BMS D1 Series User manual

BMS D1 series Biological
Microscope
INSRUCTIONS
This manual expatiates the using method, troubleshooting, and maintenance about
the microscope particularly. Please read it before using carefully. The manufacturer
is entitled to amend or improve the instrument.
The quality for this microscope is according to GB/T2985-1999 biological
microscope standard.

Contents D1 series
Before use
1. Nomenclature ……………………………………………………………………..……….1
2. Assemblage ……………………………………………………………………………….3
2-1. Assemble scheme …………………………………………………………………….....3
2-2. Assemble steps ……………………………………………………………………….....4
3. Operation………………………………………………………………………… ….……..6
3-1. Illumination …………………………………………………………………………..6
3-2. Set slides ………………………………………………………………………………..6
3-3. Focusing ………………………………………………………………………………..6
3-4. Adjust tightness of focusing………………………………………………………….....7
3-5. Adjust diopter …………………………………………………………………………..7
3-6. Adjust interpupillary distance ………… …………………………… …………………8
3-7. Adjust center of condenser ………………… ……………………………………….…8
3-8. View field diaphragm … ………………………………………………… ……………9
3-9. Aperture diaphragm………… ………………………………… ………………………9
3-10. Use objective with Oil …………………… ……………………… …………………10
3-11. Use filter ………………… ……………………………………………… ………….10
3-12. Replace fuse………………………………… ……………… ………………………10
3-13. Select light path ……………………………………………………………..……......11
3-14. Set video device ………… …………………………………………………..…….…11
3-15. Set photo adapter ………… …… …… ………………………………………..….…11
4. General specification …………………………………………………………….….… ..12
4-1. D1 series biological microscope specification …………………………………….….12
4-2. Objective parameters ……………………………………………………………….…12
5. Troubleshooting ………………………………………………………………….….… ..13

Figure 1
Before Use D1 series
1. Notice
⑴As the microscope is a precision instrument, a always
operate it with care, avoiding physical shake during the
operation.
⑵Do not let the microscope emerge in the sun directly, either
not in the high temperature, damp, dusty or acute shake
place. Make sure the worktable is horizontal.
⑶When moving the microscope, you should use both hands to
hold ①and ②as shown on figure, and lay it down
carefully. (Figure 1)
★If hold the focusing knob, eyepiece tube or stage
during moving, you will damage it.
⑷During working, the light condenser will become hot.
Please make sure enough space around ③for
cooling.(Figure 2)
⑸Connect the microscope to earth avoid lightning
strike.
⑹Make sure the power switch ④is on “0” (off) before
replacing the bulb or fuse and wait until the lamp
cools down. (Figure 3
★Bulb: 6V/20W halogen bulb(Philips 7388)
⑺It can adapt wide voltage 90~240V. There is no need
for transformer. Make sure the voltage is in this
range.
⑻Use the special plug supplied by our company.
Figure 2
Figure 3

D1 series
2.Maintenance
⑴Wipe the lens gently with a soft lens tissue. Carefully wipe off oil or fingerprints on the lens surfaces with
tissue moistened with a little of 3:7 mixture of alcohol and ether or dimethy lbenzene.
★Alcohol and ether is flammable. Don’t place these chemical near to fire or fire source. For example,
when turning on or turning off the electric device, you should use these chemical in a ventilated place.
⑵Don’t use organic solution to wipe the surfaces of the other components. Please use the neutral detergent.
⑶If the microscope damped by the liquid, you should cut off the power immediately and wipe it dry.
⑷Never disassemble or service the microscope yourself. It will influence its function or damage it.
⑸After using, cover the microscope with a dust cover.
3. Safety Sign
Sign Signification
Read manual before using. Faulty operation would lead to person hurt or
instrument trouble.
| Power swatch ON
O Power swatch OFF

1.Nomenclature D1 series
BMS D1 binocular
1
Eyepiece
Nosepiece
Objective
Stage
Condenser for Normal
illumination
Light condenser for
Normal illumination
Focusing arm
Fine adjusting
knob
Coarse
adjusting knob
Arm
Sliding binocular head

D1 series
BMS D1 trinocular
2
Tightness
adjusting knob
Gemel trinocular head
Light condenser for
Kohler illumination
Condenser for Kohler
illumination
Power switch
Lock-screw
Light adjusting
knob

2. Assemblage D1 series
2-1 Assemble scheme:
The below numbers denote the assembling older.
★Before assembling, make sure there is no dust or dirt. Assemble carefully and do not scrap any part or
touch the glass face.
3
Eyepiece
Objective
Condense
r
Plug
halogen bulb

