Baker Ruskinn PhO2xBox User manual

Affix Serial Number Sticker Here
User Manual UM-048
PhO2xBox
Cell-Culture Chamber System
PhO2xBox

i
Unit 8&9 / York Park / Bridgend, United Kingdom / CF31 3TB
+44 (0) 1656 645688 / Fax: +44 (0) 1656 667966 / sales@ruskinn.com /
www.bakerruskinn.com
CONTENTS
TABLE OF FIGURES ...................................................................................................................................I
TABLE OF TABLES ....................................................................................................................................II
INTRODUCTION.......................................................................................................................................3
SAFETY INSTRUCTIONS............................................................................................................................ 4
REGULATORY COMPLIANCE....................................................................................................................5
SYMBOLS.............................................................................................................................................6
TRANSPORT AND STORAGE ....................................................................................................................7
LOCATION AND HANDLING OF THE PHO2XBOX...................................................................................... 7
ENVIRONMENTAL OPERATING CONDITIONS..........................................................................................7
SERVICE REQUIREMENTS ........................................................................................................................8
ELECTRICAL SUPPLY REQUIREMENTS .................................................................................................8
GAS SUPPLY REQUIREMENTS..............................................................................................................8
PHO2XBOX OVERVIEW ............................................................................................................................ 9
SETTING UP THE GAS CONTROLLER AND CULTURE CHAMBER ............................................................ 13
SETTING UP THE CELL CULTURE CHAMBER ON A LABORATORY BENCH..........................................13
FOR INSTALLATION INSIDE A WORKSTATION (SCI-TIVE OR INVIVO2) OR CO2INCUBATOR..............14
USING THE GAS CONTROLLER............................................................................................................... 17
THE TOUCH SCREENS ............................................................................................................................ 19
START UP SCREENS ...........................................................................................................................19
MAIN SCREENS..................................................................................................................................19
ATMOSPHERE CONTROL SCREEN......................................................................................................20
RE-SETTING THE O2 AND CO2 VALUE DURING OPERATION............................................................... 22
WARRANTY INFORMATION...................................................................................................................26
DISPOSAL INFORMATION......................................................................................................................27
CONTACT DETAILS.................................................................................................................................28
TABLE OF FIGURES
Figure 1: Blue Tube ...............................................................................................................................10
Figure 2: Mains Lead ............................................................................................................................. 10
Figure 3: Signal Lead ............................................................................................................................. 10
Figure 4: Gas Controller (With air pump)..............................................................................................10
Figure 5: N2Gas Hose............................................................................................................................11
Figure 6: CO2Gas Hose .........................................................................................................................11
Figure 7: Cell Culture Chamber (Small QTK-A01)..................................................................................11
Figure 8: Cell Culture Chamber (Large QTK-A02).................................................................................. 11
Figure 9: Cell Culture Chamber (Large QTK-AB02)................................................................................12

ii
Unit 8&9 / York Park / Bridgend, United Kingdom / CF31 3TB
+44 (0) 1656 645688 / Fax: +44 (0) 1656 667966 / sales@ruskinn.com /
www.bakerruskinn.com
Figure 10: Rear of Cell culture chamber ...............................................................................................13
Figure 11: Rear of Cell culture multi chamber ...................................................................................... 13
Figure 12: Rear of gas controller...........................................................................................................14
Figure 13: Outside view of InvivO2 showing Universal Multi-Cable Gland ...........................................14
Figure 14: Inside view of InvivO2 showing Universal Multi-Cable Gland .............................................. 15
Figure 15: Black frame and blue plug.................................................................................................... 15
Figure 16: Multi-cable Gland.................................................................................................................16
Figure 17: Pneumatic schematic...........................................................................................................16
Figure 18: Front View............................................................................................................................17
Figure 19: Left View .............................................................................................................................. 17
Figure 20: Right View ............................................................................................................................17
Figure 21: Rear View ............................................................................................................................. 17
Figure 22: Rear View (Multi version) ....................................................................................................18
Figure 23: Top View ..............................................................................................................................18
Figure 24: Screen 1 –Logo screen ........................................................................................................19
Figure 25: Screen 2 –Main screen........................................................................................................19
Figure 25: Multi chamber version only –Chamber selection screen ...................................................20
Figure 25: Multi chamber version main screen ....................................................................................20
Figure 26: Screen 3 - Atmosphere Control Screen................................................................................20
Figure 27: Screen 4 –Level SET POINT Screen......................................................................................21
Figure 28: Screen 5 –Atmosphere Control Screen...............................................................................22
Figure 29: Screen 6 –Hypoxic Cycle Menu Settings .............................................................................22
Figure 30: Screen 7 –Trend Log............................................................................................................23
Figure 31: Screen 8 –Engineer Mode Screen .......................................................................................23
TABLE OF TABLES
Table 1: List of Symbols...........................................................................................................................6
Table 2: Electrical Service Requirements................................................................................................ 8
Table 3: Standard Gases..........................................................................................................................8

