AEMC 5233 User manual

5233
ENGLISH User Manual
DIGITAL MULTIMETER
with Non-Contact Detection

Statement of Compliance
Chauvin Arnoux®, Inc. d.b.a. AEMC®Instruments
certifies that this instrument has been calibrated
using standards and instruments traceable to
international standards.
We guarantee that at the time of shipping your
instrument has met its published specifications.
An NIST traceable certificate may be
requested at the time of purchase, or obtained
by returning the instrument to our repair and
calibration facility, for a nominal charge.
The recommended calibration interval for this
instrument is 12 months and begins on the date of
receipt by the customer. For recalibration, please
use our calibration services. Refer to our repair
and calibration section at www.aemc.com.
Serial #: ________________________________
Catalog #: 2125.65
Model #: 5233
Please fill in the appropriate date as indicated:
Date Received: _________________________________
Date Calibration Due: _______________________
Chauvin Arnoux®, Inc.
d.b.a AEMC®Instruments
www.aemc.com

Digital Multimeter Model 5233
1
Table of Contents
1. INTRODUCTION............................................................................... 3
1.1 International Electrical Symbols................................................4
1.2 DenitionofMeasurementCategories .....................................4
1.3 Receiving Your Shipment..........................................................5
1.4 OrderingInformation.................................................................5
1.4.1 Accessories and Replacement Parts............................5
2. PRODUCT FEATURES ...................................................................... 6
2.1 Description................................................................................6
2.2 Control Features .......................................................................7
2.3 Display Features.......................................................................8
2.4 Button Functions.......................................................................9
2.5 Rotary Functions.......................................................................9
3. OPERATION .................................................................................. 10
3.1 Turning the Multimeter ON......................................................10
3.2 Turning the Multimeter OFF....................................................10
3.3 Activating/Deactivating Auto-OFF...........................................10
3.4 Auto and Manual Range Selection .........................................10
3.5 Non-Contact Voltage (NCV)....................................................11
3.6 Voltage Measurement.............................................................11
3.7 Resistance Measurement .......................................................12
3.8 Continuity Test ........................................................................13
3.9 Diode Test...............................................................................14
3.10 Capacitance Test ....................................................................15
3.11 Temperature Test ....................................................................16
3.12 Current Measurement.............................................................17
4. MAINTENANCE ............................................................................. 18
4.1 Warning ..................................................................................18
4.2 Battery Replacement ..............................................................18
4.3 Fuse Replacement..................................................................19
4.4 Cleaning..................................................................................19

2
Digital Multimeter Model 5233
5. SPECIFICATIONS........................................................................... 20
Repair and Calibration...........................................................................23
Technical and Sales Assistance ............................................................23
Limited Warranty ...................................................................................24
Warranty Repairs...................................................................................24

Digital Multimeter Model 5233
3
CHAPTER 1
INTRODUCTION
Warning
This device complies with safety standard IEC-61010-1 (Ed
2–2001)forvoltagesupto1000VCATIIIor600VCATIV,atan
altitudebelow 2000m,indoors,with apollutionlevel ofnotmore
than 2.
Failure to observe the safety instructions may cause an electric
shock,re,explosion,ordestructionoftheinstrumentandofthe
installations.
• Donotusetheinstrumentinanexplosiveatmosphereorin
thepresenceofammablegasesorfumes.
• Donotusetheinstrumentonnetworksofwhichthevoltage
orcategoryexceedsthosementioned.
• Donotexceedtheratedmaximumvoltagesandcurrents
between terminals or with respect to earth/ground.
• Donotusetheinstrumentifitappearstobedamaged,
incomplete,ornotproperlyclosed.
• Beforeeachuse,checktheconditionoftheinsulationonthe
leads,housing,andaccessories.Anyelementofwhichthe
insulationisdeteriorated(evenpartially)mustbesetasidefor
repair or scrapped.
• Useleadsandaccessoriesratedforvoltagesandcategories
atleastequaltothoseoftheinstrument.
• Observetheenvironmentalconditionsofuse.
• Donotmodifytheinstrumentanddonotreplacecomponents
with “equivalents”. Repairs and adjustments must be done by
approvedqualiedpersonnel.
• Replace the battery as soon as the symbol appears
onthedisplayunit.Disconnectallleadsbeforeopeningthe
battery compartment cover.
• Use personal protective equipment when conditions require.
• Keepyourhandsawayfromunusedterminalsoftheinstrument.
• Whenhandlingprobesorcontacttips,keepyourngers
behind the guards.

