0-200% air Sat.
Accuracy
±1%
Response time
30 sec
Drift
<1% /Year
Temperature accuracy
±0.2℃
Working temperature
0~ 50℃
Storage temperature
-40 ~ 85℃
Calibration method
One-point or two-point calibration
Sensor cap lifetime
One year typical
Y600-A Instrument
Size
220 x 96 x 44mm
Weight
460g
Power
Two 18650 rechargeable batteries
Working temperature
0~50℃
Storage temperature
-40~85℃
Display
54.38 x 54.38LCD back illuminated
Data storage
Support
Pressure compensation
Built-in,auto compensation in 50~115kPa
IP rating
IP67
Regular Shutoff
Support internal storage, upto 5000 data set

- 5 -
2 Setting-up
2.1 Configuration
2.2 Installing the batteries
The instrument is designed to operate with two lithium batteries. Please follow the steps below.
1)Unscrew the screws on the battery compartment and open it. Note: The screw can't be taken
off. That means the battery compartment cannot be opened until the screw can't be turned;
2)Inspect battery compartment and install two batteries. Please pay attention to battery
polarization marks inside the battery compartment, as shown in the figure below.
3)Close the battery compartment. Note that the sealing ring cannot fall off or be uneven. Make
sure correct direction of compartment and tighten up the screws.
Item
Number
Note
Y600-A Instrument
1
Y506-A optical sensor
1
5m default,optional
18650 rechargeable battery
2
USB cable line
1
Micro USB
Wristband
1
Screwdriver
1
To open the battery compartment
O-ring
1
Sensor cap sealing part
Sponge
2
To moisturize sensor cap
Protect Cap
1
To store the sensor cap
Check the individual parts against the list of components. If anything is missing or
damaged, notify after-sales service department immediately
Note
P

- 6 -
2.3 Installing the sensor
The sensor cable has a red slot on the mini-BNC connector which must align with the connector at
the base of the instrument. Then insert gently and turn clockwise until the “click” sound is heard.
When remove the sensor, push the sensor cable connector in and turn out anticlockwise.
Install:
Remove:

- 7 -
3 Operation
Press key to turn the Y600-A on. When there is no sensor, the measurement interface will
display “No Signal!”. If the dissolved oxygen Sensor is connected, the display will show DO values
as shown in the figure below.
Keypad Fouction
1
Time(min:sec)
2
Sensor Type
(DO/CT/TUR/CHL/pH(ORP)/BGA/COD/OIW/
NH4N/TSS /MULTI)
3
Battery indicator.
If appears, charge the meter using
USB cable provided with directly or
rechargeable batteries removed from meter.
4
Sensor data:DO values in percentage unit and
in mg/L unit.
5
Temperature
1
Short press
Turn the Y600-A on
Long press
Turn the Y600-A off
2
Short press
Switch to Menu interface
3
Short press
Turn the backlight on and
off
4
Short press
Exit settings;
Return to the previous
interface
5
Short press
Scroll up the menu
selection
Increase figure
6
Short press
Scroll down the menu
selection
Decrease figure
7
Short press
Accept a menu selection
Accept setting and values
8
Short press
Switch to Calibration
interface
Long press
Switch to factory
Calibration recovery
interface
Short press:press and hold less than 2sec
Long press:press and hold more than 2sec

- 8 -
3.1 Date/Time
3.2 System
Date/Time
System
Calibration
Data Log
11:25
Menu
DO
Menu
Press the Menu key to switch to
menu interface. To select a menu
item, use the “▲▼” key to
highlight the selection and press
the “↙” button. To exit a menu or
menu item, press the ESC key.
Date: / 09 / 19
Time: 15 : 10 : 07
Save
DO
11:25
Date/Time
19
Auto Power
Air Pressure
Salinity
Probe Info
Meter Info
11:25
System
DO
Date/Time menu is where you can
setup date and time. Use the ▲or
▼keys to scroll through the
highlight data, and then press the
“↙” key to move on to next data.
Select Save and press the “↙” key
to accept desired Data/Time. Then
will shows “Save Success!”.
System menu is where you can
enquire and setup all parameters.
These include Auto Power, Air
Pressure, Salinity, Probe Info, and
Meter Info. To switch to a menu
item, use the “▲▼” keys to select
an option, and the “↙”key to
accept.

