manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. AEMC instruments
  6. •
  7. Measuring Instrument
  8. •
  9. AEMC instruments MN290 User manual

AEMC instruments MN290 User manual

99-MAN 100142.v5 04/11
OPERATION
Please make sure that you have already read and fully understand the WARNING section on page 1.
Making Measurements with the AC Current Probe Model MN290/MN291
• Connect the black lead of the current probe to “common” and the red lead to the DC voltage input on
your DMM or other voltage measuring instrument. The AC current probe has an output of 100 mV DC/A
AC. This means that for 200 A AC in a conductor around which the probe is clamped, 20 V DC will come
out of the probe leads to your DMM or instrument. Select the range which corresponds to the measured
current. If the current magnitude is unknown, start with the highest range and work down until the
appropriate range and resolution are reached. Clamp the probe around the conductor. Take the reading
on the meter and multiply it by 10 to obtain the measured current (e.g, 1000 mV DC reading = 1000 x 10
= 10,000 mA AC or 10 AC)
• For best accuracy, avoid if possible, the proximity of other conductors which may create noise.
Tips for Making Precise Measurements
• When using a current probe with a meter, it is important to select the range that provides the best
resolution. Failure to do this may result in measurement errors.
• Make sure that probe jaw mating surfaces are free of dust and contamination. Contaminants cause air
gaps between the jaws, increasing the phase shift between primary and secondary. It is very critical for
power measurement.
MAINTENANCE:
Warning
• For maintenance use only original factory replacement parts.
• To avoid electrical shock, do not attempt to perform any servicing unless you are qualified to do so.
• To avoid electrical shock and/or damage to the instrument, do not get water or other foreign agents
into the probe
Cleaning
To ensure optimum performance, it is important to keep the probe jaw mating surfaces clean at all times.
Failure to do so may result in error in readings. To clean the probe jaws, use very fine sand paper (fine
600) to avoid scratching the jaw, then gently clean with a soft oiled cloth.
REPAIR AND CALIBRATION
You must contact our Service Center for a Customer Service Authorization number (CSA#). This will
ensure that when your instrument arrives, it will be tracked and processed promptly. Please write the CSA#
on the outside of the shipping container. If the instrument is returned for calibration, we need to know if you
want a standard calibration, or a calibration traceable to N.I.S.T. (includes calibration certificate plus
recorded calibration data).
Chauvin Arnoux®, Inc. d.b.a. AEMC®Instruments
15 Faraday Drive Dover, NH 03820 USA
Tel: (800) 945-2362 (Ext. 360)
(603) 749-6434 (Ext. 360)
Fax: (603) 742-2346 or (603) 749-6309
[email protected]
(Or contact your authorized distributor)
Costs for repair, standard calibration, and calibration traceable to N.I.S.T. are available.
NOTE: All customers must obtain a CSA# before returning any instrument.
TECHNICAL AND SALES ASSISTANCE
If you are experiencing any technical problems, or require any assistance with the proper use or
application of this instrument, please call our technical hotline:
(800) 343-1391 • (508) 698-2115 • Fax (508) 698-2118
Chauvin Arnoux®, Inc. d.b.a. AEMC®Instruments
[email protected]
www.aemc.com
AC Current Probe
Models MN290 and MN291
User Manual
DESCRIPTION
The MN290/MN291 (Catalog #2115.83 and #2115.84) are the latest in compact AC Current probes.
Designed to meet the most stringent demands in industry and electrical contracting, they also meet the
latest safety and performance standards. The probes have a measurement range up to 240 Arms which
