
Page 18
13. Troubleshooting
Review the information in the table below to troubleshoot operating problems.
PROBLEM CAUSE SOLUTION
I. Optical Section:
LED operates, but eld of view remains
dark. Power supply is unplugged. Connect
Brightness is too low Set brightness to a proper level
Bertrand lens is in Remove Bertrand lens from light path
You are in a position of extinction Disengage analyzer from the light path
Field of view is obscured or not evenly
illuminated Nosepiece is not correctly engaged. Make sure that the nosepiece clicks
properly into place.
Tint plate or Bertrand lens are in a in-
termediate position. Move to a click stop
Dirt or dust is visible in the field of view. Dirt/dust on the specimen Clean the specimen
Dirt/dust on the eyepieces Clean the eyepieces
Image looks double Aperture iris diaphragm is stopped
down too far. Open aperture iris diaphragm.
The condenser is not well centered or
it is in a wrong height Set the condenser until a proper image
is achieved
Visibility is poor.
• Image is not sharp.
• Contrast is poor
• Details are indistinct
• Image glares
Nosepiece is in an incorrect position Move the nosepiece to a click stop
Aperture iris diaphragm is too closed
or too open. Adjust aperture iris diaphragm.
Dust or dirt on lenses (condenser,
objectives, eyepieces and slide) Clean thoroughly.
For transmitted light observation, the
coverglass thickness must not exceed
0.17mm
Use a coverglass with thickness
0.17mm
Focus is not even Slide holder is not at. Move the speci-
men to a at position.
One side of the image is unfocused Nosepiece is in an incorrect position Move the nosepiece to a click stop
Slide is mounted not in a at position
(tilted) Place the specimen in a at position on
the stage
Poor quality of the glass slide Use a glass slide with higher quality
Conoscopic image cannot be seen Bertrand lens is not in the light path. Insert the lens in the light path
Total extinction cannot be obtained Analyzer is not in the light path Insert analyzer in the light path.
II. Mechanical Section:
Coarse focus knob is hard to turn Tension adjustment ring is too tight Loosen tension adjustment ring
Focus is unstable Tension adjustment ring is too loose Tighten tension adjustment ring
III. Electrical Section
LED doesn’t turn on. Power supply not connected Check for proper connection
Brightness is not enough Brightness setting is too low Adjust brightness
Light blinks Power supply not well connected Check for proper connection
IV. Observation tube
Field of view of one eye does not match
that of the other. Interpupillary distance is incorrect. Adjust interpupillary distance.
Incorrect diopter adjustment. Adjust diopter.
Your view is not accustomed to micro-
scope observation. Upon looking into eyepieces, try look-
ing at overall field before concentrating
on specimen range. You may also find
it helpful to look up and into distance
for a moment before looking back into
microscope.