3
INTRODUCTION
Please read this manual carefully before using the PhO2xBox and familiarise yourself with all aspects
of using the workstation. The Baker Company (Baker) or Ruskinn Technology Ltd (Ruskinn) does not
accept responsibility for accidents to personnel or damage to the PhO2xBox workstation resulting from
incorrect use.
The PhO2xBox comprises of a Gas Controller and a Cell-Culture Chamber has been specifically designed
and developed to meet the requirements of all laboratories.
The system has built in gas control with Oxygen and Carbon Dioxide sensors to enable constant
monitoring of the atmosphere. Hypoxic conditions are created using Nitrogen (N2), Carbon Dioxide
(CO2) and Compressed Air (O2) (or 25% Oxygen in Nitrogen) to achieve the desired atmosphere. The
user can select the Oxygen concentration from 0.1% to 20.9%, and Carbon Dioxide concentrations
from 0.1% to 30.0%.
Many unique features of the PhO2xBox workstation are covered in detail in this manual. It is
recommended that the user be fully conversant with the instruction and procedures, and that the
operator familiarises themselves with all aspects and functions of the system before it is commissions
to maintain optimum performance.

4
SAFETY INSTRUCTIONS
Please read this manual carefully and familiarize yourself with all aspects of the Cell-Culture Chamber
System.
The System comprises a Gas Controller and a Cell-Culture Chamber.
Baker and/or Ruskinn does not accept responsibility for accidents to personnel or damage to the Cell-
Culture Chamber System resulting from incorrect use.
The mains appliance coupler and plug are the AC mains supply isolation device and must be easily
accessible when installed.
In case of emergency disconnect the Gas Controller from the AC Mains Outlet.
Ensure that the connecting cable is not squeezed or bent when the System is being installed or moved.
All installation work and adjustments to the System must be carried out by qualified personnel.
Work performed by persons with insufficient technical knowledge may adversely affect the
performance of the system or cause physical injury or damage to the equipment.
All servicing and repairs must be carried out by a qualified customer service engineer.
Only genuine spare parts must be used.
All gas cylinders must be adequately secured before connection to the System.
CAUTION: Asphyxiation Risk
The PhO2xBox uses Nitrogen (N2) and Carbon Dioxide (CO2) as part of normal use with the volume
released externally is inconsequential. In the event of a leak or malfunction this gas release may
become excessive. DO NOT OPERATE this unit in a SMALL ENCLOSURE such as a small room or walk-
in closet. An accidental release of Nitrogen or Carbon Dioxide could create an asphyxiating
atmosphere in a small space.
If the equipment is not use in a manor specified by the manufacture, the protection provided by the
equipment may be impaired.
Failure to adhere to these safety instructions could cause serious injury and will invalidate the
workstation warranty. Ruskinn technology limited accepts no responsibility for any accident, injury or
loss caused by unsafe operation of the workstation.

5
REGULATORY COMPLIANCE
WEEE:
This equipment must be disposed of in accordance with the Waste from Electrical and Electronic
Equipment (WEEE) Directive
This product must not be treated as household waste. Instead, it shall be handed over to an
appropriate collection point for the recycling of electrical and electronic equipment.
If in doubt, please return this equipment to Ruskinn Technology Ltd who will correctly dispose of it for
you. We strongly recommend that this product is returned to RTL at the end of its useful life.