4
Digital Multimeter Model 5233
1.1 International Electrical Symbols
Signifies that the instrument is protected by double or reinforced insulation.
This symbol on the instrument indicates a WARNING that the operator must
refer to the user manual for instructions before operating the instrument. In
this manual, the symbol preceding instructions indicates that if the instructions
are not followed, bodily injury, installation/sample and/or product damage may
result.
Compliance with the Low Voltage & Electromagnetic Compatibility European
directives (73/23/CEE & 89/336/CEE)
AC – Alternating current
AC or DC – Alternating or direct current
Risk of electric shock. The voltage at the parts marked with this symbol may be
dangerous.
Important instructions to read and understand completely.
Important information to acknowledge.
Ground/Earth symbol
In conformity with WEEE 2002/96/EC
1.2 Definition of Measurement Categories
CAT I: For measurements on circuits not directly connected to the AC
supplywalloutletsuchasprotectedsecondaries,signallevel,and
limited energy circuits.
CAT II: For measurements performed on circuits directly connected to
the electrical distribution system. Examples are measurements on
household appliances or portable tools.
CAT III:
Formeasurementsperformedinthebuildinginstallationatthedistri-
butionlevelsuchasonhardwiredequipmentinxedinstallationand
circuit breakers.
CAT IV: For measurements performed at the primary electrical supply
(<1000V)suchasonprimaryovercurrentprotectiondevices,ripple
controlunits,ormeters.

Digital Multimeter Model 5233
5
1.3 Receiving Your Shipment
Uponreceivingyourshipment,makesurethatthecontentsareconsistent
withthepackinglist.Notifyyourdistributorofanymissingitems.Iftheequip-
mentappearstobedamaged,leaclaimimmediatelywiththecarrierand
notifyyourdistributoratonce,givingadetaileddescriptionofanydamage.
Save the damaged packing container to substantiate your claim
.
1.4 Ordering Information
Multimeter Model 5233......................................................Cat. #2125.65
Includes set of two 5 ft color-coded leads (red/black) with needle tip {1000V CAT IV
15A), thermocouple Type K with adapter, set of three fuses 10A, 600V, 50kA, (fast
blow) 5x32mm, soft carrying case and a user manual.
1.4.1 Replacement Parts
Fuse-Setof10,10A,600V,50kA,(FastBlow),5x32mm
.... Cat. #2118.62
SoftCarryingCase ..............................................................Cat. #2121.54
Adapter - Banana (Male) to Mini (Female)
with K-Type Thermocouple.................................................. Cat. #2125.83
Thermocouple-Flexible(1m)K-Type58°to480°F ............ Cat. #2126.47
Setof2,Color-coded5ftleadswithsafetyneedletips(2mm)
{1000V CAT IV 15A) ..........................................................Cat. #2140.68
MultixMountingSystem.....................................................Cat. #5000.44
Order Accessories and Replacement Parts Directly Online
Check our Storefront at www.aemc.com for availability

6
Digital Multimeter Model 5233
CHAPTER 2
PRODUCT FEATURES
2.1 Description
TheModel5233isaTRMSdigitalmultimeter,speciallydesignedtocom-
binethevarious functions and measurements of the following electrical
quantities:
• Non-contactdetectionofpresenceofnetworkvoltage
(NCVfunction)
• AC voltmeter with low input impedance (voltage measurements
forelectricityandelectricalengineering)
• AC/DC voltmeter with high input impedance
(voltagemeasurementsforelectronics)
• Frequency and duty cycle measurements
• Ohmmeter
• Continuity test with buzzer
• Diode test
• Ammeter
• Capacitance meter
• Thermometerin°Cor°Fbymeasurementandlinearizationof
thevoltageacrosstheterminalsofaK-typethermocouple

