2.0 For your safety
• The Hygrotherm plus is produced according to IEC 601/
EN 60601 and listed in the following classes:
- VDE Class of protection 1
- ClassIIa(EEC93/42).
• In order to avoid the danger of an electric shock, this
device should only be connected to an earthed power
connection.
• The unit must be installed by a specialist authorized by
ATMOS(seechapter3.3).
• The Hygrotherm plus may only be used under the
supervisionofskilledstawhohavebeenauthorized
by ATMOS and trained in its operation (IEC 601-1 / EN
60601-1).
• The mains voltage indicated on the type plate must
correspond to the values of the supply network.
• Make sure prior to every application of the equipment that
it is technically safe and in proper condition. Damaged
cables must be replaced immediately!
• Correctcongurationinassemblyofcountry-specic
connections:
- green/yellow:Protectiveconductor(PE)
- blue:Neutralconductor(N)
- blackorbrown:Phase(L)
• Itisessentialthatthecountry-specicrequirementsfor
connection of medical equipment to the public drinking
water supply are considered. When in doubt, please
contact your local ATMOS partner.
• The control panel must be clearly visible and accessible
fortheuser.Ensuresucientstabilityontheinstallation
surface.
• In order to safely disconnect the device from the mains
power supply, the power cable must be removed from the
IEC power connector of the Hygrotherm!
)In the thermal process for reducing bacterial count hot
water is being ducted through the irrigation handle.
Please do not take the handle from its holder or spray
water!
)Metal parts can be hot!
)Prior to spraying, the water temperature must be checked
bytheuser(display)!
• Switchomainswitchafternishingworkinpracticeand
close water valve of supply.
• The ATMOS Hygrotherm plus may be operated only
in rooms used for medical purposes, but not in areas
subject to explosion hazards and in oxygen-rich
environments.
• The irrigation jet must not get into contact with
contaminated material.
• The irrigation jet must be disinfected after each use.
• The irrigation jet may only be used when a hose tip is
attached!
• Take caution to avoid injury to the eardrum when
introducing the water jet!
• For hygienic reasons, hose tips must be changed after
each patient. This also prevents retrograde germination
of the warm water system.
• Use only for irrigation of auditory canal!
• ATMOS is not liable for personal injury and damage to
property if
- no original ATMOS parts are being used,
- the advice for use in these operating instructions is
not being observed,
- assembly, new settings, alterations, extensions
and repairs have been carried out by personnel not
authorized by ATMOS.
• Please note:
A medical isolation transformer with earth leakage
monitor or any similar safety system acc. to EN 60 601-1
is required if several devices are connected over one
common power supply. The transformer must correspond
to the power consumption of all the devices to be
connected.
• After switching on the device or at least once a week,
theowratemustbecheckedusingavolumemeasuring
funnel. The quantity of 500 ml/min may not be exceeded.
The water jet has to be straight.