AEMC C.A 5230 User manual

1
ENGLISH User manual
Digital Multimeter
Model C.A 5230
MULTIMETER
Hz
V
V
OFF
mA
10A
MIN
MAX
RANGE
HOLD
ZERO
10A
500mA
FUSED
10Amax
FUSED
C . A 5230
!
COM
600V
CAT III
MAX
ZERO
n u F
Mk Hz
- +
HOLD
30
20
10
040
RMS MULTIMETER
mV A

2
Warning ! Please refer to the User’s Manual before using the
instrument.
In this User’s Manual, the instructions preceded by the above
symbol, should they not be carried out as shown, can result in a
physical accident or damage the instrument and the installations.
This device is protected by a double insulation or by a reinforced
insulation. No linking is required from the protection earth terminal
to ensure the electrical security.
Thank you for purchasing a C.A 5230 RMS Multimeter. To get the
best service from this instrument :
read this user manual carefully,
respect the safety precautions detailed.
SAFETY PRECAUTIONS
Respect the value and the type of fuses or there is risk of
deterioration of the instrument and invalidation of the warranty.
- Fuse 0.5A HBC (500V min - 10kA - 6.3 x 32mm)
- Fuse 12A HBC (600V min - 300kA - 10.3 x 38mm)
Never use on voltage to earth circuits above 600V, with a voltage
surge category of over III, that is to say, fixed industrial and
domestic installations (ref. IEC 664-1).
To be used indoors in environments with a maximum pollution
degree of 2 (ref. IEC 664-1), a temperature between 0 and 50°C
and relative humidity below 70%.
Use accessories which conform to safety standards
(IEC 1010-2-031), with a minimum voltage of 600V and a voltage
surge of category III.
Never open the multimeter case before disconnecting from any
electrical source.
Never connect up to a circuit to be measured if the multimeter
case is not closed correctly.
Before any measurements are taken, ensure the leads and switch
are in the correct positions.
Never measure resistances on a circuit with the power switched
on.

3
CONTENTS
1 - Presentation ............................................................................. 3
2 - Description ............................................................................... 4
3 - DC and AC voltage (V) ............................................................ 5
4 - Frequency (Hz) ....................................................................... 6
5 - Continuity and resistance (Ω/ ) ......................................... 6
6 - Diode test ( ) ........................................................................ 7
7 - Capacitance ( ) ................................................................... 8
8 - DC and AC current (A and mA) ............................................... 8
9 - General specifications ............................................................ 9
10 - To order (accessories) ........................................................ 10
11 - Warranty ............................................................................... 10
12 - Maintenance ......................................................................... 10
Technical and Sales Assistance ........................................... 11
1 - PRESENTATION
The C.A 5230 multimeter is designed for the daily needs of
professionals in electricity.
- true rms measurement (RMS)
- digital 4000 counts and 42 segment bargraph
- automatic or manual range selection
- auto off after 30 minutes without action on a control or function
- possibility of permanent operation (see HOLD)
Various measurement accessories broaden the scope of application
or provide the multimeter with new functions.
See the list of accessories and references on the “To Order” page
(Documentation upon request).
Note : Always use accessories suitable for the voltage and the
voltage surge category of the circuit to be measured (as per IEC
1010).

4
2 - DESCRIPTION
LIQUID CRYSTAL DISPLAY
4000 measurement counts - height of digits 0.5" (12mm)
Rate: 2 measurements / second
Bargraph with 42 segments / Rate: 20 measurements / second
Note: the bargraph is not operational on frequency and
capacitance measurements
Automatic display of all the symbols for commands, ranges and
functions
Use battery indication :
+
-
Low battery indication :
- the 4 on the left flashes
- bargraph totally lit plus arrow
CONTROL BUTTONS
...
/ ~ and ΩΩ
ΩΩ
Ω/
To change from AC (~) to DC (
...
) and reverse, on ammeter function.
To change from continuity test to resistance measurement Ωand
inverse, on ohmmeter function.
Note: Automatic selection of AC and when switched on.
HOLD and ZERO
To freeze the display on the last measurement (HOLD).
On capacitancemeter function , zero reset by press without
releasing.
Permanent operation: possibility of suppressing auto off, by pressing
this button without releasing before operating the switch from the
OFF position. The return to the OFF position cancels permanent
operation.
MIN - MAX
To successively display the min or max values. The instrument then
changes to manual range selection. HOLD is also displayed. Min
acquisition time 500 ms.
Note: The user can keep the min/max values in memory if you
press the HOLD button, PRIOR to disconnecting the test leads.
RANGE
Range selection: to change from automatic mode to manual mode, by
prolonged press on this button.
Note:
- automatic mode is selected when the instrument is switched on.
- this button also selects the sensitivity for the Hz frequency meter
function
To switch on and off the display backlighting.
Note: Auto off after 1 minute.

