
5
2.0 For your safety
• The Variotherm plus is produced according to IEC 601 /
EN 60601 and listed in the following classes:
- VDE Class of protection 1
- ClassIIa(93/42/EEC).
• WARNING: In order to avoid the danger of an elec-
tric shock, this device should only be connected to an
earthed power supply.
• The device must be installed by a specialist authorized by
ATMOS(seesection3.3).
• The Variotherm plus may only be used under the supervi-
sionofskilledstawhohavebeenauthorizedbyATMOS
andtrainedinitsoperation(IEC601-1/EN60601-1).
• The mains voltage indicated on the type plate must corre-
spond to the values of the supply network.
• Make sure prior to every application of the equipment that
it is technically safe and in proper condition. Damaged
cables must be replaced immediately!
• Correctcongurationinassemblyofcountry-specic
connections:
- green/yellow:protectiveconductor(PE)
- blue:neutralconductor(N)
- blackorbrown:phase(L)
• Itisessentialthatthecountry-specicrequirementsfor
connection of medical equipment to the public drinking
water supply are considered. When in doubt, please
contact your local ATMOS partner.
• The control panel must be clearly visible and accessible
fortheuser.Ensuresucientstabilityoftheinstallation
surface.
• In order to safely disconnect the device from the mains
power supply, the power cable must be removed from the
IEC power connector of the Variotherm plus!
)In the thermal process for reducing bacterial count, hot
water is conducted through the irrigation handle. Please
do not take the handle from its holder or spray water!
)Metal parts can be hot!
)Prior to spraying, the water temperature must be checked
bytheuser(display)!
• Switchothemainswitchafternishingworkintheprac-
tice and close the water valve of the water supply.
• The Variotherm plus may be operated only in rooms used
for medical purposes, but not in areas subject to explo-
sion hazards and not in oxygen-rich environments.
• All additional equipment that is connected to the analog
and digital interfaces of the device must meet the require-
mentsofrelevantENspecications(e.g.,EN60950for
data processing equipment and EN 60601 for electrical
medicalequipment).Inaddition,congurationsmust
satisfysystemspecicationEN60601-1-1.Whenaddi-
tional equipment is connected to the signal input or signal
output section on the device, the person carrying out the
connectionisdeemed“asystemcongurationoperator”
and as such is responsible for meeting the requirements
ofsystemspecicationEN60601-1-1.Ifyouhaveany
questions, please contact your local specialist supplier or
ATMOS Technical Service.
• The irrigation jet must not get into contact with contami-
nated material.
• Only use the warm water jet when a hose tip is installed!
• Take caution to avoid injury to the eardrum when insert-
ing the water jet!
• For hygienic reasons, hose tips must be changed after
each patient. This also prevents retrograde contamination
of the warm water system.
• Use only for irrigation of the auditory canal!
• ATMOS is not liable for personal injury and damage to
property if
- no original ATMOS parts are being used,
- the advice for use in these operating instructions is
not being observed,
- assembly, new settings, alterations, extensions, and
repairs have been carried out by personnel not au-
thorized by ATMOS.
• Please note:
A medical isolation transformer with earth leakage
monitor or any similar safety system acc. to EN 60601-
1 is required if several devices are connected over one
common power supply. The transformer must correspond
to the power consumption of all the devices to be con-
nected.
• After switching on the device or at least once a week,
theowratemustbecheckedusingavolumemeasuring
funnel. The quantity of 500 ml/min may not be exceeded.
The water jet has to be straight.