6
!
For your safety
2.0 For your safety
• The ATMOS patient chair E1 is designed in accordance
with IEC 601/EN 60601. The chair conforms to VDE
protection class I. It may only be connected to a correctly
installed shockproof socket.
• The chair may only be operated under constant supervision
(IEC 601-1/EN 60601-1).
• Prior to starting the chair for the rst time, check whether
the supply voltage indicated on the line voltage selector
corresponds to the value of your local mains supply.
• The power cable supplied with the chair (or an equivalent
one) must be used exclusively for connection to the mains
supply.
• Observe correct wiring when installing plugs speci c to
foreign countries:
green/yellow: ground conductor
blue: neutral conductor
black or brown: phase
• Prior to starting the chair, check whether connecting cables
show signs of damages. Damaged cables must be re-
placed.
• For the purpose of separating the chair from the mains, the
mains plug must always be pulled out of the wall socket rst
before unplugging the power cable from the chair. Never
touch power cord or plugs with wet hands.
• The ambient conditions speci ed in section 9.0 "Technical
speci cations" must be strictly observed.
• The ATMOS patient chair E1 is not intended for operation
in areas of medical rooms subject to danger of explosion.
Areas subject to explosion hazards may be created by the
use of in ammable anaesthetics, skin cleansers and disin-
fectants.
• The user must be familiar with the operation of the chair.
• ATMOS cannot be held liable for injury to persons or
damage to property if
- the parts used are not original ATMOS parts,
- the user instructions given in this manual are dis-
regarded.
• Before using the ATMOS patient chair E1, please
carefully read and observe the safety instructions and
recommendations listed in this manual.
• The electric motor is protected by an integrated thermo
protection switch. After 2,5 minutes of continued opera-
tion, the motor needs a cool down period of approx. 10
minutes. If the thermo protection switch activates, the
motor needs a cool down period of approx. 20 minutes
• The device may not be operated in:
- explosion-hazardous areas
- humid areas.
• In case of malfunction or abnormal noises, please imme-
diately disconnect the chair from the mains supply. Please
contact service technician.
• The device may only be used for the purpose described
in these instructions.