
-5 -
Note: The meter connection size and the bore size are different. For example, many of the 1” NPT
turbines have bore sizes that range from 3/8” up 1”. Be sure to use the correct bore size or the meter
could report incorrect flows and totals.
Select the Display Function– The monitor can display
RATE or TOTAL or alternate between BOTH rate and
total. At the DISPLAY prompt, press the ENTER key
once. The monitor now shows the display mode
currently in effect. If the current selection is correct,
press the ENTER key to advance to the next parameter.
To change to an alternate display mode, use the arrow
keys to scroll to the desired display mode and press
ENTER to save the choice.
Select the Rate Units of Measure – The monitor allows
the choice of many common rate units. (See the
specifications for a complete listing of the unit choices.)
At the RATE UNT prompt, press the ENTER key once.
The monitor now shows the rate units of measure the
display is currently set for. If the current selection is correct, press the ENTER key to advance
to the next parameter. To change to an alternate unit, use the arrow keys to scroll to the
desired rate unit and press ENTER to save the choice.
Select the Rate [Time] Interval– The term Rate implies that something is occurring over a
period of time. Most people are familiar with the rate of speed of a car reported in miles per
hour (MPH). The same concept holds true for a flow meter. The time choices are SEC
(seconds), MIN (minutes), HOUR (hours), and DAY (days). At the RATE INT prompt, press
the ENTER key once. The monitor now shows the time interval the display is currently set
for. If the current selection is correct, press the ENTER key to advance to the next parameter.
To change to an alternate time interval, use the arrow keys to scroll to the desired time
interval and press ENTER to save the choice.
Note: If flow rate is the only measurement of interest, skip top KFAC UNT to complete the
programming process.
Select the Total Units of Measure – If a flow amount is desirable, the units for the total must
first be chosen. The monitor allows the choice of many common totalization units. (See the
specifications for a complete listing.) At the TOTL UNT prompt, press the ENTER key once.
The monitor now shows the total units of measure the display is currently set for. If the
current selection is correct, press the ENTER key to advance to the next parameter. To change
to an alternate unit, use the arrow keys to scroll to the desired totalization unit and press
ENTER to save the choice. Note: This unit of measure does not have to reflect the rate unit
you have previously chosen. (Example: Rate Units = Gallons, Total Units = Barrels).