makes them perfect tools for measurement with DMMs without AC range, recorders. The Models
MN290/MN291 are compatible with any DC voltmeter, multimeter, or other voltage measurement
instrument that is capable of displaying 100 mV of output per amp of measured current, voltmeter
accuracy of 0.75% or better and an input impedance of ≥1 MΩ.
WARNING
The safety warnings are provided to ensure the safety of personnel and proper operation of the
instrument. Read the instruction completely.
• Use caution on any circuit: potentially high voltages and currents may be present and may pose
a shock hazard.
• Do not use the probe if damaged. Always connect the current probe to the measuring device
before it is connected around the conductor
• Do not use on non-insulated conductor with a potential to ground greater than 600V CAT III
pollution 2. Use extreme caution when clamping around bare conductors or bus bars.
• Before each use, inspect the probe; look for cracks in housing or output cable insulation.
• Do not use clamp in wet environment or in locations that hazardous gases exist.
• Do not use the probe anywhere beyond the tactile barrier.
INTERNATIONAL ELECTRICAL SYMBOLS
This symbol signifies that the current probe is protected by double or reinforced insulation. Use
only factory specified replacement parts when servicing the instrument.
This symbol signifies CAUTION! and requests that the user refer to the user manual before
using the instrument.
This is a type A current sensor. This symbol signifies that application around and removal from
HAZARDOUS LIVE conductors is permitted.
DEFINITION OF MEASUREMENT CATEGORIES
CAT I: For measurements on circuits not directly connected to the AC supply wall outlet such as
protected secondaries, signal level, and limited energy circuits.
CAT II: For measurements performed on circuits directly connected to the electrical distribution system.
Examples are measurements on household appliances or portable tools.
CAT III: For measurements performed in the building installation at the distribution level such as on
hardwired equipment in fixed installation and circuit breakers.
CAT IV: For measurements performed at the primary electrical supply (<1000V) such as on primary
overcurrent protection devices, ripple control units, or meters.
RECEIVING YOUR SHIPMENT
Upon receiving your shipment, make sure that the contents are consistent with the packing list. Notify
your distributor of any missing items. If the equipment appears to be damaged, file a claim immediately
with the carrier and notify your distributor at once, giving a detailed description of any damage.
ELECTRICAL SPECIFICATIONS
Nominal Range: 200 A
Measurement Range: 0.5 to 240 A
Transformation Ratio: DC Voltage output
Output Signal: 100 mV DC/A AC
Accuracy and Phase Shift*:
Accuracy: 0.5 to 10 A: 3.0% Reading ± 50 mV
10 to 40 A: 2.5% Reading ± 50 mV
40 to 100 A: 2.0% Reading ± 50 mV
100 to 240 A: 2% Reading
Phase Shift: Not applicable
(*Reference conditions: 18°C to 28°C, 20 to 75% RH,
external magnetic field <40 A/m, 48 to 65 Hz sine wave,
distortion factor less than 1%, no DC component, no
external current carrying conductor, test sample
centered.)
Load impedance > 1 MΩwith 2s RC filter.
Overload: 240 A for 10 min ON, 30 min OFF
Frequency Range: 40 to 10 kHz
Limit Operating Conditions:
200 A permanently to 1 kHz
Derating above 3 kHz: 200 A x (1/0.333 F), F in kHz
Crest Factor:
3 @ 200 Arms with an error (due to CF) of 3%
Working Voltage: 600 Vrms
Common Mode Voltage: 600 Vrms
Influence of Adjacent Conductor:
< 15 mA/A at 50 Hz
Influence of DC on AC signal:
DC Current < 20 A, < 5%
Influence of Conductor Position in Jaw:
0.5% of mV output @ 50/60 Hz
Influence of Frequency:
40 Hz to 1 kHz: 5% of mV output
1 kHz to 10 kHz: 12% of mV output
Influence of Load:
< 3% of mA output from 40 Hz to 1 kHz
< 12% of mA output from 1 kHz to 10 kHz
