Beko FLM SF53 Owner's manual

METPOINT®FLM SF53
FLMSF53LL220 | FLMSF53DL220 | FLMSF53LL400 | FLMSF53DL400
Thermal ow meter
Installation and operation instructions
EN - English
10-232

Installation and operating manual EN
2 METPOINT®SF53
Contents
1. Safety information.......................................................................................................................................... 4
1.1. Pictograms and symbols.......................................................................................................................... 4
1.1.1. In this documentation........................................................................................................................................................4
1.1.2. On the device.......................................................................................................................................................................4
1.2. Signal words ........................................................................................................................................... 4
1.3. Safety instructions.................................................................................................................................. 5
1.4. Transport and storage ............................................................................................................................. 6
1.5. Intended use........................................................................................................................................... 7
1.6. Warranty and liability ............................................................................................................................. 7
2. Product information ....................................................................................................................................... 8
2.1. Scope of delivery .................................................................................................................................... 8
2.2. Type plate............................................................................................................................................... 8
2.3. Product overview and description ........................................................................................................... 9
2.3.1. Identication based on product code.............................................................................................................................9
2.3.2. Product description ......................................................................................................................................................... 10
2.3.3. Operating principle.......................................................................................................................................................... 10
2.4. Control and display elements ................................................................................................................ 11
2.4.1. Model with display........................................................................................................................................................... 11
2.4.2. Model with LED ................................................................................................................................................................ 11
2.4.3. Direction of ow............................................................................................................................................................... 12
2.5. Dimensions........................................................................................................................................... 13
2.6. Technical data....................................................................................................................................... 14
2.7. Measuring ranges ................................................................................................................................. 15
3. Installation................................................................................................................................................... 17
3.1. Warning................................................................................................................................................ 17
3.1.1. Requirements for piping................................................................................................................................................. 17
3.1.2. Requirements for infeed / outfeed section................................................................................................................ 17
3.1.3. Turning the housing......................................................................................................................................................... 18
3.2. Installation ........................................................................................................................................... 19
4. Electrical installation .................................................................................................................................... 20
4.1. Pin assignment of plug-type connectors ................................................................................................ 20
4.2. Connection options............................................................................................................................... 20
4.2.1. Bidirectional RS485 bus system.................................................................................................................................... 20
4.2.2. Current output 4 ... 20 mA, 3-wire .............................................................................................................................. 21
4.2.3. MBus.................................................................................................................................................................................... 21
4.2.4. Galvanically isolated pulse output............................................................................................................................... 22
4.3. Connection of METPOINT® BDL............................................................................................................. 23
4.3.1. Bidirectional RS485 bus system.................................................................................................................................... 23
4.3.2. Current output 4 ... 20 mA, 3-wire ............................................................................................................................... 24
4.3.3. Galvanically isolated pulse output............................................................................................................................... 24
4.4. Connection to METPOINT® BDL compact............................................................................................... 25
4.4.1. Bidirectional RS485 bus system.................................................................................................................................... 25
4.4.2. Current output 4 ... 20 mA, 3-wire ............................................................................................................................... 25
4.4.3. Galvanically isolated pulse output............................................................................................................................... 26
4.5. Modbus nal termination...................................................................................................................... 26

