Cameron NUFLO MC-III User manual

NUFLOTM
MC-III™ Panel Mount
Flow Analyzer
User Manual
Manual No. 2350326-01, Rev. 02

Important Safety Information
Symbols Used in this Manual
This symbol identies information about practices or circumstances that can lead to personal
injury or death, property damage, or economic loss.
Terms Used in this Manual
Caution: Indicates actions or procedures which if not performed correctly may lead to personal
injury or incorrect function of the instrument or connected equipment.
Important: Indicates actions or procedures which may affect instrument operation or may lead to an
instrument response which is not planned.
Symbols Marked on Equipment
Attention! Refer to manual Protective (earth) ground
Technical Support Contact Information
Cameron
Measurement Systems Division
14450 John F. Kennedy Blvd.
Houston, TX 77032
Phone: 1-800-654-3760; 281-582-9500
Fax: 281-582-9599
NuFlo and MC-III are trademarks of Cameron International Corporation (“Cameron”).
Modbus is a registered trademark of the Modbus Organization, Inc.
Windows is a registered trademark of Microsoft Corporation.
Acrobat Reader is a registered trademark of Adobe Systems Incorporated.
© 2013 Cameron International Corporation (“Cameron”). All information contained in this publication is
condential and proprietary property of Cameron. Any reproduction or use of these instructions, drawings, or
photographs without the express written permission of an ofcer of Cameron is forbidden.
All Rights Reserved.
Printed in the United States of America.
Manual No. 2350326-01, Rev. 02
September 2013

iii
MC-III™ Panel Mount Flow Analyzer Table of Contents
Contents
Important Safety Information ...............................................................................................................................ii
Section 1—Introduction................................................................................................................................... 7
Operation ............................................................................................................................................................ 7
Table 1.1—MC-III Panel Mount Specications ............................................................................................. 8
LCD Display ............................................................................................................................................... 10
Keypad ....................................................................................................................................................... 10
Interface Software .......................................................................................................................................11
Power Supply ............................................................................................................................................. 12
Multipoint Linearization .............................................................................................................................. 12
Gas Volume Correction .............................................................................................................................. 12
Input Options .............................................................................................................................................. 12
Output Options ........................................................................................................................................... 12
Flow Log Archival ...................................................................................................................................... 12
Event Log Archival .................................................................................................................................... 13
Password-Protected Security ..................................................................................................................... 13
Commonly Used Functions .............................................................................................................................. 13
Reading Totals ........................................................................................................................................... 13
Saving Totals to Memory ............................................................................................................................ 13
Resetting the Total ..................................................................................................................................... 13
Viewing Daily and Hourly Logs .................................................................................................................. 14
Saving and Uploading Conguration Files ................................................................................................. 14
Exporting Log Data .................................................................................................................................... 14
Saving Log Data in a Report ...................................................................................................................... 14
Section 2—Installation................................................................................................................................... 15
Flow Meter/Magnetic Pickup Installation .......................................................................................................... 15
MC-III Installation.............................................................................................................................................. 15
Field Wiring Connections ................................................................................................................................. 17
Power Supply Wiring ........................................................................................................................................ 18
Internal Power Supply ................................................................................................................................ 18
External Power Supply .............................................................................................................................. 18
Input Wiring ...................................................................................................................................................... 19
Turbine Flow Meter (TFM) Input ................................................................................................................ 19
Pulse Input ................................................................................................................................................. 19
Remote Reset Input ................................................................................................................................... 20
Output Wiring.................................................................................................................................................... 21
Pulse Output ............................................................................................................................................. 21
Analog (4-20 mA) Rate Output ................................................................................................................. 21
Flow Meter Frequency Output ................................................................................................................... 22
RS-485 Output ........................................................................................................................................... 23
Section 3—Conguration and Operation via Keypad................................................................................. 25
Entering a Calibration Factor ........................................................................................................................... 26
Entering a Calculated Divisor ........................................................................................................................... 27
Setting Input Type and Sensitivity .................................................................................................................... 28
Conguring the Total Display ............................................................................................................................ 29
Conguring the Rate Display ............................................................................................................................ 30
Conguring the 4-20 mA Rate Output .............................................................................................................. 31
Conguring the Pulse Output ........................................................................................................................... 34
Entering the Slave Address .............................................................................................................................. 35