5
FUNCTION SELECTOR SWITCH
OFF Off position
V~ AC voltages up to 600V in rms value
V
...
DC voltages up to 600V
Hz Frequencies of voltages up to 400kHz
ΩΩ
ΩΩ
ΩContinuity sound test for a resistance R ≤40Ω, and
measurements of resistances up to 40MΩ
Note: The continuity test is selected automatically
when the instrument is switched on.
Diode test. Measurement of the voltage of the junction in
forward direction
Capacitances up to 40µF
mA Current ranges up to 400mAAC/DC (via the + terminal)
A 10AAC/DC current ranges (via the 10A terminal)
TERMINALS
Ø 4mm safety terminals
COM: common, black
+: red
10A: red lead for the 10AAC/DC ranges
3 - DC AND AC VOLTAGE
Connect the leads to the multimeter and connect it in parallel to the
circuit to be tested.
Place the selector switch on the voltmeter function VAC/DC
Automatic selection of the range: read the value measured.
Freeze the value by pressing the HOLD button.
If necessary, press the following buttons:
RANGE to select the ranges manually
MIN-MAX to read the min or max values
to light the display
VAC/DC 400mV (1) 4V 40V 400V 600V (2)
Digital 0.1mV 1mV 10mV 100mV 1V
resolution
Bargraph 10mV 100mV 1V 10V 100V
resolution
Impedance 100MΩ10MΩ9.1MΩ
Accuracy ± 0.25% of Reading ±1ct
on DC
Accuracy ± 1.5% R ± 1% of Reading ±3cts
on AC ±5cts
Permitted 600Vrms and 1000V peak
overload
(1) Accessible by the RANGE button. Min measurement : 20mV
(2) Crest factor FC
≤
3

6
4 - FREQUENCY
Connect the leads to the multimeter and connect in parallel to the
circuit to be tested.
Place the selector switch on the Hz voltage frequency function.
Automatic range selection: read the measured value
Note: The bargraph is inactive on frequency measurement
If necessary, press the following buttons:
- MIN/MAX to read the min or max values
- to light the display
Hz 100Hz 1kHz 10kHz 100kHz (2) 400kHz
Resolution 0.01Hz 0.1Hz 1Hz 10Hz 100Hz
Accuracy ± 0.1% of Reading
± 10cts ± 4cts ± 8cts ± 20cts
Permitted 500V rms and 900V peak
overload
Operating 4V to 600V
range (1)
(1) The RANGE button for this function selects the triggering threshold :
- 10mV for 100Hz, 1kHz and 10kHz ranges
- 400mV for 100kHz and 400kHz ranges
(2) Note : If the frequency applied is 99.99kHz, the decimal point oscillates between 9.999kHz
and 99.99kHz.
5 - CONTINUITY AND RESISTANCE
Never test a resistance on a live circuit
Connect the leads to the multimeter and to the terminals of the
circuit or the component to be tested.
Place the selector switch on the ohmmeter function ΩΩ
ΩΩ
Ω
Press the Ω /Ω /
Ω /Ω /
Ω / button to change from the continuity sound
test to the measurement of resistances and reverse.
Automatic range selection: read the measured value
Freeze the value, by pressing the HOLD button.
If necessary, press the following buttons:
RANGE to select the range manually
MIN/MAX to read the min or max values
to light the display