EN Installation and operating manual
METPOINT®SF53 3
5. Commissioning ............................................................................................................................................. 27
6. Conguration and operation ......................................................................................................................... 27
6.1. LED signals during operation................................................................................................................. 27
6.2. Setup menu .......................................................................................................................................... 28
6.3. Sensor Setup......................................................................................................................................... 28
6.3.1. Input interior pipe diameter.......................................................................................................................................... 29
6.3.2. Entering / changing consumption counter value ..................................................................................................... 29
6.3.3. Selecting units for consumption, ow, temperature and pressure...................................................................... 30
6.3.4. Entering reference conditions....................................................................................................................................... 31
6.3.5. Setting zero point for low-ow cut-o function....................................................................................................... 32
6.4. Modbus setup ....................................................................................................................................... 33
6.4.1. Modbus settings (2001 ... 2005)................................................................................................................................... 34
6.4.2. Values register (1001 …1500)........................................................................................................................................ 34
6.5. Pulse / alarm......................................................................................................................................... 35
6.5.1. Pulse output ...................................................................................................................................................................... 35
6.6. User Setup............................................................................................................................................ 36
6.7. Advanced ............................................................................................................................................. 36
6.8. 4 ... 20 mA ............................................................................................................................................ 37
6.9. Info ...................................................................................................................................................... 38
6.10. MBus .................................................................................................................................................. 38
6.10.1. Default communication settings................................................................................................................................ 38
6.10.2. Transferred values.......................................................................................................................................................... 38
7. Spare parts and accessories........................................................................................................................... 39
8. Maintenance and servicing............................................................................................................................ 39
9. Cleaning sensor head .................................................................................................................................... 39
10. Re-calibration............................................................................................................................................. 39
11. LED display................................................................................................................................................. 39
12. Declaration of Conformity .......................................................................................................................... 41

Installation and operating manual EN
4 METPOINT®SF53
1. Safety information
1.1. Pictograms and symbols
1.1.1. In this documentation
General instructions
Observe installation and operating instructions
General hazard symbol (danger, warning, caution)
General hazard symbol (danger, warning, caution) relating to mains voltage and powered machine parts
The packaging material is recyclable. Dispose of it according to the applicable statutory regulations.
1.1.2. On the device
General instructions
General hazard symbol (danger, warning, caution)
1.2. Signal words
DANGER Imminent danger
Consequences of non-compliance: serious or even fatal injury
WARNING Potential danger
Consequences of non-compliance: serious or even fatal injury
CAUTION Imminent danger
Consequences of non-compliance: injury and/or damage to property
NOTICE Additional notes, tips and hints
Consequences of non-compliance: Disadvantages in operation and maintenance.
No risk to persons

EN Installation and operating manual
METPOINT®SF53 5
1.3. Safety instructions
DANGER Escaping compressed gas
Risk of serious or even fatal injury from contact with escaping compressed gas or from unsecured plant
components.
• Before carrying out any assembly, installation or maintenance work, depressurize the system. All
electrical work must be carried out by authorized specialist technical personnel1.
• Use only pressure-resistant installation materials and suitable tools that are in proper working order.
• Before pressurizing the system, check all unit parts and repair them, if necessary. Open valves slowly
to prevent pressure blows during operation.
• Make sure that no persons can be injured or objects can be damaged by condensate or escaping
compressed gas.
• Protect the device parts against vibration and impact.
DANGER Mains voltage
Risk of serious or even fatal injury from electric shock when coming into contact with non-insulated,
powered components.
• For the electrical installation of the device, adhere to all applicable regulations (e.g. VDE 0100 / IEC
60364).
• Before carrying out any maintenance work, de-energize the system.
• All electrical work must be carried out by authorized specialist technical personnel1.
• The permissible operating voltage is printed on the name plate and must be strictly adhered to.
• All components of the electrical installation on site must be approved and/or bear the CE mark.
• A reliably accessible circuit breaker (e.g. power plug or switch) that shuts o all conductors must be
installed close to the unit.
WARNING Operation of unit outside limit range
If the specied limits are exceeded, there is a risk of device malfunction, potentially resulting in injury
and/or damage to property.
• The device must only be operated for the intended purpose and within the permissible limits
specied on the type plate and in the technical data.
• Use high-pressure protection above 10 bar to ensure safe installation and removal.
• Do not operate the device in connection with ammable gases.
• Strictly adhere to the prescribed operating times and maintenance intervals.
• Observe the prescribed storage and transport conditions.
• Prevent condensation on the sensor element. Ensure that the air fed through the device is free of
droplets.
1 Specialist technical personnel
Specialist technical personnel have the professional training, knowledge of measurement, control, regulation and compressed air
technology and experience and knowledge of country-specic regulations, applicable standards and directives necessary to allow
them to carry out the work described and independently recognize potential hazards. Specialized usage conditions will require
additional knowledge, for instance regarding aggressive media.