iv
Table of Contents MC-III™ Panel Mount Flow Analyzer
Entering the Baud Rate .................................................................................................................................... 35
Section 4—Conguration and Operation via Software............................................................................... 37
Installing the Software ...................................................................................................................................... 37
Accessing Help ................................................................................................................................................. 37
Connecting to the Software .............................................................................................................................. 38
Automating Functions on Software Startup ............................................................................................... 39
Changing Autorun Settings ........................................................................................................................ 40
Express Connect Option ............................................................................................................................ 40
Changing the Communications Port .......................................................................................................... 41
Software Connection in Multi-Device Network ........................................................................................... 41
Setting Log Download Preferences .................................................................................................................. 43
Conguring the MC-III Panel Mount ................................................................................................................. 44
Conguration Wizard ................................................................................................................................. 46
Table 4.1—Menus for Conguring Parameters .......................................................................................... 47
MC-III Main Screen .................................................................................................................................... 48
Buttons and Tools....................................................................................................................................... 49
System Setup ................................................................................................................................................... 52
Time/Date Synchronization ....................................................................................................................... 52
Contract Hour ............................................................................................................................................. 52
LCD Contrast Adjustment .......................................................................................................................... 52
Security Setup ............................................................................................................................................ 53
Firmware Version Number ......................................................................................................................... 53
Serial Number ............................................................................................................................................ 53
Communications Port ....................................................................................................................................... 54
Slave Address ............................................................................................................................................ 54
Baud Rate ................................................................................................................................................. 54
Bus Delay ................................................................................................................................................... 55
Bus Timeout ............................................................................................................................................... 55
Software Communication Options ............................................................................................................. 55
Wellsite Information .......................................................................................................................................... 56
Turbine Input .................................................................................................................................................... 57
Volume Display .......................................................................................................................................... 57
Rate Display ............................................................................................................................................... 57
Input Type/Sensitivity Conguration ........................................................................................................... 58
Cut-Off Thresholds ..................................................................................................................................... 58
Calculation Period ...................................................................................................................................... 58
K-Factor Entry .................................................................................................................................................. 59
K-Factor Units ............................................................................................................................................ 59
K-Factor Type............................................................................................................................................. 59
K-Factor Backup ........................................................................................................................................ 60
Gas Volume Correction (Supercompressibility Calculation) ...................................................................... 61
4-20 mA Output ................................................................................................................................................ 63
Enabling 4-20 mA Output .......................................................................................................................... 65
4-20 mA Output Testing.............................................................................................................................. 65
Pulse Output ..................................................................................................................................................... 67
Conguring Pulse Output ........................................................................................................................... 67
Pulse Output Testing .................................................................................................................................. 68
Saving and Uploading Conguration Files ....................................................................................................... 69
Saving a Conguration File ........................................................................................................................ 69
Uploading a Conguration File ................................................................................................................... 70
Advanced Access ............................................................................................................................................ 72

v
MC-III™ Panel Mount Flow Analyzer Table of Contents
Section 5—Flow Logs and Event Logs ........................................................................................................ 73
Auto-Save Log Formats.................................................................................................................................... 74
Log Directory and Filenames............................................................................................................................ 74
Flow Archive ..................................................................................................................................................... 75
Downloading Flow Logs ............................................................................................................................. 75
Viewing Trend Charts ................................................................................................................................. 77
Printing/Saving a Report ............................................................................................................................ 77
Viewing a Saved Report............................................................................................................................. 79
Exporting Flow Logs ................................................................................................................................. 81
Event Archive.................................................................................................................................................... 82
Downloading Event Logs ........................................................................................................................... 83
Printing/Saving a Report ............................................................................................................................ 84
Exporting Event Logs ................................................................................................................................ 84
Section 6 - MC-III Panel Mount Maintenance ............................................................................................... 87
Lithium Battery Replacement ........................................................................................................................... 87
Reinstalling the Velcro Strap ...................................................................................................................... 88
Circuit Assembly Replacement ......................................................................................................................... 89
Firmware Update .............................................................................................................................................. 91
Spare Parts List ................................................................................................................................................ 92
Communication Cable Assemblies ................................................................................................................... 92
Signal Cable Assemblies ................................................................................................................................. 92
Magnetic Pickups ............................................................................................................................................. 92
Appendix A—Software Program Options ...................................................................................................A-1
General Options ..............................................................................................................................................A-1
Autorun Options...............................................................................................................................................A-2
Communications Options ................................................................................................................................A-3
Express Connect Option ...........................................................................................................................A-3
Auto-Negotiate Option ..............................................................................................................................A-4
Enable Auto-Negotiated Baud Rate Option ..............................................................................................A-4
Request-to-Send (RTS) Line Option .........................................................................................................A-4
Enable Modbus Address Support for Firmware Versions 1.06 through 1.08 ............................................A-4
Timing Parameters ....................................................................................................................................A-4
Downloading Options ......................................................................................................................................A-5
Advanced Options ...........................................................................................................................................A-6
Clear EEPROM .........................................................................................................................................A-6
Automatic Data Logging ............................................................................................................................A-6
Appendix B—Lithium Battery Information..................................................................................................B-1
Lithium Battery Disposal .................................................................................................................................B-1
Transportation Information ..............................................................................................................................B-1
Material Safety Data Sheet..............................................................................................................................B-1
Appendix C—Communications Protocol ....................................................................................................C-1
Introduction ......................................................................................................................................................C-1
Supported Commands.....................................................................................................................................C-1
Data Types ......................................................................................................................................................C-1
Registers ........................................................................................................................................................C-2
System Conguration ................................................................................................................................C-3
Product Code (register 1000) ....................................................................................................................C-4
Firmware/Register Table Version Numbers (registers 1001, 1002) ..........................................................C-4
Manufacture Date/Sales Date (registers 1003, 1004) ...............................................................................C-4

