
Doc No. 7401247
MODEL 9 SERIES WELD HEAD
OPERATION MANUAL
5.0 ROTATION (SPEED) MANUAL CALIBRATION (Contd.)
5.4 Add one additional level with a duration of 5 or more seconds, with
the rotation mode set to “OFF”. This additional level (with rotation
off) will hold the electrode at its stopped position allowing the operator
to see if it traveled too far (traveling too fast) or not far enough
(traveling too slow).
5.5 Be sure the power supply is in the “TEST” mode, and run the program.
Should the electrode rotate too fast (traveling more than 360 degrees),
stopping beyond the home position, slow it down by turning the
potentiometer counter clockwise (CCW). Should the rotation be too
slow (traveling less than 360 degrees) stopping before it reaches a full
360 degrees, turn the potentiometer clockwise (CW). Continue this
procedure until the electrode stops after 360 degrees of travel.
6.0 CLAMP INSERT & GAS SEAL INSTALLATION
Clamp inserts are designed for a specific tube or pipe size. Select the
appropriate clamp for the specific tube or pipe to be welded. A set of clamps fits
one side of the weld head. Except for some special applications, or when using a
model 9E or 9ER type weld head, two sets of clamps are required. One set fits
on each side of the head.
Note: A set of clamp inserts is made from one piece, and then split into two
halves. Each half is marked with matching alphanumeric characters.
Keeping matched sets together (on the same side of the weld head) will
result in the highest level of weld joint concentricity.
6.1 Before mounting the clamp insert into the head be sure both the clamp
and weld head clamp mounting groove are clean and free of dirt, oil or
other contaminants. The clamp mounting screw need only be lightly
snug. See fig. 6.
CAUTION
Use of the wrong size clamp, or clamping the head on the curved section of an
elbow or other irregular surface can result in damage to the clamp insert, weld
head clamp latch and/or hinge.