D1 series
2-2 Assemble Steps 2-2-1 Set Condenser (Figure 4/5)
Set condenser for normal illumination
1.Turn the coarse knob with the direction showed to
make the stage to highest position. (Figure 4)
2. Turn the condenser knob with the direction showed
to make the condenser holder down to suitable
position.
3. Loose the lock screw for condenser thoroughly.
4. Slide the condenser into holder with the direction
showed till it is tied with the holder. And turn the
condenser to make the knob direct to front.
5. Tight the lock screw and use knob moving the
condenser up to highest position.
Set condenser for Kohler illumination
1.Turn the coarse knob with the direction showed to
make the stage to highest position. (Figure 5)
2. Turn the condenser knob with the direction showed
to make the condenser holder down to suitable
position.
3. Loose the lock screw for condenser thoroughly.
4. Put the screw behind condenser matching with the
slot on condenser holder and slide it into
thoroughly.
5. Tight the lock screw and use knob moving the
condenser up to highest position.
2-2-2 Assemble objectives (Figure 6)
1. Turn the coarse knob ①with the direction showed
to make the stage to suitable position. (Figure 6)
2. Assemble the objectives from low to high
magnification clockwise.
Please use objective with low magnification
(such as 4X or 10X) to find the image.
Please make sure correct position with a sound
during turning the nosepiece.
4
图3
图
14
13
5
图
4
图
4
图
4
Figure 5
Figure 4
Figure 6

D1 series
2-2-3 Set or replace bulb (Halogen 7)
◎Please use assigned Philips 7388 halogen bulb 6V20W.
In order to use the bulb longer or prevent blowout,
please do not touch it barehanded. Please swap it
with soft cloth if you leave fingerprint on the bulb.
1. Loose the lock screw ①and turn the bulb holder ②
clockwise.
2. Use gauze or others to take the bulb and set it into bulb
holder ④.
Please set the bulb softly, over press will damage it.
Use or replace bulb
During operation or just after using, the bulb and bulb
holder will be very hot. Please turn the switch to “0”
and take off the plug. After the bulb and bulb holder
cool-down, you can replace the bulb.
This microscope is also suitable for LED
illumination, the operation for LED is similar to
halogen bulb.
2-2-4 Set eyepiece (Figure 8)
1. Take off the cap on eyepiece cap ①.
2. Slide the eyepiece ②into eyepiece tube throughly.
2-2-5 Connect plug (Figure 9)
Do not press the plug too much. Be aware that
bend or enlacement will damage it.
1. Before connect the plug, please turn the switch to
“OFF”.
2. Slide the plug ①into the socket ②on microscope
and make sure good contact. (Figure 9)
3. Connect the other side of the plug to power supply
and make sure good contact.
Please always use the plug supplied by manufacturer. If
damage or lost, please use same specification one.
★It adopt wildly power supply 90-240V.
★Please connect the power supply correctly, be sure the
instruments to be connected with earth.
Figure 8
Figure 9
Figure 7

5
3. Operation D1 series
3-1 illumination (Figure 10)
1. Set the power switch ①at “ON” after connect power
suppler.
2. Use knob ②to adjust the brightness more comfortable.
Clockwise for brightness and counter-clockwise for darkness.
(Figure 10)
Under low power, the bulb will have long life.
3-2 Set slides (Figure 11)
1. Push the clamp ①.
2. Put the slides ②on specimen holder and make sure the
frontispiece up towards. Then loose clamp ①to tight the
slides. (Figure 11)
3. Adjust the knob to move the slides to center (according
to center of objective).
3-3 Focusing (Figure 12、14、15)
1. Let objective 4X move into light path.
2. Turn one dioptor ring to make it in line with
reticle on eyepiece tube. At this time, the
position is “0”. (Figure 14)
3. Turn coarse knob to find the image. (Figure 12)
4. Adjust anther dioptor to make the image clear.
(Figure 14)
5. Turn fine knob to make the image clear. (Figure
12) (Suitable for sliding head)
Or
1. Let objective 4X move into light path.
2. Turn one dioptor ring to make “0” position in
line with reticle on eyepiece tube. (Figure 15)
3. Turn coarse knob to find the image. (Figure 12)
4. Adjust anther eyepiece to make the image clear. If
not, please adjust dioptor. (Figure 15)
5. Turn fine knob to make the image clear. (Figure
12) (Suitable for gemel head)
Figure 11
Figure 12
Figure 10
Table of contents
Other BMS Microscope manuals
Popular Microscope manuals by other brands

VWR
VWR VisiScope 384 Series instruction manual

Nikon
Nikon ECLIPSE E200 POL instructions

Leica
Leica DI C800 User's manual & installation instructions

ThermoFisher Scientific
ThermoFisher Scientific Continuµm manual

ThermoFisher Scientific
ThermoFisher Scientific Continuµm manual

Olympus
Olympus SZ61 instructions




