Digital Multimeter Model 5233
7
2.2 Control Features
0600
100 500
200 400300
AUTO
AC DC MIN MAX
HOLD
P
°C°F
mVA
MkΩ
µnF
Hz%
REL
∆
2
1
3
4
5
6
Figure 2-1
1. NCV detection sensor (see § 3.5)
2. Analog and digital display (see § 2.3)
3. Function buttons (see § 2.4)
4. Rotary switch (see § 2.5)
5. Current measurement 10A terminal (see § 3.12)
6. Positive (Red) input and COM (Black) input

8
Digital Multimeter Model 5233
2.3 Display Features
Icon Function
AC Alternating Current
DC Direct Current
AUTO Automatic range selection (see § 3.4)
HOLD Freezes the display of the measurement
MAX Maximum RMS value
MIN Minimum RMS value
REL Relative value
The Over Load symbol is displayed when the signal measured
exceeds the range of the device
VVoltage
Hz Hertz
%Duty Cycle
FFarad
°C Degrees Celsius
°F Degrees Fahrenheit
AAmpere
WOhm
nPrefix “nano”
µPrefix “micro”
mPrefix “milli”
kPrefix “kilo”
MPrefix “Mega”
Continuity Beeper Enabled
Diode Test
Low Battery
P
Auto Power OFF function activated

Digital Multimeter Model 5233
9
2.4 Button Functions
Button Function
• Measurement type selection
NOTE: The DC mode is activated by default
• Activates/Deactivates Auto-OFF at start-up (see § 3.3)
• Allows manual selection of a measurement range (short press)
• Returns to Auto-range mode (long press >2s)
NOTE: Continuity and Diode modes are not Auto-ranging
• Press once to activate the MAX/MIN mode; press >2s to exit
• Once activated, press to view the MAX, MIN and current value
NOTE: MAX mode is activated by default
• Freezes/Unfreezes the display of the measured value
(short press)
• Activates/Deactivates the display backlight (long press >2s)
• Displays the frequency of the AC signal measured, along with the
duty cycle
NOTE: This is inactive in DC mode
• Displays the value relative to a reference stored when the key was
pressed
Example: If the value stored when the key was pressed equals
10V and the current value is 11.5V, the display in relative mode
will be 11.5 - 10 = 1.5V.
NOTE: Auto-range is deactivated in this mode
2.5 Rotary Functions
Range Function
OFF Powers down the multimeter
Low impedance AC voltage measurement
AC or DC voltage measurement (V)
AC or DC voltage measurement (mV)
Resistance measurement; Continuity test; Diode test
µF Capacitance measurement
°C/°F Temperature measurement
AC or DC current measurement
NCV (Non-contact Voltage) + Partial OFF mode of the multimeter
(NCV function active)

10
Digital Multimeter Model 5233
CHAPTER 3
OPERATION
3.1 Turning the Multimeter ON
Turntheswitchtotheappropriatefunction.Allsegmentsofthedisplaywill
lightforafewseconds.Thescreencorrespondingtothechosenfunction
willthenappear.Themultimeterisnowreadyformeasurements.
3.2 Turning the Multimeter OFF
Toturnthemeteroffmanually,turntheswitchtoOFF.Ifleftunusedfor
15minutes,themeterwillturnoffautomatically.At14minutes,vebeeps
warnthatthemeterisabouttobeturnedoff.Toturnbackon,pressany
button on the unit.
NOTE: The position does not completely turn the multimeter off.
It remains active for non-contact detection of the presence of network
voltage (NCV).
3.3 Activating/Deactivating Auto-OFF
Bydefault,Auto-OFFisactivatedandthe
P
symbol is displayed.
A long press on the buttonduringstart-up,whileturningtheswitch
toanyrange,deactivatestheAuto-OFFfunction.The
P
symbol is not
displayed.
3.4 Auto and Manual Range Selection
By default, the meter is in auto-range. This is indicated by the AUTO
symbolonthedisplay.Whileon,theinstrumentwillautomaticallyadjustto
the correct measurement range when taking the measurement.
TochangetherangeselectiontoManual,pressthe button.

