Adam CBC-M Series User manual

.
Adam Equipment
CBC-M SERIES
Service Manual
Software rev.: 3.06A & above
ADAM EQUIPMENT CO. LTD.
P.N. 8227, Rev. B5, October 2007
© Adam Equipment Company 2007

© Adam Equipment Company 2007

© Adam Equipment Company 2007 1
CONTENTS
1.0 INTRODUCTION.......................................................................................................2
2.0 SPECIFICATIONS....................................................................................................3
3.0 TROUBLESHOOTING..............................................................................................4
4.0 ERROR CODES .......................................................................................................5
5.0 CBC-M DESCRIPTION.............................................................................................6
6.0 LOAD CELL DAMAGE..............................................................................................6
7.0 CALIBRATION..........................................................................................................9
7.1 CALIBRATION PROCEDURE...............................................................................9
7.2 SECURITY SEALS..............................................................................................10
7.3 CALIBRATION PROCEDURE.............................................................................10
8.0 SERVICE PARAMETERS.......................................................................................11
8.1 USING “0000” TO ENTER THE CALIBRATION PARAMETER............................11
8.2 USING “2006” TO ENTER THE SERVICE PARAMETERS..................................12
8.2.1 F1 -CALIBRATION............................................................................................12
8.2.2 F2 -RESOLUTION ............................................................................................13
8.2.3 F3 –CAPACITY................................................................................................13
8.2.4 F4 -INITIAL ZERO RANGE ...............................................................................14
8.2.5 F5 -RE-ZERO RANGE......................................................................................14
8.2.6 F6 -SUCCESSIVE TARE..................................................................................14
8.2.7 F7 -A/D COUNT.................................................................................................15
8.2.8 F8 –ZERO.........................................................................................................15
8.2.9 F9 –PIN.............................................................................................................15
8.2.10 F10 –LOW VOLTAGE DETECTION..................................................................16
9.0 WIRING DIAGRAM.................................................................................................17
10.0 MECHANICAL ASSEMBLY ....................................................................................18
10.1 LIST OF MECHANICAL ASSEMBLY...................................................................19
11.0 DISPLAY ASSEMBLY.............................................................................................20
*****

1.0 INTRODUCTION
•The CBC-M series of scales provide an accurate, fast and versatile counting
and check-weighing scales.
•CBC-M scales are kilogram only scales
•There are 4 models of approved scales with capacities up to 30kg.
•They all have stainless steel weighing platforms on an ABS base assembly.
•All scales are supplied with a RS-232 bi-directional interface and real time
clock (RTC)
•All scales have sealed keypad with colour coded membrane switches and
there are three large, easy to read liquid crystal type displays (LCD). The
LCD’s are supplied with a backlight.
•The scales include automatic zero tracking, audible alarm for pre-set
weights, automatic tare, an accumulation facility that allows the count to be
stored and recalled as an accumulated total.
© Adam Equipment Company 2007 2

© Adam Equipment Company 2007 3
2.0 SPECIFICATIONS
CBC-M SERIES
Model CBC 3M CBC 6M CBC 15M CBC 30M
Maximum
Capacity 3 kg 6 kg 15 kg 30 kg
Readability 0.001 kg 0.002 kg 0.005 kg 0.01 kg
Tare Range -3 kg -6 kg -10 kg -30 kg
Repeatability
(Std Dev) 0.001 kg 0.002 kg 0.005 kg
0.01 kg
Linearity ± 0.002 kg 0.004 kg 0.01 kg
0.02 kg
Common Specifications
Stabilisation
Time 2 Seconds typical
Operating
Temperature 0°C - 40°C
32°F - 104°F
Power supply 230 VAC 50/60 Hz.
120 VAC available.
Battery Internal rechargeable battery
(~90 hours operation)
Calibration Not available
Display 3 x 6 digits LCD digital displays
Balance
Housing ABS Plastic, Stainless Steel platform
Pan Size 225 x 275mm
8.9” x 10.8”
Overall
Dimensions
(wxdxh)
315 x 355 x 110mm
12.4” x 14” x 4.3”
Net Weight 4.1 kg / 9 lb
Applications Counting Scales
Functions Parts counting, weighing, accumulating memory,
pre-set count with alarm
Interface RS-232 bi-directional interface
Date/Time Real Time Clock (RTC),
To print date and time information.
(Dates in year/month/day, day/month/year or
month/day/year formats- Battery backed)

