EMISSION STRENGTH
The in rared thermometer comes actory preset at an emission strength o 0.95.
This value can be changed by 0.10 (10E) up to 1 (100E). These changes can be
made only by specially-trained personnel. In ormation concerning emission
levels o particular materials can be received rom your local sales agent.
Tip: The touch- ree in rared thermometer should not be used or measuring
temperatures o shiny or polished metallic sur aces. To change the emission
strength, press ⑥-②* ive times -⑥ or each change o 0.01 (1 E) -②.
Touch thermometer function (COT mode)
Insert the needle sensor ① into the object being measured and press ⑦ to
display the temperature or up to 4 minutes. A ter this time, the unit will
turn o automatically to save battery power. Pressing ⑦ will stop the
measurement and display the last temperature along with the message
“Hold”. To resume measurement, press ⑦ again.
1. Do not bend or rotate the needle sensor in the wrong direction.
2. Placing too much pressure on the sensor may cause damage.
STORAGE AND CLEANING
The sensor lens is the most sensitive element o the thermometer, and as
such must be kept clean. Clean the device care ully, using a so t cloth or
cotton swab dipped in water or pharmaceutical alcohol. Wait or the lens
to dry thoroughly be ore using. Do not immerse the thermometer in
luids. Store the unit at temperatures between -4°F and +149°F
(-20°C and +65°C).