
Document No. 992053. Issue 1 –22.05.23 Page 1
1. Introduction
These instructions are intended to assist you with the use and maintenance of the split
leg positioner accessory. It is important that these instructions are read thoroughly
before using the equipment. The device will be adversely affected, and its life
expectancy reduced, if the following instructions are not observed.
We also recommend that the user familiarise themselves with the QA4 Powered Mobile
Surgery System instructions for use, document no. 992010.
If any part of the split leg positioners are damaged or faulty, the device must
be taken out of use with immediate effect and the fault reported to Anetic
Aid, your authorised dealer or maintenance department. The split leg
positioners must not be used until the damage or fault has been repaired.
1.1. Warnings & Cautions
Various warnings and cautions are made throughout these operating instructions.
A WARNING is given when the personal safety of the patient or user may be
affected and when disregarding this information could result in injury.
A CAUTION is given when special instructions must be followed. Disregarding
this information could result in permanent damage being caused to the device.
1.2. Intended Use & Contraindications
The split leg positioner accessory is intended for use within a theatre environment to
aid with the positioning and abducting of patients legs during operative procedures.
1.3. Equipment Classification
The equipment referenced in this document has been classified as a Class 1 Medical
Device, and is both CE and UKCA marked under the scope of the Medical Devices
Regulation 2017/745 and the UK Medical Devices Regulations 2002.
1.4. Putting the Split Leg Positioner into Service
It is important that the split leg positioner sections are cleaned and disinfected before
being put into service. Use this manual to check all the functions and refer to the
‘Cleaning and Disinfecting’ instructions.