- 9 -
Current: 101.6
Standard:
000.0
Save
DO
11:25
Air Pressure
1
Auto Power
In Auto Power, there are four
mode:
OFF indicates the power supply is
normally open,
When set to 5 min , 10 min, 15 min,
the instrument will shut off after 5
minutes, 10 minutes, 15 minutes
of no operation,
Use the “▲▼” keys to select the
mode, and the “↙”key to accept.
The instrument use 5 minutes by
default.
Air Pressure
This menu can do Air calibration in
kPa as it will effect DO value.
Current indicates measured air
pressure. Standard indicates the
desired pressure that to be
calibrated. Make adjustments with
the ▲or ▼arrow keys and the
“ ↙ ” key to move on to next
value. Select Save and press the
“↙” key to accept.
OFF
5 min
10 min
15 min
11:25
Auto Power
DO

- 10 -
SN: YL0219090909
HW VERSION: 1.1
SW VERSION: 2.0
11:25
Meter Info
DO
Meter Info
This menu can check information of
the instrument including SN, HW
version, SW version.
SN: YL5019081516
HW VERSION: 1.1
SW VERSION: 2.0
ID: 01
Modify
11:25
Probe Info
DO
ID:
Unit: PPT
Current: 0.0
Input: 000..0
Save
DO
11:25
Salinity
5
Salinity
This menu can setup salinity value. The
amount of DO increases as salinity level
decreases. The instrument can
compensate for the deviation of DO value
caused by salinity. Make adjustments
with the ▲or ▼keys and then press the
“↙” key to turn to next data. Select Save
and press the “↙” key to accept desired
salinity. The default value is 0 ppt.
Generally, the fresh water is 0-0.5ppt, and
the sea water is 35ppt.
Probe Info
This menu can check information of
sensor including SN, HW version,
SW version, Modbus ID. Modbus ID
can change with the ▲or ▼keys.
Select Modify and press the “↙”
key to accept new ID address.

- 11 -
3.3 Calibration
The instrument supports DO both single-point and two-points user calibration. Single-point DO
calibration means "100.0%SAT" calibration which performed in water saturated air or air saturated
water. For best results, we strongly recommend to use air saturated water solution. Two-points
calibration consist of 100.0%SAT and 0%SAT. There is no Zero calibration here.
Press and hold the “Cal” key can quickly switch to calibration interface.
Before calibration, please move protect cap with damped sponge from the sensor cap of DO
and make sure the sensor has connected to the instrument.
Two Points
Sensor Cap
11:25
Calibration
DO
One/Zero Point
Point
0%SAT DO calibration solution: take anhydrous sodium sulfite (Na2SO3) to a beaker, add 200ml of
tap water or pure water, and stir until dissolved or even precipitates crystals.
100%SAT DO calibration solution:Aerate the water with an oxygen pump and continue 15 minutes.
Then can get air saturated water solution.
* The air can be considered as saturated oxygen environment if the condition is not allowed.
Note
P

- 12 -
One-Point calibration
⚫Follow the screen instruction “In 100%”,put the DO sensor into 100% SAT solution, and press
the "↙" key to move on.
⚫Then “Waiting 3 min…”means wait for 3 minutes to stabilize the data. The measured DO value
in % unit will flash on the bottom of screen.
⚫After the data is locked, it will be displayed on the right side of the screen. And the instrument
will calculate the slope .The prompt "success" or "fail" indicates calibration result. Press the
"ESC" key to exit the calibration.
Please do not unplug the sensor during the entire calibration process.
In100%
11:25
OnePoint
DO
Click Enter P
In100%
93.62
11:25
OnePoint
DO
Waiting 3 min… P
The measured
DO value
Note
P