vi
Table of Contents MC-III™ Panel Mount Flow Analyzer
Slave Address (register 1009) ...................................................................................................................C-4
Baud Rate (register 1010) .........................................................................................................................C-4
Real Time ..................................................................................................................................................C-5
Input Conguration ....................................................................................................................................C-5
Output Conguration .................................................................................................................................C-8
Holding Registers (16-bit Mode) ...............................................................................................................C-9
Base Units/Congured Units ...................................................................................................................C-11
Conversion Factors .................................................................................................................................C-11
Polling Registers .....................................................................................................................................C-11
Pointer/Daily/Event Pointer (registers 17001 through 17006) .................................................................C-11
Real Date (registers 17007, 17008) ........................................................................................................C-11
Real Time (registers 17009, 17010) ........................................................................................................C-12
Totals ......................................................................................................................................................C-12
Pulse Output Pulses (register 17077) .....................................................................................................C-12
Analog Output Current (register 17079) ..................................................................................................C-12
Calculated K-Factor (register 17081) ......................................................................................................C-12
Holding Registers (32-bit Mode) .............................................................................................................C-12
Control Register ......................................................................................................................................C-14
Wellsite Parameters ................................................................................................................................C-15
Log Data .................................................................................................................................................C-15
Enron Registers ......................................................................................................................................C-16
Enron Hourly/Daily Record Format .........................................................................................................C-16
Enron Event Record Format ...................................................................................................................C-16
Reset Status ...........................................................................................................................................C-17
Log Capacity ...........................................................................................................................................C-17

7
MC-III™ Panel Mount Flow Analyzer Section 1
Section 1—Introduction
The NuFlo™ MC-III™ Panel Mount Flow Analyzer (Figure 1.1) packs a full spectrum of gas and liquid
measurement functionality, high-speed performance, and log archive and retrieval capabilities in an easy-
to-use totalizer. Commonly used operations can be accessed from the six-button keypad on the front of the
instrument or from the dynamic interface software, allowing you to calibrate and congure the unit quickly
and easily. The device installs easily in a ¼ DIN panel opening measuring 3.62 in. square.
Operation
The MC-III Panel Mount calculates and displays instantaneous ow rates and accumulated totals based on
a turbine ow meter input signal. The MC-III’s microprocessor circuitry counts the pulses generated by a
companion ow meter, converts that data into volume and rate values in accordance with calibration settings,
and displays the totalized data on a two-line liquid crystal display (LCD). The eight-digit top readout indicates
total ow volume; the six-digit bottom readout indicates ow rate.
With the press of a single key, totals are saved to nonvolatile memory, minimizing the risk of data loss even if
a power outage occurs.
Up to 384 daily logs, 768 hourly logs, and 345 event logs can be archived and accessed quickly on demand.
Offering a variety of user-congurable display options, input and output options, RS-485 Modbus®
compatibility, ow logging, and turbine ow meter linearization, the MC-III Panel Mount is one of the most
versatile totalizers on the market. For specications, see Table 1.1, page 8.
Figure 1.1—MC-III Panel Mount Flow Analyzer

