
V. DEGASSED WATER AND TRANSDUCER CONSIDERATIONS
Water as a Measurement Medium*
The measurements are to be performed in water because ultrasound propagation in water closely
approximates that in tissues. The ultrasonic attenuation in water can be taken as a lower limit on the
attenuation that will be encountered in the body. Large areas in the body can consist of low
attenuating material such as urine and amniotic fluid. The use of water prevents measurements in a
more highly attenuating material such as liver equivalent gels from representing the highest possible
intensities that might be encountered in the body. A measurement temperature of 24ºC. (±3ºC.) is
chosen for convenience.
Degassed Water
Ultrasound Power measurement accuracy is affected by lowering the power reading if the water
contains more than five to ten parts per million of air. To de-gas, boll distilled water for one half hour,
then seal the container tightly and place it in a refrigerator. This process should give the required ppm
02quality. (A hand-held digital dissolved oxygen test meter is available from BC Group International,
Inc. that will directly measure from 0 to 20 ppm dissolved oxygen. Ask for Model DOT-20 Dissolved
Oxygen Tester). The degassed water container should be made of glass or a two liter soft drink
plastic bottle. Polystyrene one- gallon jugs should not be used since they let oxygen enter with time
and degrade water quality. Before testing, pour water into tilted test tank with minimum amount of
turbulence. The test tank water surface will absorb oxygen and a change of degassed water is
recommended before each test. An alternate method of degassing water is to heat the water to the
boiling point, then pull a vacuum on it for five to ten minutes. Water temperature affects accuracy, so
it is recommended to use an ambient testing temperature of +70º F. to +75º F. Sonic energy agitates
the water surface through heating and scattering. The time duration of each test should be limited to a
few minutes. Prolonged testing, particularly at higher power levels, will show visible signs of air
bubbles on the transducer, target and the absorbing rubber surface.
Transducer Wetting and Placement
After tilting the transducer into the water at a 45º angle, verify that its surface is uniformly wetted if
not, wipe the surface clean. The transducer should be pointed toward and center-positioned directly
above the cone target. Small measurement variations will occur due to placement of the transducer.
Try various positions above target to verify magnitude of error due to positioning.
Sonic Radiation Levels and Absorbed Energy
Measurement of power output levels of diagnostic and therapeutic ultrasound equipment have
become increasingly important to determine exact patient exposure levels and thus avoid a potential
risk to the patient. Since the Radiation Control for Health & Safety Act of 1968 and the 1976 Medical
Device Amendments to the Food, Drug. & Cosmetic Act became effective, all manufacturers of
diagnostic and doppler ultrasound equipment are required to submit information regarding their
maximum peak and average exposure level, beam patterns, and other pertinent information.
Submitted data of the early 1980's showed that for pulse echo transducers, the Spatial Peak
Temporal Average (SPTA) was less than 10 mW/cm2to 180 mw/ cm2and the Doppler SPTA levels
were loss than 600 mW/cm2. The power levels varied over one order of magnitude without any
significant change in diagnostic performance. Exposure levels for physical therapy applicator heads
ranged from 100 mw/ cm2to 3 W/cm2.
While there are no maximum limits in the United States at the present time for diagnostic and
therapeutic applicators, there is a generally accepted safe energy guideline limit. The maximum
average power output of the transducer in the SPTA intensity of the transducer (defined as the
maximum ultrasound intensity value in space averaged over a time duration) should not exceed safe
recommended levels. Typical safe figures of SPTA are 10 mW/ cm2for echo sounding and 100 mW/
cm2for Doppler devices. The maximum ultrasound SPTA for therapeutic use is 3 W/ cm2.
* Ultrasound propagation in water closely approximates human tissue and degassed water is the generally accepted test medium for
ultrasound transducers (see AIUM/NEMA Standards Publication #UL-1-1981, SAFETY STANDARD FOR DIAGNOSTIC ULTRASOUND
EQUIPMENT).
6