
Percutaneous Leads Directions for Use
Percutaneous Leads Directions for Use
92469115-02 8 of 397
WARNING: Do not exchange the Lead stylet while the electrode array of the Lead is in the bevel of the insertion needle. If the electrode array
is in the bevel area, remove the Lead from the insertion needle before exchanging the stylet. Inserting the Lead stylet in the Lead while the
electrode array is in the bevel of the insertion needle increases the risk of Lead and tissue damage.
WARNING: If the Lead stylet is removed and reinserted, do not use excessive force when inserting the stylet into the Lead. The use of
instruments, such as forceps, to grasp the stylet during insertion is not recommended as this could result in applying excessive force and could
increase the risk of Lead and tissue damage.
10. Advance the Lead to the appropriate vertebral level under uoroscopic guidance. A sufcient length of Lead (at least 10 cm, or approximately
three vertebrae) should reside in the epidural space to aid in Lead stabilization.
11. Proceed to the instructions for connecting to the OR Cable assembly in the appropriate DFU for your SCS System, as listed in your Reference
Guide.
Peripheral Nerve Stimulation (PNS)
When using the Boston Scientic Systems for PNS, the pre-loaded, curved Lead stylet is not recommended. Replace the curved Lead stylet with a
straight Lead stylet before starting the PNS procedure. Carefully pull out the curved Lead stylet and insert the preferred straight Lead stylet. If using
the Innion CX Lead, insert the straight stylet into the Lead tail that has one marker band. If resistance is encountered while inserting the Lead stylet,
withdraw the Lead stylet approximately 3 cm, rotate the Lead and/or Lead stylet and gently advance the Lead stylet. If resistance is still encountered,
repeat this step until the straight Lead stylet can be fully inserted.
WARNING: Do not exchange or insert the Lead stylet while the electrode array of the Lead is in the bevel of the insertion needle. Inserting
the Lead stylet in the Lead while the electrode array is in the bevel of the insertion needle increases the risk of Lead and tissue damage. If the
electrode array is in the bevel area, remove the Lead from the insertion needle before exchanging the stylet.
WARNING: Do not use excessive force when inserting the Lead stylet into the Lead. The use of instruments, such as forceps, to grasp the stylet
during insertion could result in applying excessive force and increase the risk of Lead and tissue damage.
CAUTION: Proceed with care, because the proximal Lead tail is susceptible to contamination and could cause infection.
Note: If using an Innion CX Lead, see “Innion CX Lead Placement with Entrada™ Needle for PNS” on page 8.
Lead Placement for PNS
1. Position, prep, and drape the patient in the usual accepted manner. Inject a local anesthetic at the needle insertion site. Be careful to avoid
anesthetizing the target nerve(s) (minimize lateral injections, minimize injection volume).
2. With the needle stylet inserted, insert the needle provided in the Lead Kit into the subcutaneous tissue plane and advance the needle to the target
location. The needle can be advanced slightly beyond the target location to allow for more proximal repositioning during intraoperative testing.
Pinching supercial tissue away from the fascial layer helps guide the needle within the subcutaneous fat.
CAUTION: Use only an insertion needle provided by Boston Scientic. Other needles may damage the Lead. The stamped number “14” on the
needle hub, or the triangle on the hub of the curved Epimed Needle, corresponds to the orientation of the bevel.
Note: To avoid engaging the dermis, place the bevel of the needle face down from the dermal interface and pinch the skin.
CAUTION: If electrodes are too close to (or through) the fascia, the patient may experience painful direct muscle stimulation. If electrodes are too
supercial, the risk of electrode erosion is increased.
3. Remove the needle stylet from the insertion needle.
4. Slowly introduce the Lead, with straight Lead stylet, through the insertion needle to the tip of the needle.
5. Carefully withdraw the insertion needle to expose the Lead contacts by slowly pulling the needle towards the proximal end of the Lead while
keeping the Lead in the desired location.
WARNING: Only advance the needle with the needle stylet in place. Failure to do so may result in tissue coring.
CAUTION: If the Lead is exposed beyond the tip of the needle, do not advance the needle or withdraw the Lead back through the needle.
Advancing the needle after the Lead is exposed beyond the tip of the needle may cause damage to the Lead or inadequate stimulation.
CAUTION: Do not completely remove the needle. Only remove the needle enough to expose the contacts.
CAUTION: Do not withdraw the needle past the anchoring site prior to creating the anchor site incision.
6. If performing a temporary Lead Trial procedure, connect to the OR Cable assembly and External Trial Stimulator. Follow the instructions in the
appropriate DFU for your Boston Scientic System as listed in your Reference Guide.
Innion CX Lead Placement with Entrada™ Needle for PNS
1. Position, prep, and drape the patient in the usual accepted manner. Inject a local anesthetic at the needle insertion site. Be careful to avoid
anesthetizing the target nerve(s) (minimize lateral injections, minimize injection volume).
Note: An anchor must be pre-loaded on the distal end of the Innion CX Lead prior to inserting the Lead into the Entrada Needle.
2. Verify that the Entrada Needle is fully assembled by holding the sheath hub and applying forward pressure on the stylet cap cover.