8
Section 1 MC-III™ Panel Mount Flow Analyzer
Table 1.1—MC-III Panel Mount Specications
System Power Internal power supply
• 3.6 VDC, D-size lithium battery (2-year typical life)
External power supply (6 to 30 VDC at 6 mA) with internal battery backup
(reverse polarity protected)
Loop-powered (4-20 mA) with internal battery backup
(reverse polarity protected)
Loop power: 8 to 30 VDC
Load resistance: 1100 ohms @ 30 VDC
200 ohms @ 12 VDC
Operating Temperature Lithium-Powered: -40°C to 70°C (-40°F to 158°F)
LCD contrast is reduced below -20°C (-4°F)
Environmental Humidity: 0 to 90% non-condensing
Altitude: Up to 2000 m, maximum
LCD Display 8-digit Total (volume) display (7-segment characters)
6-digit Rate display (11-segment characters for easy-to-read prompts)
0.3” character height
Adjustable contrast and update period
User-selectable units of measurement (Total):
• Preprogrammed units: BBL, GAL, LIT, M3, CF, SCF, any unit x 1000
• User-dened units
User-selectable units of measurement (Rate):
• Preprogrammed units: BBL, GAL, LIT, M3, CF, SCF (per DAY, HR,
MIN, SEC), any unit x 1000 (per DAY, HR, MIN, SEC)
• User-dened units
Keypad 6-key membrane switch
Communications/
Archive Retrieval
RS-485 Modbus® communications with transfer speeds up to 115.2K (allows
full download in less than 1 minute)
Logging 384 daily logs
768 hourly logs
345 event logs
Inputs Turbine Meter Input
Congurable sensitivity adjustment via front panel
Sensitivity adjustment range: 20 mV P-P to 40 mV P-P
Frequency range: 0 to 3500 Hz
Remote Reset Input
Optically-isolated input
3.0 to 30 VDC
Pulse duration > 3 seconds to reset
Pulse Input
Optically-isolated input
3.0 to 30 VDC

9
MC-III™ Panel Mount Flow Analyzer Section 1
Table 1.1—MC-III Panel Mount Specications
Outputs Analog Output
4-20 mA, loop-powered (two-wire)
16-bit resolution
Accuracy: 0.1% of full scale @ 25°C, 50 PPM/°C temperature drift
Loop power: 8.0 to 30 VDC
Zero and full-scale engineering values congurable from front panel
RS-485 Communications
Baud rates: 300, 600, 1200, 2400, 4800, 9600, 19200, 38400, 57600 and up
to 115.2K
Volumetric Pulse Output
Solid-state relay
Output rating: 60 mA max. @ 30 VDC, on-state drop = 1.4 VDC @ 50 mA,
0.25 VDC @ 10 mA
Congurable pulse width (duration): 10 to 60,000 ms
Amp & Square (Flow Meter Frequency) Output
Open-drain transistor output of turbine meter input signal
Output rating: 50 mA @ 30 VDC, on-state drop = 0.3 VDC @ 50 mA,
0.1 VDC @ 10 mA
(Analog output and amp & square outputs cannot be used simultaneously.)
Modbus®RTU mode Modbus® supports 16-bit and 32-bit holding registers. For more
information, see Appendix C.
Enron Modbus®Flow log parameters (time stamp, period total, period run time, and supply
voltage) and download method are Enron-compatible.
System Requirements Operating System - Windows 2000 or later (Windows XP recommended)
Computer/Processor - 1 GHz or faster Pentium-compatible CPU
Memory - 128 MB of RAM
Hard Disk Space - 21 MB for program les, 30 MB for Adobe Reader,
adequate space for data les
Drive - CD-ROM for install
Display - 800 x 600 (SVGA), 16-bit (thousands of colors) color display or
greater
Browser - Internet Explorer 6 or later
Internet Connection - for web links, tech support
Communications Port - physical or virtual RS-232 compatible serial port
Key Product Features
This section presents an overview of key features of the MC-III Panel Mount. Many of these features are
discussed in more detail in Sections 3 and 4 (conguration procedures) and Section 5 (ow log archival).
Key features discussed here include:
• LCD display
• keypad
• interface software
• power supply
• calibration options