Influence of Temperature:
≤150 ppm/°K, or 0.15% of mV output per 10°K
Influence of Humidity (10 - 90% RH):
≤0.2% of mV output per 10°K @ 18°C to 28°C
MECHANICAL SPECIFICATIONS
Operating Temperature:
14° to 131°F (-10° to +55°C)
Storage Temperature:
-40° to 158°F (-40° to +70°C)
Operating Relative Humidity:
10-35°C 85% RH (without roll-off above 35°C)
Maximum Cable Diameter:
One Ø 0.78” (20 mm), bus bar 20 x 5 mm
Case Protection:
IP 40 (IEC 529)
Drop Test:
Test per IEC 68-2-32: 1.0 m drop on 38 mm of
Oak on concrete.
Mechanical Shock:
Test per IEC 68-2-27
Vibration:
Test per IEC 68-2-6
Dimensions:
5.47 x 2.00 x 1.18" (139 x 51 x 30 mm)
Weight: 180 g (6.5 oz)
Polycarbonate Material:
Jaws: Polycarbonate with 10% fiberglass charge,
Red UL 94 V0
Case: Polycarbonate 920A Gray
Opening Operations - Life: > 50,000
Output:
MN290: Two standard safety banana jacks (4 mm)
MN291: Double/reinforced insulated 5 ft. (1.5 m)
lead with safety 4mm banana plug
SAFETY SPECIFICATIONS
Electrical:
Double insulation or reinforced insulation between
the primary or secondary and the outer case of the
handle conforms to IEC 1010-2-32.
Common Mode Voltage:
600 V Category III, Pollution Degree 2
Dielectric Strength:
5550 V, 50/60 Hz between primary, secondary
and the outer case of the handle
Electromagnetic Compatibility:
EN 50081-1 Class B
EN 50082-2 Electrostatic discharge
IEC 1000-4-2
Radiated field IEC 1000-4-3
Fast transients IEC 1000-4-4
Magnetic field at 50/60 Hz IEC 1000-4-8
ORDERING INFORMATION
AC Current Probe MN290.................Cat #2115.83
AC Current Probe MN291.................Cat #2115.84
Both models include a user manual and a
product warranty and registration card.
Accessories:
Leads, set of 2, 5 ft safety
(1000V) for Model MN290................... Cat #2111.29
Banana plug adapter
(to non-recessed plug) ....................... Cat #1017.45
E192383
2.56"
(65 mm)
5.47"
(139 mm)
.83"
(21 mm)
Black
Red
.73"
2.00"
(51 mm)
(18.5 mm)
.1.18"
(30 mm)
Tactical
Barrier
(keep hands below
this barrier)
MN290 MN291
ELECTRICAL SPECIFICATIONS
Nominal Range: 200 A
Measurement Range: 0.5 to 240 A
Transformation Ratio: DC Voltage output
Output Signal: 100 mV DC/A AC
Accuracy and Phase Shift*:
Accuracy: 0.5 to 10 A: 3.0% Reading ± 50 mV
10 to 40 A: 2.5% Reading ± 50 mV
40 to 100 A: 2.0% Reading ± 50 mV
100 to 240 A: 2% Reading
Phase Shift: Not applicable
(*Reference conditions: 18°C to 28°C, 20 to 75% RH,
external magnetic field <40 A/m, 48 to 65 Hz sine wave,
distortion factor less than 1%, no DC component, no
external current carrying conductor, test sample
centered.)
Load impedance > 1 MΩwith 2s RC filter.
Overload: 240 A for 10 min ON, 30 min OFF
Frequency Range: 40 to 10 kHz
Limit Operating Conditions:
200 A permanently to 1 kHz
Derating above 3 kHz: 200 A x (1/0.333 F), F in kHz
Crest Factor:
3 @ 200 Arms with an error (due to CF) of 3%
Working Voltage: 600 Vrms
Common Mode Voltage: 600 Vrms
Influence of Adjacent Conductor:
< 15 mA/A at 50 Hz
Influence of DC on AC signal:
DC Current < 20 A, < 5%
Influence of Conductor Position in Jaw:
0.5% of mV output @ 50/60 Hz
Influence of Frequency:
40 Hz to 1 kHz: 5% of mV output
1 kHz to 10 kHz: 12% of mV output
Influence of Load:
< 3% of mA output from 40 Hz to 1 kHz
< 12% of mA output from 1 kHz to 10 kHz
Influence of Temperature:
≤150 ppm/°K, or 0.15% of mV output per 10°K
Influence of Humidity (10 - 90% RH):
≤0.2% of mV output per 10°K @ 18°C to 28°C
MECHANICAL SPECIFICATIONS
Operating Temperature:
14° to 131°F (-10° to +55°C)
Storage Temperature:
-40° to 158°F (-40° to +70°C)
Operating Relative Humidity:
10-35°C 85% RH (without roll-off above 35°C)
Maximum Cable Diameter:
One Ø 0.78” (20 mm), bus bar 20 x 5 mm
Case Protection:
IP 40 (IEC 529)
Drop Test:
Test per IEC 68-2-32: 1.0 m drop on 38 mm of
Oak on concrete.
Mechanical Shock:
Test per IEC 68-2-27
Vibration:
