Operation Quick Start
15-3. Click on “Result Name” and enter the name of
result files.
14. Cartridge Calibration:
New cartridge need to be calibrated before use.
Please follow the steps below to proceed.
14-1. Click “HV Check”.
14-2. Click “Calibrate”.
* The storage and transportation
condition may influence the Gel-
matrix and cause unstable current.
During HV Check, check current
(gray line) and see if it’s stable. if
current is unstable, please repeat
this step 2-3 times.
* For troubleshooting, please refer to cartridge unpacking
guide for details.
Bio-Fragment Analyzer: Qsep1-Plus Document No.: F0081
Ver.: ENG-A
Page: 3/3
*The “Calibrate” button will
only display after HV check
passed.
15. Click on the blank column and designate ①the sample
locations, ②test method, sample duration, runs, ③result
name and ④Para by following steps 15-1, 15-2, 15-3 and 15-4.
15-1. Click the “Sample Position”, mark the position of sample on
the plate and then click “OK”.
15-2. Click “Method” to select analysis method.
15-4. Click the icon “Para” and set the parameters.
(Baseline Factor, Peak Threshold, Calculate, etc.).
16. Click “Run” to start the process.
Brief introduction of the Signal Chart:
Fluorescence Intensity (red line)
Current Unit
Related Fluorescence Unit
Current (gray line)
Unstable
During the “HV Check”, the
last LED will show red light
Blue → SD card
Green → WiFi connectable
Red → HV check
Blue
Green
Red
Please check the Alignment Marker has been placed in
correct position.
Software will recognize peaks from Alignment Marker
signal. DO NOT use Size Marker or DNA sample doing
“Calibrate”.
* If you want to analyze the samples with Alignment Marker, please
choose the appropriate Alignment Marker and then place it in the
corresponding position. (Check ✓the box)