Installation and operating manual EN
6 METPOINT®SF53
1.4. Transport and storage
Despite our best eorts regarding packaging, etc., the device might be damaged during transport. Upon receipt, please remove all
packaging material and inspect the product for visible damage. If you detect such damage, immediately notify the carrier company
and BEKO TECHNOLOGIES GMBH or one of its agents.
CAUTION Damage caused during transport or storage
Incorrect transport or storage, or the use of unsuitable lifting equipment might cause damage to the
device.
• The device must only be transported and stored by authorized and suitably trained technical
personnel.
• If you detect any damage, do not start the device.
• Adhere to the permissible storage and transport temperatures (see technical data).
• Protect the device against direct sunlight and heat radiation.
The device must be stored in the original
packaging. Seal the packaging and store it in a
dry and frost-free room. Ensure that the storage
temperature does not exceed the limits specied
on the type plate.
Even when packaged, take suitable measures to
protect the device against the elements.
While in storage, secure the device so that it
cannot topple over or fall, and protect it against
vibration.
NOTICE Recycling of packaging material
• The packaging material is recyclable. Dispose of the packaging material according to the applicable
statutory regulations of the destination country.

EN Installation and operating manual
METPOINT®SF53 7
1.5. Intended use
The METPOINT® FLM is a thermal ow meter for the measurement of volume ow, consumption and ow velocity. By default, the
device is congured for the measurement of volume ow in m³/h, consumption in m³ and velocity in m/s.
• The METPOINT® FLM is primarily used in compressed air systems. On request, the sensor can be programmed by
BEKO TECHNOLOGIES GmbH for the measurement of other gases: Nitrogen
• The device is not suitable for operation in potentially explosive or aggressive atmospheres.
• Protect the device against direct sunlight and heat radiation.
Operate the METPOINT® FLM only for the intended purpose and within the limit range specied in the technical data. Do not
operate the unit with any media (uids, gas/vapor mixtures) other than those listed above. Any other use of the device is deemed
improper and poses a risk to persons, property and the environment.
1.6. Warranty and liability
Any liability claims shall expire if METPOINT®FLM is not used according to its intended use, or if it is operated outside of the
specifications indicated in the technical data; this includes, in particular:
• Incorrect installation, commissioning or operation; insucient maintenance
• Operation with defective components
• Non-compliance with the instructions in this document, in particular the safety instructions
• Execution of construction related interactions or modications of the device
• Non-compliance with the prescribed maintenance intervals
• Use of third-party spare parts that have not been approved by the manufacturer

Installation and operating manual EN
8 METPOINT®SF53
2. Product information
2.1. Scope of delivery
The table below shows the scope of delivery of the METPOINT®FLM.
Designation Picture
Calibration certicate
Connecting cable (5-wire)
Aligning aid
2.2. Type plate
The name plate is attached to the device housing. It contains all relevant technical data of the METPOINT® FLM. Please have these
details to hand when contacting the manufacturer or supplier:
Designation Description
METPOINT®FLM SF53 Type designation
S/N: 12579143 Serial number
P/N: 4036460 Product number
Gas: air Media
Supply: 18 ... 36 VDC Power supply rating
0.1 ... 53 scfm Min./max. measuring range
4 ... 20 mA Min./max. Data of analog output
length: 8.646 in Length of sensor tube
Pmax: 232 psi Max. permitted operating overpressure
NOTICE Handling of type plate
Do not remove or cover the name plate, and protect it against damage.
For more information regarding the symbols printed on the type plate, see “Pictograms and symbols”
on page 4.
METPOINT FLM SF53
S/N:
P/N:
Gas:
Supply:
www.beko-technologies.com
12579143
4036460
18 ... 36 VDC
Air
0.1 ... 53 SCFM
4 ... 20 mA
length: 8.646 in
Pmax: 232 psi
Made in Germany