7
5-1 Resistances
ΩΩ
ΩΩ
Ω400Ω4kΩ40kΩ400kΩ4MΩ40MΩ
Digital
resolution 0.1Ω1Ω10Ω100Ω1kΩ10kΩ
Bargraph 10Ω100Ω1kΩ10kΩ100kΩ1MΩ
resolution
Accuracy
± 1% R
± 0.5% of Reading ±3cts
± 1% R ± 1.5% R
± 4cts ±4cts ±4cts
Voltage on ≤0.4 VDC
Protection 500Vrms and 750V peak
5-2 Continuity sound test
On the 400Ωrange, emission of a continuous beep for a resistance
R ≤40Ω. Response time: 100ms
6 - DIODE TEST
Never test a diode on a live circuit
Connect the leads to the multimeter and to the terminals of the
component to be tested.
Place the selector switch on the function:
- in forward direction the display gives the value of the junction in
volts (resolution 1mV)
- accuracy: ±1% of Reading ± 2cts
- open circuit voltage: 3.2V typical
- short circuit current: 0.6mA
- in reverse direction the display indicates the open circuit voltage,
around 4V.
Freeze the value, by pressing the HOLD button.
With this function, it is possible to test, besides classic diodes,
LED’s or any other semiconductor in which the junction
corresponds to a direct voltage less than 3V.

7 - CAPACITANCE
Never make a capacitance test on a live circuit
Always respect the polarity of electrolytic condensers.
This type of condenser may be sensitive to temperature,
so avoid touching it during measurement.
♦ Connect the leads to the multimeter and to the terminals of the
condenser.
♦ Place the selector switch on the capacitance meter function
♦ Automatic range selection: read the measured value
Note: The bargraph is not operational on capacitance
Measurement
♦ Freeze the value, if necessary, by pressing the HOLD button
If necessary, press the following buttons:
♦ ZERO to set zero on the 4nF and 40nF ranges only
♦ RANGE to select the ranges manually
♦ MIN/MAX to read the min or max values
♦ to light the display
4nF 40nF 400nF 4µF
40µF(1)
Resolution 1pF 10pF 100pF 1nF 10nF
Accuracy ±2% R
±150cts ±2% of reading ± 8cts
Protection 500V mns 750 peak
(1) For a capacitance C> 20µF
the accuracy is ±5% of the reading ± 8cts
Always de-energize the circuit to be tested before
connecting the multimeter to the circuit
♦ Warning: connect the leads to the multimeter and connect in
series to the circuit
-the red lead to the + terminal up to 400mA
-the red lead to the 10A terminal, from 400mA to 10A
♦ Place the selector on the A or mA ammeter function
Press the AC/DC button (... / ~) to select AC or DC
♦ Energize the circuit
♦ Automatic range selection: read the measured value
♦ If necessary, press the following buttons
-RANGE to manually select the ranges
-MIN/MAX to read the min or max values
-to light the display
8

9
AAC/DC 40mA 400mA 10A (1)
Voltage 450mV 650mV 400mV
drop (2)
Digital 10µA 100µA 10mA
resolution
Bargraph 1mA 10mA 1A
resolution
Accuracy ± 0.8% of Reading ± 2cts ±1.2% R
on DC ± 4cts
Accuracy (3) ± 1.2% of Reading ± 4cts ± 1.5% R
on AC ± 5cts
Protection (4) Fuse 0.5A HBC Fuse 12A HBC
(1) Crest factor FC
≤
3
(2) Voltage drop at the terminals for 40mA, 400mA, 10A
(3) Use on frequency: 40Hz to 1000Hz
(4) Permitted overloads: 12A for 30 seconds for the 10A range
9 - GENERAL SPECIFICATIONS
9-1 Dimensions and Weight
2.5 x 7 x 1.7 " (64 x 177 x 42mm) 12.3 ounces (350g)
9-2 Power Supply
One battery 9V
Battery life: 300h with battery type 6 F 22
500 h with alkaline battery type 6 LF22
Low battery indication:
+
-
Auto off after 30 minutes
Note: Switch on again via the OFF position.
9.3 Buzzer
Continuous beep for the continuity test
Intermittent beep each time the switch or buttons are pressed,
and for the overload indication.
9-4 Climatic conditions
Temperature: use: 0°C to +50°C / storage : -20°C to +60°C
Relative humidity: use: ≤70% RH
storage: ≤90% RH (up to 45°C)
Altitude: use <2000m
9-5 Conformity with International Standards
9-5-1 Electrical Safety (IEC 1010-1)
Double insulation: Installation category: III
Degree of pollution: 2 Assigned voltage: 600VRMS
Note: This multimeter, of overvoltage category III satisfies the severe
requirements of reliability and availability corresponding to industrial
and domestic permanent installations (IEC 664-1).