Digital Multimeter Model 5233
11
3.5 Non-Contact Voltage (NCV)
• Turn the rotary switch to the NCV position.
• Move the Model 5233 (NCV detection sensor) close to the
potentiallyliveconductor(s)(presenceofphase).
Ifanetworkvoltageof90Vispresent,theback-lightinglightsupred,oth-
erwise,itremainsoff.
3.6 Voltage Measurement
The Model 5233 measures AC voltage at low input impedance (VLOWZ),
DC and AC voltages.
• Set the switch to , ,or . When set to the
device is in AC mode only.
• For or ,selectACorDCbypressing .Bydefault
the meter is in DC mode.
• Insert the red lead to the red “+” input jack and the black lead to
the black “COM” input jack.
• Connect the test probe tips to the sample under test.
Figure 3-1

12
Digital Multimeter Model 5233
3.7 Resistance Measurement
WARNING: When making a resistance measurement, make sure that
the power is off (de-energized circuit). It is also important that all
capacitors in the measured circuit be fully discharged.
• Turn the rotary switch to the range.
• Insert the red lead to the red “+” input jack and the black lead to
the black “COM” input jack.
• Connect the test probe tips to the sample under test.
Figure 3-2

Digital Multimeter Model 5233
13
3.8 Continuity Test
WARNING: When making a resistance measurement, make sure that
the power is off (de-energized circuit).
• Turn the rotary switch to the position.
• Press the button. The symbol is displayed.
• Insert the red lead to the red “+” input jack and the black lead to
the black “COM” input jack.
• Connect the test probe tips to the sample under test.
• The buzzer sounds when the circuit to be checked is DC or has
aresistanceoflessthan100Ω±3Ω.
Figure 3-3

14
Digital Multimeter Model 5233
3.9 Diode Test
WARNING: When making a diode measurement, make sure that the
power is off (de-energized circuit).
• Turn the rotary switch to the position.
• Press the button twice. The symbol is displayed.
• Insert the red lead to the red “+” input jack and the black lead to
the black “COM” input jack.
• Connect the test probe tips to the sample under test.
Figure 3-4

Digital Multimeter Model 5233
15
3.10 Capacitance Test
WARNING: When making a capacitance measurement, make sure that
the power is off (de-energized circuit). Observe the connection polarity
(+ to the red terminal, - to the black terminal).
• Make sure that the capacitor to be measured is discharged.
• Turn the rotary switch to the position.
• Insert the red lead to the red “+” input jack and the black lead to
the black “COM” input jack.
• Connect the test probe tips to the sample under test.
Figure 3-5

16
Digital Multimeter Model 5233
3.11 Temperature Test
• Turn the rotary switch to the position.
• Press the button to select the temperature unit and scale
(ºC/ ºF)
• Connect the temperature probe adapter to the “COM” and “+”
terminals,observingthepolarity.
• Connectthetemperatureprobetotheadapter,observingthe
polarity.
NOTE: If the probe is disconnected or open-circuit, the display unit
indicates .
Figure 3-6

Digital Multimeter Model 5233
17
3.12 Current Measurement
• Turn the rotary switch to the position.
• Select AC or DC by pressing the button.Bydefaultthe
meterisinDCmode.Dependingontheselection,thescreen
displays AC or DC.
• Insert the red lead to the “10A” input jack and the black lead to
the “COM” input jack.
• Connect the multimeter in series in the circuit.
Figure 3-7

18
Digital Multimeter Model 5233
CHAPTER 4
MAINTENANCE
4.1 Warning
• Removethetestleadsfromanyinputbeforeopeningthecase.Do
not operate the instrument without a battery case cover.
• Toavoidelectricalshock,donotattempttoperformanyservicing
unlessyouarequaliedtodoso.
• Ifthemeterisnotgoingtobeusedforalongperiodoftime,takeout
the batteries. Do not store the meter in high temperatures or high
humidity.
• Toavoidelectricalshockand/ordamagetotheinstrument,donotget
waterorotherforeignagentsintotheprobe.
4.2 Battery Replacement
• The batteries will need to be replaced when the symbol appears
on the display.
• The meter must be in the OFFpositionanddisconnectedfromany
circuit or input.
• Usingascrewdriver,unscrewthefourscrewsofthebattery
compartmentcoveronthebackofthehousing.
• Replacetheoldbatterywithonenew9Vbattery,observingthe
polarity.
• Replace the battery compartment cover and tighten the screws.