manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. CEM
  6. •
  7. Laboratory Equipment
  8. •
  9. CEM MARS 6 User manual

CEM MARS 6 User manual

Other manuals for MARS 6

3

Other CEM Laboratory Equipment manuals

CEM MARS Series User manual

CEM

CEM MARS Series User manual

CEM Discover SP-D 10/35 Configuration guide

CEM

CEM Discover SP-D 10/35 Configuration guide

CEM EDGE User manual

CEM

CEM EDGE User manual

CEM DT-982 User manual

CEM

CEM DT-982 User manual

CEM Razor User manual

CEM

CEM Razor User manual

CEM MARS 6 User manual

CEM

CEM MARS 6 User manual

CEM Liberty Blue 551290 User manual

CEM

CEM Liberty Blue 551290 User manual

CEM Razor User manual

CEM

CEM Razor User manual

CEM EasyPrep Series User manual

CEM

CEM EasyPrep Series User manual

CEM Discover Protein Hydrolysis User manual

CEM

CEM Discover Protein Hydrolysis User manual

CEM MARS 6 User manual

CEM

CEM MARS 6 User manual

CEM Q-Dry User manual

CEM

CEM Q-Dry User manual

CEM MARS 6 iWave User manual

CEM

CEM MARS 6 iWave User manual

CEM Discover Bio User manual

CEM

CEM Discover Bio User manual

CEM Discover SPS User manual

CEM

CEM Discover SPS User manual

CEM EDGE User manual

CEM

CEM EDGE User manual

CEM liberty User manual

CEM

CEM liberty User manual

CEM MARSXpress Plus User manual

CEM

CEM MARSXpress Plus User manual

CEM liberty User manual

CEM

CEM liberty User manual

CEM Smart Trac II User manual

CEM

CEM Smart Trac II User manual

CEM iPrep Series User manual

CEM

CEM iPrep Series User manual

CEM Discover SP-D User manual

CEM

CEM Discover SP-D User manual

CEM DT-986H User manual

CEM

CEM DT-986H User manual

CEM iPrep Guide

CEM

CEM iPrep Guide

Popular Laboratory Equipment manuals by other brands

Radleys Carousel 6 Plus instructions

Radleys

Radleys Carousel 6 Plus instructions

Hemotherm 400MR Operation & technical manual

Hemotherm

Hemotherm 400MR Operation & technical manual

mrclab DVO Series Operation manual

mrclab

mrclab DVO Series Operation manual

CoolLED pE-800 quick start guide

CoolLED

CoolLED pE-800 quick start guide

Luminex Guava easyCyte Maintenance Guide

Luminex

Luminex Guava easyCyte Maintenance Guide

IKA A10 basic operating instructions

IKA

IKA A10 basic operating instructions

Belden HIRSCHMANN RPI-P1-4PoE installation manual

Belden

Belden HIRSCHMANN RPI-P1-4PoE installation manual

Koehler K1223 Series Operation and instruction manual

Koehler

Koehler K1223 Series Operation and instruction manual

Globe Scientific GCM-12 quick start guide

Globe Scientific

Globe Scientific GCM-12 quick start guide

Getinge 86 SERIES Technical manual

Getinge

Getinge 86 SERIES Technical manual

CORNING Everon 6000 user manual

CORNING

CORNING Everon 6000 user manual

Biocomp GRADIENT MASTER 108 operating manual

Biocomp

Biocomp GRADIENT MASTER 108 operating manual

Cole Parmer Stuart SB3/1 instruction manual

Cole Parmer

Cole Parmer Stuart SB3/1 instruction manual

BEVS 2501/2A user manual

BEVS

BEVS 2501/2A user manual

Qiagen DML 3000 user manual

Qiagen

Qiagen DML 3000 user manual

Queensgate NANOSCAN OP400 Quick start instructions

Queensgate

Queensgate NANOSCAN OP400 Quick start instructions

Parr Instrument 4560 Operating instructions manual

Parr Instrument

Parr Instrument 4560 Operating instructions manual

Integra DOSE IT operating instructions

Integra

Integra DOSE IT operating instructions

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.

