Purchase Specications
EdgeGARD VF
EdgeGARD VF Vertical-Flow Recirculating Clean Bench
PAGE 9
1. EdgeGARD®VF vertical flow clean bench is available in 4’, 5’
and 6’ work surface widths. Each unit shall be provided with
a certified copy of the factory tests showing that filter leak
checks, electrical tests, down flow velocities, smoke patterns,
and airflow balancing have been performed. Tests to prove
Class 10 (ISO Class 4) air cleanliness for 0.5 micrometer
particles for model design shall be available if requested.
2. Hinged view screens shall be constructed of ¼ inch
shatterproof scratch and chemical resistant polycarbonate,
with a maximum opening of 30” for equipment loading.
To decrease glare and offer the best ergonomics, the view
screen shall be mounted at a 10° angle from vertical.
3. Supply filters shall be front loading and meet the zero-
probed HEPA 99.99% efficiency requirements on all particles
0.3 micron in size. System shall have washable pre-filters
to extend HEPA filter life. Magnehelic pressure gauge(s) are
provided to monitor filter loading and as a secondary means
of monitoring mass airflow.
4. The cabinet body shall be double walled. The exterior walls
shall be constructed of a single sheet of #16 gauge cold-rolled
steel with a white powder coated finish. Internal side and
back walls, as well as the work surface, shall be constructed
of #16 gauge 304 stainless steel. The side wall and the back
shall be of one-piece construction formed with radius 7/16”
corners. Either a stainless steel air diffuser or filter protector is
provided in the work area.
1. Interior Dimensions:
a) EG4VF – 46” L x 25 1/8” D x 31 13/16” H
b) EG5VF – 58” L x 25 1/8” D x 31 13/16” H
c) EG6VF – 70” L x 25 1/8” D x 31 13/16” H
5. High-velocity return air slots shall be located at the leading
edges of the work surface and side walls. Return air slots
protect against the backwash of dirty air entering the work
area when items are placed within the air stream on the work
surface.
6. Cabinet shall have a momentum air curtain down flow
velocity profile, which provides a higher airflow velocity of
100 FPM nominal behind the view screen relative to a down
flow velocity of 50-65 FPM nominal over the work surface.
7. Cabinet shall have a welded, full perimeter drain pan with
7/16” radius corners for cleaning ability with a 1” drain valve
to capture any cleaning liquids or spills.
8. Each cabinet work area shall be provided with an internally
mounted GFCI duplex 120V outlet with drip-proof cover and
circuit breaker (mounted on the left), fluorescent lighting
providing 100 foot-candles of illumination at work surface
and electronic ballasts. The cabinet shall have separate
switches for the blower and the light; if the blower is off, the
light switch shall not work.
9. The cabinet shall be capable of automatically handling a 40%
minimum increase in pressure drop across the filter without
reducing total air delivery by more than 10%. A manual speed
controller shall be capable of handling a minimum 112%
increase in pressure drop across the filter. Test data to verify
these capabilities shall be provided upon request.
10. Cabinet design shall utilize a steel plenum provided to allow
filters to be directly clamped to the plenum against a closed
cell neoprene gasket. Plenum applies force to full perimeter
of filters rather than point force. Service of the system shall be
accessible from the front.
11. System shall be factory pre-wired and have a 17-foot external
length power cord with 20 Amp plug (type NEMA 5-20P).
System design shall have easy access to electrical panel.
12. All system components shall be certified by Underwriters
Laboratory (UL61010-1 2nd Edition) for electrical, fire and
personal safety.
13. Noise levels should meet 67 dBa or lower to provide operator
comfort and reduce effect of ambient noise in the laboratory.
14. System shall be capable of moving through a standard 80”
high doorway.
15. Cabinet shall have permanently affixed label located directly
above the viewing window that has the following phrase: “Do
not use for biological, chemical or radiological work where
hazardous materials are present.”
16. The system shall be warranted for three (3) years, parts and
labor.