Programming
DSY-PM-00022-EN-07Page 12 August 2015
Select Meter’s K-Factor Unit*
At the KFacUnt prompt, press ENTER. The display shows the current K-factor unit. If the current selection is correct, press
ENTER to advance to the next parameter. To change the K-factor unit, press UP or RIGHT to scroll to the correct unit, the units
should match the units that the meter was calibrated in. Press ENTER to save and advance to the KFactor parameter.
Enters Meter’s K-Factor*
OTE:N The K-factor supplied with your meter, or calculated from calibration data, is needed to complete this step.
At the KFactor prompt, press ENTER. The most significant digit in the K-factor flashes. If the current K-factor is correct, press
ENTER to advance to the next parameter. To change the K-factor, press UP to increment the digit until it matches the meter’s
first K-factor digit. Press RIGHT to advance to the next digit. Repeat this process until all K-factor digits have been entered.
Press ENTER to save the K-factor and advance to the RateInt parameter.
OTE:N The number of digits available before and after the decimal point is determined by the bore size of the flow sensor
being used. The largest K-factors are associated with the smallest bore sizes. The maximum allowable K-factor is
99999.9. The minimum must be at least 1.000. If an out of range number is entered, the display flashes Limit and
refuses the entry.
Select Rate Interval*
At the RateInt prompt, press ENTER. The monitor flashes the current time interval. If the current selection is correct, press
ENTER to advance to the next parameter. To change to an alternate time interval, press UP or RIGHT to scroll to the correct
time interval. Press ENTER to save and advance to the RateUnt parameter.
Select Flow Rate Units*
At the RateUnt prompt, press ENTER. The monitor flashes the current rate unit. If the current selection is correct, press ENTER
to advance to the next parameter. To change to an alternate unit, press UP or RIGHT to scroll to the correct rate unit and press
ENTER to save and advance to the TotlUnt parameter.
Select Total Units of Measure*
At the TotlUnt prompt, press ENTER. The monitor flashes the current total units. If the current selection is correct, press ENTER
to advance to the next parameter. To change to an alternate unit, press UP or RIGHT to scroll to the correct totalization unit.
Press ENTER to save and advance to the TotlMul parameter.
Select Total Multiplier*
This parameter displays the accumulated flow total in multiples of 10. For example, if the optimum totalization unit is 1000
gallons, the unit total display increments by one digit for every 1000 gallons monitored. In Run mode, at 1000 gallons the total
monitor reads 1, at 3000 gallons, the total display reads 3. This feature eliminates having to look at a total, count the digits,
and mentally insert commas for each 1000 multiple.
At the TotlMul prompt, press ENTER . The monitor shows the current total multiplier. If the selection is correct, press ENTER to
advance to the next parameter. To change to an alternate multiplier, press UP or RIGHT to scroll to the correct multiplier unit
and press ENTER to and advance to the next parameter.
OTE:N If the RateUnt or TotlUnt parameter has been set to pounds or kilograms, the monitor advances to the Spec Gr
parameter. At any other setting, the monitor advances to PulsOut in Programming mode. See Totalizer Pulse Output*
on page 14.
Enter Specific Gravity Value*
Mass readings in the B3000 monitor are not temperature or pressure compensated so it is best to enter the specific gravity
of the fluid as close to the system running temperature as possible. As liquids are essentially incompressible, pressure
compensation is not necessary for liquids.
At the Spec Gr prompt, press ENTER. The most significant digit of the current specific gravity flashes. If the current specific
gravity is correct, press ENTER to advance to the next parameter. To change to an alternate specific gravity, press UP to
increment the flashing digit until you reach the first digit of the new specific gravity. Press RIGHT to move to the next digit.
When all digits have been entered, press ENTER to save and advance to the next parameter.
OTE:N If Gas was chosen as the fluid, see Gas on page 19 and follow the directions for the gas parameters.
In Programming mode, the monitor advances to the PulsOut parameter, see Totalizer Pulse Output* on page 14.