Test per IEC 68-2-6
Dimensions:
5.47 x 2.00 x 1.18" (139 x 51 x 30 mm)
Weight: 180 g (6.5 oz)
Polycarbonate Material:
Jaws: Polycarbonate with 10% fiberglass charge,
Red UL 94 V0
Case: Polycarbonate 920A Gray
Opening Operations - Life: > 50,000
Output:
MN290: Two standard safety banana jacks (4 mm)
MN291: Double/reinforced insulated 5 ft. (1.5 m)
lead with safety 4mm banana plug
SAFETY SPECIFICATIONS
Electrical:
Double insulation or reinforced insulation between
the primary or secondary and the outer case of the
handle conforms to IEC 1010-2-32.
Common Mode Voltage:
600 V Category III, Pollution Degree 2
Dielectric Strength:
5550 V, 50/60 Hz between primary, secondary
and the outer case of the handle
Electromagnetic Compatibility:
EN 50081-1 Class B
EN 50082-2 Electrostatic discharge
IEC 1000-4-2
Radiated field IEC 1000-4-3
Fast transients IEC 1000-4-4
Magnetic field at 50/60 Hz IEC 1000-4-8
ORDERING INFORMATION
AC Current Probe MN290.................Cat #2115.83
AC Current Probe MN291.................Cat #2115.84
Both models include a user manual and a
product warranty and registration card.
Accessories:
Leads, set of 2, 5 ft safety
(1000V) for Model MN290................... Cat #2111.29
Banana plug adapter
(to non-recessed plug) ....................... Cat #1017.45
E192383
2.56"
(65 mm)
5.47"
(139 mm)
.83"
(21 mm)
Black
Red
.73"
2.00"
(51 mm)
(18.5 mm)
.1.18"
(30 mm)
Tactical
Barrier
(keep hands below
this barrier)
MN290 MN291
99-MAN 100142.v5 04/11
OPERATION
Please make sure that you have already read and fully understand the WARNING section on page 1.
Making Measurements with the AC Current Probe Model MN290/MN291
• Connect the black lead of the current probe to “common” and the red lead to the DC voltage input on
your DMM or other voltage measuring instrument. The AC current probe has an output of 100 mV DC/A
AC. This means that for 200 A AC in a conductor around which the probe is clamped, 20 V DC will come
out of the probe leads to your DMM or instrument. Select the range which corresponds to the measured
current. If the current magnitude is unknown, start with the highest range and work down until the
appropriate range and resolution are reached. Clamp the probe around the conductor. Take the reading
on the meter and multiply it by 10 to obtain the measured current (e.g, 1000 mV DC reading = 1000 x 10
= 10,000 mA AC or 10 AC)
• For best accuracy, avoid if possible, the proximity of other conductors which may create noise.
Tips for Making Precise Measurements
• When using a current probe with a meter, it is important to select the range that provides the best
resolution. Failure to do this may result in measurement errors.
• Make sure that probe jaw mating surfaces are free of dust and contamination. Contaminants cause air
gaps between the jaws, increasing the phase shift between primary and secondary. It is very critical for
power measurement.
MAINTENANCE:
Warning
• For maintenance use only original factory replacement parts.
• To avoid electrical shock, do not attempt to perform any servicing unless you are qualified to do so.
• To avoid electrical shock and/or damage to the instrument, do not get water or other foreign agents
into the probe
Cleaning
To ensure optimum performance, it is important to keep the probe jaw mating surfaces clean at all times.
Failure to do so may result in error in readings. To clean the probe jaws, use very fine sand paper (fine
600) to avoid scratching the jaw, then gently clean with a soft oiled cloth.
REPAIR AND CALIBRATION
You must contact our Service Center for a Customer Service Authorization number (CSA#). This will
ensure that when your instrument arrives, it will be tracked and processed promptly. Please write the CSA#
on the outside of the shipping container. If the instrument is returned for calibration, we need to know if you