CEM Corporation
P.O. Box 200
Matthews, NC 28106
United States
Tel: (800) 726-3331 [USA & Canada]
Tel: (704) 821-7015
Fax: (704) 821-7894
e-mail: [email protected]om
www.cem.com
UNITED KINGDOM
CEM Microwave Technology Ltd.
2 Middle Slade
Buckingham Industrial Estate
Buckingham MK18 1WA
Tel: (44) 1280-822873
Fax: (44) 1280-822342
e-mail: info[email protected]
GERMANY
CEM GmbH
Carl-Friedrich-Gauss -Str. 9
47475 Kamp-Lintfort
Tel: (49) 2842-9644-0
Fax: (49) 2842-9644-11
e-mail: [email protected]
www.cem.de
ITALY
CEM S.R.L.
Via Dell’Artigianato, 6/8
24055 Cologno al Serio (Bg)
Tel: (39) 35-896224
Fax: (39) 35-891661
e-mail: info[email protected]
www.cemmicroonde.it
FRANCE
CEM µWave S.A.S.
Immeuble Ariane
Domaine Technologique de Saclay
4, rue Rene’Razel
91892 ORSAY Cedex
Tel: (33-1) 69 35 57 80
e-mail: info[email protected]
www.cemfrance.fr
JAPAN
CEM Japan K.K.
2-18-10 Takanawa
Minato-ku, Tokyo
108-0074
Tel: +81-3-5793-8542
Fax: +81-3-5793-8543
e-mail: [email protected]
www.cemjapan.co.jp
IRELAND
CEM Technology (Ireland) Ltd
Sky Business Centre
9a Plato Business Park
Damastown
Dublin 15
Tel +353 (0)1 885 1752
Fax +353 (0)1 885 1601
Email: info[email protected]
Quick Start Guide
Read Me First
600286 Rev. 2
Register your system
Get the latest information on your new MARS 6.
Visit us online at www.cem.com today to register your system.
Registration gives you access to download software updates, method
notes, replacement parts lists for your system and vessels, and more.
Go to www.cem.com
Click on Sign up
And register your new CEM product
All you need is your name, contact information, and unit serial number.
Once your product is registered, please allow up to 24 hours for your settings to up-
date. To nd your owner resources, click on the resource library tab of your product.
Already a member? Update your prole with your new serial number to gain ccess
to owner content.
Need help? Contact CEM: [email protected]
13
System Setting
Screen Context Menu
Classic Menu
One Touch Method
Back
Home
Done
Info
Run
Edit
Delete
New
Save
Add Stage
Remove Stage
Fiberoptic
Infrared
DuoTemp
Turntable
Blower
Print
Temperature
Pressure
Power
Sub
Congratulations on your
purchase of a CEM MARS 6.
The MARS 6 that you have chosen will be a valuable addition to
your laboratory.
To prepare for installing your system, a designated area with a sturdy
workbench should be arranged to allow the MARS 6 unit to be
conveniently located near a hood or other suitable fume disposal
mechanism. The system must be located within 8 feet of a hood or
exhaust line for proper ventilation.
Do not place the instrument within a fume hood. The system
comes with 8 feet of exible exhaust hose (3 1/8”OD x 2 13/16”ID).
The built-in exhaust system will push fumes through the hose at
125 CFM. To properly vent the MARS 6, your hood or exhaust line
must draw at least 125 CFM at the point of connection.
Preparing an Installation Location
1. Provide at least 8 inches space on each side and 6
inches space in the rear for proper ventilation.
Dimensions:
21 in W x 25 in D x 25 in H
Weight 140 lbs without vessels
2. Allow adequate bench space for sample handling
3. Ensure the area is free from the vibration of large
equipment and/or walk-through trac.
4. Ensure the area is away from the primary laboratory
exits and walk-through trac.
5. Location should have a temperature range of
41oF (5 oC) to 104 oF (40 oC).
6. Location should permit the instrument to be
connected to a dedicated, grounded outlet.
The MARS 6 unit requires 208/230 VAC, 60 Hz, 15A @ 230 VAC, or
200/220/240 VAC, 50 Hz, 15A@ 240 VAC, single phase power. It
requires a 250 volt-15 Amp rated receptacle (NEMA 6-15R or a 250
volt- 20 Amp rated receptacle (NEMA 6-20R). It is non-locking.
Open box. Remove foam stabilizers. Remove outer box and locate hand hold
cutouts on either side of the inner box.
Remove inner box and place
on oor in front of bench.
Open inner box and remove packing materials.
Lift MARS 6 onto bench top near
exhaust service.
Remove contents from cavity.
Push power button to turn unit on.
Attach exhaust hose and connect
to exhaust. Plug power cord into
MARS 6 and outlet.
Note: Unit requires 26”of space between
the benchtop and upper cabinets
and 6”of clearance on the sides.
123
456
789
10
Unpacking Instructions:
1. Determine whether the line
voltage in your lab is 208V or 230V.
2. Set line voltage on the MARS 6.
Restart system for the setting to activate.
Watch the Welcome Video. Select
the “Settings”menu, then
“Video”, then“Welcome” for a short
introduction to your new MARS 6.
12
11
Icon Reference Guide