10
Section 1 MC-III™ Panel Mount Flow Analyzer
• input options
• output options
• ow log archival
• password-protected security
LCD Display
The liquid crystal display (Figure 1.2, page 11) provides a simultaneous indication of accumulated total
(top readout) and ow rate (bottom readout). The eight-digit total display uses 7-segment characters to form
numbers and letters, which results in a combination of uppercase and lowercase letters. The six-digit ow rate
display uses 11-segment characters to form numbers and letters for improved readability. When the keypad
is used to calibrate the MC-III Panel Mount, the name of the menu option selected appears in the lower (rate)
display, and settings are entered in the top (total) display.
Flow volume can be measured in barrels, gallons, liters, cubic meters, cubic feet, standard cubic feet or other
user-dened units. A multiplication factor is also available for indicating ow volume in terms of 1,000 units.
The unit of measure for the Total readout and the decimal point position are selected by the operator during
calibration. If a user-dened unit is used, none of the preprogrammed volume units will be visible on the
display during operation.
Flow rate can be measured in a wide variety of preprogrammed units, or other user-dened units. The ow
rate unit of measure is selected in two steps: (1) a volume unit is chosen and (2) a time-base unit (per day, per
hour, per minute, or per second) is chosen. Users can choose any combination of preprogrammed volume and
time units in establishing the ow rate engineering unit (for example, gallons per hour, gallons per day, or
gallons per minute). Also, the volume unit used for the ow rate can be different from the volume unit used to
read Total volume. The unit of measure for the Rate readout and the decimal point position are selected by the
operator during calibration.
The daily index (Day) display is a two-digit number for selecting a daily archive log for viewing. The number
shown here represents the number of days that have passed since the log was saved. For example, an entry of
01 would yield yesterday’s log. An entry of 05 would yield the log generated 5 days ago. Up to 99 consecutive
daily logs can be viewed using the keypad. In addition, up to 384 daily logs, 768 hourly logs, and 345 event
logs can be viewed through the interface software.
The LCD contrast can be adjusted with the interface software (see Section 4).
Keypad
The six-button keypad allows users to perform a basic conguration of the instrument. Figure 1.3, page
11, summarizes the functions that can be accessed with each button. Most parameters can be congured in
seconds by selecting one of the three menu keys (K-Factor, Output, or Display), navigating settings with the
arrow buttons, and saving the selections with the Enter key.
Section 3 contains procedures for conguring the MC-III Panel Mount using the keypad. Icons of the six
buttons provide a pictorial reference to help guide users through each step of conguration.
Important: Some conguration parameters are accessible only through the interface software. See
Section 4 for instructions on conguring the instrument using the software.

11
MC-III™ Panel Mount Flow Analyzer Section 1
00000000
000000
M3
BBL
GAL
LIT
M3
BBL
GAL
LIT
/SEC
/MIN
/HR
/DAY
CF
CF
X1000
X1000
STANDARD
00
Day
Total
Rate
Figure 1.2—LCD display showing location of the Total, Rate, and daily index (Day) displays. Only the units of
measurement selected for displaying total and rate will be visible during normal operation.
Figure 1.3—MC-III Panel Mount keypad functions
Interface Software
Developed within the familiar Windows environment, the MC-III interface software is an intuitive and
easy-to-use application that provides access to all controls for setting up and operating the instrument. The
interface tailors the controls to the user’s needs, providing three options for conguring the instrument:
• a “Conguration Wizard,” which steps through the most common conguration tasks
• individual conguration menus for accessing specic settings
• an advanced menu that gives host programmers access to Modbus®registers