want a standard calibration, or a calibration traceable to N.I.S.T. (includes calibration certificate plus
recorded calibration data).
Chauvin Arnoux®, Inc. d.b.a. AEMC®Instruments
15 Faraday Drive Dover, NH 03820 USA
Tel: (800) 945-2362 (Ext. 360)
(603) 749-6434 (Ext. 360)
Fax: (603) 742-2346 or (603) 749-6309
[email protected]
(Or contact your authorized distributor)
Costs for repair, standard calibration, and calibration traceable to N.I.S.T. are available.
NOTE: All customers must obtain a CSA# before returning any instrument.
TECHNICAL AND SALES ASSISTANCE
If you are experiencing any technical problems, or require any assistance with the proper use or
application of this instrument, please call our technical hotline:
(800) 343-1391 • (508) 698-2115 • Fax (508) 698-2118
Chauvin Arnoux®, Inc. d.b.a. AEMC®Instruments
[email protected]
www.aemc.com
AC Current Probe
Models MN290 and MN291
User Manual
DESCRIPTION
The MN290/MN291 (Catalog #2115.83 and #2115.84) are the latest in compact AC Current probes.
Designed to meet the most stringent demands in industry and electrical contracting, they also meet the
latest safety and performance standards. The probes have a measurement range up to 240 Arms which
makes them perfect tools for measurement with DMMs without AC range, recorders. The Models
MN290/MN291 are compatible with any DC voltmeter, multimeter, or other voltage measurement
instrument that is capable of displaying 100 mV of output per amp of measured current, voltmeter
accuracy of 0.75% or better and an input impedance of ≥1 MΩ.
WARNING
The safety warnings are provided to ensure the safety of personnel and proper operation of the
instrument. Read the instruction completely.
• Use caution on any circuit: potentially high voltages and currents may be present and may pose
a shock hazard.
• Do not use the probe if damaged. Always connect the current probe to the measuring device
before it is connected around the conductor
• Do not use on non-insulated conductor with a potential to ground greater than 600V CAT III
pollution 2. Use extreme caution when clamping around bare conductors or bus bars.
• Before each use, inspect the probe; look for cracks in housing or output cable insulation.
• Do not use clamp in wet environment or in locations that hazardous gases exist.
• Do not use the probe anywhere beyond the tactile barrier.
INTERNATIONAL ELECTRICAL SYMBOLS
This symbol signifies that the current probe is protected by double or reinforced insulation. Use
only factory specified replacement parts when servicing the instrument.
This symbol signifies CAUTION! and requests that the user refer to the user manual before
using the instrument.
This is a type A current sensor. This symbol signifies that application around and removal from
HAZARDOUS LIVE conductors is permitted.
DEFINITION OF MEASUREMENT CATEGORIES
CAT I: For measurements on circuits not directly connected to the AC supply wall outlet such as
protected secondaries, signal level, and limited energy circuits.
CAT II: For measurements performed on circuits directly connected to the electrical distribution system.
Examples are measurements on household appliances or portable tools.
CAT III: For measurements performed in the building installation at the distribution level such as on
hardwired equipment in fixed installation and circuit breakers.
CAT IV: For measurements performed at the primary electrical supply (<1000V) such as on primary
overcurrent protection devices, ripple control units, or meters.
RECEIVING YOUR SHIPMENT
Upon receiving your shipment, make sure that the contents are consistent with the packing list. Notify
your distributor of any missing items. If the equipment appears to be damaged, file a claim immediately
with the carrier and notify your distributor at once, giving a detailed description of any damage.

