2.0 For your safety
!
For your safety
mains power supply.
• Only proper and undamaged plugs and extension cables
may be used.
• To disconnect the ATMOS®i View PRO from the power
supply, the power cable must be removed from the wall
outlet. Only then can the connection cable from the
ATMOS®i View PRO be disconnected. Never touch plug
or line with wet hands.
• Please observe the ambient conditions stated in the
technical data (chapter 9.0).
• The ATMOS®i View PRO complies with the
electromagnetic immunity requirements of standard IEC
60601-1-2 / EN 60601-1-2 „Electromagnetic Compatibility -
Medical Electrical Devices“.
• ATMOS is not liable for personal injury and damage to
property if
- no original ATMOS parts are being used,
- the advice for use in these operating instructions is not
being observed,
- assembly, new settings, alterations, extensions
and repairs have been carried out by personnel not
authorised by ATMOS.
• Unplug the device immediately if you observe fumes,
sparks or weird noises.
• After a longer use of the ATMOS®i View PRO in
connection with an ear speculum the patient may feel
dizzy!
• With every light source a warming of tissue due to
absorption may occur. Please make sure to reduce
duration of use to a minimum, to switch off the light
source when not in use and to check heat development if
necessary.
• Take into consideration, when setting up the microscope,
that the elastic force of the arm – without microscope head
– is exceedingly strong. Operate the break of the height
adjustment carefully.
• Risk of injury! Take care not to roll the mobile stand over
your feet when moving the stand.
• Please note that only PCs and monitors with IEC 606010-
1/EN 60601-1 approval may be connected to the video
outlet of the ATMOS®i View PRO supply module!
• During operation, the user is obliged to regularly check
the microscope for proper function. In the unlikely event
of failure the user must take precautions to continue the
treatment of the patient with suitable methods.
• Make sure that the unit is positioned so that all the controls
and the on/off switch are always accessible.
• To safely disconnect the unit from the mains, the power
cord must be removed from the IEC connector of the
control device!
• The ATMOS®i View PRO is a device designed in line with
IEC 60601-1-1/EN60601-1 and it is a device with protection
class I. In order to avoid the RISK of electrical shock, this
unit may only be connected to a mains supply with properly
installed earth conductor.
• Power cables, accessories and access cables need to be
checked for defects prior to setting up the ATMOS®i View
PRO. Defect cables must be replaced immediately.
• The ATMOS® i View PRO may only be operated by qualied
personnel.
• The ATMOS®i View PRO is not designed to be used
in an explosion-hazardous environment. Explosion-
hazardous areas may be caused by the use of ammable
anaesthetics, skin cleansing products and skin
disinfectants.
• The ATMOS®i View PRO may be operated only in rooms
used for medical purposes, but not in areas subject to
explosion hazards and in oxygen rich environments.
• If uids penetrated the ATMOS®i View PRO it needs to
be sent in and may only be used after the check up of an
ATMOS authorised person.
• After the transport of the ATMOS®i View PRO in
temperatures below 0°C or prior to rst start up it should
be kept at room temperature for at least six hours. If the
ATMOS®i View PRO is not acclimatised it may not be used.
• Do not plug in electric connections (plug, socket) under
the use of force. If this is not possible check whether the
plug ts the socket. If you should ascertain a defect in the
connection you should have it repaired by our service.
• Never look straight into the sun with lenses or eye lenses.
• Always make sure that you do not blind patients with the
light source! Watch out that patients do not look directly into
the light source!
Never look directly into the light source!
> Damage to the eyes due to the strong glare.
• Please pay attention to the period tests in chapter 6
„Service and maintenance“ on page 24.
• Prior to every use the microscope suspension (all joints
included) need to be checked for safe connections.
• Take care that the patient does not touch the device or have
any contact with it.
• Please note that only the ATMOS®Strobo 21 LED may be
connected to the strobe port of the ATMOS®i View PRO!
• Please observe the EMC Directives. Failure to follow this
guideline can result in a hazard.
• Dispose of wrappings accordingly.
• Before connecting the ATMOS i View PRO it needs to
be checked whether the requested mains voltage of the
ATMOS®i View PRO matches the mains voltage of the