EN Installation and operating manual
METPOINT®SF53 9
2.3. Product overview and description
2.3.1. Identication based on product code
FLM
Sensor length (L)
Model with display
Interface: RS485
Operating pressure: 232 psi
SensorFlow (model designation)
FlowMeasurement (product designation)
SF 53 DL220
L

Installation and operating manual EN
10 METPOINT®SF53
2.3.2. Product description
The METPOINT® FLM thermal ow meter measures the volume ow, which forms the basis for intelligent energy management.
It can be used to identify potential savings, overloads and weak points in a system to improve its eciency. By measuring the
actual ow to the various production units, operators are in a position to make decisions based on facts. At the same time,
the METPOINT® FLM lets them know whether there are any leaks in their system. The METPOINT® FLM thus provides all the
information operators need to correctly dimension and congure their system systems and components for improved eciency.
The device is equipped with a Modbus RTU(RS485) interface, a 4 ... 20 mA current output, as well as a galvanically insulated pulse
output and an optional MBus interface.
2.3.3. Operating principle
Two temperature sensors are installed in series in the direction of ow. The rst temperature sensor measures the current process
temperature, while the second sensor is electrically heated to a temperature that is exactly 40 K above the temperature measured
by the rst sensor. As the mass or volume ow increases, the temperature sensors would normally cool o, but the electric heater
of the second heater prevents that from occurring.
The electric energy required to maintain the temperature dierence is directly proportional to the mass ow. As the mass ow
increases, so dies the heating power, which is then converted into the appropriate measured value. Taking into account the inside
diameter of the pipe, the METPOINT® FLM determines the exact mass ow.

EN Installation and operating manual
METPOINT®SF53 11
2.4. Control and display elements
2.4.1. Model with display
NOTICE Additional information
For more information regarding the operation of the device, see “Conguration and operation” on page
27.
2.4.2. Model with LED
There is an LED on the top of the METPOINT® FLM housing, which is used to indicate the calibration time. The light ashes after
15 months to indicate there is an upcoming re-calibration. The ashing LED does not inuence measured values. The measurement
signal is still output.
The time interval can be adjusted as requested by the customer in the manufacturing plant.
ME TPOINT®F LM
FLOW MEASURING
HW:1.00 SW:1.00
AIR
395,38
160515
m3/h
m3
Plug-type connector A
Plug-type connector B
METPOINT®FLM
C
A
L
I
B
R
A
T
I
O
N

Installation and operating manual EN
12 METPOINT®SF53
2.4.3. Direction of ow
The direction of ow is indicated by the arrows (1) on the housing and on the probe tube of the METPOINT®FLM.
1
NOTICE Additional information
If necessary, turn the housing (e.g. to change the direction of ow through the device). For more
information, see “Turning the housing” on page 18.

EN Installation and operating manual
METPOINT®SF53 13
2.5. Dimensions
Dimensions
Model with display Model with LED
AG½" (ISO 228/1)
B (in) 16.34 (Standard) 16.48 (Standard)
C (in) 3.15
D (in) Ø 0.46
E (in) 8.66 (Standard), optional: 15.75
F (in) 3.70
G (in) 4.02 4.15
F
F
E
D
E
D
C
A
B
G
A
B
G
C
F
F
E
D
E
D
C
A
B
G
A
B
G
C

Installation and operating manual EN
14 METPOINT®SF53
2.6. Technical data
Technical data
SF53
Max. operating pressure 232 psi, optional 725 psi
Measuring technique Calorimetric
Operating temperature Sensor tube and ttings: -30 ... +140 °C (-22 ... +284 °F)
Housing: -30 ... +80 °C (-22 ... +176 °F)
Measured parameters m³/h (factory settings)
On the display version, the following units can be chosen:
m³/min, l/min, l/s, ft/min, cfm, m/s, kg/min, kg/s
Sensor Pt45, Pt1000
Media Compressed air, nitrogen
Humidity of medium max. 90 % rH (no droplets)
Power supply 18 ... 36 VDC
Power consumption max. 5 W
Digital output RS485 (Modbus RTU)
Analog output 4 ... 20 mA (max. load < 500 Ω)
Pulse output Floating switch contact
Passive: max. 48 VDC 150 mA
1 pulse per m³ or per liter
Unit adjustable at display
Accuracy ± 1.5 % of measured value
± 0.3 % of nal value
Display/indicator Display: TFT 1.8" (resolution: 220 x 167) or service LED
Screw tting G½ (ISO 228/1)
Material Sensor tube and ttings: 1.4301 stainless steel
Housing: Powder-coated aluminum
Flange: 1.4404 (DIN EN 1092-1)