12
Section 1 MC-III™ Panel Mount Flow Analyzer
For instructions on installing the interface software and entering conguration parameters via the interface
software, see Section 4.
Power Supply
The MC-III Panel Mount is shipped with a lithium battery. Alternately, the MC-III Panel Mount may be
powered by an external power source; in this case, the lithium battery provides a backup power supply,
signicantly extending battery service life. Low-power microprocessor technology enables the MC-III Panel
Mount to operate approximately 2 years on a single lithium battery.
External Power Supply. The device will automatically switch to battery power when external power is lost. It
is recommended that the external power supply be an approved ELV source, insulated from the AC main by
double/reinforced insulation per CSA C22.2 No. 61010-1-04 / UL61010-1 – 2nd Edition.
Wiring diagrams for connecting an external power supply are provided in Section 2.
Multipoint Linearization
The MC-III interface software supports up to 12 calibration points in addition to single-point calibration based
on the K-factor provided with the turbine ow meter. See K-Factor Type, page 59 for more information.
Gas Volume Correction
Gas turbine meters are calibrated in actual cubic feet (ACF), and measure gas in actual cubic feet. In some
applications, a user may benet from referencing gas measurements back to standard conditions by measuring
in terms of standard cubic feet (SCF). The MC-III Panel Mount makes this process quick and easy, by using
xed average values for the owing gas temperature and owing gas pressure. See Gas Volume Correction
(Supercompressibility Calculation), page 61 for more information.
Input Options
The ow meter signal can be obtained from a magnetic pickup or a pre-amplier device. The sensitivity of
the ow meter input may be adjusted with the instrument keypad or the interface software. See Section 2 for
wiring diagrams. See Sections 3 and 4 for conguration procedures.
Output Options
The MC-III Panel Mount standard circuitry provides:
• a scaled pulse output representing an increment in volume for each pulse
• a loop-powered 4-20 mA output representing the ow rate
• a ow meter frequency output for use with remote equipment to derive ow rate and volume
• an RS-485 output for communication with interface software or other telemetry equipment
When the 4-20 mA rate output feature is used, the MC-III Panel Mount is powered by the current loop, and
the lithium battery is used as a backup supply.
The pulse output and 4-20 mA output features should be turned off when not required for reduced current
consumption. See Section 2 for wiring diagrams. See Sections 3 and 4 for conguration procedures.
Flow Log Archival
The MC-III Panel Mount saves up to 384 daily logs and 768 hourly logs in nonvolatile memory. By
connecting with the interface software, users can download the logs for viewing and/or printing in tabular
format or in a trend chart.
Users can also export daily and hourly logs to a spreadsheet. For more information, see Section 5.

13
MC-III™ Panel Mount Flow Analyzer Section 1
Event Log Archival
The MC-III Panel Mount saves up to 345 user event logs. Event logs are generated to track user changes such
as
K-factor changes, input setting changes, power-on reset and “watch-dog” reset, ow cut-off and frequency
cut-off.
By connecting with the interface software, users can download the logs for viewing and/or printing in tabular
format. In addition to showing old and new values, each event log is time-stamped, and includes the register
associated with the change. For more information, see Section 5.
Password-Protected Security
A keypad security access code prevents unauthorized personnel from altering the calibration or accumulated
volume data in the instrument. The security feature may be disabled if this protection is not required.
Password-protected security access is enabled using the interface software. When this feature is enabled,
the user will be prompted for a password when attempting to enter any menu from the keypad. For more
information, see Security Setup, page 53.
Commonly Used Functions
While the functions of the MC-III Panel Mount are too numerous to mention, some of the most commonly
used functions are detailed in this section. They include:
• reading the rate and accumulated total
• saving totals to memory
• resetting the total
• viewing daily and hourly logs
• saving and uploading conguration les
• exporting log data to spreadsheet
• saving log data in a report
Reading Totals
Current totals can be viewed from the LCD on the front of the MC-III Panel Mount or from the interface
software (MC-III Main screen). The software calculates the ow total and updates the LCD display every 4
seconds, by default. The user can adjust the calculation period with the interface software. See Calculation
Period, page 58, for more information.
Saving Totals to Memory
Hourly and daily totals are automatically saved to nonvolatile memory. A user may also save an accumulated
total at any time by pressing ENTER (SAVE) on the keypad. In the event of a power failure, the last saved
total will be displayed on the LCD when power is restored.
Important: Always save the accumulated total before replacing batteries.
Resetting the Total
Totals can be reset to zero using the keypad, the interface software, or a pulse from an external device.
• To reset the total with the keypad, press LEFT ARROW (LOG) and ENTER (SAVE) keys simultaneously.