© Adam Equipment Company 2007 4
3.0 TROUBLESHOOTING
1. Unit does not
turn on Check whether the scale is plugged into the power supply properly.
Check the operation of the charging circuit.
Check whether the battery is charged- LED should turn to green from red.
Check the adapter output.
.
2. The scale
weighs but is
unstable
Air drafts or vibration or unstable table.
Pan rubbing against case or not installed correctly.
Scale not installed properly.
Improper connections on ADC circuit.
3. Scale shows
wrong weight Scale not installed correctly.
Check the Calibration.
To check whether a part of product weighed is trapped between the pan
and the scale.
Incorrect adjustment of the mechanical stops.
4. An error “E4”
displays on
the screen
Load cell damage.
Scale not installed properly.
Incorrect adjustment of the mechanical stops.
Factory calibration has been tampered with- check the ADC counts and
re-calibrate.

© Adam Equipment Company 2007 5
4.0 ERROR CODES
ERROR CODE DESCRIPTION POSSIBLE CAUSES
Err 4 Initial Zero is greater than
allowed (4% of maximum
capacity) when power is turned
on or when the [Zero/Enter] key
is pressed.
Weight on the pan when turning
the scale on.
Excessive weight on the pan when
zeroing the scale.
Improper calibration of the scale.
Damaged load cell.
Damaged Electronics.
Err5 Keyboard error. Improper operation of the scale.
Err6 A/D count is not correct when
turning the scale on. Platform is not installed.
Load cell is damaged.
Electronics is damaged.
Err7 Scale is not stable when setting
the percentage weighing. The scale is not stable.
Improper operation of the scale.
FAILHor FAILL Calibration error. Improper calibration (should be
within +10% of the factory
calibration). The old calibration
data will be retained until the
calibration process is complete.
Err9 Scale is unstable. There is vibration or draft making
the scale unstable.
Electronics may be damaged.

© Adam Equipment Company 2007 6
5.0 CBC-M DESCRIPTION
The CBC-M scales have an enclosure with all components mounted within it. To gain
access to the components remove the 5 screws securing the cover to the base.
The basic unit consists of:
9Base
9Load Cell frame
9Power switch
9Main PCB assembly
9Battery
9Display PCB assembly
9Keypad
All models of CBC-M are similar except the selection of load cells and the program.
Normally if a problem is found with a circuit board the most cost effective method of
solving the problem is to replace the circuit board.
The main PCB assembly includes:
9A/D converter (AU1),
9Microprocessor (DU1) and
9Power supply/battery charger circuits (DU5, T2).
6.0 LOAD CELL DAMAGE
The most common reason a scale fails is that the load cell has been damaged. The
damage can be from 2 primary causes. The first is physical damage due to an overload or
an impact from the side and the second is damage due to the environment, such as
moisture, extreme heat or a cut cable if the cables are exposed.
PRINCIPLE OF OPERATION
The stain gauge load cell is a method of measuring the amount of stress put onto a metal
structure due to a weight being supported by the structure. The stress is measured using
4 strain gage sensors mounted in a particular way on the metal structure.
These 4 strain gauges are connected in such a way that they form a wheatstone bridge.
See figure below. In the simplest load cells there are only the 4 resistances of the strain