EN Installation and operating manual
METPOINT®SF53 15
2.7. Measuring ranges
The METPOINT® FLM can measure ow velocities up to 185.0 m/s and is precongured for an inside pipe diameter of 53.1 mm. At
the analog output of 4 ... 20 mA, this corresponds to:
Rated
diameter Øinterior Volume ow (measuring range limit value in
scfm) V max.
Zoll mm in Luft * Luft ** N2** m/s ft/s
1/4" 6.0 0.24 5.53 5.12 5.12 185.0 606.95
10.0 0.39 17.54 16.13 16.13 185.0 606.95
15.0 0.59 45.73 42.02 42.02 185.0 606.95
1/2" 16.1 0.63 53.56 49.26 49.26 185.0 606.95
3/4" 21.7 0.85 104.65 96.23 96.23 185.0 606.95
1" 25.0 0.98 143.55 132.02 132.02 185.0 606.95
26.0 1.02 156.09 143.55 143.55 185.0 606.95
27.3 1.07 173.45 159.5 159.5 185.0 606.95
28.5 1.12 190.29 174.98 174.98 185.0 606.95
30.0 1.18 212.54 195.41 195.41 185.0 606.95
1 1/4" 32.8 1.29 257.03 236.37 236.37 185.0 606.95
36.0 1.42 312.83 287.64 287.64 185.0 606.95
36.3 1.43 318.48 292.88 292.88 185.0 606.95
1 1/2" 39.3 1.55 376.57 346.32 346.32 185.0 606.95
40.0 1.57 390.64 359.21 359.21 185.0 606.95
41.9 1.65 428.72 394.23 394.23 185.0 606.95
43.1 1.70 457.5 420.72 420.72 185.0 606.95
45.8 1.80 519.24 477.45 477.45 185.0 606.95
2" 50.0 1.97 623.42 573.33 573.33 185.0 606.95
51.2 2.02 654.56 601.88 601.88 185.0 606.95
53.1 2.09 704.88 648.2 648.2 185.0 606.95
54.5 2.15 743.43 683.69 683.69 185.0 606.95
57.5 2.26 877.92 807.35 807.35 185.0 606.95
60.0 2.36 908.82 835.78 835.78 185.0 606.95
64.2 2.53 1044.31 960.38 960.38 185.0 606.95
2 1/2" 65.0 2.56 1071.8 985.63 985.63 185.0 606.95
70.3 2.77 1258.32 1157.14 1157.14 185.0 606.95
71.1 2.80 1287.1 1183.63 1183.63 185.0 606.95
76.1 3.00 1478.04 1359.2 1359.2 185.0 606.95
* According to DIN 1945 / ISO 1217 (20°C, 1000 mbar), with compressed air.
** Adjustment for DIN 1343: 0°C, 1013.25 mbar