System Setting
Screen Context Menu
Classic Menu
One Touch Method
Back
Home
Done
Info
Run
Edit
Delete
New
Save
Add Stage
Remove Stage
Fiberoptic
Infrared
DuoTemp
Turntable
Blower
Print
Temperature
Pressure
Power
Sub
Congratulations on your
purchase of a CEM MARS 6.
The MARS 6 that you have chosen will be a valuable addition to
your laboratory.
To prepare for installing your system, a designated area with a sturdy
workbench should be arranged to allow the MARS 6 unit to be
conveniently located near a hood or other suitable fume disposal
mechanism. The system must be located within 8 feet of a hood or
exhaust line for proper ventilation.
Do not place the instrument within a fume hood. The system
comes with 8 feet of exible exhaust hose (3 1/8”OD x 2 13/16”ID).
The built-in exhaust system will push fumes through the hose at
125 CFM. To properly vent the MARS 6, your hood or exhaust line
must draw at least 125 CFM at the point of connection.
Preparing an Installation Location
1. Provide at least 8 inches space on each side and 6
inches space in the rear for proper ventilation.
Dimensions:
21 in W x 25 in D x 25 in H
Weight 140 lbs without vessels
2. Allow adequate bench space for sample handling
3. Ensure the area is free from the vibration of large
equipment and/or walk-through trac.
4. Ensure the area is away from the primary laboratory
exits and walk-through trac.
5. Location should have a temperature range of
41oF (5 oC) to 104 oF (40 oC).
6. Location should permit the instrument to be
connected to a dedicated, grounded outlet.
The MARS 6 unit requires 208/230 VAC, 60 Hz, 15A @ 230 VAC, or
200/220/240 VAC, 50 Hz, 15A@ 240 VAC, single phase power. It
requires a 250 volt-15 Amp rated receptacle (NEMA 6-15R or a 250
volt- 20 Amp rated receptacle (NEMA 6-20R). It is non-locking.
Open box. Remove foam stabilizers. Remove outer box and locate hand hold
cutouts on either side of the inner box.
Remove inner box and place
on oor in front of bench.
Open inner box and remove packing materials.
Lift MARS 6 onto bench top near
exhaust service.
Remove contents from cavity.
Push power button to turn unit on.
Attach exhaust hose and connect
to exhaust. Plug power cord into
MARS 6 and outlet.
Note: Unit requires 26”of space between
the benchtop and upper cabinets
and 6”of clearance on the sides.
123
456
789
10
Unpacking Instructions:
1. Determine whether the line
voltage in your lab is 208V or 230V.
2. Set line voltage on the MARS 6.
Restart system for the setting to activate.
Watch the Welcome Video. Select
the “Settings”menu, then
“Video”, then“Welcome” for a short
introduction to your new MARS 6.
12
11
Icon Reference Guide
CEM Corporation
P.O. Box 200
Matthews, NC 28106
United States
Tel: (800) 726-3331 [USA & Canada]
Tel: (704) 821-7015
Fax: (704) 821-7894
e-mail: [email protected]om
www.cem.com
UNITED KINGDOM
CEM Microwave Technology Ltd.
2 Middle Slade
Buckingham Industrial Estate
Buckingham MK18 1WA
Tel: (44) 1280-822873
Fax: (44) 1280-822342
e-mail: info[email protected]
GERMANY
CEM GmbH
Carl-Friedrich-Gauss -Str. 9
47475 Kamp-Lintfort
Tel: (49) 2842-9644-0
Fax: (49) 2842-9644-11
e-mail: [email protected]
www.cem.de
ITALY
CEM S.R.L.
Via Dell’Artigianato, 6/8
24055 Cologno al Serio (Bg)
Tel: (39) 35-896224
Fax: (39) 35-891661
e-mail: info[email protected]
www.cemmicroonde.it
FRANCE
CEM µWave S.A.S.
Immeuble Ariane
Domaine Technologique de Saclay
4, rue Rene’Razel
91892 ORSAY Cedex
Tel: (33-1) 69 35 57 80
e-mail: info[email protected]
www.cemfrance.fr
JAPAN
CEM Japan K.K.
2-18-10 Takanawa
Minato-ku, Tokyo
108-0074
Tel: +81-3-5793-8542
Fax: +81-3-5793-8543
e-mail: [email protected]
www.cemjapan.co.jp
IRELAND
CEM Technology (Ireland) Ltd
Sky Business Centre
9a Plato Business Park
Damastown
Dublin 15
Tel +353 (0)1 885 1752
Fax +353 (0)1 885 1601
Email: info[email protected]
Quick Start Guide
Read Me First
600286 Rev. 2
Register your system
Get the latest information on your new MARS 6.
Visit us online at www.cem.com today to register your system.
Registration gives you access to download software updates, method
notes, replacement parts lists for your system and vessels, and more.
Go to www.cem.com
Click on Sign up
And register your new CEM product
All you need is your name, contact information, and unit serial number.
Once your product is registered, please allow up to 24 hours for your settings to up-
date. To nd your owner resources, click on the resource library tab of your product.
Already a member? Update your prole with your new serial number to gain ccess
to owner content.
Need help? Contact CEM: [email protected]om
13