6
Symbols
Before using the PhO2xBox, please ensure that you are familiar with the symbols.
Symbol
Meaning
Refer to user manual.
~
Alternating current
O
Off
I
On
Functional Earth Connection
Protective Earth Connection
Caution, do not remove covers. No end user serviceable parts behind covers.
Please refer to this manual in all cases where this symbol appears, in order to
find out the nature of the Potential Hazard and actions to be taken in order to
avoid the Hazard.
Warning, this equipment contains high voltage circuitry.
Contains material or substances that may be hazardous to human health. Please
refer to your local biohazardous material handling procedure for further advice
on the handling and disposal of these items.
PhO2xBox contains hazardous components and must not be disposed of at a
household waste site. Instead it should be taken to the appropriate collection
point for the recycling of electrical and electronic equipment.
USB socket
Date of manufacture in format
YYYY MM
Table 1: List of Symbols

7
TRANSPORT AND STORAGE
When not in use, the PhO2xBox controller box & cell culture chamber must only be stored within a
temperature of between 0°C and 30°C
Storage outside of this range may damage the gas controller.
LOCATION AND HANDLING OF THE PHO2XBOX
The PhO2xBox should only be installed or relocated by a qualified engineer. To arrange installation or
relocation please contact your local distributor.
The mains appliance coupler and plug are the AC mains supply isolation device and must be easily
accessible.
ENVIRONMENTAL OPERATING CONDITIONS
The PhO2xBox should only be operated under the following environmental conditions:
•Temperature –Between 15⁰C and 30⁰C
•Humidity –Between ambient and 90% RH, Non-Condensing
The workstation must be located in a well-ventilated area.

8
SERVICE REQUIREMENTS
Electrical Supply Requirements
The Gas Controller must be connected to a mains power supply. A power cord is supplied to connect
the workstation to the mains supply. If an alternative power cord is used is must be rated appropriately
for the power requirements of the Gas Controller, refer to Table 2. The Gas Controller must be
connected to a protective earth.
To ensure safe operation of the system, it must be connected to a supply of the correct voltage and
frequency as stated on the rating label shown.
The mains supply voltage fluctuations must not exceed +/- 10% of the nominal mains voltage.
Voltage Range
Frequency
Rated Current
Rated Power
220±20V
50Hz
0.1A
35W
110±10V
60Hz
0.1A
25W
Table 2: Electrical Service Requirements
Gas Supply Requirements
The gas regulator should provide a minimum supply pressure of 2 bar gauge. The maximum supply
pressure permissible is 3 bar gauge. See Table 4 below. A supply pressure greater than this will damage
internal components of the gas mixer and will invalidate the warranty.
Gas
Symbol
Specification
Regulator Output
Pressure
Nitrogen
N2
Oxygen Free
(Industrial or medical)
2 to 3 bar
Carbon Dioxide
CO2
100%
(Industrial or medical)
2 to 3 bar
Table 3: Standard Gases
For the gas supply we recommend direct connections of the gas cylinder regulator to ensure minimal
pressure drops and flow rate restrictions. Only the above stated gases are to be used with the system.
Failure to comply with this may cause the product to become hazardous.
Only the above stated gases are to be used with the workstation. Failure to comply with this may
cause the product to become hazardous.

9
PHO2XBOX OVERVIEW
Background
The System comprises a Cell Culture Chamber , which comes in two sizes (Small and Large) and
different colours (Black & Transparent), and a Gas Controller. The Gas Controller is connected to the
Cell Culture Chamber. The Cell Culture Chamber can hold cell culture flasks and 6, 12, 24 or 96 well
plates. The gas concentration of the Cell Culture Chamber can be changed automatically by using the
touchscreen on the Gas Controller.
Hypoxic conditions are created inside the Culture Chamber by using Nitrogen (N2), Carbon Dioxide
(CO2) and laboratory air (which is sucked into the Gas Controller via an Air pump). The user can select
the Oxygen concentration from 0.3% to 20.0%, and the Carbon Dioxide concentration from 0.3% to
20.0%. The Cell Culture Chamber is well-sealed so it can reach the desired culture conditions quickly.
Built in Oxygen and Carbon Dioxide sensors enable constant monitoring of the atmosphere inside the
Cell Culture Chamber.
The Cell Culture Chamber can be placed into a CO₂ incubator, CO₂/O₂ incubator or Physoxia/Hypoxia
workstation.
CONTENTS (QTK-AO1/AO2)
1 mains lead
1 signal lead
1 blue tube (1m)
1 black gas hose (CO2) (3m)
1 blue gas hose (N2) (3m)
1 Cell Culture Chamber (Large QTK-AO2 or Small QTK-AO1)
Large: W x D x H (mm)= 360 x 300 x 230
Small: W x D x H (mm)= 340 x 300 x 160
1 Gas Controller (with air pump built in)