gages to consider. Normally without a load on the load cell all the resistances are the
same. However when the load cell has a weight on it 2 of the strain gages will be in
compression (A and D) and the resistance will decrease and 2 will be in tension (B and C)
and their resistance will increase.
350 ohm
350 ohm
350ohm
350ohm
AB
CD
+E
-S
-E
+S
BALANCED BRIDGE
10 VOLT POWER
SUPPLY
Vout = O VOLTS,
BETWEEN +S AND -s
351ohm
349ohm
351ohm
349ohm
AB
CD
+E
-S
-E
+S
UNBALANCED BRIDGE
FULL CAPACITY LOAD
ON THE LOAD CELL
10 VOLT POWER
SUPPLY
Vout = 0.028 VOLTS (28mv),
BETWEEN +S AND -s
This will cause the bridge to become unbalanced. When the bridge has a voltage across it
from +E to -E then the signal output at +S and -S will show a voltage of zero volts with no
load and a small voltage proportional to the load as the load is increased. Typical load
cells show 20-30mv of signal if the excitation voltage is 10volts and the load cell is fully
loaded.
Most load cells used in scales are not as simple as this example. They have additional
resistance elements added to compensate for temperature variations and to set the
outputs to correct voltages. The circuit of a typical load cell is shown below-
© Adam Equipment Company 2007 7

© Adam Equipment Company 2007 8
A method to check the basic function and integrity of a load cell can be done using a
ohmmeter and voltmeter with up to a 10volt power supply.
Before the load cell is connected to the power supply use the ohmmeter to measure the
resistance between the wires. Typical resistance values are:
+E to -E 410 ohms ±30 ohms
+S to -S 350 ohms ± 2 ohms
+E to +S or any other combination similar approximately 270 ohms.
If the load cell has Sense connections in addition they are connected to the excitation
internal to the load cell.
Connect the load cell +E and -E to a power supply, typically 10 volts. Never use more
than 12 volts as it may damage the load cell.
Measure the voltage between +S and -S.
With no load the voltage should be approximate 0mv ±3mv.
With a mass on the load cell the output voltage should increase. The amount by which it
increases will be dependent upon the sensitivity of the load cell, capacity of the load cell,
the excitation voltage and the amount of mass placed on the load cell. Most load cells
have a sensitivity of either 2mv/V or 3mv/V.
The expected change to the output is:
Vout = (2mv/V) * Vext * Mass
capacity of load cell
For example a 2mv/V load cell of 30Kg capacity with 20Kg placed on it. Vext is 5 volts
would give a output of approximately:
Vout =
(2mv/V) * Vext * Mass = 2mv/V * 5V * 20Kg = 6.7mv
capacity of load cell 30kg
If the load cell has been damaged the no load voltage will likely be greater than 3mv or the
loaded voltage will be grossly different from the expected value. Remember the loaded
voltage will be offset by the amount of zero load voltage measured.

The load cell can be tested while it is connected to the A/D converter circuit board, using
the scale power supply for excitation voltage.
TESTING LOAD CELLS IN THE SCALE
The signal from the load cells is amplified by the circuits on the A/D circuit board. The
signal needs to be amplified to make it acceptable for the A/D converter used. The CBW
scales amplify the input by 200 and change the no load voltage to about +2.5V. The
amplified and offset voltage is measured at pin 6 of IC7, the INA118P amplifier IC. This
voltage is sent to the filter and then to the A/D converter.
7.0 CALIBRATION
7.1 CALIBRATION PROCEDURE
The scales are sealed to prevent unauthorised calibration.
Contact Adam Equipment or your supplier for more details.
The calibration of the CBC-M scale is
accomplished by removing the label at
the base of the scale which covers a hole
through which the PCB can be accessed.
If your scale is not provided with this
hole, you need to break the security
seals and remove the top cover to gain
access to the circuit boards inside. See
the figure on the security seal below.
WARNING: CALIBRATION OF THE SCALES MAY MAKE IT ILLEGAL TO USE THE
SCALES FOR SALES OF GOODS. CONTACT YOUR TRADING STANDARDS
OFFICE FOR FURTHER ASSISTANCE.
© Adam Equipment Company 2007 9

