12
8.0 Cleaning / Disinfection
The way the suction device is used determines its reliability and
safety. These hygiene measures are indispensable for protecting
the patient and the user and for maintaining a safe and reliable
suction device.
When the suction device is used on another patient or in case
the device has been oversucked a professional reprocessing
bythemanufacturer,byacertiedATMOSpartnerorbya
specialist, who is authorised by ATMOS, in compliance with
Medical Devices Operator Ordinance, Medical Devices Act and
BV-Med rules is required (please see chapter 9.2 reprocessing).
8.1 Basic information
●The following cleaning measures must only be performed if the
device was used. When the device is only rarely used then a
function check must be conducted at least every three months.
Only when you adhere to these points the function of the device
can be guaranteed.
●We recommend you to document any maintenance work and
also any exchange of parts.
●Please always wear disposable gloves for any work you
perform.
●Prior to cleaning the device please remove and dispose all
disposable parts like filter, fingertip and catheter.
●Prior to cleaning please remove the mains cable from the
device.
●The described measures for cleaning and disinfection
do not replace valid rules for operating the device!
●Please observe the instructions for use prescribed by the
manufacturers of disinfectants. Especially regarding
concentration and regarding their suitability for use.
●Attention: The lid parts and silicone hoses might get dyed by
some disinfectants; a fact which does not take effect on the
attributes of the materials.
●Basically all parts (canister, cover, overflow safety and
hose) which come into contact with suction material must be
cleaned and disinfected. Single-use parts like, for example,
filters, catheter, fingertip... must be exchanged as soon as
the device is used in another patient. Please see the different
instructions for cleaning (chapter 8.3).
●In case the suction device is used in one patient only device
and accessories should, for hygienic reasons, be cleaned
and disinfected. Please see the different instructions for
cleaning (chapter 8.3).
●Cleaning in an automatic cleaner and disinfector is also possible
(hose connector, secretion canister and canister lid).
Thermal disinfection is carried out at 93° C.
8.1.1 Bacteriallter
● Exchangethebacteriallterwhenchangingthepatient.ATMOS
recommends:Replacethebacteriallterafter14days,evenif
there is no patient change.
● Replacethebacteriallterwithanewlterifitisdiscoloured,
contaminated or oversucked.
● Payattentiontostoringasucientnumberofreplacement
bacteriallters.
8.1.2 Suction hose, hose connector and
vacuum hose
● Suctionhoseandhoseconnectormustalwaysbedisinfected
with an instrument disinfectant recommended on page 14 in
case the device is used in another patient. Prior to disinfection
or autoclaving the parts must be rinsed with clear water for at
least10secondsinordertoincreasecleaningeciency.
Please observe the instructions for use of the relevant
disinfectant solution!
8.1.3 Fingertip
●Fingertip is not included in delivery, please order
separately.
●Prior to using the device in another patient the
ngertipmustbeexchanged.
●In case the device is used in one patient only we
recommend for hygienic reasons the exchange of
thengertipeveryday.
8.1.4 Secretion canister
● Thesecretioncanistermustalwaysbe
disinfected with an instrument disinfectant
recommended on page 14 in case the device is
used in another patient. Prior to disinfection resp.
autoclaving please take care to empty the canister
and to rinse it with clear water in order to increase
cleaningeciency.
Please observe the instructions for use of the
relevant disinfectant solution!
● Incasethedeviceisusedinonepatientonlywe
recommend disinfection of the canister every day
as described above.
● Forhygienicreasonswerecommendtoemptythe
secretion canister after every suction procedure
and to rinse it with clear water.
Information on how to remove the canister you
willndinchapter5.2onpage9.
● Incasethepartsareusedinonepatientonly
we recommend an exchange every 4 weeks.
● Inadditionwerecommendtothoroughlyrinse
hose, hose connector and vacuum hose with clear
water and to disinfect them at least once a day as
described above.
!
!
!
8.1.5 Canister lid
●The canister lid must always be disinfected with an
instrument disinfectant recommended on page 14 in
case the device is used in another patient. Prior to
disinfection please take care to remove the bacterial
lterfromthelidandtodisassemblethelidintoits
singlecomponents(lid,oatballandhose
connector). Prior to disinfection or autoclaving the parts
must be rinsed with clear water for at least 10 seconds
inordertoincreasecleaningeciency.Afterwardsthe
parts have to be disinfected.
Fig 17.