14
Section 1 MC-III™ Panel Mount Flow Analyzer
• To reset the total with the interface software, double-click the MC-III icon on the computer desktop and
wait for the software to connect to the instrument; then select MC-III Main from the Device Autorun Op-
tions screen, and click on the “Reset Flow Total” button in the lower right corner of the Main screen.
• To reset the total remotely, wire a switch or a pulse-generating device to the remote reset input connector
on the MC-III board. See Remote Reset Input, page 20.
Viewing Daily and Hourly Logs
Each day, as the user-dened contract hour passes, a daily ow total is saved to nonvolatile memory. Hourly
logs are also automatically saved. A total of 384 daily logs and 768 hourly logs are accessible for viewing and
exporting using the interface software. See Section 5 for details.
Up to 99 daily ow log totals can be viewed from the LCD. Hourly ow log totals are accessible only through
the interface software.
To view daily ow totals from the LCD, perform the following steps:
1. Press the LEFT ARROW (LOG) key on the keypad. The words “Daily Volume Archive” will scroll across
the bottom of the LCD and the day index will display “01.” The daily ow total recorded at the last con-
tract hour will appear at the top of the LCD.
The index number represents the number of days previous to the current date. For example, yesterday’s
totals are read by entering an index of “01”; totals from two days previous are read by entering “02.”
2. Press UP ARROW (TEST) to increment the index (01, 02, 03...); press LEFT ARROW (LOG) to decre-
ment the index (01, 99, 98...).
3. Press Enter (Save) to exit the Daily Volume Archive menu. (After 2 minutes of inactivity, the Daily Vol-
ume Archive menu will time out and the total readout will be restored automatically.)
Saving and Uploading Conguration Files
The MC-III interface software allows users to save an unlimited number of conguration les to their
computer. In the event that a conguration setting gets changed unintentionally or a user simply wants to
restore the settings he used previously, the user can upload the conguration le and resume operation within
minutes. The upload function also allows a conguration le to be loaded quickly into multiple devices. The
default directory for saving conguration les is C:\NuFlo log data\MC-III. However, users can specify a
different location, if desired.
For complete information, see Saving and Uploading Conguration Files, page 69.
Exporting Log Data
Flow logs and event logs can be directly exported to an .xls or .csv le. For complete information, see
Exporting Flow Logs, page 81, and Exporting Event Logs, page 84. The default directory for exported
logs is C:\NuFlo log data\MC-III. However, users can specify a different location, if desired.
Saving Log Data in a Report
Flow logs can be saved in a report format that can be loaded back into the software for viewing or printing
at a later time. For complete information, see Printing/Saving a Report, page 77. The default directory for
log reports is C:\NuFlo log data\MC-III\<WELL NAME>. However, users can specify a different location, if
desired.

15
MC-III™ Panel Mount Flow Analyzer Section 2
Section 2—Installation
The MC-III™ Panel Mount is fully assembled at the time of shipment and ready for installation. Before
attempting to install the MC-III Panel Mount, make sure the ow meter and magnetic pickup are installed.
Flow Meter/Magnetic Pickup Installation
Install the ow meter and magnetic pickup as follows:
1. Install the turbine ow meter in the ow line.
2. Lightly grease the threads on both ends of the magnetic pickup, taking care to keep grease off of the con-
nector contacts. If the connector is plastic, apply grease only to the end that threads into the meter.
3. Install the magnetic pickup in the ow meter.
4. After the ow meter and magnetic pickup are installed in the ow line, mount the MC-III Panel Mount
ow analyzer as described below.
MC-III Installation
The MC-III mounts in a standard ¼ DIN square opening (3.62” X 3.62”) in a vertical panel. To install,
reference the dimensions in Figure 2.1, page 16.
Studs (#6-32 X ¾”) and nuts (#6-32) with integral lockwasher are supplied for mounting the instrument.
Additional customer-provided mounting fasteners such as washers or clamps are required to secure the device
from the back side of the panel. The style will vary with the installation and the thickness of the panel used.

16
Section 2 MC-III™ Panel Mount Flow Analyzer
SIDE VIEW
TOP VIEW
3.10
[78.74]
STUDS (4)
#6-32
3.10
[78.74]
BETWEEN
STUDS
Figure 2.1—Typical mount dimensions in inches (millimeters)

17
MC-III™ Panel Mount Flow Analyzer Section 2
Field Wiring Connections
All eld wiring connects to the circuit assembly mounted on the back side of the switchplate.
Wiring Procedure
To wire the MC-III Panel Mount for operation, complete the following eld connections:
1. Connect the lithium battery connector to the J1 receptacle on the circuit assembly.
2. Connect wiring for external power, if appropriate. See Figure 2.2, page 18.
3. Connect the ow meter or pulse input wiring to terminal block TB1. See Figure 2.3 or Figure 2.4, page
19.
4. Connect wiring for the remote reset input to terminal block TB1, if appropriate. See Figure 2.5 or Figure
2.6, page 20.
5. Connect wiring for output signals, if appropriate. See Figure 2.7, page 21, through Figure 2.12, page
24.
6. Recalibrate the MC-III Panel Mount (if necessary).
7. If external and internal power supplies were removed, reset the clock to ensure that the time stamps in the
log data are accurate. The clock is reset using the MC-III interface software. See Time/Date Synchroniza-
tion, page 52.