10
Figure 1: Blue Tube
Figure 2: Mains Lead
Figure 3: Signal Lead
Figure 4: Gas Controller (With air pump)

11
Figure 5: N2Gas Hose
Figure 6: CO2Gas Hose
Figure 7: Cell Culture Chamber (Small QTK-A01)
Figure 8: Cell Culture Chamber (Large QTK-A02)

12
Figure 9: Cell Culture Chamber (Large QTK-AB02)

13
SETTING UP THE GAS CONTROLLER AND CULTURE CHAMBER
Setting up the cell culture chamber on a laboratory bench
The Cell Culture Chamber can be placed adjacent to or on top of the Gas Controller.
Connect the blue tube to InA on the Cell Culture Chamber and OutA on the Gas Controller.
Connect the signal lead to InB on the Cell Culture Chamber and OutB on the Gas Controller.
Figure 10: Rear of Cell culture chamber
Figure 11: Rear of Cell culture multi chamber

14
Figure 12: Rear of gas controller
Use the blue hose to connect N2on the Gas Controller to the N2gas cylinder and the black hose to
connect CO2on the Gas Controller to the CO2cylinder.
Connect the mains lead to the right port shown on Gas Controller.
Open gas cylinders and turn the gas controller power on.
For installation inside a workstation (SCI-tive or InvivO2) or CO2Incubator
Place Cell Culture Chamber in desired Workstation or Incubator.
Place the Gas Controller adjacent to the Workstation or Incubator.
Locate Universal Multi-Cable Gland on left side panel of Workstation or Incubator.
Figure 13: Outside view of InvivO2 showing Universal Multi-Cable Gland

15
Figure 14: Inside view of InvivO2 showing Universal Multi-Cable Gland
Figure 15: Black frame and blue plug
Unscrew inside black frame of the Universal Multi-Cable Gland and remove the blue plug. The Plug
can contain up to 6 pre-drilled holes depending on how it was ordered. Additional holes can also be
drilled by user.
Connect the blue tube to InA on the Cell Culture Chamber and OutA on the Gas Controller. The blue
tube has to be put through the Universal Multi-Cable gland.
Connect the signal lead to InB on the Cell Culture Chamber and OutB on the Gas Controller. The lead
has to be put through the Universal Multi-Cable gland.
Ensure that the blue tube and signal cable goes through the Universal Multi-Cable Gland into the plug
from the inside.

16
Figure 16: Multi-cable Gland
Use the blue hose to connect N2on the Gas Controller to the N2gas cylinder and the black hose to
connect CO2on the Gas Controller to the CO2cylinder.
Connect the mains lead to the right port shown on Gas Controller.
Figure 17: Pneumatic schematic
Open gas cylinders and turn the Gas Controller power on.

17
USING THE GAS CONTROLLER
Gas Controller
The Gas Controller is shown as follows.
Figure 18: Front View
Figure 19: Left View
Figure 20: Right View
Figure 21: Rear View

18
Figure 22: Rear View (Multi version)
Figure 23: Top View

19
THE TOUCH SCREENS
Start up screens
When the unit is first powered a Logo Screen (Screen 1) will appear and then will be replaced by the
Main Screen (Screen 2).
Figure 24: Screen 1 –Logo screen
Main screens
From this menu you will be able to access the Main screen below.
Figure 25: Screen 2 –Main screen
The multi chamber version chamber selection must first be made.
Table of contents
Other Baker Laboratory Equipment manuals

Baker
Baker ChemoSHIELD CS500 User manual

Baker
Baker SterilGARD III Advance User manual

Baker
Baker InvivO2 1000 User manual

Baker
Baker ChemGARD FH4 User manual

Baker
Baker SterilGARD SG303 User manual

Baker
Baker SterilGARD SG404 User manual

Baker
Baker BD InoqulA User manual

Baker
Baker SterilGARD e3 SG404 User manual

Baker
Baker EdgeGARD HF User manual

Baker
Baker Bugbox Ax User manual
Popular Laboratory Equipment manuals by other brands

Agilent Technologies
Agilent Technologies 1290 Infinity II user manual

Nüby
Nüby IDVP50 manual

Thermo Scientific
Thermo Scientific BF51842C Installation and operation manual

Gilson
Gilson CENTRY 103 Minicentrifuge user guide

Nichiryo
Nichiryo Nichipet Premium user manual

VWR
VWR VAPOUR-Line lite instruction manual