7.2 SECURITY SEALS
To seal the scale a lead-wire seal can be
used as shown below. The base has a
metal pin that protrudes through the
cover. The security seal may be threaded
through this metal pin, making it
impossible to remove the cover without
damaging either the security seal or the
enclosure.
Metrology labels and additional security
measures may be added to the scale as
required by the national legislation.
An alternative method of sealing the scales is to have a seal covering the joint between the
base and cover. If this seal is broken the scale must be sealed by the relevant authorities,
using either the lead wire seal or an acceptable seal between the cover and base.
7.3 CALIBRATION PROCEDURE
The scales are sealed to prevent unauthorised calibration.
CBC-M scales need to be re-calibrated if the variation in gravity takes place due to change
in location or if the calibration done earlier was not correct.
The calibration is carried out by removing the label placed on the hole at the base of the
scale to gain access to the PCB inside.
The EC approved models should follow the requirements of the Metrological authorities to
have the scale verified for use after calibration.
If calibration is not possible, the load cells or PCBs may have been damaged. Contact
Adam Equipment or the supplier.
© Adam Equipment Company 2007 10

8.0 SERVICE PARAMETERS
8.1 USING “0000” TO ENTER THE CALIBRATION PARAMETER
Calibration can be enabled by placing a jumper into the pins on the main PCB (marked as
position J4) through a hole provided on the base of the scale. After the jumper has been
placed in the position shown, apply power to the scale.
Turn the scale upside down.
Remove the seal covering the hole to access
the PCB.
Place a jumper through the hole into the pins. If
a jumper is not available, short the pins.
Turn on the scale using the switch on the base.
“Pn”
When “Pn” is displayed.
Enter “0 0 0 0” and press [Tare]
“UnLoAd ”
Empty the pan by removing the load, if there is
any and press [Tare]
“LoAd” “6” “KiLoS” Load the requested calibration weight and press
[Tare]
“SPAn” “PASS” If Calibration is complete, “SPAn PASS” will be
displayed. Remove the calibration weight.
Or, “SPAn” “FAiLEd ” This means calibration has failed. Remove the
calibration weight and repeat the process.
“JP On” Remove the jumper or shorting of the pins
whichever is used. The scale will return to
normal weighing.
© Adam Equipment Company 2007 11

© Adam Equipment Company 2007 12
8.2 USING “2006” TO ENTER THE SERVICE PARAMETERS
After the jumper has been placed in the position shown, apply power to the scale.
The display will ask for a code number, “Pn “ on the Weight Display.
Enter the number 2006 when “Pn “ is displayed and then press [Tare].
The displays will show the first parameters, called “F1” “CAL”.
To select another parameter press the [U.Wt./Units] key to advance through the
parameters.
Press the [Tare] key to enter a parameter.
To exit a parameter, press the [Zero] key.
The Weight window will show the parameter number and the Unit Price and Count
windows will show the word describing the function.
When a parameter is entered by pressing the [Tare] key, the displays will guide you
through the parameter selected and the options available.
The parameters available are:
“F1” “CAL” To enter the Calibration
“F2” “rES” Resolution selection
“F3” “CAPA” Select capacity
“F4” “init” “ZEro” Initial Zero Range
“F5” “rE” “Zero” Re-Zero range
“F6” “SCSivE” “tArE” Successive Tare Enable
“F7” “Ad” “CoUnt” Display the A/D counts
“F8” “Zero” Select the zero mode
“F9” “Pin” To set the password
“F10” “Lvd” “mode” Low voltage detection
8.2.1 F1 -CALIBRATION
To enter the calibration parameter, press the [Tare] key when “F1” is displayed.
The display will instruct you to remove any weight from the scale, “UnloAd”. Press [Tare].

© Adam Equipment Company 2007 13
The display will tell you to add weight to the scale: “LoAd” “ 04” “kiloS”
Add the weight shown, wait for stability then press the [Tare] key.
The display will show “SPAn” “PASS” if the calibration is OK. Or it will show “SPAn” “FAIL” if
calibration could not be completed correctly.
The display will then show “JP” “On” indicating the jumper is still in place. Remove the
weight. Switch off the scale and switch it on again to continue with the other Service
parameters.
8.2.2 F2 -RESOLUTION
NOTE: This value can not be changed for the CBC-M scales.
8.2.3 F3 – CAPACITY
To enter this parameter, press the [Tare] key when “F3” is shown.
The display will show the current capacity.
Press the [U.Wt./Units] key to change the value.
Press [Tare] to accept the displayed value.
NOTE: The scale must be originally built for the capacity selected. The 3kg unit uses the
5kg load cells, 6 kg unit uses 10kg load cells, the 15 kg unit uses 20 kg load cells and the
30 kg unit uses 35 kg load cells.
Press [Zero] to return to weighing.