- 13 -
Two-points calibration
First performed 100% SAT calibration, then start calibrating 0% SAT. Follow the screen instruction
“In 0%”,put the DO sensor into 0% SAT solution, and press the "↙" key to move on. Wait for 3
minutes to stabilize the data. The prompt "success" or "fail" indicates calibration result after
calculating the user calibration parameters. Press the "ESC" key to exit.
In 100%
Success
91.56
In 0%
Fail
20.56
11:25
TwoPoints
DO
Locked value at
0%
Calibration
failed
Locked value at
100%
In 100%
Success 91.56
91.56
11:25
OnePoint
DO
Cali Done!
Locked value at
100%

- 14 -
3.4 Data Log
Data Store
The Y600-A instrument has 4096 measurements that can be logged to memory and can be
programmed to store measurements in two logging formats. When set to Data Log/Data
Store/Store 1 Times, the user samples start to be logged a time. When set to Data Log/Data
Store/Store 50 Times, the instrument will continue logging 50 times with an interval of 2 seconds.
K4 : 010939
K5 : 829976
K6 : 804008
K7 : 051952
11:25
Write Para
DO
Save
Write Para
11:25
Sensor Cap
DO
Read Para
Sensor Cap
To update the sensor cap, you need
to input correct coefficients specific
to the sensor cap. There are K0 to
K7 total 8 parameters.
Select Read Para and press the “↙”
key to switch to read sensor cap
parameters interface. These consist of K0-
K7 8 parameters. To change the value,
highlight the Write Para option and press
the “↙” key. Make adjustments with
the ▲or ▼keys and then press the “↙”
key to move on to next data. Select Save
and press the “↙” key to accept all new
coefficients.

- 15 -
Data View
Data Delete
*Select Yes to delete all data. Please choose carefully.
Max Size : 4096
Use Size : 54
Current Sel : 54
11:25
Data View
DO
190920 15:32:37
Type : DO
SN: YL0119081516
DT: 190920 15:32:37
Pressure: 101.13
Salinity: 0.00
Celsius: 27.82
Percent : 99.81
mg/L: 7.79
11:25
Data View
DO
The stored measurements are
named after date and time the
samples were taken. For example,
190920 15:32:37 means the sample
item 54 is stored at 15:32:37 on
September 20, 2019. Use the ▲or
▼keys to scroll through the other
Current Sel and press the “↙” key to
view all the measurement
parameters.
Max Size
Indicates the numbers of
measurements to be logged.
Use Size
Indicates the numbers of
measurements have been logged.
Current Sel
Indicates the selected sample item
that contains the data to be
recalled. As shown in below
picture, currently, the sample item
54 is read.
.

- 16 -
Cal
Connecting to a Computer
The Y600-A instrument comes with special software called MeterRead, which is designed to
provide you access to data memory. The software is stored in the USB shipped with the goods.
Follow the instructions to install the software on your computer and connect the supplied
computer interface USB cable to the Y600-A and to a computer USB port. Then the COM PORT can
identify the port. Select the download data path and click Load. The file format is. CSV.
Remember in measurement interface where press the “Cal” key to switch to calibration
interface.
The software cannot recognize the port only when the instrument in shut off state.
NO
DO
11:2
Reset
YES
Press the “Cal” key to switch to
calibration interface. Please see
chapter 3.3 in detail.
Press and hold the “Cal” key to
switch to factory Calibration
recovery interface where restore k
= 1, b =0.
Note
P
Note
P
注意
P