Installation and operating manual EN
16 METPOINT®SF53
Rated
diameter Øinterior Volume ow (measuring range limit value in
scfm) V max.
Zoll mm in Luft * Luft ** N2** m/s ft/s
3" 80.0 3.15 1637.36 1505.7 1505.7 185.0 606.95
82.5 3.25 1741.31 1601.28 1601.28 185.0 606.95
84.9 3.34 1844.07 1695.81 1695.81 185.0 606.95
90.0 3.54 2074.8 1907.99 1907.99 185.0 606.95
4" 100.0 3.94 2564.55 2358.37 2358.37 185.0 606.95
107.1 4.22 2945.18 2708.34 2708.34 185.0 606.95
110.0 4.33 3106.87 2857.02 2857.02 185.0 606.95
5" 125.0 4.92 4016.75 3693.74 3693.74 185.0 606.95
133.7 5.26 4595.32 4225.81 4225.81 185.0 606.95
6" 150.0 5.91 5791.02 5325.39 5325.39 185.0 606.95
159.3 6.27 6531.39 6006.2 6006.2 185.0 606.95
182.5 7.19 8582.58 7892.47 7892.47 185.0 606.95
190.0 7.48 9302.53 8554.51 8554.51 185.0 606.95
8" 200.0 7.87 10319.83 9489.99 9489.99 185.0 606.95
206.5 8.12 11001.52 10116.89 10116.89 185.0 606.95
10" 250.0 9.84 16143.98 14845.82 14845.82 185.0 606.95
260.4 10.25 17535.97 16125.91 16125.91 185.0 606.95
12" 300.0 11.81 23275.01 21403.45 21403.45 185.0 606.95
309.7 12.19 24804.43 22809.92 22809.92 185.0 606.95
339.6 13.37 29825.18 27426.96 27426.96 185.0 606.95
388.8 15.31 41377.78 38050.61 38050.61 185.0 606.95
500.0 19.69 64652.8 59454.12 59454.12 185.0 606.95
600.0 23.62 93100 85613.94 85613.94 185.0 606.95
700.0 27.56 126719.44 116530.04 116530.04 185.0 606.95
800.0 31.50 165511.13 152202.51 152202.51 185.0 606.95
900.0 35.43 209475.07 192631.32 192631.32 185.0 606.95
1000.0 39.37 258611.13 237816.44 237816.44 185.0 606.95
* According to DIN 1945 / ISO 1217 (20°C, 1000 mbar), with compressed air.
** Adjustment for DIN 1343: 0°C, 1013.25 mbar

EN Installation and operating manual
METPOINT®SF53 17
3. Installation
3.1. Warning
DANGER Escaping compressed gas
Risk of serious or even fatal injury from contact with escaping compressed gas or from unsecured plant
components.
• Before carrying out any assembly, installation or maintenance work, depressurize the system. Such
work must be carried out by authorized specialist technical personnel only (see “Safety instructions”
on page 5).
• Use high-pressure protection above 145 psi to ensure safe installation and removal.
• Tighten the clamping sleeve with a torque of 14.75 ... 22.13 ft lb (wrench size 27).
• Use only pressure-resistant installation materials and suitable tools that are in proper working order.
• Before pressurizing the system, check all unit parts and repair them, if necessary. Open valves slowly
to prevent pressure blows during operation.
3.1.1. Requirements for piping
• Correctly dimensioned gaskets.
• Correctly installed and aligned anges and gaskets.
• Dierences in pipe diameters at joints should not exceed 1 mm. For more information, see ISO 14511.
• Clean, properly installed pipes.
3.1.2. Requirements for infeed / outfeed section
The following table shows the required infeed sections, depending on the direction of ow.
Table of additional inlet pipe sections required in specic cases
Flow obstruction upstream of measuring
section
Minimum length of inlet pipe section
(L1)
Minimum length of outlet pipe
section (L-L1)
Slight bend
(angle < 90°) 12 x D 5 x D
Reduction
(pipe diameter becomes smaller towards
measuring section)
15 x D 5 x D
Enlargement
(pipe diameter becomes larger towards
measuring section)
15 x D 5 x D
90° elbow section
or T-piece 15 x D 5 x D
2x 90° elbow sections
in a plane 20 x D 5 x D
2x 90° elbow sections
change of direction in 3 dimensions 35 x D 5 x D
Shut-o valve 45 x D 5 x D
5 x D
12 x D 5 x D
15 x D 5 x D
15 x D
5 x D
15 x D
5 x D
20 x D 5 x D
35 x D
5 x D
45 x D
NOTICE Deviating measurements
The above values are required minimum values. If the
recommended settling sections cannot be implemented, the
measuring deviations might be much greater.