© Adam Equipment Company 2007 14
8.2.4 F4 -INITIAL ZERO RANGE
To enter this parameter, press the [Tare] key when “F4” is shown.
The display will show the current initial zero range.
Press the [U.Wt./Units] key to change the value and press [Tare] to accept the value.
Press [Zero] to return to weighing.
8.2.5 F5 -RE-ZERO RANGE
To enter this parameter, press the [Tare] key when “F5” is shown.
The display will show the current re-zero range.
Press the [U.Wt./Units] key to change the value.
Press [Tare] to accept the value.
Press [Zero] to return to weighing.
8.2.6 F6 -SUCCESSIVE TARE
To enter this parameter, press the [Tare] key when “F6” is shown.
The display will show if the successive tare is on or off.
Press the [U.Wt./Units] key to change the value.
Press [Tare] to accept the displayed value.
Press [Zero] to return to weighing.

© Adam Equipment Company 2007 15
8.2.7 F7 -A/D COUNT
To enter this parameter, press the [Tare] key when “F7” is shown.
This parameter allows you to view the A/D counts from the internal A/D converter. This can
be an aid to service.
Press the [Tare] key to return to the PARAMETER menu.
Press the [Zero] key to return to weighing.
Typical value at zero is 30,000-90,000 (approx.)
Typical value at full capacity is 500,000 (approx.)
8.2.8 F8 –ZERO
This parameter allows the normal method of setting zero.
To enter this parameter, press the [Tare] key when “F8” is shown.
The display will show the current settings, “modE1” or “modE2”.
Press the [U.Wt./Units] key to change and press [Tare] to accept the value.
Press [Zero] to return to weighing.
8.2.9 F9 –PIN
This parameter allows setting of password to enter these Service Parameters.
To enter this parameter, press the [Tare] key when “F9” “Pin” is shown.
The “WEIGHT” display will show "Pin1"
Enter the new password. The “Unit Weight” display will show dashes. Press [Tare].

© Adam Equipment Company 2007 16
The “WEIGHT” display will change to "Pin 2 ", Enter the password again and press [Tare].
The display will show "donE" indicating the new password has been accepted and will
return to the menu. “F9” “Pin” is shown. Record the new password in a secured place.
Press [Zero] to return to weighing.
8.2.10 F10 –LOW VOLTAGE DETECTION
This parameter allows detection of low voltage when the battery wears down.
To enter this parameter, press the [Tare] key when “F10” is shown.
The display will show if the LVD Mode is set to on or oFF.
Press the [U.Wt./Units] key to change the value.
Press [Tare] to accept the displayed value.
Press [Zero] to return to weighing.
NOTE 1: TO SECURE THE SCALE AFTER CALIBRATION, IT IS NECESSARY TO
REMOVE THE JUMPER AND RE-SEAL THE SCALE.
NOTE 2: The scale has a special feature for displaying x10 the value of the weight which
is a useful tool during testing. Press the [00] key for more than 3 seconds. It will show one
extra decimal place making it a high resolution display. Pressing the [00] key again for
more than 3 seconds will bring the display back to the normal resolution.

9.0 WIRING DIAGRAM
© Adam Equipment Company 2007 17

10.0 MECHANICAL ASSEMBLY
© Adam Equipment Company 2007 18
This manual suits for next models
4
Table of contents
Other Adam Scale manuals

Adam
Adam Swift SWZ Series User manual

Adam
Adam GBK Mplus Series User manual

Adam
Adam AGF Series User manual

Adam
Adam TBB User manual

Adam
Adam GBK 16a User manual

Adam
Adam LBX C Series User manual

Adam
Adam WBW 2L User manual

Adam
Adam CPWplus 35 User manual

Adam
Adam Luna LAB Series User manual

Adam
Adam ACBplus Series User manual