18
Section 2 MC-III™ Panel Mount Flow Analyzer
Power Supply Wiring
Internal Power Supply
The MC-III Panel Mount is shipped with a 3.6-V lithium battery.
Low-power microprocessor technology enables the MC-III Panel Mount analyzer to operate approximately
2 years on a single lithium battery. The lithium battery is strongly recommended for use in extreme
temperatures (below -20°C).
Users can power the instrument from an external power supply or a 4-20 mA current loop, and use the lithium
as a backup power supply. The use of an alternate power source extends battery life and helps ensure that
timekeeping and volume accumulation will not be interrupted during a power failure.
External Power Supply
The MC-III Panel Mount can be connected to a remote power supply by a two-conductor cable (Figure 2.2).
The power supply and cable must be capable of supplying 6 to 30 VDC @ 10 mA. This capability is available
only if the 4-20 mA rate output is not used.
External Power Supply. The device will automatically switch to battery power when external power is lost. It
is recommended that the external power supply be an approved ELV source, insulated from the AC main by
double/reinforced insulation per CSA C22.2 No. 61010-1-04 / UL61010-1 – 2nd Edition.
Caution: When using the amp & square output with the external power supply, make sure the
power supplies for both features share a common negative (-) terminal or that they are
totally isolated from each other, since both share a common negative (-) connection.
TB3
TB1
PULSE
INPUT
RESET
INPUT
TFM
A&S
GND
EXT POWER
TB2
RS485
SLAVE
4-20
OUT
PULSE
OUT
6-30VDC
POWER
SUPPLY
6 to 30
VDC
J1
J2
RESET
SWITCH
BATTERY
Figure 2.2—External power supply wiring

19
MC-III™ Panel Mount Flow Analyzer Section 2
Input Wiring
Turbine Flow Meter (TFM) Input
The TFM input provides the turbine ow meter input signal generated by a magnetic pickup, enabling the
MC-III Panel Mount to calculate and display instantaneous ow rates and accumulated totals.
6-30VDC
TB1
PULSE
INPUT
RESET
INPUT
TFM
A&S
GND
TB3
EXT POWER
TURBINE
MAGNETIC PICKUP
J1
J2
RESET
SWITCH
BATTERY
A
B
Figure 2.3—Flow meter input wiring
Pulse Input
The pulse input provides an optically isolated input in systems where a preamplier is inserted between the
sensor and the MC-III Panel Mount.
TB3
TB1
PULSE
INPUT
RESET
INPUT
TFM
A&S
GND
EXT POWER
6-30VDC
J1
J2
RESET
SWITCH
BATTERY
PULSE INPUT
3 TO 30 VDC
Figure 2.4—Pulse input wiring
For a list of recommended pickup adapters,
see the parts list on 92.
Vturbine max = 5.0 Vdc or peak
Voc = 1.0 Vdc
Isc = 5.0 mA max
Ca = 30 µF max
La = 2 H max

20
Section 2 MC-III™ Panel Mount Flow Analyzer
Remote Reset Input
The remote reset input allows the operator to reset the accumulated volume on the MC-III Panel Mount to
zero with a switch or a pulse. This input is optically isolated. The input is shown connected in two ways, with
a power supply and switch in a remote location (Figure 2.5), and with a pulse generator in a remote location
(Figure 2.6).
The reset input or reset pulse must be active for 3 seconds to clear the total.
TB1
PULSE
INPUT
RESET
INPUT
TFM
A&S
GND
TB3
EXT POWER
6-30VDC
J1
J2
RESET
SWITCH
BATTERY
POWER SUPPLY
3 to 30 VDC
Figure 2.5—Reset input wiring (switch/power supply)
TB1
PULSE
INPUT
RESET
INPUT
TFM
A&S
GND
TB3
EXT POWER
6-30VDC
J1
J2
RESET
SWITCH
BATTERY
RESET PULSE
3 TO 30 VDC
Figure 2.6—Reset input wiring (pulse)
Table of contents
Other Cameron Measuring Instrument manuals

Cameron
Cameron NUFLO Scanner 2000 microEFM Instructions for use

Cameron
Cameron Scanner 3100 User manual

Cameron
Cameron Barton 199 User manual

Cameron
Cameron 351 User manual

Cameron
Cameron NUFLO MC-III EXP User manual

Cameron
Cameron LEFM 200 User manual

Cameron
Cameron NUFLO EZ-IN Series User manual

Cameron
Cameron Scanner 3100 EFM User manual

Cameron
Cameron NUFLO 1141C User manual

Cameron
Cameron NUFLO Scanner 2000 microEFM User manual