- 17 -
4 Maintenance
4.1 The instrument maintenance
The 600-A instrument require very little maintenance to keep it performing well. If the instrument
case needs cleaning, wipe it off with a soft cloth. Please keep the battery compartment dry during
using.
4.2 DO maintenance
Optical dissolved oxygen sensor is a maintenance free sensor, however to ensure its accuracy
specifications, we strongly recommend to perform routine maintenance based on following
schedule.
Maintenance
Frequency Recommended
Sensor clean up
Every 30 days
Sensor and DO sensor cap inspection
Every 30 days
DO sensor cap replacement
Every year
Sensor user calibration
Every 3~6 months
Moisturizing DO membrane surface
Every 30 days to Keep DO sensor storage cap wet
Maintenance Schedule:
1) Clean the sensor case:Use clean water to wash the sensor case. Then gently wipe off with
a lens cleaning tissue that has been moistened. For some stubborn dirt, please use gentle
home dish cleaner and a soft brush.
2) Clean the sensor cap surface:Please move protect cap from the sensor cap, rinse any fouling
with clean water. Please gently wipe off with a lens cleaning tissue that has been moistened if
necessary. Then install the protect cap. Do not scratch the DO membrane with nails or sharp
objects. That will cause DO membrane damage and the sensor will not work permanent!
3) Clean the sensor cap internal surface:If water vapor or dust intrudes into the sensor cap, the
cleaning steps are as follows:
i. Carefully remove the sensor cap
ii. Use clean water to wash the sensor cap internal surface
iii. You can use gentle home dish cleaner if necessary
iv. Then Followed with clean water wash
v. Dry up with lens cleaning tissue
4) Store DO Sensor Cap:Keep DO sensor storage cap wet to moisturize DO membrane surface.
Store the sensor cap inside protect cap with a piece of damped sponge. Please take off the
protect cap and keep it properly during use. Note: if the sponge is dry and lack of water,

- 18 -
deionized water or purified water can be added for wetting; if DO membrane is kept in a dry
state for a long time, it will lead to inaccurate measurement and greatly shorten the life of DO
membrane.
5) Inspect sensor cable:The cable should not be tightened in use; otherwise there is a risk of
breaking the wire. The sensor will not work normally;
6) The sensor case is inspected for damage due to corrosion or other reasons.
DO Sensor Cap replacement
1)Please turn off the power and unplug the sensor from the instrument before perform sensor
cap replacement;
2)Hold the sensor by one hand, and remove the sensor cap from the sensor by rotating the sensor
cap counterclockwise using the other hand. DO NOT USE ANY TOOLS.
3)Inspect the O-ring on the probe carefully. If damaged, replace it. Do not use tools to remove or
install o-rings.
4) Ensure the O-ring installed is clean.
5) Hold the sensor by one hand, and install the new sensor cap clockwise using the other hand.
And dry the inside surface with clean tissue. DO NOT USE ANY TOOLS.
6)After replaced, set the new characteristic parameters that came with the product to the
instrument. Please see chapter 3.3 in detail.
Do not touch the DO membrane with hand.
Do not apply any mechanical stress directly to the DO membrane (pressure, scratches, etc.)
Note
P
注意
P
注意
P
注意
P
注意
P
注意
P

- 19 -
5 Common Errors and Trouble Shooting
Table 5-1 lists the symptoms, possible causes, and recommended solutions for common problems
encountered with the Y600-A instrument. If your symptom is not listed, or if none of the solutions
solves your problem, please contact us.
Table 5-1 Troubleshooting
Symptom
Possible Causes
Corrective Action
No data displayed on
Meter (if sensors are
connected)
Connection Issue
Reconnect sensor to the meter
No batteries or batteries are low
Charge the meter using USB cable
provided with directly or
rechargeable batteries removed
from meter.
Cable problem
Contact Yosemite Technologies
Customer Service
Bad DO sensor cap or loose cap
Reinstall sensor cap or replace
Unstable DO value
Dirty sensor cap
Clean sensor cap and agitate the
sensor to eliminate air bubbles
during measurement.
Bad sensor cap
Replace sensor cap
Sensor cap expired
Table of contents