Installation and operating manual EN
18 METPOINT®SF53
3.1.3. Turning the housing
If the direction of ow changes, the housing can be turned by loosening 4 grub screws with 0.059 in hexagon socket to the desired
position. Then tighten the grub screws again by hand.
NOTICE Risk of damage to device
• Ensure that the connecting pipes are correctly mounted and that the gaskets are installed properly.
• Then align the sensor in the direction of ow using the alignment aid.

EN Installation and operating manual
METPOINT®SF53 19
3.2. Installation
The sensor is installed by means of a ½, DN 15 ball valve (min. Ø 0.591 in).
Install the straight tting with the O-ring (G½ thread, wrench size 32) in the connecting nozzle. Ensure that the assembly is
pressure-tight.
Place the sensor head at the center of the tube and aligned it to the direction of ow. To do this, the sensor tube is equipped with a
depth scale, arrows indicating the direction of ow and an aligning aid. After the sensor is correctly positioned and aligned, tighten
the slotted spring pin with a torque of 14.75 ... 22.13 ft lb (wrench size 27).
When tightening the straight tting and the slotted spring pin, take care not to change the alignment of the sensor. Otherwise,
you might need to correct the sensor position and alignment.
The angle of the sensor should not deviate by more than ± 2° from the ideal position. Larger angles might result in inaccurate
measurement.
The high-pressure protection device (Art. No. 4025892) must be used above a working overpressure of >145 psi.
This allows for assembly under pressure and allows the sensor to be attached securely to the measurement point.
NOTICE Additional information
Further information on mounting the high-pressure protection device is available from the enclosed
installation and operating manual.
y
160
170
180
x
dA Circlip
Engraved depth scale
for precise installation
Installation depth = x + y
dA = exterior diameter
x = dA
2

Installation and operating manual EN
20 METPOINT®SF53
4. Electrical installation
4.1. Pin assignment of plug-type connectors
Pin assignment of plug-type connector A, M12 x 1, 5-pin, A-coded (according to EN 61076-2-101)
Pin assignment of connector
Transmitter side
Pin assignment of connector
Socket side
Pin assignment of connector
Screw side
Pin assignment of plug-type connector B, M12 x 1, 5-pin, A-coded (according to EN 61076-2-101)
Pin assignment of connector
Transmitter side
Pin assignment of connector
Socket side
Pin assignment of connector
Screw side
4.2. Connection options
4.2.1. Bidirectional RS485 bus system
Connection by means of plug-type connector A.
Connection diagram for METPOINT®SF53 and METPOINT®BDL
q
METPOINT®FLM SFxx
1
2
4
3
RS485
A
B
V
U
+
V
U
+
V
U
-
V
-U
RS485
MODBUS
+
-
V
Pin assignment - sensor Function Wire color
PIN-1 + UVPlus (+) connection, power supply brown
PIN-2 Bus A (+) Non-inverted signal (+) from RS485 interface white
PIN-3 - UVMinus (-) connection, power supply blue
PIN-4 Bus B (-) Inverted signal (-) from RS485 interface black
Other manuals for FLM SF53
1
This manual suits for next models
4
Table of contents
Other Beko Measuring Instrument manuals

Beko
Beko FLM SF53 User manual

Beko
Beko METPOINT BDL compact User manual

Beko
Beko METPOINT User manual

Beko
Beko DP 109 User guide

Beko
Beko METPOINT FLM SF13 User manual

Beko
Beko METPOINT BDL portable User guide

Beko
Beko BEKOMAT 12 KW User guide

Beko
Beko METPOINT OCV User guide

Beko
Beko METPOINT OCV User manual
Popular Measuring Instrument manuals by other brands

Tektronix
Tektronix IsoVu TIVP02 Quick reference guide

Burkert
Burkert 8077 operating instructions

ATORN
ATORN DFT Series Operator's manual

BRONKHORST
BRONKHORST ES-FLOW ES Series instruction manual

Bushnell
Bushnell YARDAGE PRO SPORT 450 owner's manual

Extech Instruments
Extech Instruments SDL800-NIST user guide