10
9-5-2 Electromagnetic Compatibility: Conforms to CE
Emission (EN 50081-1) Immunity (EN 50082-1)
9-5-3 Mechanical protection
Degree of watertightness (IEC 529): protection index IP 40
10 - TO ORDER
Use the designations and references below.
C.A 5230 .................................................................... Cat. #2116.76
Sold with shockproof case, a pair of leads with probes attached,
a 9V battery and this user’s manual
Accessories
AC current probe model MN 251 (200AAC) ............ Cat. #2115.77
AC current probe model MD 303 (500AAC) ............ Cat. #1201.21
AC current probe model SR 652 (1000AAC) .......... Cat. #2113.46
AC/DC MicroProbe model K110
(from 100µAAC/DC to 300mAAC / 450mADC) ............... Cat. #2111.73
AC/DC current probe model MR410
(400AAC / 600ADC) ..................................................... Cat. #1200.70
AmpFLEX model 1000-24-1-1 (1000AAC) ............. Cat. #2112.39
Replacement leads .................................................. Cat. #2118.66
Soft carrying case .................................................. Cat. #2118.65
Replacement shockproof case ............................... Cat. #2980.15
Fuse set of 10, 0.5A ............................................... Cat. #2970.28
Fuse set of 10, 12A ................................................ Cat. #2970.21
11 - WARRANTY
Our guarantee is applicable for three years after the date on which
the equipment is made available (copy available on request).
12 - MAINTENANCE
For maintenance, use only specified spare parts.
The manufacturer will not be held responsible for any
accident occurring following a repair done other than
by its factory or approved repairers.
12-1 Replacing the battery and the fuses
The multimeter must be disconnected from any
electrical source.
Place the selector switch on the “OFF” position.
Unscrew the tool release screws and remove the lower half of
the case.

11
Replace the dead battery by a 9V battery (6F22 or 6LF22)
Replace the defective fuses respecting their value and their type
:
- Fuse 0.5A HBC (500V - 10kA - min 6.3 x 32mm)
- Fuse 12A HBC (600V min - 300kA - 10.3 x 38mm)
12-2 Storage
If the multimeter is not to be used for a period of 60 days or more,
remove the battery and store separately.
12-3 Cleaning
The multimeter must be disconnected from any
electrical source.
To clean the case, use a cloth slightly moistened with soapy water.
Rinse with a damp cloth. Then, dry rapidly with a cloth or in a hot air
stream.
12-4 Repair and calibration
To guarantee that your instrument complies with the factory
specifications, we recommend that the Model C.A 5230 be submitted
to our factory service center at one-year intervals for recalibration,
or as required by other standards.
For instrument repair and/or calibration, please call our factory, toll-
free, at (800) 945-AEMC (800-945-2362):
CHAUVIN ARNOUX, Inc.
d.b.a. AEMCInstruments
15 Faraday Drive
Dover, NH 03820 USA
Tel : (800) 954-2362
(603) 749-6434
Fax : (603) 742-2346
(Or contact your authorized distributor)
Costs for repair, standard calibration, and calibration traceable to
N.I.S.T. are available upon request. Overseas customers must receive
written authorization before returning any instrument.
12-5 Technical and Sales Assistance
If you are experiencing any technical problems, or require any
assistance with the proper use or application of this instrument,
please call our technical hotline:
CHAUVIN ARNOUX, Inc.
d.b.a. AEMCInstruments
200 Foxborough Blvd.
Foxborough, MA 02111 USA
Tel : (800) 343-1391
(508) 698-2115
Fax : (508) 698-2118
www.aemc.com
Table of contents
Other AEMC Multimeter manuals