This manual suits for next models

3

Other AEMC instruments Measuring Instrument manuals

AEMC instruments MN02 User manual

AEMC instruments

AEMC instruments MN02 User manual

AEMC instruments MR525 User manual

AEMC instruments

AEMC instruments MR525 User manual

AEMC instruments JM835A User manual

AEMC instruments

AEMC instruments JM835A User manual

AEMC instruments FlexProbe 24-3001 User manual

AEMC instruments

AEMC instruments FlexProbe 24-3001 User manual

AEMC instruments MN185 User manual

AEMC instruments

AEMC instruments MN185 User manual

AEMC instruments 1015 User manual

AEMC instruments

AEMC instruments 1015 User manual

AEMC instruments Simple Logger AmpFlex AL24-2500 User manual

AEMC instruments

AEMC instruments Simple Logger AmpFlex AL24-2500 User manual

AEMC instruments Digital FlexProbes 400D-10 User manual

AEMC instruments

AEMC instruments Digital FlexProbes 400D-10 User manual

AEMC instruments CA811 User manual

AEMC instruments

AEMC instruments CA811 User manual

AEMC instruments MD305 User manual

AEMC instruments

AEMC instruments MD305 User manual

AEMC instruments MN134 User manual

AEMC instruments

AEMC instruments MN134 User manual

AEMC instruments MN379T User manual

AEMC instruments

AEMC instruments MN379T User manual

AEMC instruments Chauvin Arnoux 6412 User manual

AEMC instruments

AEMC instruments Chauvin Arnoux 6412 User manual

AEMC instruments OX5042 User manual

AEMC instruments

AEMC instruments OX5042 User manual

AEMC instruments CA7027 User manual

AEMC instruments

AEMC instruments CA7027 User manual

AEMC instruments 6292 User manual

AEMC instruments

AEMC instruments 6292 User manual

AEMC instruments 5070 User manual

AEMC instruments

AEMC instruments 5070 User manual

AEMC instruments Clamp-On 3705 User manual

AEMC instruments

AEMC instruments Clamp-On 3705 User manual

AEMC instruments CA846 User manual

AEMC instruments

AEMC instruments CA846 User manual

AEMC instruments Measure Up PowerPad III User manual

AEMC instruments

AEMC instruments Measure Up PowerPad III User manual

AEMC instruments 502 User manual

AEMC instruments

AEMC instruments 502 User manual

AEMC instruments MN261 User manual

AEMC instruments

AEMC instruments MN261 User manual

AEMC instruments 6527 User manual

AEMC instruments

AEMC instruments 6527 User manual

AEMC instruments PowerPad IV 8345 User manual

AEMC instruments

AEMC instruments PowerPad IV 8345 User manual

Popular Measuring Instrument manuals by other brands

Fema S40-C1 user manual

Fema

Fema S40-C1 user manual

Solinst 101 P7 SC1000 quick start guide

Solinst

Solinst 101 P7 SC1000 quick start guide

horiba LAQUAact D-75 Quick manual

horiba

horiba LAQUAact D-75 Quick manual

elsner elektronik KNX eTR M1 Technical specifications and installation instructions

elsner elektronik

elsner elektronik KNX eTR M1 Technical specifications and installation instructions

Meec tools 010989 operating instructions

Meec tools

Meec tools 010989 operating instructions

Gammon GTP-9330 Series Installation and operation

Gammon

Gammon GTP-9330 Series Installation and operation

Keysight M1970 Series user guide

Keysight

Keysight M1970 Series user guide

Leica NA320 user manual

Leica

Leica NA320 user manual

NI cDAQ-9178 USER GUIDE AND SPECIFICATIONS

NI

NI cDAQ-9178 USER GUIDE AND SPECIFICATIONS

Bamo BAMOWIZ OX 452 user manual

Bamo

Bamo BAMOWIZ OX 452 user manual

ETI 7000 operating instructions

ETI

ETI 7000 operating instructions

Ecotech Serinus 40 user manual

Ecotech

Ecotech Serinus 40 user manual

Co2meter SAN-20 user manual

Co2meter

Co2meter SAN-20 user manual

ZYLIA ZR-1 Reference manual

ZYLIA

ZYLIA ZR-1 Reference manual

Lascar Electronics PanelPilot SGD 24-M-IP quick start guide

Lascar Electronics

Lascar Electronics PanelPilot SGD 24-M-IP quick start guide

GLI International F53 operating manual

GLI International

GLI International F53 operating manual

National Instruments NI PXI-562X user manual

National Instruments

National Instruments NI PXI-562X user manual

PCB Piezotronics IMI SENSORS 602D91 Installation and operating manual

PCB Piezotronics

PCB Piezotronics IMI SENSORS 602D91 Installation and operating manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.